Common questions about changing a registered medical device in India. Each question separates published rules from details still awaiting verification.

Structured regulatory requirements

Explore all 65 source-linked facts →

24 of 25 research dimensions currently covered · 25 official sources · latest verification 2026-10-21

Market entry

D01

Authority & regime

Verified
Competent authority
CDSCO is the Central Licensing Authority (CLA) for import of all device classes, manufacture of Class C and Class D, clinical investigation of investigational devices, and clinical performa…
1 source-linked fact
Checked 2026-08-25
Open details →
D02

Legal basis & guidance

Verified
Chapter IIIA registration of all notified devices and transition
From 1 April 2020 all devices notified under s.3(b) of the Act must be registered with the CLA on the Online System for Medical Devices (registration voluntary for 18 months, then mandatory…
Primary rules and commencement
Medical Devices Rules, 2017 (notified as G.S.R. 78(E) dated 31 January 2017) under the Drugs and Cosmetics Act, 1940 govern medical devices and IVDs; the rules came into force on 1 January…
2 source-linked facts
Checked 2026-08-25
Open details →
D03

Device definition & scope

Verified
All-device definition
Since 1 April 2020 every device instrument, apparatus, appliance, implant, material or other article, used alone or in combination, including software or an accessory, not achieving its pri…
Devices regulated as drugs
Medical devices are regulated as 'drugs' under s.3(b)(iv) of the Drugs and Cosmetics Act, 1940: devices intended for internal or external use in diagnosis, treatment, mitigation or preventi…
2 source-linked facts
Checked 2026-08-25
Open details →
D05

Registration pathways

Verified
Class A nonsterile/nonmeasuring route boundary
Class A non sterile and non measuring devices have a separate online registration framework under Rules19G 19M and the stated Eighth Schedule exemptions. The licensed major/minor PAC and co…
Class A non-sterile non-measuring registration route
Class A devices that are non sterile and non measuring are exempt from the licensing regime: manufacturer and importer register them on the Online System for Medical Devices with self certi…
4 source-linked facts
Checked 2026-10-21
Open details →
D06

Reliance & recognition

Verified
Free sale certificate reliance for import
Where a free sale certificate has been issued by the national regulatory authority of Australia, Canada, Japan, an EU country, the United Kingdom or the United States, the import licence is…
4 source-linked facts
Checked 2026-09-15
Open details →
D07

Economic-operator roles

Verified
IVD subsequent importer has a same-manufacturer condition
IVD FAQ Addendum02 Q7 directs a new authorised agent to the Subsequent Importer portal application only for an already licensed IVD where the legal and actual manufacturing site remain the…
New Indian agent for an already licensed device
A subsequent importer is a new Indian authorised agent seeking an import licence for a device already licensed to another agent with the same legal and actual manufacturer/site. The MD FAQ…
4 source-linked facts
Checked 2026-10-21
Open details →
D08

Establishment, import & distribution

Verified
Sale registration certificate (Form MD-42)
Any person selling, stocking, exhibiting or offering for sale or distributing devices must obtain a registration certificate: application in Form MD 41 to the State Licensing Authority, gra…
1 source-linked fact
Checked 2026-08-25
Open details →
D09

Dossier requirements

Verified
Brand identity in an Indian import application
The MD FAQ says that, for an Indian agent owned brand registered under the Trade Marks Act, 1999 and entered in Form MD 14, the brand must be present on the FSC issued to the legal manufact…
Materials for an import label PAC
The official Import License(Form MD 14) PAC label table marks the old/new label comparison with reasons, revised label and PAC application as mandatory. It marks an existing licence/retenti…
7 source-linked facts
Checked 2026-10-21
Open details →
D10

Official forms & portals

Verified
Import application form and portal
Import licence applications are made in Form MD 14 on the SUGAM online portal, with the licence issued as Form MD 15; state side applications use the Ministry's identified online portal.
1 source-linked fact
Checked 2026-08-25
Open details →

Submission & evidence

D11

Label, IFU & language

Verified
IVD label identity and importer details
For an IVD under the Medical Devices Rules, the label must identify the device, manufacturer and actual manufacturing premises and carry the applicable particulars in Rule 44. For imported…
Electronic instructions for use
Electronic instructions for use are expressly recognised: G.S.R. 30(E) inserted 'or electronic instructions for use' into rule 3(v), rule 59(3)(iii) and the Fourth and Seventh Schedules.
4 source-linked facts
Checked 2026-10-21
Open details →
D12

