RDC 665/2022 — Good Manufacturing Practice (GMP) for Medical Devices

Link directly to a passage with its anchor — for example #art-1 — to open it highlighted. This reference version is generated from the authority’s publication and does not replace it. Language: English.

COLLEGIATE BOARD RESOLUTION – RDC NO. 665 OF 30 MARCH 2022 Provides for the Good Manufacturing Practices for Medical Products and In Vitro Diagnostic Products. The Collegiate Board of Directors of the Brazilian Health Regulatory Agency, in the use of the attributions vested in it under Article 15, items III and IV, and Article 7, items III and IV of Law no. 9,782 of 26 January 1999, and item VI, paragraph 1 of Article 187 of the Internal Regulation approved by Collegiate Board Resolution – RDC no. 585 of 10 December 2021, adopts the following Collegiate Board Resolution, as decided upon in Extraordinary Meeting – Rextra No. 6, held on 30 March 2022, and I, Director-President, determine its publication.

Chapter I — INITIAL PROVISIONS

Section I · Objective

art-1

Article 1

This regulation provides for the Good Manufacturing Practices (GMP) for Medical Products and In Vitro Diagnostic Products, establishing the requirements that describe the GMP for methods and controls used in the design, purchasing, manufacturing, packaging, labeling, storage, distribution, installation, and technical assistance applicable to the manufacture of medical products and in vitro diagnostic products.

art-1-p1

§ 1

The requirements referred to in the caption of this article are intended to ensure that medical products and in vitro diagnostic products are safe and effective.

art-1-p2

§ 2

This Resolution incorporates into the national legal system the Resolution issued by the Mercosur’s Common Market Group (GMC, in Portuguese) No. 20 of 17 November 2011, MERCOSUR/GMC/RES. No. 20/11, "Mercosur Technical Regulation on Good Manufacturing Practices for Medical Products and In Vitro Diagnostic Products (Revoking GMC Resolutions No. 04/95, 38/96, 65/96, and 131/96)".

Section II · Scope

art-2

Article 2

This regulation applies to manufacturers, distributors, storers, and importers of medical products and in vitro diagnostic products that are commercialized in Brazil.

art-2-p1

§ 1

When the manufacturers referred to in the caption of this article conclude that certain requirements established in this Resolution are not applicable to their processes, they must document the justification for such understanding.

art-2-p2

§ 2

The distributors of medical products and in vitro diagnostic products must comply with the following requirements of this Resolution, at least: I – Chapters I, VII, and VIII, in full; II – Chapter II, in full, except Section IV; III – Chapter III,

Section I

; IV – Chapter V, articles 67, 68, 69, 70, 71, 72, 73, 74, 75, 76, and 77, in addition to Section IV; and V – Chapter VI, in full, except Article 119.

art-2-p3

§ 3

The storers of medical products and in vitro diagnostic products must comply with the following requirements of this Resolution, at least: I – Chapters I and VII, in full; II – Chapter II, in full, except Section IV; III – Chapter III,

Section I

; IV – Chapter V, articles 67, 68, 69, 70, 71, 72, 73, 74, 75, 76, and 77; and V – Chapter VI, in full, except Article 119.

art-2-p4

§ 4

The importers of medical products and in vitro diagnostic products must comply with the following requirements of this Resolution, at least: I – Chapters I, II, VII, VIII, and IX in full; II – Chapter III,

Section I

and Section III; III – Chapter IV, Article 63, items III, IV, and V; IV – Chapter V, articles 67, 68, 69, 70, 71, 72, 73, 74, 75, 76, 77, 85, 86, and 87, in addition to Sections III and IV; and V – Chapter VI, in full, except Article 119.

art-2-p5

§ 5

Companies that carry out more than one activity must comply with the specific requirements defined for each activity.

art-2-p6

§ 6

The minimum requirements to be complied with, defined in Paragraphs 2, 3, and 4 of this Article, are applicable to distributors, storers, and importers, even if the provisions mention the word “manufacturer” only.

