16. Whether any product used in determining the presence of histamine, substances ,
Microbial detection in food & food products, animal feeds, liquor (wine, beer),
environmental samples like water, soil etc. and is not intended for In-vitro diagnostic
use in humans or animals, are being regulated under the provision of Medical Devices
Rules, 2017?
Ans: No.
17. Will products such as RUO – Research Use Only, Q.C material for accreditation,
panel for Q.C testing & product used for food, water, sterility testing used by various
industry for Q.C etc., and is not intended to be used in human or animals for
diagnosis of any disease or disorder, be regulated under MDR, 2017 ?
Ans: No
18. Will products such as microbiological culture media, stains indicators and reagents
used for food and water testing and is not intended to be used in human or animals
for diagnosis of any disease or disorder be regulated under MDR, 2017 ?
Ans: No
19. Whether Specimens collection tubes (vacuum type or not) without needle used for the
collection of Blood, Urine, Stool, Sputum, Semen, etc., for purpose of specimens
collection are being regulated under the provision of Medical Devices Rules, 2017?
Ans: Yes. Specimen collection tubes are regulated under the definition of ―specimen receptacle‖
as specified in Sub-clause (zs) of Rule 3 of MDR-2017 and are classified as Class A as per First
Schedule, Part II (2(v)(3)) of MDR-2017.
20. Whether IVDs for HBsAg, HIV and HCV approved to manufacture or import by the CLA
or SLA, as the case may be, permitted to use for both the purposes; for blood
screening and diagnostic.
Ans: Yes. In – Vitro Diagnostic devices for HBsAg, HIV and HCV manufactured / Imported
under valid license issued by the CLA or SLA, may also be used in Blood Bank, as the
criteria like Sensitivity (%) and Specificity (%) for evaluation of the HBsAg, HIV and HCV
diagnostic kits for the Transfusion purpose (Blood Banks) and Diagnostic purpose are
same, Provided the manufacturer claims in the product labels or in the IFU that the product
is intended both the purposes; for blood screening and diagnostic.
21. Which IVD reagents/kits are prohibited in India?
Ans: (1) Serodiagnostic test kits for diagnosis of tuberculosis are prohibited to Import,
Manufacture, Sale, Distribution and Use in the country under Section 10A and Section 26A
of the Drugs and Cosmetics Act, 1940 Gazette notification(s) GSR432(E) & GSR433(E)
dated June 7, 2012.
(2) Antibody Detecting Rapid Diagnostic Tests for routine diagnosis of malaria are
prohibited to Import, Manufacture, Sale, Distribution and Use in the country under Section
26A and Section 10A of the Drugs and Cosmetics Act, 1940, Gazette notification(s)
GSR1352(E) dated March 23, 2018 and GSR1074(E) dated October 30, 2018 respectively.