Pure Global AI
MDD · structured requirements

🇭🇰 Hong Kong IVD regulatory data

Explore the IVD requirements currently available for Hong Kong, organized by topic and linked to official sources.

1 facts1 dimensions1 official sourceslatest verification 2026-09-15
Explore questions and sources →

Common questions about changing a registered IVD in Hong Kong. Each question separates published rules from details still awaiting verification.

Reliance & recognition

Which foreign decisions or evidence can be reused, with what effect and conditions?

Compare countries

Existing records include acceptance, restrictions and evidence reuse. The applicable product scope and conditions are preserved below; a reference does not by itself confer market access.

Hong Kong approvals and evidence in other markets (1 records)

Prior approvals and evidence for Hong Kong (14 records)

Hong KongUS FDA

Prior approval / evidence record

FDA 510(k)/PMA/De Novo

Authority

FDA 510(k)/PMA/De Novo

Accepted

Yes

Benefit

Satisfies D002 in place of an MDACS Conformity Assessment Certificate; a Certificate to Foreign Government may be requested

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Hong KongSource authority not specified

Prior approval / evidence record

Expedited Approval Scheme (GN-02 item E002)

Name

Expedited Approval Scheme (GN-02 item E002)

Description

Streamlined review, applied by default unless the applicant opts out

Timeline Reduction

Streamlined review rather than a published shorter clock

Eligibility
  • Applicant is an LRP with a valid LRP listing number
  • No reported deaths or serious injury associated with the device, locally or worldwide
  • No active recalls, field safety corrective actions or adverse events, locally or worldwide
  • Every device in the application is supported by two or more valid, independent marketing approvals from the D002 jurisdictions

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Hong KongSource authority not specified

Prior approval / evidence record

Class II

Id

class-ii

Class Name

Class II

Risk Level

medium

Pathway

MDACS listing (Form MD101 via MDIS)

Official Timeline
Min

12

Max

12

Unit

weeks

Verification Status

official-verified

Official Source
Title

MDD Hong Kong GN-02:2026 — listing Class II/III/IV general medical devices (12-week vetting and approval)

Url

https://www.mdd.gov.hk/filemanager/common/mdacs/GN-02-E.pdf

Accessed Date

2026-08-30

Realistic Timeline
Min

3

Max

4

Unit

months

Prior Approval Benefits
  • A reference-market approval removes the need for an MDACS Conformity Assessment Certificate
Notes
  • GN-02 clause 5.3.1: vetting and approval should normally be completed within 12 weeks, counted from submission of the application and all required supporting information — including anything requested mid-assessment, so a question resets the practical start
  • Acknowledgement of receipt goes to the registered email; GN-02 clause 5.2.1 tells applicants to chase MDD only if nothing arrives within 2 weeks
  • Failure to supply requested supporting information within the stipulated timeframe closes the application outright rather than pausing it
  • Appeals against rejection, conditional-approval conditions or delisting go to the Medical Device Administration Appeal Committee within 14 working days, and the outcome is notified within 4 weeks of a complete appeal (GN-02 clauses 6.1 and 6.5)
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Hong KongSource authority not specified

Prior approval / evidence record

Class III

Id

class-iii

Class Name

Class III

Risk Level

high

Pathway

MDACS listing (Form MD101 via MDIS)

Official Timeline
Min

12

Max

12

Unit

weeks

Verification Status

official-verified

Official Source
Title

MDD Hong Kong GN-02:2026 — listing Class II/III/IV general medical devices (12-week vetting and approval)

Url

https://www.mdd.gov.hk/filemanager/common/mdacs/GN-02-E.pdf

Accessed Date

2026-08-30

Realistic Timeline
Min

3

Max

5

Unit

months

Prior Approval Benefits
  • A reference-market approval removes the need for an MDACS Conformity Assessment Certificate
Notes
  • Same 12-week GN-02 clause 5.3.1 clock as Class II and Class IV — MDD does not band the listing period by risk class
  • The applicant either holds an MDACS Conformity Assessment Certificate from an MDD-recognised body (D001) or submits marketing approvals from the named reference jurisdictions (D002); one or the other is mandatory
  • Where an MDSAP certificate is used as the QMS evidence, it must be paired with a marketing approval from a jurisdiction that itself recognises MDSAP
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Prior approval / evidence record

