Pure Global AI
MDA · structured requirements

🇲🇾 Malaysia IVD regulatory data

Explore the IVD requirements currently available for Malaysia, organized by topic and linked to official sources.

14 facts7 dimensions9 official sourceslatest verification 2026-10-21
Explore questions and sources →

Common questions about changing a registered IVD in Malaysia. Each question separates published rules from details still awaiting verification.

Change control

Does the event require internal documentation, notice, prior approval, transfer, or a new application?

Compare countries

Transfer of the registration between ARs

answers.ar-holder-transfer
Fact-level verified

GD0041 provides a registration-ownership application by the new AR after registration and after pending applications are completed or withdrawn. The effective transfer date is MDA approval, not the manufacturer appointment-letter date. Its current/new AR forms and unchanged manufacturer field do not establish unrestricted substitution of a different foreign legal manufacturer.

Conditions
  • Distinguish a real new AR/registration-holder legal person from a same-entity rename and from a new foreign manufacturer.

Logo and other label edits

answers.artwork-other-labels
Fact-level verified

A logo or layout edit must be checked against GD0020 5.6.3. Other label-information changes can be Category 3; the listed no-CN exceptions cover rephrasing/reordering IFU information, adding/removing languages not required by MDA, and distributor information that does not affect registered information. A new device name or registered identifier is assessed separately under the Category 2 row; a logo edit is not automatically exempt.

Combining changes and pending applications

answers.cn-bundles-and-sequence
Fact-level verified

The official FAQ permits any Category 2/3 changes for one medical device in one CN. Combining2-50 devices requires similar changes and is limited to the listed types, including specified facility/QMS/manufacturer-name or administrative-address changes. A pending CN for the same device must finish before another CN can be submitted. List every related change in Annex B; this permission does not combine an ownership transfer automatically.

Manufacturer name and administrative office changes

answers.manufacturer-office-cat3
Fact-level verified

GD0020 Table 2 5.6.4(c) assigns manufacturer registered-name or administrative-office-address changes, and manufacturing-site-name-only changes with no site-address change, to Category 3. These changes may be implemented after complete documents are submitted. A physical manufacturing-site move has a different Category 2 branch; it is not assumed for an office-only change.

Name and registered identifier changes

answers.name-identifier-cat2
Fact-level verified

A change to a registered device name or identifier with no change to intended use or technical specifications is Category 2 under GD0020 Table 1 5.5.6(d). A proprietary-name change caused by acquisition or merger is also listed there. Obtain MDA evaluation and endorsement before implementation and market placement. An additional device under another proprietary name may instead trigger new registration under 5.3(f).

Conditions
  • The common registered-device rules expressly include IVDs; IVD-specific technical changes retain their separate categories.

Dossier requirements

Which evidence, local deltas, formats, and signatures are required?

Compare countries

AR transfer documents, fee and timing

answers.ar-transfer-materials-fee-clock
Fact-level verified

GD0041 requires Annex A signed by the new AR, its establishment licence, manufacturer LOA under GD0027 and a manufacturer request naming the devices and appointment date before submission. The current AR supplies Annex B, surrenders certificates and transfers distribution/complaints records. The guide gives approximately 30 working days from complete submission and RM 500 per registration ID; original certificate validity continues from its initial registration date.

Conditions
  • The guide describes registration-holder/AR transfer; it is not a complete foreign-manufacturer succession dossier.
  • Approximate timing is not a guarantee.

Documents for manufacturer name or office CN

answers.manufacturer-office-documents
Fact-level verified

The Category 3 manufacturer-information row calls for a declaration of conformity, the manufacturer undertaking to support devices already supplied under its former name where applicable, amended labels, an updated QMS certificate or relevant official evidence of the change, and annexes. The undertaking is post-market support evidence; it does not grant an unrestricted old-label sell-through period.

Evidence for a name or identifier CN

answers.name-identifier-documents
Fact-level verified

The Category 2 name/identifier row requires a declaration of conformity, a manufacturer declaration that the device is unchanged including intended use and technical specifications, updated configurations identifying affected devices, labels showing every amended section and annexes. The summary must state present/proposed information and reasons for all related changes.

Fees

What regulator fee applies to this class, route, event, and date?

Compare countries

Change-notification fees by category and ID

answers.cn-specific-fees
Fact-level verified

The official CN fee table charges Category 2 RM 50/500/1000/1500 for Classes A/B/C/D respectively, or RM 2500 for a device containing a medicinal product; Category 3 is RM 30 across the merged class column. For a single ID with both categories, add the category fees; for multiple IDs multiply each applicable category fee by its ID count. The printed three-Class B-ID Category 2+3 example totals RM 1590.

Conditions
  • These are CN fees, separate from new-registration, CAB, ownership-transfer or export-certificate charges.

Label, IFU & language

Which label and IFU elements, languages, symbols, and translation controls apply?

