Medical Device Coordination Group Document MDCG 2018-1 Rev.4
Each of the abovementioned certificates shall identify and cover all devices
associated with the same Basic UDI-DI, that is referred to in that certificate.
The association between different Basic UDI-DIs, where applicable, shall be
identified through the technical dossiers.
In accordance with Annex IV of the two Regulations, the declaration of conformity
shall contain the Basic UDI-DI and the product and trade name, product code,
catalogue number or other unambiguous reference allowing identification and
traceability of the device covered by the EU declaration of conformity.
Changes of UDI-DI
A new UDI—DI shall be required whenever there is a change that could lead to
misidentification of the device and/or ambiguity in its traceability. In particular, a new
UDI-DI shall be required in the case of any change of the following elements: name
or trade name, device version or model, labelled as single use, packaged sterile,
need for sterilization before use, quantity of devices provided in a package, critical
warnings or contra-indications (e.g. containing latex or DEHP), CMR/Endocrine
disruptors1 2.
A UDI-DI shall be associated with one and only one Basic UDI-DI.
Regarding changes to the specific data elements listed below3, the following
considerations should be noted:
1. Is the Device Directly Marked (Yes/No)
The database design will force the creation of a new UDI-DI only if there is a change
from Yes to No and not vice versa for this data element.
2. Manner in which production of the device Is controlled (expiry date or
manufacturing date, lot number, serial number, software identification) - Type
of UDI-PI
As long as there is no change to the label, a change to this data element will not
require the allocation of a new UDI-DI.
1
It should be noted that a new regulatory decision classifying an existing product as CMR/Endocrine
disruptor might not result in a new UDI-DI for products already containing that substance. The decision on
whether to assign a new UDI-DI should be based on the conformity assessment of the product with regard
to the impact of the information provided and the significance of the change.
2
Specific attention shall be paid to the fact that changes of colour or language might also require a new UDI-
DI when those changes might lead to misidentification of product or change the product
safety/performance. For example:
A- Change of colour coding of e.g. connectors, latex-free surgical gloves, blood tubes
B- Two identical self-testing devices, that exist in parallel and cannot be substituted due to local labelling
requirements (IVD Article 10(10) of Regulation 746/2017), requires different UDI-DIs
Specifications of EU designated issuing entity should be used as a reference source to identify other
possible examples.
3
Please refer to Guidance UDIWG 2018-1 on "UDI database. Definitions, descriptions and formats of the UDI
core elements’ available at https://ec.europa.eu/docsroom/documents/28669