China NMPA IVD approval: recognised reference approval (GN-06)

reliance.china-nmpa-reliance-route-ivd
Fact-level verified
Statement

GN-06:2026 permits an NMPA registration certificate, with associated Product Technical Requirements where applicable, as a reference marketing approval supporting MDACS listing of Hong Kong Class B/C/D IVDs under D002. The Hong Kong class must be assessed separately. D003 requires the signed MDIVD-CCL; an EU ER/GSPR checklist is an alternative only with qualifying EU approval evidence. Local listing and remaining requirements still apply.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

NMPA Class II/III registration certificate; associated Product Technical Requirements where applicable

Target Jurisdiction Id

hong-kong

China Classes In Scope

NMPA-registered IVD; separately determine whether the product falls within Hong Kong Class B/C/D

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • MDACS listing remains voluntary and requires a Hong Kong Local Responsible Person.
  • Determine the Hong Kong risk class independently; Chinese and Hong Kong classes are not automatic equivalents.
  • Submit the NMPA registration certificate and associated Product Technical Requirements where applicable, explicitly covering all product models/codes.
  • Certificate of Free Sale may be requested.
  • One reference approval is not sufficient for the Expedited Approval Scheme: the separate E002 requirements include two independent approvals and the LRP/safety conditions.
  • Submit MDIVD-CCL; an EU ER/GSPR checklist is an alternative only when valid EU marketing approval certificates accompany the submission.

Effective from 2026-07-31

Independent review of the current official PDF on 2026-09-15: D002 identifies Chinese Mainland NMPA registration certificates; D003 retains the essential-principles checklist. No automatic risk-class mapping or retired checklist exemption is established by these items.

Hong KongSource authority not specified

Prior approval / evidence record

Class IV

Id

class-iv

Class Name

Class IV

Risk Level

highest

Pathway

MDACS listing (Form MD101 via MDIS)

Official Timeline
Min

12

Max

12

Unit

weeks

Verification Status

official-verified

Official Source
Title

MDD Hong Kong GN-02:2026 — listing Class II/III/IV general medical devices (12-week vetting and approval)

Url

https://www.mdd.gov.hk/filemanager/common/mdacs/GN-02-E.pdf

Accessed Date

2026-08-30

Realistic Timeline
Min

3

Max

6

Unit

months

Prior Approval Benefits
  • A reference-market approval removes the need for an MDACS Conformity Assessment Certificate
Notes
  • Same 12-week GN-02 clause 5.3.1 clock; Class IV was the first class MDACS opened to listing when the system was phased in from 2004
  • Approval may be conditional — commonly on annual post-market surveillance reports — and breaching a condition can lead to delisting
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Hong KongNMPA · US FDA

Prior approval / evidence record

"Novo Lane" innovative device priority (GN-02 item E001)

Name

"Novo Lane" innovative device priority (GN-02 item E001)

Description

Prioritised processing for first-of-its-kind devices, devices approved through another regulator’s innovation channel (NMPA 創新器械名錄, FDA De Novo) within 5 years, devices with AI features approved within 5 years, or devices whose core technology was patented within 5 years

Timeline Reduction

Priority in the processing queue; MDD publishes no shorter target period, and the device still needs the same D001 or D002 evidence

Eligibility
  • At least one of the four E001 innovation criteria
  • Conformity Assessment Certificate or reference marketing approval still required

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Prior approval / evidence record

NMPA (Chinese Mainland)

Authority

NMPA (Chinese Mainland)

Accepted

Yes

Benefit

Registration certificate satisfies D002; pairs with a GB/T 42061 certificate or manufacturing licence as QMS evidence

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Hong KongEU conformity assessment

Prior approval / evidence record

EU CE Mark

Authority

EU CE Mark

Accepted

Yes

Benefit

MDR certificates, or MDD certificates with a Regulation (EU) 2023/607 extension letter, satisfy D002; an EU ER or GSPR checklist may then replace the MD-CCL

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Related structured datasets

Classification systems · 1 record
Open Hong Kong IVD risk classification →
Id

hong-kong:ivd

Jurisdiction Id

hong-kong

Domains
  • ivd
Name

Hong Kong IVD risk classification

Classes
  • Id

    a

    Label

    IVD Class A

    Normalized Risk Level

    low

  • Id

    b

    Label

    IVD Class B

    Normalized Risk Level

    medium

  • Id

    c

    Label

    IVD Class C

    Normalized Risk Level

    high

  • Id

    d

    Label

    IVD Class D

    Normalized Risk Level

    highest

Regulator

MDD

Official Domains
  • mdd.gov.hk
  • mdco.gov.hk
  • dh.gov.hk
Source Status

official-source-verified

Source Title

MDD — Overview of the Hong Kong Medical Device Listing Process (scope of MDACS and device classification)

