Searchable PDF text layer
Addendum to FAQ on Medical Devices Rules, 2017
1. Whether the manufacturer or importer need to conduct the Biocompatibility
test on the all applied products?
➢ The biocompatibility is required for the devices which comes in direct or
indirect contact of human body. The manufacturer shall generate the
Biocompatibility data on the applied device as per applicable standard.
These tests need to be conducted by the manufacturer as a part of the
Material validation initially and shall revalidate if there are any changes to
the manufacturer of the material or change in the process or supplier.
The Biocompatibility testing of medical devices need to be viewed
rationally and shall be tested as per IS/ISO 10993 series of standards.
➢ However, if the manufacturer is already carried out the Biocompatibility
test on the particular category of the devices and subsequently applied
for approval of the similar devices which is being manufactured by using
the same manufacturing process, same supplier / vendor of raw materials
of the particular devices, in such cases the manufacturer may submit the
initial biocompatibility data generated on the particular category of the
devices along with the justification & biological risk assessment reports
of the applied devices for consideration.
➢ Further, if the material is used for a product with different invasiveness,
the additional tests which need to be conducted as per the requirements
listed in IS/ISO10993-part 1 standard need to be conducted and validated.
➢ However, the manufacturer shall perform a fore cause testing based on
adverse event reported to the them or based on trending result of the
complaints received by them.
2. Whether the manufacturer can release the sterile medical device to the market
based on parametric release?
➢ As per Clause 7.5.2.2 of Fifth Schedule of Medical Devices Rules, 2017
which prescribes the requirements of Quality Management System for
manufacture, storage, installation and distribution of Medical Devices,
the manufacturer shall establish documented procedures for the
Page 2 of 7
Medical Devices Division, Central Drugs Standard Control Organization,
Ministry of Health and Family Welfare, Govt. of India