Compare countries

Labels must stay aligned with the registered device

answers.labels-and-ivd-details
Fact-level verified

GD0026 Seventh Edition requires device name/model and manufacturer information, plus the applicable local manufacturer/AR contact and IVD particulars. Labelling compliance may be completed after import or manufacture but before market placement; added AR/importer/translation labels must preserve the specified original traceability/date information. Changes to electronic labels follow GD0020, and corresponding paper IFUs must remain aligned.

Conditions
  • Bahasa Malaysia is required for home-use devices; English applies to other devices.
  • Relabelling permission is not a blanket old-stock grace or permission to market changed Category 2 devices before endorsement.

Reliance & recognition

Which foreign decisions or evidence can be reused, with what effect and conditions?

Compare countries

Existing records include acceptance, restrictions and evidence reuse. The applicable product scope and conditions are preserved below; a reference does not by itself confer market access.

Malaysia approvals and evidence in other markets (0 records)

No linked record is currently maintained. This does not establish that a pathway is unavailable.

Prior approvals and evidence for Malaysia (9 records)

MalaysiaUS FDA

Prior approval / evidence record

FDA 510(k)

Authority

FDA 510(k)

Accepted

Yes

Benefit

Accepted as reference-country evidence supporting the conformity assessment, but the MDA publishes no shorter period for it

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Prior approval / evidence record

Malaysia-China Phase 2 joint evaluation: excludes already registered devices

reliance.china-nmpa-bilateral-programme
Fact-level verified
Statement

Following the 2025 IVD reliance pilot (Phase 1, held), MDA and NMPA launched Pilot Phase 2, the Malaysia-China Joint Evaluation Pilot Programme, from 1 July to 30 September 2026 under the GHWP Common Evaluation Reliance Practice (CERP) framework. Phase 2 replaces one-way reliance with simultaneous, collaborative assessment by both regulators of devices not yet registered in either country: Chinese manufacturers submit to NMPA under its Special Channel Pilot Programme and Malaysian manufacturers submit to MDA for full conformity assessment, with each authority evaluating and issuing its own certificate. This is not a pathway for a product that already holds a China NMPA registration certificate.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

No prior registration certificate in either China or Malaysia (eligibility prerequisite)

Target Jurisdiction Id

malaysia

China Classes In Scope

China-made NMPA Class II or III general medical devices and IVDs (via NMPA 'Special Channel Pilot Programme'); Malaysian-made Class B/C/D GMD and IVDs (via MDA full conformity assessment)

Edge Type

Bilateral Programme

Market Access Strength

none

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Pilot window 1 July - 30 September 2026
  • Manufacturer must be based in, and own and operate, the manufacturing facility in China (for China-made) or Malaysia (for Malaysian-made)
  • Device must not already hold a registration certificate in Malaysia or China
  • Not open to third-party brand owners (rebranders, relabelers, contract assemblers)
  • Malaysian route: establishment licence, premarket screening by MDA, CAB full conformity assessment per MDA/GD/0031, MeDC@St submission, 5-year MDAR listing
  • China-made route: submission to NMPA (contact info@ChinaMedDevice.com per announcement)

Effective from 2026-07-01 · Effective until 2026-09-30

Source passage: The second phase introduces a Joint Evaluation approach. Unlike the previous reliance model, this phase enables collaborative, simultaneous assessment by both the MDA and NMPA. ... Device Classes: Class B, C, or D (GMD or IVD) [Malaysian-Made] / Class II or Class III (GMD or IVD) [China-Made]. Prior Registration Status: The device must not have obtained registration certificates in Malaysia or China.

Prior approval / evidence record

ASEAN Medical Device Directive (AMDD) harmonisation

Name

ASEAN Medical Device Directive (AMDD) harmonisation

Description

The AMDD gives the ten member states one device definition, one set of classification rules, one set of essential principles and one dossier structure, the Common Submission Dossier Template. It is not a reliance route. The ACCSQ Medical Device Product Working Group states in terms that the Directive is "not intended to confer mutual recognition" and that each member state retains the sovereign right to make its own registration decisions, so a registration in one ASEAN market authorises nothing in another and shortens no national review period.

Timeline Reduction

None — the saving is in dossier preparation, since one CSDT core can be reused across member states with national addenda

Eligibility
  • Any device filed in an ASEAN member state that has transposed the AMDD into national law

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Prior approval / evidence record

ASEAN member-state registration

Authority

ASEAN member-state registration

Accepted

No

Benefit

There is no ASEAN mutual recognition agreement for device registration — the AMDD harmonises the dossier, not the decision

Timeline Savings

None

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

MalaysiaUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA · UK MHRA · Singapore HSA · TFDA (jurisdiction not specified in this entry)

Prior approval / evidence record

Class B

Id

class-b

Class Name

Class B

Risk Level

medium

Pathway

Registration

Official Timeline
Min

30

Max

30

Unit

working-days

Verification Status

official-verified

Official Source
Title

MDA/GD/0070 — registration submission guide, cl. 6.4 turnaround

Url

https://www.mda.gov.my/index.php/documents/ukk/3766-final-medical-device-registration-submission-guide-for-conformity-assessment-by-way-of-verification-process-and-submission-of-application-in-medc-st-second-edition