Source Url

https://www.mdd.gov.hk/filemanager/common/information-publication/Overview%20of%20Hong%20Kong%20Medical%20Device%20Listing%20Process.pdf

Verified On

2026-08-25

Source Supports
  • ivd-class-axis
Notes
  • Classification rules for IVD medical devices are set out in MDACS technical reference TR-006; Class A IVDMDs are outside the MDACS listing scope.
Registration pathways, prior approvals and planning · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Hong Kong registration →
Country

Hong Kong

Jurisdiction Id

hong-kong

Country Code

HK

Flag

🇭🇰

Regulator

Medical Device Division

Regulator Abbrev

MDD

Region

asia-pacific

Tier

2

Risk Classification

4-class (I, II, III, IV)

Classes
  • Id

    class-ii

    Class Name

    Class II

    Risk Level

    medium

    Pathway

    MDACS listing (Form MD101 via MDIS)

    Official Timeline
    Min

    12

    Max

    12

    Unit

    weeks

    Verification Status

    official-verified

    Official Source
    Title

    MDD Hong Kong GN-02:2026 — listing Class II/III/IV general medical devices (12-week vetting and approval)

    Url

    https://www.mdd.gov.hk/filemanager/common/mdacs/GN-02-E.pdf

    Accessed Date

    2026-08-30

    Realistic Timeline
    Min

    3

    Max

    4

    Unit

    months

    Prior Approval Benefits
    • A reference-market approval removes the need for an MDACS Conformity Assessment Certificate
    Notes
    • GN-02 clause 5.3.1: vetting and approval should normally be completed within 12 weeks, counted from submission of the application and all required supporting information — including anything requested mid-assessment, so a question resets the practical start
    • Acknowledgement of receipt goes to the registered email; GN-02 clause 5.2.1 tells applicants to chase MDD only if nothing arrives within 2 weeks
    • Failure to supply requested supporting information within the stipulated timeframe closes the application outright rather than pausing it
    • Appeals against rejection, conditional-approval conditions or delisting go to the Medical Device Administration Appeal Committee within 14 working days, and the outcome is notified within 4 weeks of a complete appeal (GN-02 clauses 6.1 and 6.5)
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

  • Id

    class-iii

    Class Name

    Class III

    Risk Level

    high

    Pathway

    MDACS listing (Form MD101 via MDIS)

    Official Timeline
    Min

    12

    Max

    12

    Unit

    weeks

    Verification Status

    official-verified

    Official Source
    Title

    MDD Hong Kong GN-02:2026 — listing Class II/III/IV general medical devices (12-week vetting and approval)

    Url

    https://www.mdd.gov.hk/filemanager/common/mdacs/GN-02-E.pdf

    Accessed Date

    2026-08-30

    Realistic Timeline
    Min

    3

    Max

    5

    Unit

    months

    Prior Approval Benefits
    • A reference-market approval removes the need for an MDACS Conformity Assessment Certificate
    Notes
    • Same 12-week GN-02 clause 5.3.1 clock as Class II and Class IV — MDD does not band the listing period by risk class
    • The applicant either holds an MDACS Conformity Assessment Certificate from an MDD-recognised body (D001) or submits marketing approvals from the named reference jurisdictions (D002); one or the other is mandatory
    • Where an MDSAP certificate is used as the QMS evidence, it must be paired with a marketing approval from a jurisdiction that itself recognises MDSAP
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

  • Id

    class-iv

    Class Name

    Class IV

    Risk Level

    highest

    Pathway

    MDACS listing (Form MD101 via MDIS)

    Official Timeline
    Min

    12

    Max

    12

    Unit

    weeks

    Verification Status

    official-verified

    Official Source
    Title

    MDD Hong Kong GN-02:2026 — listing Class II/III/IV general medical devices (12-week vetting and approval)

