Pure Global AI
SAHPRA · structured requirements

🇿🇦 South Africa IVD regulatory data

Explore the IVD requirements currently available for South Africa, organized by topic and linked to official sources.

14 facts8 dimensions13 official sourceslatest verification 2026-10-21
Explore questions and sources →

Common questions about changing a registered IVD in South Africa. Each question separates published rules from details still awaiting verification.

Change control

Does the event require internal documentation, notice, prior approval, transfer, or a new application?

Compare countries

Successive amendments and renewal sequence

amendment.and.notification.sequence
Fact-level verified

All intended amendment changes must be stated in the cover letter and the updated full product listing includes approved products, with products intended for removal omitted as instructed. A further amendment may not be submitted while a prior amendment is pending. Product-list notifications may accompany renewal if the licence particulars and approved class do not change; amendments to company/entity/address/AR/holder/activities/personnel or new class must be filed after renewal is complete and the latest licence is issued.

Conditions
  • The cover-letter rule does not authorize one cross-regime filing/fee/clock for a new legal entity plus NIRMED or CFS changes.

Establishment amendment trigger and materials

establishment.amendment.trigger.and.documents
Fact-level verified

An establishment-licence amendment is required for changes to licence-holder particulars, an unapproved device class/category, responsible personnel/authorised representative, activities or special storage conditions reflected on the approved licence. MD10 requires a cover letter detailing the amendment, the latest complete application in Excel and signed/initialled PDF, proof of amendment payment, AR CV, updated quality manual or wholesaler site master file, current licence, and an ISO13485 certificate matching the licensed entity/address if available under the stated prerequisite. Submit by the designated medical-device administration email.

Conditions
  • This is establishment licensing, not a universal per-product marketing-authorization variation.
  • Apply the current Class A exemption before assuming an establishment licence is required.

Class C/D originating-approval changes

foreign.approval.change.reporting
Fact-level verified

For a Class C or D device/IVD, the licence holder must submit full information within 30 days of a modification or a change in the status/certification of the originating approval supporting its licence application. Recall in that originating market requires immediate notification and recall action; withdrawal requires immediate notification and an agreed withdrawal plan. A corporate, brand or REF project must identify whether its supporting foreign approval actually changes.

Conditions
  • This is an applicant reporting duty, not an SAHPRA review time or an automatic sell-through allowance.

Additional listed-electronic-product licence gate

listed.electronic.product.change.route
Fact-level verified

If the product is a listed Group III electronic product, the NIRMED guidance requires a new licence application with compliance evidence for a manufacturer/brand change, a new brand/model name, or a different model. Administrative address/contact updates can be sent under the stated existing-licence detail-update route. Assess the actual R1302 listed category: the schedule includes blood-gas analysers, so the gate cannot be excluded solely because a product is an IVD. Local manufacture also includes assembly, preparation, processing, repair or maintenance under the Hazardous Substances Act.

Conditions
  • This is an additional conditional regime, separate from an establishment licence.
  • A pure orderable REF change is not automatically a licensed-model name change.
  • The current medical-manufacture application acceptance/form remains an authority interpretation issue; non-medical 01C, legacy MAN or X-ray-only XR24D must not be substituted as a confirmed current medical form.

Product-list-only notification route and materials

product.list.notification.gate
Fact-level verified

A product-list update may use notification only where no other particulars of the issued establishment licence change and the approved classes remain unaffected. Original-manufacturer/supplier details and same-class product additions may fall within that gate. Submit the signed cover letter, AR declaration that licence/class details are unaffected, latest full application in Excel and PDF, current licence, retention-payment proof, supporting product documents and matching ISO13485 certificate if available. If licence particulars are affected, apply for amendment instead.

Conditions
  • The notification cannot absorb a change of local legal entity/address, AR, activity, personnel or approved class.
  • COVID-related updates have a separate Q&A amendment qualification.

Notification implementation window and no application fee

product.list.notification.ten.day.condition
Fact-level verified

For a product-list-only notification within MD09’s conditions, no application fee is applicable. If SAHPRA has not responded or communicated an outcome within 10 working days after submission, the organisation may implement the change only if it does not constitute an amendment or otherwise affect particulars reflected on the licence. SAHPRA may require reclassification as an amendment; ten working days is not deemed approval for a change outside the notification gate.

Conditions
  • Do not apply this window to an establishment amendment, true successor licence or NIRMED application.

Dossier requirements

Which evidence, local deltas, formats, and signatures are required?