Accessed Date

2026-08-25

Realistic Timeline
Min

3

Max

6

Unit

months

With Prior Approval
Min

3

Max

6

Unit

months

Prior Approval Benefits
  • EU CE Mark
  • MHLW
  • TGA
  • Health Canada
  • FDA
  • MHRA
  • HSA
  • TFDA
Notes
  • MDA/GD/0070 cl. 6.4: Class B registration turn-around 30 working days from complete documents and cleared payment
  • ASEAN harmonization applies
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Regulatory reference

China NMPA approval: not reference

reliance.china-nmpa-not-reference
Fact-level verified
Statement

Outside the time-limited bilateral pilots, MDA's verification route (abbreviated CAB conformity assessment) is available only for devices holding a qualifying approval from the authorities enumerated in MDA/GD/0070 Annex 1 Table 5: EU Notified Bodies, Japan MHLW, Australia TGA, Health Canada, US FDA, UK MHRA, Singapore HSA and Thai FDA. China's NMPA is not in the table, so an NMPA-only device must undergo full conformity assessment.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

malaysia

China Classes In Scope

Class B/C/D devices and IVDs under the standard conformity-assessment-by-verification route; NMPA approvals not eligible

Edge Type

No Reliance Route

Market Access Strength

none

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Recognised authorities per Table 5: EU NB, Japan MHLW (Ninsho/Shonin), Australia TGA (ARTG), Health Canada licence, US FDA 510(k)/PMA, UK MHRA (UKCA / EC-UKNI), Singapore HSA (SMDR), Thai FDA (Class 2-3 notified / Class 4 licensed)
  • Apply the current verification eligibility and exclusions in MDA/GD/0070. The September 2025 revision removed the one-year prior marketing requirement; do not reintroduce it.
  • This conclusion is limited to the ordinary verification pathway, not every possible bilateral programme.

Effective from 2025-10-13

Source passage: Types of approvals eligible for conformity assessment by way of verification process

MalaysiaUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA · UK MHRA · Singapore HSA · TFDA (jurisdiction not specified in this entry)

Prior approval / evidence record

Class C

Id

class-c

Class Name

Class C

Risk Level

high

Pathway

Registration

Official Timeline
Min

30

Max

30

Unit

working-days

Verification Status

official-verified

Official Source
Title

MDA/GD/0070 — registration submission guide, cl. 6.4 turnaround

Url

https://www.mda.gov.my/index.php/documents/ukk/3766-final-medical-device-registration-submission-guide-for-conformity-assessment-by-way-of-verification-process-and-submission-of-application-in-medc-st-second-edition

Accessed Date

2026-08-25

Realistic Timeline
Min

3

Max

6

Unit

months

With Prior Approval
Min

3

Max

6

Unit

months

Prior Approval Benefits
  • EU CE Mark
  • MHLW
  • TGA
  • Health Canada
  • FDA
  • MHRA
  • HSA
  • TFDA
Notes
  • MDA/GD/0070 cl. 6.4: Class C registration turn-around 30 working days from complete documents and cleared payment
  • Clinical data may be required
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

MalaysiaUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA · UK MHRA · Singapore HSA · TFDA (jurisdiction not specified in this entry)

Prior approval / evidence record

Class D

Id

class-d

Class Name

Class D

Risk Level

highest

Pathway

Full Registration

Official Timeline
Min

30

Max

30

Unit

working-days

Verification Status

official-verified

Official Source
Title

MDA/GD/0070 — registration submission guide, cl. 6.4 turnaround

Url

https://www.mda.gov.my/index.php/documents/ukk/3766-final-medical-device-registration-submission-guide-for-conformity-assessment-by-way-of-verification-process-and-submission-of-application-in-medc-st-second-edition

Accessed Date

2026-08-25

Realistic Timeline
Min

3

Max

6

Unit

months

With Prior Approval
Min

3

Max

6

Unit

months

Prior Approval Benefits
  • EU CE Mark
  • MHLW
  • TGA
  • Health Canada
  • FDA
  • MHRA
  • HSA
  • TFDA
Notes
  • MDA/GD/0070 cl. 6.4: Class D registration turn-around 30 working days from complete documents and cleared payment
  • Clinical evidence required
Clinical Trial Required

Yes

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

MalaysiaEU conformity assessment

Prior approval / evidence record

EU CE Mark

Authority

EU CE Mark

Accepted

Yes

Benefit

Accepted as reference-country evidence supporting the conformity assessment, but the MDA publishes no shorter period for it

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Review clocks

What review, validation, and applicant stop-clock rules apply?

Compare countries

CN review targets and implementation points

answers.cn-specific-times
Fact-level verified

GD0020 Table 5 lists30 working days for a single submission ID in Category 2 or3,60 working days for a single ID combining2+3, and60 working days for multiple IDs. These targets start with a complete application and supporting documents. Category 2 still requires endorsement before implementation; Category 3 may be implemented on complete submission rather than waiting for that administrative target.

