4. Definitions
Active Medical Device Any medical device, operation of which depends on a source of electrical energy or any source of power other than that directly generated by the human body or gravity and which acts by converting this energy. Medical devices intended to transmit energy, substances or other elements between an active medical device and the patient, without any significant change, are not considered to be active medical devices. Standalone software is considered to be an active medical device. (GHTF/SG1/N77:2012) Adaptable Medical Device A medical device that meets the following requirements: • it is mass-produced; and • it is adapted, adjusted, assembled, or shaped at the point of care, in accordance with the manufacturer's validated instructions, to suit an individual patient's specific anatomo- physiologic features prior to use. (IMDRF/PMD WG/N49 FINAL: 2018). Authorized Healthcare Professional An authorized healthcare professional is a person legally entitled to provide health services in the applicable jurisdiction. (IMDRF/ PMD WG/N58 FINAL: 2023) Clinical Data Safety, clinical performance and/or effectiveness information that is generated from the clinical use of a medical device. (IMDRF MDCE WG/N56 FINAL:2019) Clinical Evaluation A set of ongoing activities that use scientifically sound methods for the assessment and analysis of clinical data to verify the safety, clinical performance and/or effectiveness of the device when used as intended by the manufacturer. (IMDRF MDCE WG/N56 FINAL:2019) Clinical Evidence The clinical data and its evaluation pertaining to a medical device. (IMDRF MDCE WG/N56 FINAL:2019) Clinical Investigation Any systematic investigation or study in or on one or more human subjects, undertaken to assess the safety, clinical performance and/or effectiveness of a medical device. (IMDRF MDCE WG/N56 FINAL:2019) Clinical Performance The ability of a medical device to achieve its intended clinical purpose as claimed by the manufacturer. (IMDRF MDCE WG/N56 FINAL:2019) Comparable Device A medical device with related function chosen by the manufacturer to inform the clinical evaluation of the device in question. (IMDRF MDCE WG/N56 FINAL:2019) Conformity Assessment The systematic examination of evidence generated and procedures undertaken by the manufacturer, under requirements established by the Regulatory Authority, to determine that a medical device is safe and performs as intended by the manufacturer and, therefore, conforms to the Essential Principles of Safety and Performance for Medical Devices. (GHTF/SG1/N78:2012) Conformity Assessment Body (CAB) A body, other than a Regulatory Authority, engaged in determining whether the relevant requirements in technical regulations or standards are fulfilled. (GHTF/SG1/N78:2012) Custom-made Medical Device A medical device that, at a minimum, meets the following requirements: • it is intended for the sole use of a particular individual (which could be a patient or healthcare professional); and • it is specifically made in accordance with a written request of an authorized professional, which gives, under their responsibility, specific design characteristics; even though the design may be developed in consultation with a manufacturer; and • it is intended to address the specific anatomo-physiological features or pathological condition of the individual for whom it is intended. NOTE 1: Medical devices that are patient-matched, adaptable, or mass-produced shall not be custom- made. NOTE 2: A custom-made device is intended for a case where an individual's specific needs cannot be met or cannot be met at the appropriate level of performance, by an alternative device available on the market. (IMDRF/PMD WG/N49 FINAL: 2018) Direct Clinical Evidence For the purposes of this document, direct clinical evidence is defined as evidence derived from an evaluation of clinical data pertaining to the subject device. Effectiveness The ability of a medical device to achieve clinically meaningful outcome(s) in its intended use as claimed by the manufacturer. (IMDRF MDCE WG/N56 FINAL:2019) Expected Lifetime/Expected Service Life Time-period specified by the manufacturer during which the medical device or IVD medical device is expected to maintain safe and effective use. NOTE 1: The expected lifetime can be determined by stability or by other methods. NOTE 2: Maintenance, repairs, or upgrades (e.g., safety or cybersecurity modifications) can be necessary during the expected lifetime. (IMDRF/GRRP WG/N52) Harm Physical injury or damage to the health of people or damage to property or the environment. (GHTF/SG1/N77:2012) Hazard Potential source of harm. (GHTF/SG1/N77:2012) Implantable Device Any device, including those that are partially or wholly absorbed, which is intended: • to be totally introduced into the human body or, • to replace an epithelial surface or the surface of the eye, by surgical intervention which is intended to remain in place after the procedure. Any device intended to be partially introduced into the human body through surgical intervention and intended to remain in place after the procedure for at least 30 days is also considered an implantable device. (GHTF/SG1/N77:2012) In Vitro Diagnostic (IVD) Medical Device Means a medical device, whether used alone or in combination, intended by the manufacturer for the in-vitro examination of specimens derived from the human body solely or principally to provide information for diagnostic, monitoring or compatibility purposes. NOTE 1: IVD medical devices include reagents, calibrators, control materials, specimen receptacles, software, and related instruments or apparatus or other articles and are used, for example, for the following test purposes: diagnosis, aid to diagnosis, screening, monitoring, predisposition, prognosis, prediction, determination of physiological status. NOTE 2: In some jurisdictions, certain IVD medical devices may be covered by other regulations. (GHTF/SG1/N071:2012) Indications for Use A general description of the disease or condition the medical device or IVD medical device will diagnose, treat, prevent, cure, or