Compare countries

CFS after identity/product changes

cfs.local.manufacturer.and.exempt.route
Fact-level verified

The 2026 CFS guideline applies to South African manufactured medical devices including IVDs for export, with a licensed-manufacturer route and a specific exempt Class A route. Submit GLF-MD-21A, destination/purpose cover letter, current manufacturer/product/GMDN/class particulars, licence and product listing where applicable, supporting conformity/ISO/essential-principles evidence as required and payment proof. Exempt Class A applicants instead supply the exemption declaration, classification rationale and labels/IFU/marketing evidence. Current manufacturer and product identities must match the requested certificate; a foreign product import change alone does not establish eligibility for a South African CFS.

Conditions
  • Confirm the manufacturer’s actual South African activity, Class A exemption details and destination needs.
  • The guideline’s Class A wording uses and/or in places; do not silently broaden eligibility for a borderline sterile/measuring portfolio.

Economic-operator roles

Which manufacturer, representative, holder, applicant, importer, and distributor roles apply?

Compare countries

Same company amendment versus a new legal licensee

same.entity.versus.new.company
Fact-level verified

A company/establishment/entity-name change where the company registration remains the same is an establishment amendment. A South African establishment licence is not transferable to another company: each new applicant must obtain its own licence. An overseas original-manufacturer change that leaves all local licence particulars unchanged must instead be assessed under the product-list gate, with its updated supporting evidence; these are distinct roles.

Conditions
  • A change of contact or authorised representative is not proof of a change of legal entity.
  • If an actual product registration certificate exists, assess its separate statutory transfer route; an establishment amendment fee is not a product-certificate transfer fee.

Establishment, import & distribution

Which establishment licences, importer permissions, and distribution controls apply?

Compare countries

Class A establishment-licence exemption

class.a.establishment.exemption.condition
Fact-level verified

GN5855 of 12 February 2025 exempts the specified establishments dealing only in the described Class A medical devices/IVDs from section22C(1) establishment licensing for three years and repeals the 2024 exemption. Its operative text uses the wording “do not have measuring properties/characteristics and /or are non-sterile”; the 2026 CFS guide gives a separate exempt-applicant evidence route. Check the actual sterile/measuring characteristics and authority treatment rather than treating all Class A products or all IVDs as exempt.

Conditions
  • Conservative clear case: only non-measuring, non-sterile Class A portfolio. Borderline and/or cases need specific interpretation.
  • The exemption is not an exemption from label, vigilance or a triggered listed-electronic-product licence.

Fees

What regulator fee applies to this class, route, event, and date?

Compare countries

Establishment amendment fee gate

amendment.fee.versus.notification
Fact-level verified

A change that affects the approved establishment licence is a fee-bearing amendment under MD10 and the current gazetted establishment-licence amendment tariff. MD09’s qualified product-list-only notification carries no application fee. New-company establishment licensing, approved-licence issuance/collection, retention and CFS are distinct chargeable events; none is a universal product-registration transfer tariff.

Conditions
  • Use the separately referenced exact fee records for amounts; do not sum them into an automatic combined-case fee.

Label, IFU & language

Which label and IFU elements, languages, symbols, and translation controls apply?

Compare countries

IVD label identity and separate catalogue field

ivd.label.change.identities
Fact-level verified

An IVD label must carry the name/trade name, product description/intended use, catalogue code where applicable, manufacturer name/business address, holder-of-registration name/business address and the other prescribed traceability/safety particulars. It must be in at least English. Catalogue code is an expressly distinct label field; an orderable REF does not by itself establish a new model or UDI. Reagents separately supplied from a kit must individually comply.

Conditions
  • Check product-specific UDI/GS1 requirements separately rather than treating REF and UDI as synonyms.

Reliance & recognition

Which foreign decisions or evidence can be reused, with what effect and conditions?

Compare countries

Existing records include acceptance, restrictions and evidence reuse. The applicable product scope and conditions are preserved below; a reference does not by itself confer market access.

South Africa approvals and evidence in other markets (0 records)

No linked record is currently maintained. This does not establish that a pathway is unavailable.

