Medical Device Coordination Group Document MDCG 2022-6
1 Introduction
Article 110(3) of Regulation (EU) 2017/746 on in vitro diagnostic medical devices (IVDR), as
amended by Regulation (EU) 2022/1121, states that under certain conditions the following
devices may be placed on the market or put into service after the date of application of the
IVDR, i.e. 26 May 2022, until the end of the different transition periods specified in Article
110(3) IVDR:
devices which have a valid certificate issued by a notified body under the Directive
98/79/EC on in vitro diagnostic medical devices (IVDD), and
devices for which a declaration of conformity was drawn up prior to 26 May 2022 in
accordance with the IVDD and for which the conformity assessment procedure
pursuant to the IVDR (contrary to the IVDD) requires the involvement of a notified
body.
In line with Q&A 9 of the CAMD’s FAQ – IVDR Transitional provisions2, the certificates
covered by Article 110(3) of the IVDR include all certificates which are commonly issued by
notified bodies3 with reference to the IVDD.
The conditions for the application of the transitional provisions in Article 110(3) IVDR are that
the devices continue to comply with the IVDD4 and that there are no significant changes in
the design or intended purpose of the device after the date of application of the IVDR.
Therefore, it is important for manufacturers and notified bodies to have a clear understanding
as to what changes to design or intended purpose would be considered ‘significant’ under
Article 110(3) IVDR.
To the extent that devices subject to the transitional provisions in Article 110(3) IVDR are
covered by certificates issued by a notified body in accordance with the IVDD, it is essential
for IVDs to be placed on the market that those certificates remain valid5 following changes
that are not significant with regard to design or intended purpose and that the required
appropriate surveillance is carried out by the notified body that issued the certificate.
Manufacturer and respective notified body should agree on the latter on a contractual basis.
This guidance document is based on MDCG 2020-3 Guidance on significant changes
regarding the transitional provision under Article 120 of the MDR with regard to devices
covered by certificates according to MDD or AIMDD. Where appropriate, this guidance
1 OJ L 19, 28.1.2022 p.3.
2 CAMD Transition Sub Group, FAQ – IVDR Transitional provisions, V1.0, 17.1.2018.
3 According to Article 110(1) IVDR, from 26 May 2022 any publication of a notification in respect of a notified
body in accordance with Directive 98/79/EC becomes void. Irrespective of this, the term ‘notified body’ is used
throughout this document also for notified bodies previously designated under Directive 98/79/EC.
4 However, the requirements of IVDR relating to post-market surveillance, market surveillance, vigilance,
registration of economic operators and of devices shall apply to the devices and replace the corresponding
requirements in the Directive.
5 This does not exclude the possibility that during the transition period an EC certificate for the manufacturer’s
approved quality system issued in accordance with Annex IV or Annex VII IVDD, which has become invalid, is
replaced by a EU QMS certificate issued in accordance with Annex IX, chapter 1, IVDR, provided that the EC
design examination certificate issued under Annex IV, section 4, IVDD or EC type-examination certificate
issued under Annex V IVDD remains valid.