mitigate, including a description of the patient population for which the medical device or IVD medical device is intended. (IMDRF/GRRP WG/N52) Instructions for Use Information provided by the manufacturer to inform the device user of the medical device’s intended purpose and proper use and of any precautions to be taken. (GHTF/SG1/N70:2011) Intended Use/ Purpose The objective intent regarding the use of a product, process or service as reflected in the specifications, instructions and information provided by the manufacturer. NOTE 1: The intended use/intended purpose are also part of promotional or sales materials or statements, although these materials lie outside the scope of this document. NOTE 2: The intended use can include the indications for use. (IMDRF/GRRP WG/N52) Kits Kits are a collection of products, including medical devices, that are packaged together to achieve a common intended use and is being distributed as a medical device. These could also be called procedure packs or convenience kits. NOTE: Jurisdictions may differ in their definition of kit. (IMDRF/UDI WG/N7FINAL:2013) Label Written, printed, or graphic information either appearing on the medical device itself, or on the packaging of each unit, or on the packaging of multiple devices. NOTE: The definition above refers to the human readable label. (GHTF/SG1/N70:2011) Labelling The label, instructions for use, and any other information that is related to identification, technical description, intended purpose and proper use of the medical device, but excluding shipping documents. (GHTF/SG1/N70:2011) Life-cycle All phases in the life of a medical device, from the initial conception to final decommissioning and disposal. (GHTF/AHWG-GRM/N1R13:2011) Manufacturer Means any natural or legal person with responsibility for design and/or manufacture of a medical device with the intention of making the medical device available for use, under his name; whether or not such a medical device is designed and/or manufactured by that person himself or on his behalf by another person(s). Notes: 1. This 'natural or legal person' has ultimate legal responsibility for ensuring compliance with all applicable regulatory requirements for the medical devices in the countries or jurisdictions where it is intended to be made available or sold, unless this responsibility is specifically imposed on another person by the Regulatory Authority (RA) within that jurisdiction. 2. The manufacturer's responsibilities are described in other GHTF guidance documents. These responsibilities include meeting both pre-market requirements and post-market requirements, such as adverse event reporting and notification of corrective actions. 3. 'Design and/or manufacture', as referred to in the above definition, may include specification development, production, fabrication, assembly, processing, packaging, repackaging, labelling, relabelling, sterilization, installation, or remanufacturing of a medical device; or putting a collection of devices, and possibly other products, together for a medical purpose. 4. Any person who assembles or adapts a medical device that has already been supplied by another person for an individual patient, in accordance with the instructions for use, is not the manufacturer, provided the assembly or adaptation does not change the intended use of the medical device. 5. Any person who changes the intended use of, or modifies, a medical device without acting on behalf of the original manufacturer and who makes it available for use under his own name, should be considered the manufacturer of the modified medical device. 6. An authorized representative, distributor or importer who only adds its own address and contact details to the medical device or the packaging, without covering or changing the existing labelling, is not considered a manufacturer. 7. To the extent that an accessory is subject to the regulatory requirements of a medical device, the person responsible for the design and/or manufacture of that accessory is considered to be a manufacturer. (GHTF/SG1/N055:2009) Medical Device Production System (MDPS) A medical device production system (MDPS) is a combination of the resultant medical device and the medical device production process (MDPP) elements. The elements of an MDPP includes the raw materials, software and digital files, main production and post-processing (if applicable) equipment, and operating instructions intended to be used by specific end users at a healthcare facility (HCF), to produce a specific type of medical device for treating the patients of the HCF. • An MDPS includes the resultant medical device it is intended to produce and the intended use for the device is validated in accordance with safety and performance requirements in the relevant regulatory jurisdiction. • An MDPS classification should be determined by the risk-based classification of the resultant medical device it is intended to produce, which may include consideration of any additional or likely foreseeable risks that may arise as a result of the operation of the MDPS. • An MDPS may require the use of ancillary equipment, human factors considerations, technical capability requirements, or other specified input and design limit controls; however, all components must be validated as a production process to consistently produce the resultant medical device with the use of the supplied operating instructions. (IMDRF/ PMD WG/ N58 FINAL: 2023) Patient-matched Medical Device A medical device that meets the following requirements: • it is matched to a patient's anatomy within a specified design envelope using techniques such as scaling of the device based on anatomic references, or by using the full anatomic features from patient imaging; and • it is typically produced in a batch through a process that is capable of being validated and reproduced; and • it is designed and produced under the responsibility of a manufacturer even though the design may be developed in consultation with an authorized healthcare professional. Note 1: A written request from an authorized healthcare professional may be present; but is