Prior approvals and evidence for South Africa (5 records)

Regulatory reference

China NMPA approval: not reference

reliance.china-nmpa-not-reference
Pending verification
Statement

SAHPRA recognises six jurisdictions for premarket evidence: Australia, Brazil, Health Canada, Europe, Japan and USFDA, plus WHO prequalification. Establishment-licence applications must annex 'Evidence of premarket registration from one of the 6 jurisdictions recognised by SAHPRA', and Certificate of Free Sale applications for Class C and D devices must show pre-market approval/registration or emergency-use authorisation from at least one of the six or WHO PQ. China NMPA is not one of the six, so an NMPA-only Chinese device cannot satisfy the Class C/D evidence requirement.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

south-africa

China Classes In Scope

NMPA Class II/III certificates; target South Africa Class C and D devices and IVDs (licence listing and Certificate of Free Sale)

Edge Type

No Reliance Route

Market Access Strength

none

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Class C and D devices/IVDs listed in a CFS application (3.1.28).
  • ISO 13485 certification of the original manufacturer and a declaration of conformity with the Essential Principles are also required.
  • CFS processed within 15 days; valid at most one year.
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Effective from 2025-05-21

Source passage: For medium to high risk (Class C) and high risk (Class D) medical devices listed in the Certificate of Free Sale application: o Evidence of pre-market approval/registration/evidence of emergency use authorisation for each listed medical device/s from at least one of the six jurisdictions recognised by SAHPRA (Australia, Brazil, Health Canada, Europe, Japan, USFDA) or pre-qualification by the World Health [Organization] o Evidence of ISO13485 certification of the original manufacturer

Regulatory reference

China NMPA approval: outside SAHPRA reliance-guideline recognised authorities

reliance.china-nmpa-not-reference-reliance-guideline
Pending verification
Statement

SAHPRA's Medical Devices Reliance Guideline (SAHPGL-MD-22 v1, 25 February 2026; consulted as the November 2025 consultation text SAHPGL-MD-19_v1) grants abridged review only to devices already approved by a Recognised Regulatory Authority with which SAHPRA aligns (list to be published on SAHPRA's agreements/MoUs page), by a WHO Listed Authority (WHO permanent WLA list), under WHO CRP/PQ/EUL, or under the AMA continental procedure; a device without prior approval from any listed recognised authority goes to the full evaluation route. China NMPA is neither on the WHO WLA list nor named as an RRA, and SAHPRA's agreements page currently lists no China arrangement.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

south-africa

China Classes In Scope

NMPA Class II/III certificates; target Class B, C and D product registration and licence listing (reliance pathways)

Edge Type

No Reliance Route

Market Access Strength

none

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Verification of sameness with the product approved by the reference authority is mandatory under every reliance pathway.
  • Not applicable to Class A non-sterile/non-measuring devices or emergency-use registrations.
  • MDSAP audit reports and locally recognised CAB ISO 13485 audits are recognised for QMS.
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Effective from 2026-02-25

Source passage: A medical device that has not obtained any prior approval from any of the listed medical device recognised regulatory authorities at the point of application will be subject to the full evaluation route. [...] i) RECOGNISED REGULATORY AUTHORITIES An application must have been approved by one or more of the RRAs with which SAHPRA aligns itself. The list of RRA is available on the SAHPRA website. [...] iv) WHO Listed Authority (WLA) https://www.who.int/publications/m/item/list-of-who-listed-authorities-wlas

South AfricaEU conformity assessment

Prior approval / evidence record

CE Mark (EU MDR)

Authority

CE Mark (EU MDR)

Accepted

Yes

Benefit

A CE certificate is the technical dossier and QMS evidence SAHPRA works from on the reliance pathways.

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

South AfricaSource authority not specified

Prior approval / evidence record

Recognised regulatory authority (reliance guideline)

Authority

Recognised regulatory authority (reliance guideline)

Accepted

Yes

Benefit

Prior approval by a recognised authority is what makes the abridged, recognition, work-sharing and verification pathways available at all, and a device without one falls out of reliance into full evaluation. What it does not buy is a shorter clock, because SAHPRA has not published one for either route.

Timeline Savings

None published — neither the reliance pathways nor full evaluation carries a period

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Prior approval / evidence record

FDA 510(k)/PMA

Authority

FDA 510(k)/PMA

Accepted

Yes

Benefit

A US clearance or approval serves the same role as a recognised-authority approval on the reliance pathways.

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Review clocks

What review, validation, and applicant stop-clock rules apply?

Compare countries

CFS clock and validity

cfs.issue.time.and.validity
Fact-level verified

Where all CFS requirements are met, SAHPRA’s 2026 guideline states issue within 15 working days. Incomplete applications are not finalised until deficiencies are addressed; the applicant must respond within two working days and only two review opportunities are allowed. The certificate is valid for 12 months from issue. This CFS clock does not govern amendment, product registration, NIRMED or corporate-succession decisions.