Section III · Definitions

art-3

Article 3

For the purposes of this Resolution, the following definitions are adopted: I – technical assistance: maintenance or repair of a finished product, in order to restore it to its specifications; II – quality audit: an established, systematic, and independent examination of a manufacturer's entire quality system, performed at regular intervals and with sufficient frequency to ensure that the quality system’s activities and its results comply with the procedures specified in its quality system; III – component: raw material, substance, part, piece, software, hardware, packaging, label, or instruction for use, used during the manufacture of a medical product or in vitro diagnostic product, intended to be included as part of the finished product; IV – design input: description of physical attributes, indication of use, performance, compatibility, safety, efficacy, ergonomics, usability, information retrieved from previous designs, and risk management results, amongst other requirements for medical products or in vitro diagnostic products, that are used as the basis for a design development; V – design output: results from the work at each stage of the design development process, including its final result, which, when concluded, is the basis for the device master record (DMR); VI – damage: a person’s physical lesion or health injury or a property or environmental deterioration; VII – specifications: requirements that products, components, production activities, technical assistance, services, quality system, or any other activity must comply with; VIII – to establish: to define, to document either in writing or electronically, and implement; IX – manufacturer: any person who design, manufacture, assemble, or process a finished product, including those who execute sterilization, labeling, and packaging activities by contract; X – executive management: a company’s senior management team, responsible for providing resources and with authority to establish or modify the company's quality policy and system; XI – risk management: policies, procedures, and management practices applied systematically to the tasks of analysis, assessment, control, and risk monitoring associated to a given product or process; XII – lot or batch: quantity of a product elaborated in a manufacturing or sterilization cycle, the essential characteristic of which is homogeneity; XIII – manufacturing material: material or substance used during the manufacturing process or to support this process, including cleaning agents, molding releasing agents, lubricating oils, sterilizing agents, or other subproducts from the manufacturing process; XIV – non-conformity: failure to comply with a previously specified requirement; XV – serial number or batch: unique combination of letters and/ or numbers, from which the complete purchase, manufacture, packaging, labeling, and distribution history of the finished products may be determined; XVI – hazard: potential source of harm; XVII – quality policy: totality of an organization's intentions and guidelines related to quality, expressed by the executive management; XVIII – special process: any process the results of which cannot be fully verified by subsequent inspections and tests; XIX – production: all activities involved in the manufacturing of a given product, from the receipt of components, through processing and packaging, until the obtention of the finished product; XX – finished product: any product or accessory already packaged and labeled and ready for use; XXI – quality: all the aspects and characteristics that allow a medical product or in vitro diagnostic product to meet suitability, safety, and performance requirements; XXII – complaint: written, oral or electronic communication, related to non-acceptance of a product’s identity, quality, durability, reliability, safety, efficacy, or performance; XXIII – record: document set down in writing or electronically, which evidences data, facts, specific events, and the results achieved regarding compliance with the quality system procedures and standards; XIV – device history record: compilation of records comprising the complete production history of a finished product; XV – design history record: compilation of documents comprising the complete design history of a finished product; XVI – device master record (DMR): compilation of documents including specifications, instructions, and procedures for the obtention of a finished product, in addition to its installation, technical assistance, and maintenance; XVII – rework: partial or total manufacturing activities intended to correct non-conformities of a component, intermediary product, or finished product, so that it complies with the specifications defined in the DMR; XVIII – design review: documented systematic and complete examination performed during the design development to assess its compliance to the design development planning and the objectives established; XIX – risk: combination of occurrence probability and damage severity; XXX – quality system: organizational structure, responsibilities, procedures, specifications, processes, and resources required for quality management; XXXI – validation: proof by analysis and objective evidence that the requirements defined for a given purpose consistently lead to the expected result; XXXII – verification: proof by analysis and presentation of objective evidence that the requirements specified have been met, including the process of assessing the results of an activity to determine compliance with the specifications established; XXXIII – lifespan: manufacturer’s estimated time length during which a product correctly performs its originally intended functions.

art-3-p1

§ 1

The procedures referred to in item II of the caption of this article must be implemented efficiently and also be adequate so the quality system’s objectives can be achieved.

art-3-p2

§ 2

The quality audit referred to in item II of the caption of this article differs from other quality system’s activities required in this Resolution.

art-3-p3

§ 3

When applied to a project, the validation mentioned in item XXXI of the caption of this article means establishing and documenting objective evidence that the product specifications meet the user's needs and its intended use.

art-3-p4

§ 4

When applied to a process, the validation referred to in item XXXI of the caption of this article means establishing and documenting objective evidence that the process will consistently produce a result that meets the predetermined specifications.

Chapter II — QUALITY SYSTEM GENERAL REQUIREMENTS

Section I · General requirements

art-4

Article 4

Each manufacturer must establish and maintain a quality system to ensure that the requirements of this Resolution are met and the products manufactured are safe, effective, and adequate for the intended use.

art-4-sole

Sole paragraph

As part of quality system activities referred to in the caption of this article, each manufacturer must: I – establish and maintain effective quality system instructions and procedures in accordance with the requirements in this Resolution; and II – establish procedures to meet the legal provisions determined in the health legislation in force.

Section II · Management responsibility Subsection I Quality policy

art-5

Article 5

The executive management of each manufacturer must establish its quality policy and objectives, which must be measurable and consistent with the policy established.

art-6

Article 6

The executive management must keep the quality policy at all levels of the organization.

art-7

Article 7

The executive management must ensure that the quality policy is described in a quality manual and that this policy is understood by all employees who may affect or influence the quality of a product. Subsection II Organization and responsibilities

art-8

Article 8

Each manufacturer must: I – establish and maintain an adequate organizational structure, expressed by an organizational chart, with sufficient personnel to ensure that the products are manufactured in accordance with the requirements in this Resolution; II – establish the responsibility, authority, and interrelation of all personnel who manages, executes, and verifies quality-related work, with the required independence to carry out their duties; and III – establish verification functions, provide adequate resources, and assign trained personnel to perform verification activities.