    Url

    https://www.mdd.gov.hk/filemanager/common/mdacs/GN-02-E.pdf

    Accessed Date

    2026-08-30

    Realistic Timeline
    Min

    3

    Max

    6

    Unit

    months

    Prior Approval Benefits
    • A reference-market approval removes the need for an MDACS Conformity Assessment Certificate
    Notes
    • Same 12-week GN-02 clause 5.3.1 clock; Class IV was the first class MDACS opened to listing when the system was phased in from 2004
    • Approval may be conditional — commonly on annual post-market surveillance reports — and breaching a condition can lead to delisting
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

Expedited Pathways
  • Name

    "Novo Lane" innovative device priority (GN-02 item E001)

    Description

    Prioritised processing for first-of-its-kind devices, devices approved through another regulator’s innovation channel (NMPA 創新器械名錄, FDA De Novo) within 5 years, devices with AI features approved within 5 years, or devices whose core technology was patented within 5 years

    Timeline Reduction

    Priority in the processing queue; MDD publishes no shorter target period, and the device still needs the same D001 or D002 evidence

    Eligibility
    • At least one of the four E001 innovation criteria
    • Conformity Assessment Certificate or reference marketing approval still required
  • Name

    Expedited Approval Scheme (GN-02 item E002)

    Description

    Streamlined review, applied by default unless the applicant opts out

    Timeline Reduction

    Streamlined review rather than a published shorter clock

    Eligibility
    • Applicant is an LRP with a valid LRP listing number
    • No reported deaths or serious injury associated with the device, locally or worldwide
    • No active recalls, field safety corrective actions or adverse events, locally or worldwide
    • Every device in the application is supported by two or more valid, independent marketing approvals from the D002 jurisdictions
Prior Approvals
  • Authority

    FDA 510(k)/PMA/De Novo

    Accepted

    Yes

    Benefit

    Satisfies D002 in place of an MDACS Conformity Assessment Certificate; a Certificate to Foreign Government may be requested

  • Authority

    EU CE Mark

    Accepted

    Yes

    Benefit

    MDR certificates, or MDD certificates with a Regulation (EU) 2023/607 extension letter, satisfy D002; an EU ER or GSPR checklist may then replace the MD-CCL

  • Authority

    Health Canada

    Accepted

    Yes

    Benefit

    Medical Device Licence satisfies D002

  • Authority

    Australia TGA

    Accepted

    Yes

    Benefit

    ARTG certificate satisfies D002

  • Authority

    Japan MHLW / registered certification body

    Accepted

    Yes

    Benefit

    Approval or certification satisfies D002

  • Authority

    NMPA (Chinese Mainland)

    Accepted

    Yes

    Benefit

    Registration certificate satisfies D002; pairs with a GB/T 42061 certificate or manufacturing licence as QMS evidence

  • Authority

    Singapore HSA

    Accepted

    Yes

    Benefit

    Record of approval satisfies D002

  • Authority

    South Korea MFDS

    Accepted

    Yes

    Benefit

    Product Licence satisfies D002; pairs with a KGMP certificate as QMS evidence

Key Requirements
  • Local Responsible Person incorporated or business-registered in Hong Kong, either the manufacturer or backed by a designation letter from it
  • Class I general medical devices and Class A IVDs cannot be listed at all — GN-01 clause 4.2.1 confines listing to Class II/III/IV devices and Class B/C/D IVDs, so there is no clock to publish for the lowest class and its absence is not a data gap
  • Chinese/English labelling per TR-005
  • QMS evidence: ISO 13485 from a GACI-signatory-accredited body, GB/T 42061, an NMPA manufacturing licence, MDSAP, or KGMP — each with its own paired marketing-approval requirement
  • Either an MDACS Conformity Assessment Certificate (D001) or reference-market marketing approvals (D002)
  • Essential Principles Conformity Checklist MD-CCL, or an EU ER/GSPR checklist where EU approvals are submitted, with supporting evidence due within 2 weeks of a request
  • MDACS is an administrative, voluntary system pending statutory control; listing is nonetheless required for supply into public healthcare institutions
Local Rep Required

Yes

Local Rep Name

Local Responsible Person (LRP)

Mdsap Accepted

Yes

Mdsap Benefit

Accepted as QMS evidence when paired with a marketing approval from a jurisdiction that recognises MDSAP

Registration Validity

The listing does not expire, but conditional approvals carry ongoing obligations such as annual post-market surveillance reports, and non-compliance can result in delisting