QMS, audit & inspection

Verified
QMS evidence for market access
The Fifth Schedule QMS (design, manufacture, packaging, labelling, testing, installation, servicing) applies to manufacturers; for import and Chapter IIIA registration a notarised ISO 13485…
Notified Body audit timing
For Class A manufacture no site audit is required before grant; the registered Notified Body must audit within 120 days after grant. For Class B the Notified Body audit precedes grant: audi…
2 source-linked facts
Checked 2026-08-25
Open details →
D13

Standards, testing & ancillary approvals

Unresolved
CDSCO electrical/EMC evidence
Electrical safety
WPC ETA
Wireless / radio
TEC MTCTE
Telecom terminal
7 maintained overlays
Open details →
D15

Review clocks

Verified
Class A · Registration
45 days
Class B · Registration
45-140 days
Class C · Registration with clinical data
90 days
4 maintained pathways · 4 related facts
Checked 2026-10-21
Open details →
D16

Approval validity & renewal

Verified
Perpetual licence validity
Manufacturing licences (Forms MD 5/MD 6/MD 9/MD 10) remain valid in perpetuity subject to payment of the licence retention fee before completion of every five year period, unless suspended…
Lapse on non-payment
If the retention fee is not paid within the permitted late fee period, the licence is deemed to have been cancelled (and a fresh licence is then required).
3 source-linked facts
Checked 2026-08-26
Open details →

Lifecycle & post-market

D17

Change control

Verified
IVD constitution change is a fresh licence route
For an import licence, a qualifying change in the constitution of the authorised agent or overseas manufacturer that executed the power of attorney requires written notice within 30 days an…
IVD brand approval and actual-site rename
For already approved Class C/D IVDs, CDSCO FAQ Addendum02 Q24 directs a separate online application for endorsement of a brand or additional brand and CLA approval. Q19 says a name only cha…
7 source-linked facts
Checked 2026-10-21
Open details →
D18

UDI, listing & traceability

Verified
IVD UDI commencement is order dependent
The substituted Rule 46 requires the unique device identification from the date and in the manner specified by a Central Government order. The substituted provision itself does not establis…
UDI deferred pending Central Government order
Rule 46 (as substituted by G.S.R. 918(E) with effect from 31 December 2021) requires every device approved for manufacture or import to bear a unique device identification from a date and i…
2 source-linked facts
Checked 2026-10-21
Open details →
D19

Clinical investigation & performance

Verified
No-predicate device permission
Import or manufacture of a device without a predicate requires prior permission after clinical investigation (application Form MD 26, permission Form MD 27, decision within 120 days extenda…
Clinical investigation permission
No sponsor may conduct a clinical investigation of an investigational device in human participants except under CLA permission (application Form MD 22 with Seventh Schedule data; permission…
2 source-linked facts
Checked 2026-08-25
Open details →
D20

Special access & derogation

Verified
Custom-made device exemption
Custom made devices are exempt from the licensing chapters (IV and V) where made to a duly qualified medical practitioner's written prescription, under his responsibility, for the sole use…
Government hospital import of investigational devices
Small quantities of investigational devices not otherwise importable but approved in the country of origin may be imported for treatment of a patient with a life threatening disease, diseas…
2 source-linked facts
Checked 2026-08-25
Open details →
D21

Manufacturer PMS

Verified
Periodic Safety Update Reports
Holders of Form MD 27 no predicate permission must submit Periodic Safety Update Reports to the CLA from the date of market launch: every six months for the first two years, then annually f…
1 source-linked fact
Checked 2026-08-25
Open details →
D22

Vigilance & incident reporting

Verified
Foreign regulatory action reporting by importers
Import licence holders must inform the CLA forthwith, and in all circumstances within fifteen days, of any administrative action abroad taken on account of an adverse reaction market withdr…
Materiovigilance Programme of India
The Materiovigilance Programme of India (MvPI), approved by the Ministry of Health and Family Welfare on 6 July 2015, collects device adverse events: the Indian Pharmacopoeia Commission is…
3 source-linked facts
Checked 2026-08-25
Open details →
D23

