Medical Device Coordination Group Document MDCG 2020-3 Rev.1
MDCG 2020-3 revision 1 changes
Adjustments all over the document to align it to Regulation (EU) 2023/607 and
guidance MDCG 2022-2
1 Introduction
Article 120(3) of the Medical Device Regulation (EU) 2017/745 (MDR), last amended by
Regulation (EU) 2023/607, states that devices which continue to comply with the AIMDD or MDD
may be placed on the market or put into service until 31 December 20271 or 31 December 20282,
as applicable, provided the conditions set out in Article 120(3c) MDR are fulfilled.
These devices are called ‘legacy devices’ and in line with MDCG Guidance Document 2021-253,
‘legacy devices’ should be understood as devices, which, in accordance with the MDR’s
transitional provisions, are placed on the market after the MDR’s date of application (i.e. 26 May
2021) if certain conditions are fulfilled. Those devices can be:
devices which are class I devices under Directive 93/42/EEC (MDD), for which an EC
declaration of conformity was drawn up prior to 26 May 2021 and for which the conformity
assessment procedure under the MDR requires the involvement of a notified body;
devices covered by a valid EC certificate issued in accordance with Directive 90/385/EEC
(AIMDD) or the MDD prior to 26 May 2021.4
The conditions are set out in Article 120(3c) MDR and include, among others, that legacy
devices must continue to comply with the AIMDD/MDD5, as applicable, and that there are no
significant changes in the design or intended purpose of the device6. Therefore, it is important for
manufacturers and notified bodies to have a clear understanding as to what changes to design or
intended purpose would be considered ‘significant’.
1 In case of all class III devices and class IIb implantable devices except sutures, staples, dental fillings, dental braces,
tooth crowns, screws, wedges, plates, wires, pins, clips and connectors.
2 In case of class IIb devices other than those referred to in footnote n.1, class IIa devices, and class I devices placed
on the market in sterile condition or having a measuring function.
3 MDCG 2021-25: Regulation (EU) 2017/745 - application of MDR requirements to ‘legacy devices’ and to devices
placed on the market prior to 26 May 2021 in accordance with Directives 90/385/EEC or 93/42/EEC.
https://health.ec.europa.eu/system/files/2021-10/md_mdcg_2021_25_en_0.pdf
4 See question 1 in Q&A on practical aspects related to the implementation of Regulation (EU) 2023/607,
https://health.ec.europa.eu/document/download/592008f6-3456-4afb-a13a-
733a87da1b00_en?filename=mdr_proposal_extension-q-n-a_0.pdf
5 See Article 120(3c), point (a) MDR; according to Article 120(3d) MDR, the requirements relating to post-market
surveillance, market surveillance, vigilance, registration of economic operators and of devices shall apply to the
devices and replace the corresponding requirements of the Directives.
6 See Article 120(3c), point (b) MDR.