Source provenance & time

Which original source, locator, and effective period support each fact, and when was it last verified?

Compare countries

GD0020 remains the conditional current basis

answers.current-transition-and-renewal
Fact-level verified

MDA announcement1799 makes GD0072 implementation conditional on the formal launch of MeDC@St3.0. The 14 October2026 re-registration guide still describes3.0 as under development, with a December2026 target and a further announcement before implementation. Use GD0020 for the recorded current CN answer unless a formal launch changes that condition. The automated2.0+ renewal route requires no active/pending CN or COO for the device.

Conditions
  • A development target or visible3.0 login page is not the formal implementation event.
  • Recheck a later launch announcement before an actual filing.

Related structured datasets

Classification systems · 1 record
Open Risk classes A through D. →
Id

malaysia:medical-device

Jurisdiction Id

malaysia

Domains
  • medical-device
  • ivd
Name

Risk classes A through D.

Classes
  • Id

    a

    Label

    Class A

    Normalized Risk Level

    low

  • Id

    b

    Label

    Class B

    Normalized Risk Level

    medium

  • Id

    c

    Label

    Class C

    Normalized Risk Level

    high

  • Id

    d

    Label

    Class D

    Normalized Risk Level

    highest

Regulator

MDA

Official Domains
  • mda.gov.my
Source Status

official-source-verified

Source Title

Medical Devices Regulations 2012 — First Schedule (Rules of Classification of Medical Device, paras 3-4, Appendices 1-2)

Source Url

https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

Verified On

2026-08-25

Source Supports
  • class-axis
  • classification-rules
  • ivd-class-axis
  • ivd-classification-rules
Notes
  • First Schedule para 3(1): the manufacturer is responsible for classification; para 3(2): four classes A-D by risk to patients, users and other persons.
  • Appendix 1 general-device rules (non-invasive, invasive, active, special): highest class applies when multiple rules match; software driving a device takes the device's class; standalone software may be a device; accessories classified separately. Appendix 2 IVD rules: Class D for blood/transfusion screening agents and congenital-disorder screening; Class C for self-testing.
  • Act 737 s. 3(1): classification done by the establishment based on risk level, intended use and vulnerability of the human body; disputes with a conformity assessment body go to the Authority (decision within 30 days per reg. 3(3) MDR 2012).
Registration pathways, prior approvals and planning · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Malaysia registration →
Country

Malaysia

Jurisdiction Id

malaysia

Country Code

MY

Flag

🇲🇾

Regulator

Medical Device Authority

Regulator Abbrev

MDA

Region

asia-pacific

Tier

2

Risk Classification

4-class (A, B, C, D)

Classes
  • Id

    class-a

    Class Name

    Class A

    Risk Level

    low

    Pathway

    Registration

    Official Timeline
    Min

    14

    Max

    20

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    MDA/GD/0070 — registration submission guide, cl. 6.4 turnaround

    Url

    https://www.mda.gov.my/index.php/documents/ukk/3766-final-medical-device-registration-submission-guide-for-conformity-assessment-by-way-of-verification-process-and-submission-of-application-in-medc-st-second-edition

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    1

    Max

    3

    Unit

    months

    Notes
    • MDA/GD/0070 cl. 6.4: Class A registration turn-around 14-20 working days from submission of complete documents and cleared payment
    • Simplified registration
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    No

  • Id

    class-b

    Class Name

    Class B

    Risk Level

    medium

    Pathway

    Registration

    Official Timeline
    Min

    30

    Max

    30

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    MDA/GD/0070 — registration submission guide, cl. 6.4 turnaround

    Url

    https://www.mda.gov.my/index.php/documents/ukk/3766-final-medical-device-registration-submission-guide-for-conformity-assessment-by-way-of-verification-process-and-submission-of-application-in-medc-st-second-edition

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    3

    Max

    6

    Unit

    months

    With Prior Approval
    Min

    3

    Max

    6

    Unit

    months

    Prior Approval Benefits
    • EU CE Mark
    • MHLW
    • TGA
    • Health Canada
    • FDA
    • MHRA
    • HSA
    • TFDA
    Notes
    • MDA/GD/0070 cl. 6.4: Class B registration turn-around 30 working days from complete documents and cleared payment
    • ASEAN harmonization applies
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

  • Id

    class-c

    Class Name

    Class C

    Risk Level

    high

    Pathway

    Registration

    Official Timeline
    Min

    30

    Max

    30

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    MDA/GD/0070 — registration submission guide, cl. 6.4 turnaround

    Url

    https://www.mda.gov.my/index.php/documents/ukk/3766-final-medical-device-registration-submission-guide-for-conformity-assessment-by-way-of-verification-process-and-submission-of-application-in-medc-st-second-edition

