Process Validation is establishing documented evidence that a manufacturing process consistently produces a product that meets predetermined specifications and quality attributes.
Complete Guide to Process Validation
Process Validation is a critical quality assurance activity required by regulatory authorities worldwide to ensure that manufacturing processes consistently produce medical devices that meet predetermined specifications. It provides documented evidence through the collection and evaluation of data from the process design stage through commercial production.
Three stages of process validation:
1. Process Design (Stage 1)
- Define the commercial manufacturing process
- Identify critical process parameters (CPPs)
- Identify critical quality attributes (CQAs)
- Establish process controls and acceptance criteria
- Risk assessment using FMEA or similar tools
2. Process Qualification (Stage 2)
- Installation Qualification (IQ) - Verify equipment is installed correctly according to manufacturer specifications, all utilities are connected properly, and safety features are functional
- Operational Qualification (OQ) - Demonstrate equipment operates consistently within specified parameters across the anticipated operating ranges
- Performance Qualification (PQ) - Confirm the process consistently produces acceptable product under normal operating conditions using production materials, environment, and operators
3. Continued Process Verification (Stage 3)
- Ongoing monitoring and data collection
- Statistical process control (SPC)
- Trend analysis and continuous improvement
- Revalidation when changes occur
21 CFR 820.75 requirements:
The FDA's Quality System Regulation requires manufacturers to:
- Establish and maintain procedures for validating device design and production processes
- Validate processes when results cannot be fully verified by inspection and test
- Document validation activities and results
- Approve processes according to established procedures
Types of manufacturing processes requiring validation:
- Sterilization processes - EtO, gamma radiation, e-beam, steam
- Aseptic processing - Sterile filling and assembly
- Molding and extrusion - Plastic component manufacturing
- Welding and bonding - Permanent joining processes
- Coating processes - Drug-eluting coatings, antimicrobial coatings
- Software processes - Automated manufacturing control systems
Validation protocol elements:
1. Objective and scope
2. Responsibilities
3. Equipment and materials
4. Process description and parameters
5. Acceptance criteria
6. Test methods and procedures
7. Data collection and analysis plan
8. Deviations and corrective actions
9. Approval signatures
When revalidation is required:
- Significant process changes
- Equipment replacement or modification
- Changes in raw materials or suppliers
- Transfer to new manufacturing site
- Deviations or out-of-specification results
- Periodic revalidation (typically every 1-3 years)
Relationship to other standards:
- ISO 13485 - Clause 7.5.6 requires validation of processes for production and service provision
- EU MDR Annex I - General Safety and Performance Requirements include process validation
- MDSAP - Process validation required across all participating countries
Documentation requirements:
All validation activities must be thoroughly documented in:
- Validation master plan
- Validation protocols
- Validation reports
- Standard operating procedures (SOPs)
- Change control records
- Training records
Statistical considerations:
Process validation typically requires:
- Minimum of 3 consecutive successful production runs
- Sample sizes determined by statistical power analysis
- Use of control charts and capability studies (Cpk)
- Demonstration of process capability and reproducibility
Process validation is not a one-time event but a lifecycle activity that begins during process design and continues throughout the commercial life of the product.
Related Terms
More Compliance & Standards
View allThe FDA's Quality System Regulation (QSR) that establishes current Good Manufacturing Practice (cGMP) requirements for medical device manufacturers in the United States.
A systematic, independent examination of a Quality Management System to determine whether quality activities and results comply with planned arrangements and whether these arrangements are implemented effectively.
A systematic approach to investigating, correcting, and preventing quality problems in medical device manufacturing and operations.
A mandatory conformity marking for medical devices sold in the European Economic Area, indicating compliance with EU health, safety, and environmental requirements.