art-9

Article 9

Each manufacturer’s executive management team must assign a person from the executive management team itself, who, regardless of other activities, has authority and responsibility to: I – ensure that the quality system requirements are established and maintained in accordance with this Resolution; II – report the quality system performance assessment to the executive management for review and provide information on the quality system improvement.

art-9-sole

Sole paragraph

The assignment referred to in the caption of this article must be documented. Subsection III Management review

art-10

Article 10

The executive management of each manufacturer must assess the quality system adequacy and effectiveness at defined intervals and with sufficient frequency to ensure that the quality system meets the requirements in this Resolution and the objectives of the quality policy established.

art-11

Article 11

The management review must be conducted in accordance with the review procedures established and the results from each quality system review must be documented.

art-12

Article 12

Matters related to audit results, post-marketing information, process performance and product compliance, status of corrective and preventive actions, alterations that can affect the quality system or a product compliance, and regulatory requirements, amongst others, must be considered for management review.

Section III · Personnel

art-13

Article 13

Each manufacturer must have sufficient personnel with instruction, experience, training, and practice compatible with their job description, in order to ensure that all activities provided for in this Resolution are correctly performed.

art-14

Article 14

Descriptions must be kept defining authority, responsibility, and the necessary requirements of all personnel for the various tasks of the company.

art-15

Article 15

Each manufacturer must ensure that all personnel are trained to perform their assigned tasks properly.

art-15-p1

§ 1

The training referred to in the caption of this article must be conducted by qualified personnel, in accordance with the established procedures, to ensure that employees have an adequate understanding of their jobs and of the requirements in this Resolution that are applicable to their activities.

art-15-p2

§ 2

As part of the training mentioned in the caption of this article, all employees must be warned about product defects that may occur as a result of incorrect execution of their specific duties.

art-15-p3

§ 3

Personnel training must be documented.

art-16

Article 16

Each manufacturer must ensure that any external consultant that advises on methods or controls used for a product’s design, purchase, manufacture, packaging, labeling, storage, installation, or technical assistance activities has sufficient qualifications – education, training, and experience – to advise on matters for which the consultant was hired.

art-17

Article 17

The consultant’s hiring process must be conducted in accordance with the purchase control requirements determined in this Resolution.

Section IV · Risk management

art-18

Article 18

Each manufacturer must establish and maintain a continuous risk management process that covers the entire life cycle of a medical device or in vitro diagnostic product, from conception to discontinuation, in order to: I – identify the associated hazards; II – estimate and assess the risks involved; III – control the associated risks; and IV – assess the effectiveness of the controls established.

art-19

Article 19

The continuous risk management process must include the following elements: I – analysis; II – assessment; III – control; and IV – risk monitoring.

art-20

Article 20

The company's executive management team must designate the responsible professionals, establish the policy to determine the risk acceptability criteria, and also determine a periodic review of risk management activities, to ensure the adequacy and effectiveness of such activities.

Section V · Purchase controls

art-21

Article 21

Each manufacturer must establish and maintain procedures to ensure that components, manufacturing materials, and finished products manufactured, processed, labeled, or packaged by third parties, or stored by them under contract, comply with the specifications.

art-21-sole

Sole paragraph

Each manufacturer must ensure that the services performed by third parties referred to in the caption of this article comply with the specifications established.

art-22

Article 22

Each manufacturer must establish and maintain, according to the impact on the quality of the finished product, criteria for the assessment of suppliers, specifying the requirements that must be complied with by suppliers, including quality requirements.

art-23

Article 23

Each manufacturer must assess and select potential suppliers, according to their ability to meet the previously established requirements, maintaining a record of approved suppliers.

art-23-sole

Sole paragraph

The records of and results from supplier assessment must be kept.

art-24

Article 24

An agreement must be documented in which the suppliers agree to notify the manufacturer of any alteration in the product or service, so that the manufacturer can determine if the alteration affects the quality of the finished product.

art-25

Article 25

Each manufacturer must keep records of the purchase orders that clearly describe or refer to specifications, including quality requirements for components, manufacturing materials, finished products, or services requested or contracted.

art-26

Article 26

Each manufacturer must review and approve purchase orders prior to their release.

art-27

Article 27

Approval of purchase orders, including the date and the handwritten or digital signature of the person responsible for them, must be documented.