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    3

    Max

    6

    Unit

    months

    With Prior Approval
    Min

    3

    Max

    6

    Unit

    months

    Prior Approval Benefits
    • EU CE Mark
    • MHLW
    • TGA
    • Health Canada
    • FDA
    • MHRA
    • HSA
    • TFDA
    Notes
    • MDA/GD/0070 cl. 6.4: Class C registration turn-around 30 working days from complete documents and cleared payment
    • Clinical data may be required
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

  • Id

    class-d

    Class Name

    Class D

    Risk Level

    highest

    Pathway

    Full Registration

    Official Timeline
    Min

    30

    Max

    30

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    MDA/GD/0070 — registration submission guide, cl. 6.4 turnaround

    Url

    https://www.mda.gov.my/index.php/documents/ukk/3766-final-medical-device-registration-submission-guide-for-conformity-assessment-by-way-of-verification-process-and-submission-of-application-in-medc-st-second-edition

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    3

    Max

    6

    Unit

    months

    With Prior Approval
    Min

    3

    Max

    6

    Unit

    months

    Prior Approval Benefits
    • EU CE Mark
    • MHLW
    • TGA
    • Health Canada
    • FDA
    • MHRA
    • HSA
    • TFDA
    Notes
    • MDA/GD/0070 cl. 6.4: Class D registration turn-around 30 working days from complete documents and cleared payment
    • Clinical evidence required
    Clinical Trial Required

    Yes

    Local Testing Required

    No

    Gmp Required

    Yes

Expedited Pathways
  • Name

    ASEAN Medical Device Directive (AMDD) harmonisation

    Description

    The AMDD gives the ten member states one device definition, one set of classification rules, one set of essential principles and one dossier structure, the Common Submission Dossier Template. It is not a reliance route. The ACCSQ Medical Device Product Working Group states in terms that the Directive is "not intended to confer mutual recognition" and that each member state retains the sovereign right to make its own registration decisions, so a registration in one ASEAN market authorises nothing in another and shortens no national review period.

    Timeline Reduction

    None — the saving is in dossier preparation, since one CSDT core can be reused across member states with national addenda

    Eligibility
    • Any device filed in an ASEAN member state that has transposed the AMDD into national law
Prior Approvals
  • Authority

    FDA 510(k)

    Accepted

    Yes

    Benefit

    Accepted as reference-country evidence supporting the conformity assessment, but the MDA publishes no shorter period for it

    Timeline Savings

    None published

  • Authority

    EU CE Mark

    Accepted

    Yes

    Benefit

    Accepted as reference-country evidence supporting the conformity assessment, but the MDA publishes no shorter period for it

    Timeline Savings

    None published

  • Authority

    ASEAN member-state registration

    Accepted

    No

    Benefit

    There is no ASEAN mutual recognition agreement for device registration — the AMDD harmonises the dossier, not the decision

    Timeline Savings

    None

Key Requirements
  • Malaysia Authorized Representative (MAR)
  • English/Malay labeling
  • ISO 13485 certification
  • Free Sale Certificate
  • MDAR registration
  • Applicant clocks (MDA/GD/0070 cl. 6.2-6.3): mandatory information within 30 working days of request (refusal ground); other additional information within 90 working days
  • Establishment licence processed in 14-21 working days; CAB verification-assessment stages per Annex 2 Table 6
Local Rep Required

Yes

Local Rep Name

Malaysia Authorized Representative (MAR)

Mdsap Accepted

No

Mdsap Benefit

Not formally recognized but ISO 13485 accepted

Registration Validity

5 years

Verification Status

official-verified

Official Sources
  • Title

    MDA/GD/0070 — registration submission guide, cl. 6.4 turnaround

    Url

    https://www.mda.gov.my/index.php/documents/ukk/3766-final-medical-device-registration-submission-guide-for-conformity-assessment-by-way-of-verification-process-and-submission-of-application-in-medc-st-second-edition

    Accessed Date

    2026-08-25

  • Title

    ACCSQ Medical Device Product Working Group — Implementing the ASEAN Medical Device Directive: "Not intended to confer mutual recognition; each member state retains sovereign right for registration decisions"

    Url

    https://www.ahwp.info/sites/default/files/6%20Implementing%20the%20ASEAN%20Medical%20Device%20Directive%201.0.pdf

    Accessed Date

    2026-08-30

Reviewed On

2026-08-30

Last Updated

2026-08-30

Process
Template Id

default

Status

planning-template

Usage Rule

Maintained planning steps and estimates, not verified regulator deadlines.

Steps
  • Step

    1

    Title

    Pre-Submission Preparation

    Description

    Gather technical documentation, quality management system (QMS) records, and clinical data required by MDA. Appoint a Malaysia Authorized Representative (MAR).

    Duration

    1-3 months

    Icon

    📋

  • Step

    2

    Title

    Application Submission

    Description

    Submit complete registration dossier to MDA including device description, risk analysis, and testing reports.

    Duration

    1-2 weeks

    Icon

    📤

  • Step

    3

    Title

    Administrative Review

    Description

    MDA reviews application for completeness. May request additional documents or clarifications.

