3.0.
3.1 Active Medical Device: Any medical device, operation of which depends on a source of electrical energy or any source of power other than that directly generated by the human body or gravity and which acts by converting this energy. Medical devices intended to transmit energy, substances or other elements between an active medical device and the patient, without any significant change, are not considered to be active medical devices. Standalone software is considered to be an active medical device. (GHTF/SG1/N77:2012) 3.2 Analytical Performance of an IVD Medical Device: The ability of an IVD medical device to detect or measure a particular analyte. (GHTF/SG5/N6:2012) 3.3 Appropriately Reduce [Risks]: The reduction of risk to an acceptable level as determined by the manufacturer and regulatory authority (reducing risk as low as reasonably practicable, reducing risk as low as reasonably achievable, or reducing risk as far as possible) without adversely affecting the benefit-risk ratio. 3.4 Conformity Assessment Body (CAB): A body other than a Regulatory Authority engaged in determining whether the relevant requirements in technical regulations or standards are fulfilled. (IMDRF/GRRP WG/N040:2017) 3.5 Clinical Data: Safety and/or performance information that are generated from the clinical use of a medical device. (GHTF/SG5/N1R8:2007) 3.6 Clinical Evaluation: The assessment and analysis of clinical data pertaining to a medical device to verify the clinical safety and performance of the device when used as intended by the manufacturer. (GHTF/SG5/N1R8:2007) 3.7 Clinical Evidence: The clinical data and the clinical evaluation report pertaining to a medical device. (GHTF/SG5/N1R8:2007) 3.8 Clinical Evidence for an IVD Medical Device: All the information that supports the scientific validity and performance for its use as intended by the manufacturer. (GHTF/SG5/N6:2012) 3.9 Clinical Investigation: Any systematic investigation or study in or on one or more human subjects, undertaken to assess the safety and/or performance of a medical device. Explanation: This term is synonymous with ‘clinical trial’ and ‘clinical study’. (GHTF/ IMDRF GRRP WG/N47 FINAL: 2018 SG5/N1R8) 3.10 Clinical Performance: The ability of a medical device to achieve clinical outcome(s) in its intended purpose as claimed by the manufacturer. (Modified from GHTF/SG5/N1R8:2007) 3.11 Clinical Performance of an IVD Medical Device: The ability of an IVD medical device to yield results that are correlated with a particular clinical condition/physiological state in accordance with target population and intended user. (Modified from GHTF/SG5/N6:2012) NOTE 1: Clinical performance can include diagnostic sensitivity and diagnostic specificity based on the known clinical/physiological state of the individual, and negative and positive predictive values based on the prevalence of the disease. 3.12 Effective: The ability of a medical device or IVD medical device to provide clinically significant results in a significant portion of the target population. NOTE: This ability is assessed in situations where the medical device or IVD medical device is used for its intended uses and conditions of use and accompanied by adequate directions for use and warnings against unsafe use. 3.13 Expected Lifetime/Expected Service Life: Time period specified by the manufacturer during which the medical device or IVD medical device is expected to maintain safe and effective use. NOTE 1: The expected lifetime can be determined by stability. NOTE 2: Maintenance, repairs, or upgrades (e.g. safety or cybersecurity modifications) can be necessary during the expected lifetime. 3.14 Expiry Date/Expiration Date: Upper limit of the time interval during which the safety and performance characteristics of a material stored under specified conditions can be assured. NOTE 1: This also applies to medical devices whose physical, chemical or functional properties are maintained during a specified and known period, such as for capital equipment. NOTE 2: Expiry dates are assigned to IVD reagents, calibrators, control materials and other components by the manufacturer, based on experimentally determined stability properties. (Modified from ISO 18113-1:2009) 3.15 Harm: Injury or damage to the health of people, or damage to property or the environment. (ISO/IEC Guide 51:2014) 3.16 Hazard: Potential source of harm. (ISO/IEC Guide 51:2014) 3.17 Indications for Use: A general description of the disease or condition the medical device or IVD medical device will diagnose, treat, prevent, cure, or mitigate, including a description of the patient population for which the medical device or IVD medical device is intended. IMDRF GRRP WG/N47 FINAL: 2018 3.18 Intended Use / Intended Purpose: The objective intent regarding the use of a product, process or service as reflected in the specifications, instructions and information provided by the manufacturer. (Modified from GHTF/SG1/N77:2012) NOTE: The intended use can include the indications for use. 3.19 Instructions for Use: Information provided by the manufacturer to inform the device user of the medical device’s intended purpose and proper use and of any precautions to be taken. (GHTF/SG1/N70:2011) NOTE: Instructions for use can also be referred to as “package insert.” 3.20 In Vitro Diagnostic (IVD) Medical Device: ‘In Vitro Diagnostic (IVD) medical device’ means a medical device, whether used alone or in combination, intended by the manufacturer for the in-vitro examination of specimens derived from the human body solely or principally to provide information for diagnostic, monitoring or compatibility purposes. NOTE 1: IVD medical devices include reagents, calibrators, control materials, specimen receptacles, software, and related instruments or apparatus or other articles and are used, for example, for the following test purposes: diagnosis, aid to diagnosis, screening, monitoring, predisposition, prognosis, prediction, determination of physiological status. NOTE 2: In some jurisdictions, certain IVD medical devices may be covered by other regulations. (GHTF/SG1/N071:2012) 3.21 Label: Written, printed, or graphic information either appearing on the medical device itself, or on the packaging of each unit, or on the packaging of multiple devices. (GHTF/SG1/N70:2011) NOTE: The definition above refers to the human readable label. 