Chapter III — QUALITY DOCUMENTS AND RECORDS

Section I · General requirements

art-28

Article 28

Each manufacturer must establish and maintain document control procedures to ensure that all the documents referred to in this Resolution are correct and adequate for the intended use and are understood by everyone who may affect or influence the quality of a product.

art-29

Article 29

Each manufacturer must designate personnel to assess and approve all documents established in this Resolution for adequacy before their issuance.

art-29-sole

Sole paragraph

The approval referred to in the caption of this article, including the date and handwritten or digital signature of the person responsible for approving the documents, must be recorded.

art-30

Article 30

Each manufacturer must ensure that all documents are updated and available at the application points and that all unnecessary or obsolete documents are withdrawn from use or protected from unintended use.

art-31

Article 31

Alterations related to quality system specifications, methods, or procedures must be assessed, documented, reviewed, and approved by personnel whose job and level of responsibility are equivalent to those who performed the original review and approval.

art-32

Article 32

Each manufacturer must keep records of document alterations that must include: I – description of the alteration; II – identification of the altered documents; III – identification of the affected documents; IV – identification of the person responsible for the alteration; V – date when the alteration was approved; VI – date when the alteration becomes effective.

art-33

Article 33

A list of documents in force must be maintained to identify the documents’ current status and ensure that only updated and approved documents are in use.

art-34

Article 34

All quality documents and records must be legible and stored in a way to minimize damage, prevents loss, and allow quick recovery.

art-35

Article 35

All digital documents and records must have backup copies.

art-36

Article 36

Documents and records considered confidential by the manufacturer may be signalized to alert the competent health authority.

art-37

Article 37

All necessary documents and records related to a product must be kept for a period of time equivalent to the product’s lifespan, counting from the date of its distribution and in no situation this period can be shorter than two years.

Section II · Device history record

art-38

Article 38

Each manufacturer must keep device history records.

art-39

Article 39

Each manufacturer must establish and maintain procedures to ensure that the device history records are kept for each batch or series, in order to confirm that the products were manufactured in accordance with the device master record and the requirements in this Resolution.

art-40

Article 40

The device history records must include or refer to the following information: I – manufacturing date; II – components used; III – quantity manufactured; IV – results from inspections and tests; V – parameters of special processes; VI – quantity released for distribution; VII – labeling; VIII – identification of serial number or production batch; and IX – product final release.

Section III · Inspection and test records

art-41

Article 41

Each manufacturer must keep a record of the results from established inspections and tests, when these are directly related to product critical quality attributes.

art-42

Article 42

The records of established inspections and tests must include the acceptance criteria, results, equipment/ instrument used, and the date, and handwritten or digital signature of the person responsible for them.

Chapter IV — DESIGN CONTROL AND DEVICE MASTER RECORD (DMR)

Section I · Design control

art-43

Article 43

Each manufacturer must establish and maintain product design control procedures to ensure that the requirements specified for the design are complied with.

art-44

Article 44

Each manufacturer must establish and maintain plans that describe or refer to design and development activities, as well as the personnel responsible for each activity.

art-44-p1

§ 1

The plans referred to in the caption of this article must include any interaction between the various organizational and technical groups somehow involved in the design.

art-44-p2

§ 2

The plans referred to in the caption of this article must be assessed, updated, and approved during design development.

art-45

Article 45

Each manufacturer must establish and maintain procedures to ensure that product requirements are appropriate and meet its intended use, including user and patient needs, as well as the applicable legal and regulatory requirements.

art-45-sole

Sole paragraph

The procedures referred to in the caption of this article must include a mechanism that allows incomplete, ambiguous, or conflicting requirements to be identified and addressed.

art-46

Article 46

Design input data must be documented, assessed, and approved by a qualified person assigned for such task.

art-47

Article 47

The approval of design requirements, including the date and handwritten or digital signature by the person responsible for it, must be documented.

art-48

Article 48

Each manufacturer must establish and maintain procedures for product design verification.

art-48-p1

§ 1

The design verification must be performed by a designated person and must ensure that the design output data match the design input data.

art-48-p2

§ 2

The design verification results, including the identification of the design verified, the verification methods, the date, and the name of the person in charge of the verification, must be documented in the design history record.

art-49

Article 49

Each manufacturer must define and document the design output data in a way that allows the assessment of the design's conformity to the requirements established as input data.

art-49-p1

§ 1

The design output data must meet the input data requirements, include the acceptance criteria, and identify the design characteristics that are essential for the product intended use.

art-49-p2

§ 2

The design output data must be documented, reviewed, and approved before its release.

art-50

Article 50

Each manufacturer must establish and maintain procedures to ensure that the assessment of design results are planned, conducted, and documented during the various stages of design development.

art-50-sole

Sole paragraph

The procedures referred to in the caption of this article must ensure that representatives of all activities directly related to the design stage being reviewed are involved, as well as representatives from related areas and the specialists needed.

art-51

Article 51

The design review results must be documented in the design history record.

art-52

Article 52

Each manufacturer must establish and maintain procedures to ensure that product design is correctly translated into production specifications.

art-53

Article 53

Each manufacturer must establish and maintain a procedure to validate the product design.

art-54

Article 54

Design validation must be performed under predetermined operational conditions, during the initial manufacture of batches or units.

art-55

Article 55

Design validation must ensure that the product meets the user’s needs and the indication of use, and it must include trials of products under real or simulated use conditions.

art-56

Article 56

Design validation must include software validation, where appropriate.

art-57

Article 57

The project’s validation results, including identification, methods, date, and handwritten or digital signature of those responsible, must be documented in the design history record.

art-58

Article 58

Whenever applicable, stability studies must be performed during design validation.