Conditions
  • Use the current CFS tariff separately; a changed CFS is not automatic approval of the underlying establishment/product change.

Related structured datasets

Classification systems · 1 record
Open Risk classes A through D. →
Id

south-africa:medical-device

Jurisdiction Id

south-africa

Domains
  • medical-device
  • ivd
Name

Risk classes A through D.

Classes
  • Id

    a

    Label

    Class A

    Normalized Risk Level

    low

  • Id

    b

    Label

    Class B

    Normalized Risk Level

    medium

  • Id

    c

    Label

    Class C

    Normalized Risk Level

    high

  • Id

    d

    Label

    Class D

    Normalized Risk Level

    highest

Regulator

SAHPRA

Official Domains
  • sahpra.org.za
Source Status

official-source-verified

Source Title

SAHPRA SAHPGL-MD-04 v5 — Guideline for Classification of Medical Devices and IVDs (Regulation 11(1), GN R1515)

Source Url

https://www.sahpra.org.za/wp-content/uploads/2025/03/SAHPGL-MD-04_v5-Guideline-for-Classification-of-MD-and-IVDs.pdf

Verified On

2026-08-25

Source Supports
  • class-axis
  • classification-rules
  • ivd-class-axis
  • ivd-classification-rules
Notes
  • Regulation 11(1) GN R1515 (GG 40480): Class A low, B low-moderate, C moderate-high, D high, where risk relates to the patient, user or public health; SAHPGL-MD-04 v5 (27 Feb 2025) supplies the technical classification rules.
  • Classification is the manufacturer's/distributor's responsibility using SAHPRA-supplied rules based on intended use, risk, invasiveness and duration; where inconclusive or between classes the higher class applies (reg 11(4)); groups/systems/procedure packs take the highest classification of any component; accessories are classified separately.
  • IVD technical rules: self-testing IVDs Class C unless otherwise specified; specified high-risk purposes/markers Class D; certain listed IVDs Class A; any IVD not mentioned in rules 1-7 defaults to Class B.
Registration pathways, prior approvals and planning · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open South Africa registration →
Country

South Africa

Jurisdiction Id

south-africa

Country Code

ZA

Flag

🇿🇦

Regulator

South African Health Products Regulatory Authority

Regulator Abbrev

SAHPRA

Region

africa

Tier

3

Risk Classification

4-class (A, B, C, D) — EU-aligned (licence-based system)

Classes
  • Id

    establishment-licence

    Class Name

    Medical device establishment licence

    Pathway

    SAHPRA establishment licensing

    Official Timeline
    Min

    6

    Max

    8

    Unit

    weeks

    Realistic Timeline
    Min

    2

    Max

    4

    Unit

    months

    Notes
    • SAHPRA communication MD019 §3: establishment licence applications are processed within 6-8 weeks, and the same figure is repeated in the licensing FAQ
    • An expedited pathway aiming to conclude within 10-15 working days exists but is scoped to devices and IVDs required during the COVID-19 pandemic under MD005, so it is not generally available
    • The clock is conditional on the applicant: a deficiency observation letter must be answered within 10 working days on the standard route or 2 working days on the expedited one, only two response cycles are permitted, and a third round or a missed deadline ends the licensing process with no recommendation
    • One clock covers the whole route and it does not vary by device class
    Gmp Required

    Yes

    Verification Status

    official-verified

    Official Source
    Title

    SAHPRA MD019 v1 (22 July 2020) — processing of medical device establishment licence applications: 6-8 weeks, expedited pathway 10-15 working days, observation-letter responses in 10 working days (2 on the expedited route), maximum two response cycles

    Url

    https://www.sahpra.org.za/wp-content/uploads/2020/07/MD019-Processing-of-licence-applications-v1-22072020.pdf

    Effective Date

    2020-07-22

    Accessed Date

    2026-08-30

  • Id

    establishment-licence-renewal

    Class Name

    Medical device establishment licence renewal

    Official Timeline
    Min

    4

    Max

    8

    Unit

    weeks

    Realistic Timeline
    Min

    2

    Max

    4

    Unit

    months

    Notes
    • SAHPRA licence-renewal Q&A §3.5: processing "may be for a period of at least 4 to 8 weeks", so the lower bound is a floor rather than a promise
    • The renewal application must be filed at least 90 days before the licence expiry date, and licensees whose licence had already lapsed at publication were given a grace period of up to 90 days
    • Product-list changes can ride along with the renewal without a separate amendment as long as the licensed device class does not change; address, company-name, activity, Class D additions and responsible-person changes are amendments that carry an extra fee and can only be filed after the renewed licence has been issued
    Gmp Required