Verification Status

official-verified

Official Sources
  • Title

    MDD Hong Kong GN-02:2026 — listing Class II/III/IV general medical devices (12-week vetting and approval)

    Url

    https://www.mdd.gov.hk/filemanager/common/mdacs/GN-02-E.pdf

    Accessed Date

    2026-08-30

  • Title

    MDD Hong Kong GN-01:2025-1 — overview of the MDACS, listing scope and LRP obligations

    Url

    https://www.mdd.gov.hk/filemanager/common/mdacs/GN-01-E.pdf

    Accessed Date

    2026-08-30

Reviewed On

2026-08-30

Last Updated

2026-08-30

Process
Template Id

default

Status

planning-template

Usage Rule

Maintained planning steps and estimates, not verified regulator deadlines.

Steps
  • Step

    1

    Title

    Pre-Submission Preparation

    Description

    Gather technical documentation, quality management system (QMS) records, and clinical data required by MDD. Appoint a Local Responsible Person (LRP).

    Duration

    1-3 months

    Icon

    📋

  • Step

    2

    Title

    Application Submission

    Description

    Submit complete registration dossier to MDD including device description, risk analysis, and testing reports.

    Duration

    1-2 weeks

    Icon

    📤

  • Step

    3

    Title

    Administrative Review

    Description

    MDD reviews application for completeness. May request additional documents or clarifications.

    Duration

    2-4 weeks

    Icon

    🔍

  • Step

    4

    Title

    Technical Evaluation

    Description

    In-depth technical assessment of device safety and performance. MDD may issue questions.

    Duration

    12 weeks

    Icon

    ⚙️

  • Step

    5

    Title

    Registration Approval

    Description

    Upon successful review, MDD issues Hong Kong market authorization certificate.

    Duration

    1-2 weeks

    Icon

    ✅

Government fee schedules · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Hong Kong fees →
Jurisdiction Id

hong-kong

Country

Hong Kong

Country Code

HK

Authority

MDD

Low Risk Local
Min

0

Max

0

Currency

HKD

Per Product

Yes

High Risk Local
Min

0

Max

0

Currency

HKD

Per Product

Yes

Details
  • Id

    mdacs-device-listing-application

    Item

    MDACS device-listing application

    Currency

    HKD

    Amount Local

    0

    Amount Local Formatted

    HKD 0

    Frequency

    Per application

    Notes

    GN-01 section 5.4 states that the Government does not charge for inclusion in the Lists of Medical Devices or an application for inclusion.

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mdd.gov.hk/filemanager/common/mdacs/GN-01-E.pdf

    Source Title

    Hong Kong MDACS GN-01: no government device-listing application fee

    Verified On

    2026-08-25

Notes
  • The government application charge is HKD 0.
  • Conformity-assessment-body certification and other private costs are separate.
Last Updated

2026-08-25

Verification Status

official-verified

Official Sources
  • Title

    Hong Kong MDACS GN-01: no government device-listing application fee

    Url

    https://www.mdd.gov.hk/filemanager/common/mdacs/GN-01-E.pdf

    Accessed Date

    2026-08-25

Reviewed On

2026-08-25

Amount Status

exact-current

Range Available

Yes

Additional certifications and licences · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Hong Kong certifications →
Slug

hong-kong

Market

Hong Kong

Confidence

low

Schemes
  • Id

    medical-electrical-evidence

    Name

    Medical electrical evidence

    Category

    Medical Electrical Safety

    Status

    Included In Medical Registration

    Object

    Finished Device

    Note

    MDACS application checklist item C015: risk analysis per ISO 14971 or equivalent standards with a report or comprehensive summary, plus the list of international/national standards applied; product-specific statutory overlays apply under Cap 303/138/137/134/106; MDD maintains a recognised standards list (RS-01).