not mandatory. Note 2: The number and type of design inputs in consultation with a healthcare professional may vary depending on the medical devices to be manufactured. Note 3: The design must remain within the validated parameters of the specified design envelope. (IMDRF/PMD WG/N49 FINAL: 2018) Performance The ability of a medical device to achieve its intended purpose as stated by the manufacturer. Performance may include both clinical and technical aspects. (IMDRF GRRP WG/N47 FINAL: 2018) Personalized Medical Device (PMD) A generic term to describe any of the types of medical devices that are intended for a particular individual, which could be either a custom-made, patient-matched, or adaptable medical device. (IMDRF/PMD WG/N49 FINAL: 2018) Software used as part of production rather than software that meets the definition of a medical device in its own right. Post-market clinical follow-up study A study carried out following marketing authorization intended to answer specific questions (uncertainties) relating to safety, clinical performance and/or effectiveness of a device when used in accordance with its labelling. (IMDRF MDCE WG/N65FINAL:2021) Process Validation Establishing by objective evidence that a process consistently produces a result or product meeting its predetermined requirements. (GHTF/SG3/N99-10:2004 (Edition 2)) Quality Management System Management system to direct and control an organization with regard to quality. (GHTF/SG3/N19:2012) Regulatory Authority (RA) A government body or other entity that exercises a legal right to control the use or sale of medical devices within its jurisdiction, and may take enforcement action to ensure that medical products marketed within its jurisdiction comply with legal requirements. (GHTF/SG1/N78:2012) Residual Risk Risk remaining after protective measures have been taken. (GHTF/SG3/N15R8:2005) Risk Combination of the probability of occurrence of harm and the severity of that harm. (GHTF/SG1/N77:2012) Risk Analysis Systematic use of available information to identify hazards and to estimate the risk. (GHTF/SG3/N15R8:2005) Risk Assessment Overall process comprising a risk analysis and a risk evaluation. (GHTF/SG3/N15R8:2005) Risk Control Process through which decisions are reached and protective measures are implemented for reducing risks to, or maintaining risks within, specified levels. (GHTF/SG3/N15R8:2005) Risk Evaluation Judgment, on the basis of risk analysis, of whether a risk which is acceptable has been achieved in a given context based on the current values of society. (GHTF/SG3/N15R8:2005) Risk Management The systematic application of management policies, procedures and practices to the tasks of analyzing, evaluating, controlling and monitoring risk. (GHTF/SG3/N15R8:2005) Safety Acceptability of risks as weighed against benefits, when using the medical device according to the manufacturer's labelling. (IMDRF MDCE WG/N56FINAL:2019) Software as a Medical Device (SaMD) The term "Software as a Medical Device" (SaMD) is defined as software intended to be used for one or more medical purposes that perform these purposes without being part of a hardware medical device. NOTES: • SaMD is a medical device and includes in-vitro diagnostic (IVD) medical device. • SaMD is capable of running on general purpose (non-medical purpose) computing platforms • "without being part of" means software not necessary for a hardware medical device to achieve its intended medical purpose; • Software does not meet the definition of SaMD if its intended purpose is to drive a hardware medical device. • SaMD may be used in combination (e.g., as a module) with other products including medical devices; • SaMD may be interfaced with other medical devices, including hardware medical devices and other SaMD software, as well as general purpose software • Mobile apps that meet the definition above are considered SaMD. (IMDRF/SaMD WG/N10 FINAL:2013) Specified Design Envelope Minimum and maximum dimensions, mechanical performance limits, and other relevant factors that characterize a medical device for production purposes, which may be based on a standard device template model. (IMDRF/PMD WG/N49 FINAL: 2018) Technical Documentation The documented evidence, normally an output of the quality management system that demonstrates conformity of a device to the Essential Principles of Safety and Performance of Medical Devices. (GHTF/SG1/N78:2012) Unique Device Identification (UDI) The UDI is a series of numeric or alphanumeric characters that is created through a globally accepted device identification and coding standard. It allows the unambiguous identification of a specific medical device on the market. The UDI is comprised of the UDI-DI (device identifier) and UDI-PI (production identifier). NOTE: The word "Unique" does not imply serialization of individual production units. (IMDRF/UDI WG/N7FINAL:2013) User The person, either professional or lay, who uses a medical device. The patient may be the user. (GHTF/SG1/N70:2011) “Computing platforms” include hardware and software resources (e.g. operating system, processing hardware, storage, software libraries, displays, input devices, programming languages etc.). “Operating systems” that SaMD require may be run on a server, a workstation, a mobile platform, or other general purpose hardware platform. Validation Confirmation through provision of objective evidence that the requirements for a specific intended use or application have been fulfilled. NOTE 1: The term "validated" is used to designate the corresponding status. NOTE 2: The use conditions for validation can be real or simulated. (GHTF/SG3/N18:2010) Verification Confirmation through provision of objective evidence that specified requirements have been fulfilled. NOTE 1: The term "verified" is used to designate the corresponding status. NOTE 2: Confirmation can comprise activities such as: • performing alternative calculations, • comparing a new design specification with a similar proven design specification, undertaking tests, performing demonstrations, and reviewing and approving documents prior to issue. (GHTF/SG3/N18:2010)