    Yes

    Verification Status

    official-verified

    Official Source
    Title

    SAHPRA Questions & Answers: License Renewal v1 (4 February 2022) — renewal filed at least 90 days before expiry, processing takes at least 4-8 weeks, annual retention fees apply

    Url

    https://www.sahpra.org.za/wp-content/uploads/2022/02/License-Renewal-Q-and-A-to-support-communication-to-industry_Final.pdf

    Effective Date

    2022-02-04

    Accessed Date

    2026-08-30

Key Requirements
  • Local establishment (manufacturer, importer, distributor, or wholesaler) must hold a SAHPRA licence and appoint an authorised representative resident in South Africa
  • Product registration is no longer purely prospective: the Medical Devices Reliance Guideline effective 31 October 2025 sets out abridged, recognition, work-sharing and verification pathways for registering Class B, C and D devices and IVDs against the technical dossier guidelines, and SAHPRA ran a voluntary registration feasibility study in 2024
  • What registration still lacks is a clock. The reliance guideline publishes no processing period, and it points forward to a Guideline on Medical Device Registration Process that will carry the document requirements, so the step that will eventually decide market entry is currently unbounded
  • Class A non-measuring and non-sterile devices, and any device without prior approval from a recognised regulatory authority, fall outside the reliance route and face full evaluation
  • Certificate of Free Sale (CFS) issued within 15 days, valid 1 year
Prior Approvals
  • Authority

    Recognised regulatory authority (reliance guideline)

    Accepted

    Yes

    Benefit

    Prior approval by a recognised authority is what makes the abridged, recognition, work-sharing and verification pathways available at all, and a device without one falls out of reliance into full evaluation. What it does not buy is a shorter clock, because SAHPRA has not published one for either route.

    Timeline Savings

    None published — neither the reliance pathways nor full evaluation carries a period

  • Authority

    CE Mark (EU MDR)

    Accepted

    Yes

    Benefit

    A CE certificate is the technical dossier and QMS evidence SAHPRA works from on the reliance pathways.

    Timeline Savings

    None published

  • Authority

    FDA 510(k)/PMA

    Accepted

    Yes

    Benefit

    A US clearance or approval serves the same role as a recognised-authority approval on the reliance pathways.

    Timeline Savings

    None published

Local Rep Required

Yes

Local Rep Name

South African Licensed Establishment

Mdsap Accepted

Yes

Mdsap Benefit

Accepted as QMS evidence

Registration Validity

Each establishment licence carries its own expiry date printed on the licence; SAHPRA does not publish a single statutory term. Renewal must be filed at least 90 days before that date and annual retention fees are payable in the interim

Last Updated

2026-08-30

Verification Status

partially-verified

Reviewed On

2026-08-30

Official Sources
  • Title

    SAHPRA MD019 v1 (22 July 2020) — processing of medical device establishment licence applications: 6-8 weeks, expedited pathway 10-15 working days, observation-letter responses in 10 working days (2 on the expedited route), maximum two response cycles

    Url

    https://www.sahpra.org.za/wp-content/uploads/2020/07/MD019-Processing-of-licence-applications-v1-22072020.pdf

    Effective Date

    2020-07-22

    Accessed Date

    2026-08-30

  • Title

    SAHPRA Questions & Answers: License Renewal v1 (4 February 2022) — renewal filed at least 90 days before expiry, processing takes at least 4-8 weeks, annual retention fees apply

    Url

    https://www.sahpra.org.za/wp-content/uploads/2022/02/License-Renewal-Q-and-A-to-support-communication-to-industry_Final.pdf

    Effective Date

    2022-02-04

    Accessed Date

    2026-08-30

  • Title

    SAHPRA Medical Devices Reliance Guideline (SAHPGL-MD-19 v1, effective 31 October 2025) — reliance pathways for establishment listing, Class B/C/D product registration, ISO 13485 audits and clinical trials; publishes no processing periods

    Url

    https://www.sahpra.org.za/wp-content/uploads/2025/11/SAHGL-MD-19_v1-MEDICAL-DEVICES-RELIANCE-GUIDELINE-document-for-comment.pdf

    Effective Date

    2025-10-31

    Accessed Date

    2026-08-30

  • Title

    SAHPRA FAQs — medical device establishment licence processed in 6-8 weeks

    Url

    https://www.sahpra.org.za/faqs-medical-devices-licencing/

    Accessed Date

    2026-08-25

Process
Template Id

default

Status

planning-template

Usage Rule

Maintained planning steps and estimates, not verified regulator deadlines.