    Duration

    2-4 weeks

    Icon

    🔍

  • Step

    4

    Title

    Technical Evaluation

    Description

    In-depth technical assessment of device safety and performance. MDA may issue questions.

    Duration

    14-20 working days

    Icon

    ⚙️

  • Step

    5

    Title

    Registration Approval

    Description

    Upon successful review, MDA issues Malaysia market authorization certificate.

    Duration

    1-2 weeks

    Icon

    ✅

Government fee schedules · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Malaysia fees →
Country

Malaysia

Jurisdiction Id

malaysia

Country Code

MY

Authority

MDA

Low Risk Local
Min

1250

Max

1250

Currency

MYR

Per Product

Yes

High Risk Local
Min

2500

Max

3750

Currency

MYR

Per Product

Yes

Details
  • Id

    application-fee-class-a

    Item

    Registration application fee - Class A

    Currency

    MYR

    Amount Local

    500

    Amount Local Formatted

    MYR 500

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://portal.mda.gov.my/index.php/documents/ukk/3859-pua-330-2025-2/file

    Source Title

    Malaysia Medical Device (Amendment) Regulations 2025, P.U. (A) 330

    Source Effective Date

    2026-01-01

    Verified On

    2026-08-25

  • Id

    application-fee-class-b

    Item

    Registration application fee - Class B

    Currency

    MYR

    Amount Local

    250

    Amount Local Formatted

    MYR 250

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    application-fee-class-c

    Item

    Registration application fee - Class C

    Currency

    MYR

    Amount Local

    500

    Amount Local Formatted

    MYR 500

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    application-fee-class-d

    Item

    Registration application fee - Class D

    Currency

    MYR

    Amount Local

    750

    Amount Local Formatted

    MYR 750

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    registration-fee-class-a

    Item

    Registration fee (on registration) - Class A

    Currency

    MYR

    Amount Local

    750

    Amount Local Formatted

    MYR 750

    Frequency

    Per registration, 5-year validity

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://portal.mda.gov.my/index.php/documents/ukk/3859-pua-330-2025-2/file

    Source Title

    Malaysia Medical Device (Amendment) Regulations 2025, P.U. (A) 330

    Source Effective Date

    2026-01-01

    Verified On

    2026-08-25

  • Id

    registration-fee-class-b

    Item

    Registration fee - Class B

    Currency

    MYR

    Amount Local

    1000

    Amount Local Formatted

    MYR 1,000

    Frequency

    Per registration, 5-year validity

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    registration-fee-class-c

    Item

    Registration fee - Class C

    Currency

    MYR

    Amount Local

    2000

    Amount Local Formatted

    MYR 2,000

    Frequency

    Per registration, 5-year validity

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    registration-fee-class-d

    Item

    Registration fee - Class D

    Currency

    MYR

    Amount Local

    3000

    Amount Local Formatted

    MYR 3,000

    Frequency

    Per registration, 5-year validity

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    registration-fee-medicinal-product-device

    Item

    Registration fee - device incorporating a medicinal product

    Currency

    MYR

    Amount Local

    5000

    Amount Local Formatted

    MYR 5,000

    Frequency

    Per registration, 5-year validity

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    establishment-application-per-role

    Item

    Establishment-licence application fee (per role: manufacturer, AR, distributor, importer)

    Currency

    MYR

    Amount Local

    250

    Amount Local Formatted

    MYR 250

    Frequency

    Per role

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    establishment-licence-manufacturer

    Item

    Establishment licence - manufacturer

    Currency

    MYR

    Amount Local

    4000

    Amount Local Formatted

    MYR 4,000

    Frequency

    Per licence

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    establishment-licence-authorised-representative

    Item

    Establishment licence - authorised representative

    Currency

    MYR

    Amount Local

    4000

    Amount Local Formatted

    MYR 4,000

    Frequency

    Per licence

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    establishment-licence-distributor

    Item

    Establishment licence - distributor

    Currency

    MYR

    Amount Local

    2000

    Amount Local Formatted

    MYR 2,000

    Frequency

    Per licence

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    establishment-licence-importer

    Item

    Establishment licence - importer

    Currency

    MYR

    Amount Local

    2000

    Amount Local Formatted

    MYR 2,000

    Frequency

    Per licence

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    establishment-renewal-application-per-role

    Item

    Establishment-licence renewal application fee

    Currency

    MYR

    Amount Local

    200

    Amount Local Formatted

    MYR 200

    Frequency

    Per role

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    establishment-renewal-manufacturer-or-ar

    Item

    Establishment-licence renewal - manufacturer or authorised representative

    Currency

    MYR

    Amount Local

    2000

    Amount Local Formatted

    MYR 2,000

    Frequency

    Per renewal

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    establishment-renewal-distributor-or-importer

    Item

    Establishment-licence renewal - distributor or importer

    Currency

    MYR

    Amount Local

    1000

    Amount Local Formatted

    MYR 1,000

    Frequency

    Per renewal

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    cab-registration-application