art-59

Article 59

Each manufacturer must ensure that the design is released for production only after approval by the personnel designated by the manufacturer for such task.

art-59-p1

§ 1

The designated personnel, referred to in the caption of this article, must review all records required for the design history record, in order to ensure that it is complete and the final design is compatible with the approved plans, before its release.

art-59-p2

§ 2

The release referred to in the caption of this article must be documented, including the date and the handwritten or digital signature of the person responsible for it.

art-60

Article 60

Each manufacturer must establish and maintain procedures for the identification, documentation, validation, review, and approval of design alterations prior to their implementation, including a risk assessment as within the risk management process.

art-61

Article 61

Each manufacturer must establish and maintain a design history record for each product.

art-61-sole

Sole paragraph

The design history record must contain or refer to all records necessary to demonstrate that the design was developed in accordance with the design plan approved and the requirements in this Resolution.

Section II · Device Master Record (DMR)

art-62

Article 62

Each manufacturer must keep device master records (DMRs).

art-63

Article 63

The DMR for each type of product must include or refer to the following information: I – product specifications, including the respective designs, composition, formulation, component specifications, software design specifications, and their source codes; II – production process specifications, including infrastructure specifications, equipment, production methods and instructions, and environmental production specifications; III – packaging and labeling specifications, including methods and processes used; IV – inspection and testing procedures, with the respective acceptance criteria; and V – Installation, maintenance, and technical assistance methods and procedures.

Chapter V — PROCESS AND PRODUCTION CONTROLS

Section I · General requirements

art-64

Article 64

Each manufacturer must design, conduct, control, and monitor all production processes to ensure that the manufactured product complies with its specifications.

art-65

Article 65

Each manufacturer must establish and maintain process control procedures that describe the process controls necessary to ensure compliance with product specifications.

art-65-sole

Sole paragraph

Process controls must be established at any stage where deviation from product specifications may occur as a result of the manufacturing process.

art-66

Article 66

Process controls must include: I – documented instructions, standard operational procedures, and methods that define and control the form of production, installation, and maintenance; II – monitoring and control of process parameters; III – compliance with technical regulations, standards, or reference codes; and IV – instructions for approval of process start.

art-67

Article 67

The company's facilities must be properly designed in order to: I – ensure the adequate flow of people; II – allow the performance of all operations; and III – prevent exchange or contamination of components, manufacturing materials, intermediate products, and finished products, and ensure the correct handling of such materials.

art-68

Article 68

Each manufacturer must provide suitable environmental conditions for production operations, in a way to prevent contamination or other adverse effects on the product.

art-68-sole

Sole paragraph

For the purposes of the provisions in the caption of this article, the correct functioning of the environmental control systems established must be monitored, and the corresponding records must be kept.

art-69

Article 69

Each manufacturer must establish and maintain appropriate cleaning and sanitization procedures, as well as a schedule that meets the requirements in the manufacturing process specifications.

art-69-sole

Sole paragraph

Each manufacturer must ensure that the personnel involved in cleaning activities understand the cleaning and sanitization procedures.

art-70

Article 70

Each manufacturer must ensure that personnel who are in contact with the product or its environment are clean, healthy, and dressed appropriately for the activity to be performed.

art-71

Article 71

Any person that, after clinical exam or supervisor observation, might be considered in a health condition that could affect the product, must be suspended from manufacturing operations until his or her health condition is considered adequate.

art-71-sole

Sole paragraph

Personnel must be instructed to report to supervisors whenever they have a health condition that could affect the product.

art-72

Article 72

Each manufacturer must restrict food and beverage consumption to specific locations, so it does not affect production areas.

art-73

Article 73

Each manufacturer must establish and maintain procedures to prevent the contamination of equipment, components, manufacturing materials, intermediate products, and finished products with cleaning and disinfecting materials, including hazardous substances or contaminants generated by the manufacturing process.

art-74

Article 74

A pest control program must be established, and it must ensure that, whenever chemical agents are used, such agents do not affect the quality of the product.

art-75

Article 75

The treatment and disposal of waste, chemical effluents, and by-products must occur in accordance with the applicable legislation in force.

art-76

Article 76

Biological safety standards must be followed in cases where there is a biological risk.

art-77

Article 77

Each manufacturer must ensure compliance with the applicable standards related to workers' health, including the use of personal protective equipment compatible with the work processes performed.

art-78

Article 78

Each manufacturer must ensure that all equipment used in the manufacturing process is suitable for its intended use and correctly designed, built, and installed to facilitate maintenance, adjustment, cleaning, and use.

art-79

Article 79

Each manufacturer must establish and maintain a program for equipment maintenance, adjustment and, when necessary, cleaning, to ensure that all manufacturing specifications are complied with.

art-79-sole

Sole paragraph

The maintenance program must be easily accessible to the personnel responsible for the equipment maintenance and operation.

art-80

Article 80

Maintenance activities must be recorded, with the date of performance and the identification of the personnel responsible for the task.

art-81

Article 81

Each manufacturer must ensure that any acceptable tolerances or inherent limitations are posted in a visible location or near the equipment requiring periodic adjustments, or readily available to the personnel responsible for such adjustments.

art-82

Article 82

Each manufacturer must establish and maintain procedures for the use and removal of manufacturing materials, to ensure that such materials are removed from the product or limited to a specified amount that does not adversely affect the product’s quality.

art-83

Article 83

Special processes must be conducted in accordance with established procedures and parameters to ensure compliance with the specifications.

art-83-sole

Sole paragraph

The critical parameters of special processes must be monitored and recorded in the device history record.