3.22 Labeling: the label, instructions for use, and any other information that is related to identification, technical description, intended purpose and proper use of the medical device, but excluding shipping documents. (GHTF/SG1/N70:2011) NOTE 1: Labeling can also be referred to as “information supplied by the manufacturer.” NOTE 2: Labeling can be in printed or electronic format and may either physically accompany the medical device or direct the user to where the labeling information can be accessed (such as through a website). 3.23 Lay User: Individual who does not have formal training in a relevant field or discipline. (Modified from GHTF/SG1/N045:2008) NOTE 1: Principles for lay person(s) may also apply to self-testing for a medical device or IVD medical device. NOTE 2: For an IVD medical device used outside of a laboratory setting, the user of the IMDRF GRRP WG/N47 FINAL: 2018 IVD medical device will be considered a lay user. NOTE 3: For an IVD medical device for self-collection/self-testing, a self-tester is considered a lay user. 3.24 Life-Cycle: All phases in the life of a medical device, from the initial conception to final decommissioning and disposal. (ISO/IEC Guide 51:2014) 3.25 Manufacturer: “Manufacturer” means any natural or legal person1 with responsibility for design and/or manufacture of a medical device with the intention of making the medical device available for use, under their name; whether or not such a medical device is designed and/or manufactured by that person themselves or on their behalf by another person(s). (GHTF/SG1/N055:2009) NOTE 1: This ‘natural or legal person’ has ultimate legal responsibility for ensuring compliance with all applicable regulatory requirements for the medical device in the countries or jurisdictions where it is intended to be made available or sold, unless this responsibility is specifically imposed on another person by the Regulatory Authority (RA) within that jurisdiction. NOTE 2: The manufacturer’s responsibilities are described in other GHTF guidance documents. These responsibilities include meeting both pre-market requirements and post- market requirements, such as adverse event reporting and notification of corrective actions. NOTE 3: ‘Design and/or manufacture’, as referred to in the above definition, may include specification development, production, fabrication, assembly, processing, packaging, repackaging, labeling, relabeling, sterilization, installation, or remanufacturing of a medical device; or putting a collection of devices, and possibly other products, together for a medical purpose. NOTE 4: Any person who assembles or adapts a medical device that has already been supplied by another person for an individual patient, in accordance with the instructions for use, is not the manufacturer, provided the assembly or adaptation does not change the intended use of the medical device. NOTE 5: Any person who changes the intended use of, or modifies, a medical device without acting on behalf of the original manufacturer and who makes it available for use under his own name, should be considered the manufacturer of the modified medical device. NOTE 6: An authorised representative, distributor or importer who only adds its own address and contact details to the medical device or the packaging, without covering or changing the existing labeling, is not considered a manufacturer. NOTE 7: To the extent that an accessory is subject to the regulatory requirements of a medical device2, the person responsible for the design and/or manufacture of that accessory is considered to be a manufacturer. 1 The term “person” that appears here and in the other definitions of this document, includes legal entities such as a corporation, a partnership or an association. 2 See GHTF/SG1/N29 Information Document Concerning the Definition of the Term “Medical Device” IMDRF GRRP WG/N47 FINAL: 2018 3.26 Medical Device: Any instrument, apparatus, implement, machine, appliance, implant, reagent for in vitro use, software, material or other similar or related article, intended by the manufacturer to be used, alone or in combination, for human beings, for one or more of the specific medical purpose(s) of: • diagnosis, prevention, monitoring, treatment or alleviation of disease, • diagnosis, monitoring, treatment, alleviation of, or compensation for, an injury, • investigation, replacement, modification, or support of the anatomy, or of a physiologi- cal process, • supporting or sustaining life, • control of conception, • cleaning, disinfection or sterilization of medical devices, • providing information by means of in vitro examination of specimens derived from the human body; and does not achieve its primary intended action by pharmacological, immunological, or metabolic means, in or on the human body, but which may be assisted in its intended function by such means. NOTE 1: Products which may be considered to be medical devices in some jurisdictions but not in others include: • disinfection substances, • aids for persons with disabilities, • devices incorporating animal and/or human tissues, • devices for in-vitro fertilization or assisted reproduction technologies. (Modified from GHTF/SG1/N071:2012) NOTE 1: For clarification purposes, in certain regulatory jurisdictions, devices for cosmetic/aesthetic purposes are also considered medical devices. NOTE 2: For clarification purposes, in certain regulatory jurisdictions, the commerce of devices incorporating human tissues is not allowed. 