Steps
  • Step

    1

    Title

    Pre-Submission Preparation

    Description

    Gather technical documentation, quality management system (QMS) records, and clinical data required by SAHPRA. Appoint a South African Licensed Establishment.

    Duration

    1-3 months

    Icon

    📋

  • Step

    2

    Title

    Application Submission

    Description

    Submit complete registration dossier to SAHPRA including device description, risk analysis, and testing reports.

    Duration

    1-2 weeks

    Icon

    📤

  • Step

    3

    Title

    Administrative Review

    Description

    SAHPRA reviews application for completeness. May request additional documents or clarifications.

    Duration

    2-4 weeks

    Icon

    🔍

  • Step

    4

    Title

    Technical Evaluation

    Description

    In-depth technical assessment of device safety and performance. SAHPRA may issue questions.

    Duration

    6-8 weeks

    Icon

    ⚙️

  • Step

    5

    Title

    Registration Approval

    Description

    Upon successful review, SAHPRA issues South Africa market authorization certificate.

    Duration

    1-2 weeks

    Icon

    ✅

Government fee schedules · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open South Africa fees →
Jurisdiction Id

south-africa

Country

South Africa

Country Code

ZA

Authority

SAHPRA

Low Risk Local
Min

15600

Max

26200

Currency

ZAR

Per Product

No

High Risk Local
Min

15600

Max

26200

Currency

ZAR

Per Product

No

Details
  • Id

    new-establishment-licence-manufacturer

    Item

    New medical-device/IVD establishment licence - manufacturer

    Currency

    ZAR

    Amount Local

    26200

    Amount Local Formatted

    ZAR 26,200

    Frequency

    Per establishment licence

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sahpra.org.za/wp-content/uploads/2025/02/Regulations-Regarding-Fees-Payable-In-Terms-Of-The-Provisions-Of-The-Medicines-And-Related-Substances-Act-1965-Act-No.-101-Of-1965.pdf

    Source Title

    South Africa Government Gazette 52106: SAHPRA fees payable

    Source Effective Date

    2025-02-12

    Verified On

    2026-08-25

  • Id

    new-establishment-licence-distributor-wholesaler

    Item

    New medical-device/IVD establishment licence - distributor or wholesaler

    Currency

    ZAR

    Amount Local

    15600

    Amount Local Formatted

    ZAR 15,600

    Frequency

    Per establishment licence

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sahpra.org.za/wp-content/uploads/2025/02/Regulations-Regarding-Fees-Payable-In-Terms-Of-The-Provisions-Of-The-Medicines-And-Related-Substances-Act-1965-Act-No.-101-Of-1965.pdf

    Source Title

    South Africa Government Gazette 52106: SAHPRA fees payable

    Source Effective Date

    2025-02-12

    Verified On

    2026-08-25

  • Id

    renewal-establishment-licence-manufacturer

    Item

    Establishment-licence renewal - manufacturer

    Currency

    ZAR

    Amount Local

    22900

    Amount Local Formatted

    ZAR 22,900

    Frequency

    Per renewal

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sahpra.org.za/wp-content/uploads/2025/02/Regulations-Regarding-Fees-Payable-In-Terms-Of-The-Provisions-Of-The-Medicines-And-Related-Substances-Act-1965-Act-No.-101-Of-1965.pdf

    Source Title

    South Africa Government Gazette 52106: SAHPRA fees payable

    Source Effective Date

    2025-02-12

    Verified On

    2026-08-25

  • Id

    renewal-establishment-licence-distributor

    Item

    Establishment-licence renewal - distributor

    Currency

    ZAR

    Amount Local

    13100

    Amount Local Formatted

    ZAR 13,100

    Frequency

    Per renewal

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sahpra.org.za/wp-content/uploads/2025/02/Regulations-Regarding-Fees-Payable-In-Terms-Of-The-Provisions-Of-The-Medicines-And-Related-Substances-Act-1965-Act-No.-101-Of-1965.pdf

    Source Title

    South Africa Government Gazette 52106: SAHPRA fees payable

    Source Effective Date

    2025-02-12

    Verified On

    2026-08-25

  • Id

    renewal-establishment-licence-wholesaler

    Item

    Establishment-licence renewal - wholesaler

    Currency

    ZAR

    Amount Local

    13100

    Amount Local Formatted

    ZAR 13,100

    Frequency

    Per renewal

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sahpra.org.za/wp-content/uploads/2025/02/Regulations-Regarding-Fees-Payable-In-Terms-Of-The-Provisions-Of-The-Medicines-And-Related-Substances-Act-1965-Act-No.-101-Of-1965.pdf