    Source Binding

    unresolved

  • Id

    ofca-technical-compliance

    Name

    OFCA technical compliance

    Category

    Radio Spectrum

    Status

    Mandatory If Feature

    Object

    Radio Equipment

    Note

    OFCA HKCA 10XX list (read 2026-08-30) includes HKCA 1052 Issue 4 (06/2022) Performance Specification for Medical Implant Communication Systems. OFCA's ISMEM consumer page requires an Industrial, Scientific and Medical Electronic Machine Licence (HK$80/year) for a person to possess and use ISM equipment that intentionally generates and emits radio waves for industrial, scientific or medical purposes (examples named: RF probes, RF cannulae, magnetic stimulator, medical thermage machine, magnetic resonance tomography), and that equipment must be type-approved against HKCA 1007 under the Compulsory Certification Scheme. SSAC Paper 4/2026 (OFCA, for discussion 14 July 2026) restates that equipment in the scope of HKCA 1007 is classified under the Compulsory Certification Scheme of HKTEC and must be certified before use. Issue 6 of HKCA 1007 was still a proposal on that paper, so Issue 5 remains the live specification cited on the ISMEM page.

    Source Binding

    unresolved

  • Id

    ofca-voluntary-certification-label

    Name

    OFCA voluntary certification/label

    Category

    Radio Spectrum

    Status

    Voluntary Or Procurement

    Object

    Exempt Or Licence Free Apparatus

    Note

    OFCA Short Range Devices page: Bluetooth in 2400–2483.5 MHz at up to 100 mW e.r.p. is exempt from licensing under the Telecommunications (Telecommunications Apparatus) (Exemption from Licensing) Order (Cap. 106Z) and certification against HKCA 1035 is voluntary. Generic SRD in 433.00–434.79 MHz at up to 10 mW e.r.p. sits under the Class Licence for Short Range Devices with voluntary certification against HKCA 1061. This is the opposite of the ISMEM/HKCA 1007 compulsory route.

    Source Urls
    • https://www.ofca.gov.hk/en/consumer_focus/guide/help_for_consumers/information_on_radio_applications/short_range_devices/index.html
    Source Binding

    candidate

  • Id

    dealer-licence

    Name

    Dealer licence

    Category

    Telecom Network Attachment

    Status

    Mandatory If Feature

    Object

    Dealer Activity

    Source Binding

    unresolved

  • Id

    radiation-authorization

    Name

    Radiation authorization

    Category

    Ionizing Radiation

    Status

    Mandatory If Feature

    Object

    Product Or Use Site

    Note

    MDACS product-specific overlays name Cap 303 Radiation Ordinance. Hong Kong e-Legislation is a JavaScript application; the Cap 303 section text was not recovered from this egress, so the licence trigger is not claimed from the Ordinance body. Retry condition: a reachable HKeL print of Cap 303 ss. 7–9.

    Source Binding

    unresolved

Official Sources
  • Authority

    OFCA

    Topic

    telecommunications standards information notes

    Url

    https://www.ofca.gov.hk/en/industry_focus/standards/tel_standards/telecom/information_notes/index.html

  • Authority

    OFCA

    Topic

    certification bodies and voluntary labelling

    Url

    https://www.ofca.gov.hk/en/industry_focus/standards/tel_standards/telecom/list_of_certification_bodies/index.html

  • Authority

    Medical Device Division

    Topic

    GN-02 application documents ISO 14971 risk analysis checklist

    Url

    https://www.mdd.gov.hk/filemanager/common/mdacs/GN-02-E.pdf

  • Authority

    OFCA

    Topic

    HKCA 10XX list including HKCA 1052 Medical Implant Communication Systems

    Url

    https://www.ofca.gov.hk/en/industry_focus/standards/tel_standards/hkca/radio_equipment_specifications/index.html

  • Authority

    OFCA

    Topic

    ISMEM licence and compulsory certification against HKCA 1007

    Url

    https://www.ofca.gov.hk/en/consumer_focus/guide/help_for_consumers/information_on_radio_applications/industrial_scientific_and_medical_electronic_mach/index.html

  • Authority

    OFCA

    Topic

    Short Range Devices — Cap. 106Z exemption and voluntary HKCA 1035/1061 certification

    Url

    https://www.ofca.gov.hk/en/consumer_focus/guide/help_for_consumers/information_on_radio_applications/short_range_devices/index.html

  • Authority

    OFCA

    Topic

    SSAC Paper 4/2026 — HKCA 1007 remains compulsory certification under HKTEC

    Url

    https://www.ofca.gov.hk/filemanager/ofca/en/content_751/SSAC_Paper_4_2026.pdf

Keep researching this market
View the Hong Kong resource hub →
Compare regulatory data across all countries →
DJ Fang
Need to apply Hong Kong's requirements to a product?
Confirm the current MDD pathway, evidence, and local-role details for the actual device before filing.
Pricing