3.27 Near-Patient Testing: Testing that is performed near a patient and outside of centralized laboratory testing facilities. NOTE 1: Users of near-patient testing can include lay or professional users. NOTE 2: This is not intended to refer to sample collection procedures. NOTE 3: In certain regulatory jurisdictions, this is also referred to as Point of Care Testing. 3.28 Normal Use: operation, including routine inspection and adjustments by any user, and IMDRF GRRP WG/N47 FINAL: 2018 stand-by, according to the instructions for use or in accordance with generally accepted practice for those medical devices or IVD medical devices provided without instructions for use. (modified from ISO 62366-1:2015). NOTE 1: Normal use should not be confused with intended use. While both include the concept of use as intended by the manufacturer, intended use focuses on the medical purpose while normal use incorporates not only the medical purpose, but maintenance, transport, etc. as well. NOTE 2: Use error can occur in normal use. NOTE 3: Medical devices and IVD medical devices that can be used safely without instructions for use are exempted from having instructions for use by some authorities with jurisdiction. | Product to be used for the containment, protection, handling, delivery, storage, transport and presentation of goods, from raw materials to processed goods, from the | producer to the user or consumer, including processor, assembler or other intermediary. | ( | 3.30 Patient: An individual under the care of a healthcare provider who may benefit from the action of a medical device. A patient may also be a user of a medical device. 3.31 Performance: The ability of a medical device to achieve its intended purpose as stated by the manufacturer. Performance may include both clinical and technical aspects. 3.32 Performance Evaluation of an IVD Medical Device: Assessment and analysis of data to establish or verify the scientific validity, the analytical and, where applicable, the clinical performance of an IVD medical device. 3.33 Performance of an IVD Medical Device: The ability of an IVD medical device to achieve its intended use/intended purpose as claimed by the manufacturer. The performance of an IVD medical device consists of the analytical and, where applicable, the clinical performance supporting the intended use of the IVD medical device. (GHTF/SG5/N6:2012) 3.34 Regulatory Authority (RA): A government body or other entity that exercises a legal right to control the use or sale of medical devices within its jurisdiction, and that may take enforcement action to ensure that medical products marketed within its jurisdiction comply with legal requirements. (IMDRF/GRRP WG/N040:2017) 3.35 Risk: Combination of the probability of occurrence of harm and the severity of that harm. (ISO/IEC Guide 51:2014) 3.36 Risk Analysis: Systematic use of available information to identify hazards and to estimate the risk. (ISO/IEC Guide 51:2014) 3.37 Risk Assessment: Overall process comprising a risk analysis and a risk evaluation. IMDRF GRRP WG/N47 FINAL: 2018 (ISO/IEC Guide 51:2014) 3.38 Risk Evaluation: Procedure based on the risk analysis to determine whether tolerable risk has been exceeded. (ISO/IEC Guide 51:2014) 3.39 Safety: Freedom from unacceptable risk. (ISO/IEC Guide 51:2014) 3.40 Self-Testing: A medical device or IVD medical device used by a lay user who is responsible for collecting the data or specimen, by themselves and on themselves, relying solely on the instructions provided by the manufacturer. This use can also include performing the test and interpreting the results by themselves and on themselves. 3.41 Shelf-Life: Period of time until the expiry date during which a medical device in its original packaging maintains its stability under the storage conditions specified by the manufacturer. NOTE: Stability (3.38) and expiry date (3.11) are related concepts (Modified from ISO 18113-1:2009) 3.42 Stability: Ability of a medical device and IVD medical device to maintain its safety and performance characteristics within the manufacturer’s specifications over a specified period of time. NOTE 1: Stability applies to - Sterile and non-sterile medical devices whose physical, chemical or functional properties may be altered or compromised over a stated time interval; - IVD reagents, calibrators and controls, when stored, transported and used in the conditions specified by the manufacturer, - Reconstituted lyophilized materials, working solutions and material removed from sealed containers, when prepared, used and stored according to the manufacturer’s instructions for use, - Measuring instruments or measuring systems after calibration. NOTE 2: Stability of an IVD reagent or measuring system is normally quantified with respect to time and specified conditions, - In terms of the duration of a time interval over which a measured property changes by a stated amount or - In terms of the change of a property under specified conditions. (Modified from ISO 18113-1:2009) 3.43 State of the Art: Developed stage of technical capability at a given time as regards products, processes and services, based on the relevant consolidated findings of science, technology and experience. NOTE1: The state of the art embodies what is currently and generally accepted as good practice in technology and medicine. The state of the art does not necessarily imply the most technologically advanced solution. The state of the art described here is sometimes IMDRF GRRP WG/N47 FINAL: 2018 referred to as the “generally acknowledged state of the art”. (Modified from ISO/IEC Guide 2:2004) 3.44 User: The person, professional or lay, who uses a medical device. The patient may be that user. (GHTF/SG1/N070:2011)