    Source Title

    South Africa Government Gazette 52106: SAHPRA fees payable

    Source Effective Date

    2025-02-12

    Verified On

    2026-08-25

  • Id

    annual-licence-retention

    Item

    Annual establishment-licence retention

    Currency

    ZAR

    Amount Local

    4400

    Amount Local Formatted

    ZAR 4,400

    Frequency

    Annual, by the last working day of June

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sahpra.org.za/wp-content/uploads/2025/02/Regulations-Regarding-Fees-Payable-In-Terms-Of-The-Provisions-Of-The-Medicines-And-Related-Substances-Act-1965-Act-No.-101-Of-1965.pdf

    Source Title

    South Africa Government Gazette 52106: SAHPRA fees payable

    Source Effective Date

    2025-02-12

    Verified On

    2026-08-25

  • Id

    establishment-licence-amendment

    Item

    Establishment-licence amendment

    Currency

    ZAR

    Amount Local

    5500

    Amount Local Formatted

    ZAR 5,500

    Frequency

    Per amendment

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sahpra.org.za/wp-content/uploads/2025/02/Regulations-Regarding-Fees-Payable-In-Terms-Of-The-Provisions-Of-The-Medicines-And-Related-Substances-Act-1965-Act-No.-101-Of-1965.pdf

    Source Title

    South Africa Government Gazette 52106: SAHPRA fees payable

    Source Effective Date

    2025-02-12

    Verified On

    2026-08-25

  • Id

    licence-issuance-collection

    Item

    Approved licence issuance / collection

    Currency

    ZAR

    Amount Local

    3500

    Amount Local Formatted

    ZAR 3,500

    Frequency

    Per licence

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sahpra.org.za/wp-content/uploads/2025/02/Regulations-Regarding-Fees-Payable-In-Terms-Of-The-Provisions-Of-The-Medicines-And-Related-Substances-Act-1965-Act-No.-101-Of-1965.pdf

    Source Title

    South Africa Government Gazette 52106: SAHPRA fees payable

    Source Effective Date

    2025-02-12

    Verified On

    2026-08-25

  • Id

    certificate-of-free-sale

    Item

    Certificate of Free Sale

    Currency

    ZAR

    Amount Local

    1460

    Amount Local Formatted

    ZAR 1,460

    Frequency

    Per certificate

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sahpra.org.za/wp-content/uploads/2025/02/Regulations-Regarding-Fees-Payable-In-Terms-Of-The-Provisions-Of-The-Medicines-And-Related-Substances-Act-1965-Act-No.-101-Of-1965.pdf

    Source Title

    South Africa Government Gazette 52106: SAHPRA fees payable

    Source Effective Date

    2025-02-12

    Verified On

    2026-08-25

  • Id

    clinical-trial-application

    Item

    Clinical-trial application - safety and efficacy, including devices/IVDs

    Currency

    ZAR

    Amount Local

    33700

    Amount Local Formatted

    ZAR 33,700

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sahpra.org.za/wp-content/uploads/2025/02/Regulations-Regarding-Fees-Payable-In-Terms-Of-The-Provisions-Of-The-Medicines-And-Related-Substances-Act-1965-Act-No.-101-Of-1965.pdf

    Source Title

    South Africa Government Gazette 52106: SAHPRA fees payable

    Source Effective Date

    2025-02-12

    Verified On

    2026-08-25

  • Id

    inspection-charge-out

    Item

    Inspection charge-out rate

    Currency

    ZAR

    Amount Local

    1660

    Amount Local Formatted

    ZAR 1,660

    Frequency

    Per hour, per inspector

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sahpra.org.za/wp-content/uploads/2025/02/Regulations-Regarding-Fees-Payable-In-Terms-Of-The-Provisions-Of-The-Medicines-And-Related-Substances-Act-1965-Act-No.-101-Of-1965.pdf

    Source Title

    South Africa Government Gazette 52106: SAHPRA fees payable

    Source Effective Date

    2025-02-12

    Verified On

    2026-08-25

Notes
  • These are establishment-licensing charges, not a universal per-product registration tariff.
  • South Africa’s device product-registration framework remains route-dependent.
  • Fees are set by GG 52106 GN 5856 of 12 February 2025: the new-licence row is R26,200 for a manufacturer licence and R15,600 for a distributor or wholesale licence.
Last Updated