Recall, FSCA & market surveillance

Verified
Recall duty and authority directions
A manufacturer or authorised agent who has reason to believe a placed device may be unsafe must immediately initiate procedures to withdraw it from the market and patients, indicate the rea…
1 source-linked fact
Checked 2026-08-25
Open details →
D24

Advertising, claims & promotion

Verified
Definitions used by the Drugs and Magic Remedies Act
Under the Drugs and Magic Remedies (Objectionable Advertisements) Act 1954, 'advertisement' includes any notice, circular, label, wrapper or other document and any announcement made orally…
Misleading drug advertisements
An advertisement relating to a drug is prohibited if it directly or indirectly gives a false impression regarding the drug's true character, makes a false claim, or is otherwise false or mi…
5 source-linked facts
Checked 2026-08-26
Open details →
D25

Submission communications & appeal

Verified
IVD PAC reminders and product-count advice
The 13 March2026 IVD FAQ gives a30 day first reminder interval for PAC responses and further reminders at30 day intervals; nonresponse after the fourth reminder leads to deemed withdrawal.…
Appeal against import refusal
An applicant whose import licence application is rejected may appeal to the Central Government within forty five days; the Government decides within ninety days of the appeal. Documents fou…
3 source-linked facts
Checked 2026-10-21
Open details →

CDSCO registration fees in USD · reviewed 2026-08-25

Fee lineAmountProvenance
Import licence - Class A manufacturing siteUSD 1,000Verified
Import licence - Class A deviceUSD 50Verified
Import licence - Class B manufacturing siteUSD 2,000Verified
Import licence - Class B deviceUSD 1,000Verified
17 more fee lines on the full schedule · 21 of 21 rows source-verified

Central Drugs Standard Control Organisation (CDSCO) review clocks and planning ranges · 4-class (A, B, C, D) classification

Class / pathwayReview clockPlanning totalProvenance
Class A(Registration)

Licence granted or rejected within 45 days of application (Rule 20(4)); Notified Body audit happens within 120 days after licence grant, not before (Rule 20(4) proviso (ii))

GHTF approval provides significant advantage

45 days6-9 monthsVerified
Class B(Registration)

Rule 20(6) chain: Notified Body audit within 90 days of application, NB report to the licensing authority within 30 days of the audit, licence decision within 20 days of the audit report (140 days total)

Licence grant-or-reject clock is 45 days where it applies (Rule 20(4))

NABL testing may be required

45-140 days8-12 monthsVerified
Class C(Registration with clinical data)

Rules 21(4)/25(1): CDSCO document scrutiny within 45 days (restartable on deficiency rectification), then licence decision within 45 days of the site-inspection report

Clinical data required

90 days1-1.5 yearsVerified
Class D(Registration with extensive clinical)

Rules 21(4)/25(1): CDSCO document scrutiny within 45 days (restartable on deficiency rectification), then licence decision within 45 days of the site-inspection report

Indian clinical trials may be required

90 days1.3-2 yearsVerified
  1. 1Pre-Submission Preparation(1-3 months)
  2. 2Application Submission(1-2 weeks)
  3. 3Administrative Review(2-4 weeks)
  4. 4Technical Evaluation(45 days)
  5. 5Registration Approval(1-2 weeks)

Official CDSCO rules and guidance · 36 of 36 documents linked to official sources

Acts and RulesMedical Devices Acts and Rules Collection
GeneralVerified
CDSCO Medical DevicesCentral Drugs Standard Control Organisation - Medical Devices
GeneralVerified
Device ClassificationMedical Device Classification in India
ClassificationVerified
G.S.R. 102(E)Medical Devices (Amendment) Rules, 2020
RegistrationVerified

Computed from 71,807 official CDSCO license records · data as of 2026-08-26

71,807
Licenses
3,500
License Holders
Continue with this market

Turn India research into practical work

The target market is carried into each tool. Add the product, evidence, or regulatory change you want to assess.

View every AI workpaper →
Country hubs in other markets

Other South & Southeast Asia markets

DJ Fang
Entering the India market?
Local representation, CDSCO submissions, and lifecycle maintenance — handled with an expert.
Pricing