Section II · Controls for packaging, labeling, and instructions for use

art-84

Article 84

Each manufacturer must establish procedures for packaging, in a way to protect the product from any alteration, damage, or contamination during the stages of processing, storage, handling, and distribution.

art-85

Article 85

Each manufacturer must establish and maintain procedures to ensure the integrity and prevent accidental mixing of labels, instructions for use, packaging materials, or identification tags.

art-86

Article 86

Each manufacturer must ensure that labels are designed, printed and, where applicable, applied so that they remain legible and adhered to the product during processing, storage, handling, and use.

art-87

Article 87

Labels and instructions for use must not be released for use until an authorized person has examined their compliance with the information they contain.

art-87-p1

§ 1

The approval of labels and instructions for use must be documented in the device history record, including the date, name and handwritten or digital signature of the person responsible for it.

art-87-p2

§ 2

In case of importers, the approval documentation referred to in Paragraph 1 of this article may be recorded in a specific document instead of the device history record.

Section III · Inspection and tests

art-88

Article 88

Each manufacturer must establish and maintain inspection and testing procedures, or other verification methods, to ensure product compliance with specified requirements throughout the manufacturing process.

art-89

Article 89

The conformity with the specified requirements must be assessed upon the receipt of components and manufacturing materials, as well as during intermediate stages of production and at the final approval of the finished product.

art-89-p1

§ 1

The results from the activities referred to in the caption of this article must be documented, including the conclusion – either approval or rejection.

art-89-p2

§ 2

The manufacturer must delegate the authority and responsibility for carrying out the activities referred to in the caption of this article.

art-90

Article 90

Components and manufacturing materials received, as well as components, intermediate products, and returned products, must not be used or processed until their compliance with the established requirements has been verified.

art-91

Article 91

Each manufacturer must establish and maintain procedures for the retention of components, manufacturing materials, intermediate products, and returned products, until inspections, tests, or other established verifications have been performed and documented.

art-92

Article 92

Finished products may only be released when the activities specified in the DMR have been completed and the associated documentation and data have been reviewed by a designated person to ensure that all acceptance criteria have been met.

art-92-sole

Sole paragraph

The release of finished products must be documented, including the date and handwritten or digital signature of the person responsible for the task.

Section IV · Measurement and testing equipment

art-93

Article 93

Each manufacturer must ensure that all measurement and testing equipment, including mechanical, automated, or electronic equipment, is adequate for its intended purpose and capable of producing valid results.

art-94

Article 94

Each manufacturer must establish and maintain procedures to ensure that measurement and testing equipment is routinely calibrated, inspected, and controlled.

art-95

Article 95

Each manufacturer must establish and maintain calibration procedures that include specific guidelines and precision and accuracy limits, as well as instructions for corrective actions when the precision and accuracy limits are not achieved.

art-96

Article 96

Calibration must be carried out by personnel who have the necessary education, training, practice, and experience.

art-97

Article 97

Measurement and testing equipment must be identified to allow the calibration status to be determined.

art-98

Article 98

Each manufacturer must establish and maintain calibration standards for measurement equipment that are traceable to official national or international standards.

art-98-sole

Sole paragraph

The manufacturer must establish and maintain its own calibration standard when there is no applicable calibration standard available.

art-99

Article 99

Each manufacturer must ensure that records are kept including calibration dates, the measurements obtained, name of the person responsible for the task, and the date scheduled for the next calibration.

art-99-p1

§ 1

The manufacturer must keep the records referred to in the caption of this article.

art-99-p2

§ 2

The records referred to in the caption of this article must be available to the personnel that uses the equipment and to those responsible for its calibration.

art-100

Article 100

Each manufacturer must establish and maintain procedures to ensure that handling, preservation, and storage of test, inspection, and measurement equipment are carried out in a manner that preserves its accuracy and adequacy for use.

art-101

Article 101

Each manufacturer must protect facilities and inspection, test, and measurement equipment, including testing hardware and software, from adjustments that could invalidate calibration.

art-102

Article 102

Each manufacturer must establish procedures to assess the impact of results from previous measurements, when non-conformities are found in measurement and testing equipment, and the result of such assessment must be documented.