2026-08-25

Verification Status

official-verified

Official Sources
  • Title

    South Africa Government Gazette 52106: SAHPRA fees payable

    Url

    https://www.sahpra.org.za/wp-content/uploads/2025/02/Regulations-Regarding-Fees-Payable-In-Terms-Of-The-Provisions-Of-The-Medicines-And-Related-Substances-Act-1965-Act-No.-101-Of-1965.pdf

    Effective Date

    2025-02-12

    Accessed Date

    2026-08-25

  • Title

    SAHPRA current medical-device establishment-licensing fee guide

    Url

    https://www.sahpra.org.za/wp-content/uploads/2023/02/SAHPGL-MD-07_v4-Guideline-on-Questions-and-Answers-Licensing-of-Medical-Device-Establishments.pdf

    Effective Date

    2025-05-21

    Accessed Date

    2026-08-25

Reviewed On

2026-08-25

Amount Status

exact-current

Range Available

Yes

Additional certifications and licences · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open South Africa certifications →
Slug

south-africa

Market

South Africa

Confidence

low

Schemes
  • Id

    sahpra-electrical-emc-evidence

    Name

    SAHPRA electrical/EMC evidence

    Category

    Medical Electrical Safety

    Status

    Included In Medical Registration

    Object

    Finished Device

    Source Urls
    • https://www.sahpra.org.za/radiation-control/
    • https://www.sahpra.org.za/wp-content/uploads/2025/03/SAHPGL-MD-04_v5-Guideline-for-Classification-of-MD-and-IVDs.pdf
    Source Binding

    candidate

  • Id

    icasa-type-approval

    Name

    ICASA Type Approval

    Category

    Radio Spectrum

    Status

    Mandatory If Feature

    Object

    Covered Radio Or Terminal Equipment

    Source Urls
    • https://www.icasa.org.za/pages/type-approval
    Source Binding

    candidate

  • Id

    nrcs-letter-of-authority

    Name

    NRCS Letter of Authority

    Category

    General Electrical Components

    Status

    Mandatory If Feature

    Object

    Product Under Compulsory Specification

    Source Urls
    • https://www.nrcs.org.za/business-units/electro-technical
    Source Binding

    candidate

  • Id

    sahpra-radiation-control

    Name

    SAHPRA Radiation Control

    Category

    Ionizing Radiation

    Status

    Mandatory If Feature

    Object

    Radiation Product Importer Or Use Site

    Note

    Hazardous Substances Act 15/1973 licences administered by SAHPRA Radiation Control: Group III electronic generators of ionising radiation (section 3; Regulation R1332 of 1973), Group IV radioactive sources (section 3A; Regulations R246 and R247 of 1993) and listed electronic products under the Schedule for Listed Electronic Products (Regulation R1302 of 1991), with licences for the import, manufacture, sale and use of listed products.

    Source Urls
    • https://www.sahpra.org.za/radiation-control/
    Source Binding

    reviewed

  • Id

    legal-metrology

    Name

    Legal Metrology

    Category

    Legal Metrology

    Status

    Mandatory If Feature

    Object

    Regulated Measuring Instrument

    Note

    Measuring devices must display South African legal units of measurement (or units acceptable to the Authority) per Essential Principle 10; class A devices with a measuring function require conformity-assessment evidence beyond the South African Declaration of Conformity (SAHPGL-MD-04 v5 §3.3).

    Source Binding

    unresolved

Official Sources
  • Authority

    ICASA

    Topic

    type approval

    Url

    https://www.icasa.org.za/pages/type-approval

  • Authority

    NRCS

    Topic

    electrotechnical compulsory specifications and Letter of Authority

    Url

    https://www.nrcs.org.za/business-units/electro-technical

  • Authority

    SAHPRA

    Topic

    radiation control licence categories hazardous substances act

    Url

    https://www.sahpra.org.za/radiation-control/

  • Authority

    SAHPRA

    Topic

    SAHPGL-MD-04 v5 measuring function classification requirements

    Url

    https://www.sahpra.org.za/wp-content/uploads/2025/03/SAHPGL-MD-04_v5-Guideline-for-Classification-of-MD-and-IVDs.pdf

Keep researching this market
View the South Africa resource hub →
Compare regulatory data across all countries →
DJ Fang
Need to apply South Africa's requirements to a product?
Confirm the current SAHPRA pathway, evidence, and local-role details for the actual device before filing.
Pricing