    Item

    Conformity-assessment-body registration application

    Currency

    MYR

    Amount Local

    1500

    Amount Local Formatted

    MYR 1,500

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    cab-registration

    Item

    Conformity-assessment-body registration

    Currency

    MYR

    Amount Local

    8000

    Amount Local Formatted

    MYR 8,000

    Frequency

    Per registration

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    export-permit

    Item

    Export permit

    Currency

    MYR

    Amount Local

    100

    Amount Local Formatted

    MYR 100

    Frequency

    Per permit

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    appeal

    Item

    Appeal

    Currency

    MYR

    Amount Local

    250

    Amount Local Formatted

    MYR 250

    Frequency

    Per appeal

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

  • Id

    advertising-approval

    Item

    Advertising-approval processing fee

    Currency

    MYR

    Amount Local

    1000

    Amount Local Formatted

    MYR 1,000

    Frequency

    Per advertisement

    Notes

    Medical Device (Advertising) Regulations 2019 reg. 2(3)(b).

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Source Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Verified On

    2026-08-25

Notes
  • Fees follow the Medical Device Regulations 2012 Fifth Schedule as amended by P.U.(A) 330/2025 from 1 January 2026: Class A application rose from RM100 to RM500 and a Class A registration fee of RM750 was introduced.
  • Application and registration fees are cumulative per device; establishment licences are billed per role.
  • Conformity-assessment-body charges are separate commercial fees.
Last Updated

2026-08-25

Verification Status

official-verified

Official Sources
  • Title

    Malaysia Medical Device (Amendment) Regulations 2025, P.U. (A) 330

    Url

    https://portal.mda.gov.my/index.php/documents/ukk/3859-pua-330-2025-2/file

    Effective Date

    2026-01-01

    Accessed Date

    2026-08-25

  • Title

    Malaysia Medical Device Regulations 2012, Fifth Schedule fee table

    Url

    https://www.mda.gov.my/documents/dmdocuments/Medical%20Device%20Regulations%202012.pdf

    Accessed Date

    2026-08-25

Reviewed On

2026-08-25

Amount Status

exact-current

Range Available

Yes

Additional certifications and licences · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Malaysia certifications →
Slug

malaysia

Market

Malaysia

Confidence

low

Schemes
  • Id

    mda-electrical-emc-evidence

    Name

    MDA electrical/EMC evidence

    Category

    Medical Electrical Safety

    Status

    Included In Medical Registration

    Object

    Finished Device

    Note

    MDA/GD/0070 Annex 4: active devices need electrical-safety, mechanical/environmental-protection and EMC evidence (IEC 60601-1 series test summaries, reports or certificates); QMS certificates per ISO 13485 or equivalents (MDSAP, US FDA QSR, Japan MHLW Ordinance 169; ISO 9001 only for Class A empty gas cylinders); clinical investigational plans per ISO 14155; Class A QMS verification covers sterility maintenance and metrology compliance for measuring functions (Third Schedule MDR 2012 para 6(2)(a)).

    Source Urls
    • https://www.mda.gov.my/index.php/documents/ukk/3766-final-medical-device-registration-submission-guide-for-conformity-assessment-by-way-of-verification-process-and-submission-of-application-in-medc-st-second-edition
    Source Binding

    candidate

  • Id

    mcmc-sirim-communications-product-certification

    Name

    MCMC/SIRIM communications product certification

    Category

    Radio Spectrum

    Status

    Mandatory If Feature

    Object

    Communications Multimedia Or Hybrid Product

    Source Urls
    • https://www.sirim-qas.com.my/service/communication-multimedia-and-hybrid-product/
    Source Binding

    candidate

  • Id

    jakim-medical-device-halal-certification

    Name

    JAKIM medical-device halal certification

    Category

    Halal And Material Origin

    Status

    Voluntary Or Procurement

    Object

    Eligible Material Risk Device

    Source Urls
    • https://www.halal.gov.my/index.php?content_id=202203306243ae3925019&data=bW9kdWxlcy9jb250ZW50X2RldGFpbHM7Ozs%3D&page_title=Announcement
    Source Binding

    candidate

  • Id

    atom-class-c-irradiating-apparatus-licence

    Name

    Atom Malaysia Class C irradiating-apparatus licence

    Category

    Ionizing Radiation

    Status

    Mandatory If Feature

    Object

    Product And Use Site

    Note

    Act 304 and the Radiation Protection (Licensing) Regulations 1986 regulate radiation generators; Atom Malaysia lists Class C for irradiating apparatus. Read the 2011 Act reprint together with Act A1779: its section 9 updates Act 304 section 12(3), including the eligible medical professionals, while sections 13–14 replace the Board with the Director General in sections 15–16. Applications for a medical purpose are referred to the Director General of Health; doubtful medical/non-medical classification is referred to the Advisory Council under section 16(4a). The current licence catalogue lists an RM 15 application fee and a licence fee by category. Act A1779 received royal assent on 20 November 2025. P.U. (B) 425/2025 commenced it on 1 December 2025 except sections 10, 15, 17, 18, 33 and 53. The new import/export/transit permit provisions are among those exceptions; do not treat their enactment or the general commencement date as proof that a specific permit provision is in force.