Section V · Validation

art-103

Article 103

Special processes must be validated according to protocols previously established and the validation results, including the date and identification of the person responsible for its approval, must be recorded.

art-104

Article 104

Analytical methods, auxiliary systems to support the manufacturing process or environmental control, automated computerized systems, and software that may adversely affect product quality or the quality system must be validated.

art-105

Article 105

Each manufacturer must establish procedures to verify periodically its processes, analytical methods, auxiliary systems to support the manufacturing process or environmental control, automated computerized systems, and validated software and, when applicable, establish the frequency for revalidation.

art-106

Article 106

Each manufacturer must establish a procedure to control alterations in auxiliary systems, software, equipment, processes, methods, or other alterations that may influence the quality of products, including a risk assessment within the risk management process.

art-106-p1

§ 1

The procedure referred to in the caption of this article must describe the actions to be taken, including, when applicable, the need for requalification or revalidation.

art-106-p2

§ 2

The alterations mentioned in the caption of this article must be formally requested, documented, and approved before implementation.

Chapter VI — HANDLING, STORAGE, DISTRIBUTION, AND TRACEABILITY

Section I · Handling

art-107

Article 107

Each manufacturer must establish and maintain procedures to ensure that inversions (exchanges), damage, deterioration, or other adverse effects that affect components, manufacturing materials, intermediate products, finished products, and quality control samples do not occur during any stage of handling.

art-108

Article 108

Each manufacturer must establish and maintain procedures to identify the conformity of components, manufacturing materials, intermediate products, and finished products to ensure that only those duly approved are used or distributed.

art-109

Article 109

Article 107 and Article 108 of this Resolution must ensure that components, manufacturing materials, intermediate products, or finished products: I – are not used or distributed, when their quality or the “fit for use” status has been lost over time; II – closest to the expiration date are distributed or used first; and III – are not distributed or used when expired.

Section II · Storage and distribution

art-110

Article 110

Each manufacturer must establish and maintain procedures to identify components, manufacturing materials, intermediate products, finished products, and quality control samples, in a way to prevent inversions (exchanges) during storage.

art-111

Article 111

Components, manufacturing materials, intermediate products, finished products, and quality control samples must be stored in physical and environmental conditions that prevent damage, deterioration, or other adverse effects during the period in which they remain in storage.

art-112

Article 112

Each manufacturer must keep distribution records that include or refer to: I – the consignee’s name and address; II – the identification and quantity of products shipped, with the shipment date; and III – any numerical control used for traceability.

Section III · Identification, traceability, and non-conformities

art-113

Article 113

Each manufacturer must establish and maintain procedures to identify components, manufacturing materials, intermediate products, and finished products during all stages of storage, production, distribution, and installation, in order to avoid confusion and to ensure the correct fulfillment of orders.

art-114

Article 114

Each manufacturer must identify each product unit, lot, or batch with a serial or batch number, and such identification must be included in the device history record.

art-114-sole

Sole paragraph

In case of distributors, storers, and importers, the identification referred to in the caption of this article may be recorded in a specific document instead of the device history record.

art-115

Article 115

Each manufacturer must establish and maintain procedures to ensure that components, manufacturing materials, intermediate products, finished products, and returned products, which are not in conformity with the established requirements, are not used or installed inadvertently.

art-115-sole

Sole paragraph

The procedures referred to in the caption of this article must contain instructions for the identification, documentation, assessment, segregation, and disposal of non-conforming components, manufacturing materials, intermediate products, and finished products.

art-116

Article 116

The assessment of non-conforming components, manufacturing materials, intermediate products, and finished products must include the need for investigation and notification of the people and/ or organizations involved in the non-conformity.

art-116-sole

Sole paragraph

The assessment results and eventual investigations referred to in the caption of this article must be recorded.

art-117

Article 117

Responsibility for review and authority to dispose of non-conforming components, manufacturing materials, intermediate products, finished products, and returned products must be defined.

art-118

Article 118

The review and disposal process for non-conforming components, manufacturing materials, intermediate products, finished products, and returned products must be described in an established procedure.

art-118-p1

§ 1

The disposal of the products referred to in the caption of this article must be documented, and a record of the decision rationale and a handwritten or digital signature of the person or people responsible for the disposal must be kept.

art-118-p2

§ 2

In case of authorization to use the products referred to in the caption of this article, the decision must be based on a technically justifiable risk assessment.

art-119

Article 119

Each manufacturer must establish and maintain procedures for the rework, reinspection, and reassessment of intermediate or finished products after rework, in order to ensure the products meet their original specifications.

art-119-sole

Sole paragraph

Activities related to rework and reassessment of products referred to in the caption of this article, including issues related to rework activities, must be documented in the device history record.