    Source Urls
    • https://www.atom.gov.my/en/license-application
    • https://www.atom.gov.my/wp-content/uploads/2024/04/Act-304-Atomic-Energy-Licensing-Act-1984.reprint-2011.pdf
    • https://www.atom.gov.my/wp-content/uploads/2025/12/A1779-ATOMIC-ENERGY-LICENSING-AMENDMENT-ACT-2025.pdf
    • https://www.atom.gov.my/wp-content/uploads/2025/12/PUB-425_2025-Penetapan-Tarikh-Permulaan-Kuat-Kuasa.pdf
    Source Binding

    reviewed

  • Id

    mda-establishment-import-licence

    Name

    MDA establishment import licence

    Category

    Import Conformity

    Status

    Mandatory At Import

    Object

    Commercial Importer

    Note

    Medical Device Act 2012 (Act 737) s.15 requires an establishment licence for importing, exporting or placing registered medical devices on the market. Under MDA General Order No. 1 of 2024 (single-license policy), new applications have been limited to one establishment role since 16 February 2024, with full policy implementation from 1 July 2024. An authorised representative (AR) that also imports must obtain the corresponding importer licence; an AR licence does not automatically authorise importing under the single-role policy. Existing multiple-role licences transition to separate licences at renewal during the 2024–2027 implementation period; check the roles and validity of the actual licence. Importers are appointed by the AR to import its registered devices. Applications go through MeDC@St and the applicable GDPMD requirements must be met.

    Source Urls
    • https://portal.mda.gov.my/index.php/industry/establishment-licence/establishment-license-information
    • https://www.mda.gov.my/index.php/faq/single-license-policy-implementation
    Source Binding

    reviewed

Official Sources
  • Authority

    SIRIM QAS

    Topic

    communications, multimedia and hybrid product certification

    Url

    https://www.sirim-qas.com.my/service/communication-multimedia-and-hybrid-product/

  • Authority

    JAKIM

    Topic

    medical-device halal certification scheme

    Url

    https://www.halal.gov.my/index.php?content_id=202203306243ae3925019&data=bW9kdWxlcy9jb250ZW50X2RldGFpbHM7Ozs%3D&page_title=Announcement

  • Authority

    MDA

    Topic

    MDA/GD/0070 conformity assessment verification submission guide

    Url

    https://www.mda.gov.my/index.php/documents/ukk/3766-final-medical-device-registration-submission-guide-for-conformity-assessment-by-way-of-verification-process-and-submission-of-application-in-medc-st-second-edition

  • Authority

    Jabatan Tenaga Atom (Atom Malaysia)

    Topic

    Act 304 licence classes including Class C irradiating apparatus via elesen-eSPP

    Url

    https://www.atom.gov.my/en/license-application

  • Authority

    MDA

    Topic

    Establishment licence — Act 737 s.15(1) import/export/place-in-market requirement and importer role

    Url

    https://portal.mda.gov.my/index.php/industry/establishment-licence/establishment-license-information

  • Authority

    MDA

    Topic

    MDA establishment import licence — single-license policy, separate AR/importer roles, implementation dates and renewal transition

    Url

    https://www.mda.gov.my/index.php/faq/single-license-policy-implementation

  • Authority

    Jabatan Tenaga Atom (Atom Malaysia)

    Topic

    Act 304 radiation-generator licensing — principal Act, read with Act A1779 and its commencement notice

    Url

    https://www.atom.gov.my/wp-content/uploads/2024/04/Act-304-Atomic-Energy-Licensing-Act-1984.reprint-2011.pdf

  • Authority

    Jabatan Tenaga Atom (Atom Malaysia)

    Topic

    Act A1779 radiation-generator licensing amendments — sections 9, 13 and 14 update Act 304 sections 12, 15 and 16

    Url

    https://www.atom.gov.my/wp-content/uploads/2025/12/A1779-ATOMIC-ENERGY-LICENSING-AMENDMENT-ACT-2025.pdf

  • Authority

    Jabatan Tenaga Atom (Atom Malaysia)

    Topic

    Act A1779 commencement — P.U. (B) 425/2025 excludes sections 10, 15, 17, 18, 33 and 53 from 1 December 2025 commencement

    Url

    https://www.atom.gov.my/wp-content/uploads/2025/12/PUB-425_2025-Penetapan-Tarikh-Permulaan-Kuat-Kuasa.pdf

Keep researching this market
View the Malaysia resource hub →
Compare regulatory data across all countries →
DJ Fang
Need to apply Malaysia's requirements to a product?
Confirm the current MDA pathway, evidence, and local-role details for the actual device before filing.
Pricing