Chapter VII — CORRECTIVE AND PREVENTIVE ACTIONS

Section I · General requirements

art-120

Article 120

Each manufacturer must establish and maintain procedures to: I – analyze processes, work operations, quality audit reports, quality records, technical assistance records, complaints, returned products, and other quality data sources, in order to identify existing and potential causes for non-conformities related to a product, process, or quality system; II – investigate the cause for non-conformities related to a product, process, or quality system; III – identify and execute the necessary actions to prevent the occurrence, correct the problem, and prevent the recurrence of non-conformities; IV – verify or validate the corrective action effectiveness and ensure that such corrective action does not adversely affect the product; V – record the activities related to corrective and preventive actions; VI – ensure that information about quality problems or non-conforming products is properly disseminated amongst those directly involved in maintaining product quality or preventing the occurrence of such problems; VII – submit relevant information about quality issues identified, as well as preventive and corrective actions to the executive management for awareness and follow-up, and also to the competent health authority, when applicable; and VIII – determine recall of products and other field actions that are relevant in case of products already distributed.

art-120-p1

§ 1

The analysis referred to in item I of this article must be based on a valid statistical technique to detect recurring quality problems, when applicable.

art-120-p2

§ 2

In order to comply with the provisions in item IV of this article, any alteration made must follow the alteration control procedures and validation protocols established, when applicable.

Section II · Management of complaints

art-121

Article 121

Each manufacturer must establish and maintain procedures to receive, examine, assess, investigate, and file complaints, ensuring that: I – complaints are received, documented, examined, assessed, investigated, and filed by a formally designated unit; II – complaints are notified to the competent health authority, when applicable; III – complaints are examined to verify the need for an investigation; IV – all complaints involving a possible non-conformity of a product are examined, assessed, and investigated; V – when an investigation is conducted, records are kept, including the following information: a) name of the product; b) date the complaint was received; c) any control number used; d) complainant’s name, address, and telephone number; e) nature of the complaint; and f) investigation date and results, including actions taken.

art-121-p1

§ 1

When the investigation referred to in item III of this article is not conducted, the unit must record the reason why the investigation was not carried out and the names of the people responsible for the decision not to investigate.

art-121-p2

§ 2

When any complaint referred to in item IV of this article is related to death, injury, or threat to public health, it must be immediately examined, assessed, and investigated.

Section III · Quality audit

art-122

Article 122

Each manufacturer must conduct and document quality audits to assess the quality system's compliance with the requirements established.

art-123

Article 123

Quality audits must be carried out by trained personnel, in accordance with the audit procedures established, and who have no direct responsibility for the matters being audited.

art-123-sole

Sole paragraph

Those responsible for conducting the quality audit cannot have direct responsibility for the matters being audited.

art-124

Article 124

Those responsible for the audited areas must be notified about non-conformities identified.

Chapter VIII — INSTALLATION AND TECHNICAL ASSISTANCE

art-125

Article 125

Each manufacturer must establish and maintain adequate instructions and procedures for the correct installation of products.

art-126

Article 126

During the product installation, either by the manufacturer or its authorized representative, there must be a verification whether the product’s operation meets the criteria established.

art-126-sole

Sole paragraph

The results from the verification referred to in the caption of this article must be recorded.

art-127

Article 127

Each manufacturer must ensure that installation instructions and procedures are distributed with the product or are otherwise available to the person responsible for the product installation.

art-128

Article 128

Each manufacturer must establish and maintain procedures to ensure that finished products submitted to technical assistance by the manufacturer or his representative meet the specifications.

art-129

Article 129

Each manufacturer must establish and maintain procedures to ensure that technical assistance records are maintained and include the following information: I – the product submitted to technical assistance; II – the control number used; III – the technical assistance date; IV – the technical assistance provider´s identification; V – the description of the service performed; and VI – the results from inspections and tests to approve the service.

art-130

Article 130

Each manufacturer must assess the technical assistance records periodically.

art-130-sole

Sole paragraph

In the cases where the assessment referred to in the caption of this article identifies failure trends, which represent danger, or records involving death or serious injury, a corrective/ preventive action must be initiated in accordance with the requirements in this Resolution.

Chapter IX — STATISTICAL TECHNIQUES

art-131

Article 131

Each manufacturer must establish and maintain procedures to identify valid statistical techniques to verify the quality system performance and the process ability to meet the specifications established.

art-132

Article 132

Sampling plans must be recorded in written and based on valid statistical logics.

art-133

Article 133

Each manufacturer must establish and maintain procedures to ensure that sampling methods are adequate for their intended use and that they are regularly reviewed.

art-134

Article 134

The review of sampling plans must consider the occurrence of product nonconformities, quality audit reports, complaints, and other indicators.

Chapter X — FINAL PROVISIONS

art-135

Article 135

The documentation that proves compliance with the requirements established in this Resolution must be available whenever requested by health surveillance authorities.

art-136

Article 136

Failure to comply with the provisions contained in this Resolution shall be considered an infraction of health regulations, pursuant to Law No. 6,437 of 20 August 1977 and its updates, without prejudice to the applicable civil, administrative, and criminal liabilities.

art-137

Article 137

The following regulations are hereby revoked: I – Collegiate Board Resolution – RDC No. 16 of 28 March 2013, published in the Federal Official Gazette No. 61 of 1 April 2013, Section 1, page 75; and II – Normative Instruction – IN No. 8 of 26 December 2013, published in the Federal Official Gazette No. 252 of 30 December 2013, Section 1, page 758.

art-138

Article 138

This Resolution enters into force on 2 May 2022. ANTONIO BARRA TORRES This content does not replace the one published in the certified version.