🇭🇺 Hungary IVD regulatory data

Explore the IVD requirements currently available for Hungary, organized by topic and linked to official sources.

22 facts15 dimensions11 official sourceslatest verification 2026-09-15

Changes after registration in Hungary

Source-linked rules for common changes to a registered IVD. Read the scope and conditions with each rule; unverified details are listed separately.

Does changing a device brand name or logo require an update?

A change to the registered brand name and a change to logo artwork can follow different rules. The legal manufacturer, device design and intended purpose remain the same.

A source-verified answer for this specific scenario is not yet published.

How do I update a manufacturer's name without changing the legal entity?

The same legal manufacturer changes its name, potentially alongside the device brand. A transfer to another legal entity is a separate question; permission to combine filings needs its own evidence.

A source-verified answer for this specific scenario is not yet published.

What changes when the manufacturer's office address changes?

An office address change for the same legal entity, potentially alongside name and brand changes. This does not cover moving a factory or manufacturing site, and a combined filing cannot be assumed.

A source-verified answer for this specific scenario is not yet published.

Does a catalogue or reference code change affect registration or UDI?

Check which identifier actually changes. A commercial catalogue, REF or order code is not automatically a device model, registered identifier, GTIN or UDI-DI. Rules for one identifier do not establish the route for another.

A source-verified answer for this specific scenario is not yet published.

Can several changes be filed together?

Changes to brand, manufacturer name, office address and reference codes may occur together, including with a real transfer to another legal manufacturer, registration holder or owner. Assess transfer with each single change, and transfer with brand plus reference, manufacturer name plus reference, or office address plus reference changes: seven transfer combinations in total. Evidence must establish the exact combination and the order of filings and approvals. These combinations and their sequence are not all verified by the rules below; a rule for one change does not prove that a single application, fee or review period covers the combination.

A source-verified answer for this specific scenario is not yet published.

Approval validity & renewal

What expires, what is retained periodically, and when must renewal start?

Compare countries

Authority & regime

Which authority and binding regime govern this product, task, and date?

Compare countries

Competent authority

authority.competent-authority
Fact-level verified

NNGYK (Nemzeti Népegészségügyi és Gyógyszerészeti Központ, National Center for Public Health and Pharmacy), through its Medical Device Department (Orvostechnikai Főosztály, site still branded OGYÉI), is the competent authority for medical devices and IVDs. Duties include market surveillance, NOR registration, EUDAMED data entry, clinical investigation authorization, incident-reporting oversight and free-sale certificate (FSC) issuance.

Conditions
  • Applies to all device classes and to IVDs.

Baseline fact confirmed: site branding remains OGYÉI but the department operates as an NNGYK organizational unit; contact mailboxes use @nngyk.gov.hu (amd@, nor@, fsc@, ifv@, amd.vig@, implantatum@).

Change control

Does the event require internal documentation, notice, prior approval, transfer, or a new application?

Compare countries

Actor data changes within 7 days

change.actor-data-7-days
Fact-level verified

Economic operators must report any change in their registered EUDAMED Actor data within 7 days of occurrence, via cégkapu e-paper referencing the original case number (fee 21,600 Ft, fee table II.5), and must simultaneously make the change in EUDAMED.

Conditions
  • Implements MDR Article 31(4) / IVDR Article 28(4) nationally.

Clinical investigation & performance

Which clinical or IVD performance evidence or study is required and accepted?

Compare countries

Dossier requirements

Which evidence, local deltas, formats, and signatures are required?

Compare countries

EUDAMED Actor registration uploads

dossier.eudamed-actor-documents
Fact-level verified

EUDAMED Actor registration requires uploading a declaration named 'Declaration on Information Security Responsibilities', the LAA designation documents, and for non-EU manufacturers an extract of the authorised-representative contract.

Conditions
  • Uploads are made during Actor-module registration via the EUDAMED webgate with EU Login.

Economic-operator roles

Which manufacturer, representative, holder, applicant, importer, and distributor roles apply?

Compare countries

Fees

What regulator fee applies to this class, route, event, and date?

Compare countries

Fee payment mechanics

fees.payment-mechanics
Fact-level verified

Administrative service fees for fee-charged device/IVD procedures must be transferred to the NNGYK bank account from 31 July 2023; the payment reference must start with 'ORVOSTECHNIKA' followed by the case (APP) number, and proof of payment must be attached to the submission.

Conditions
  • Account 10032000-00290438-00000000, IBAN HU55 1003 2000 0029 0438 0000 0000, BIC HUSTHUHB; fee proof for some procedures goes to amd@nngyk.gov.hu.

Effective from 2023-07-31

Distinct legal rule on payment mechanics; the individual fee amounts are recorded as typed findings under government-fees.

Official forms & portals

Which current form, portal, account, and submission package must be used?

Compare countries

e-ügyintézés submission channel

portals.e-ugyintezes-channel
Fact-level verified

National filings are submitted through Hungarian electronic administration (e-ügyintézés under 2015. évi CCXXII. törvény, using Company Gate/cégkapu); the deviation-permit application document must be completed and sent via e-ügyintézés, with large files via transfer.ogyei.gov.hu.

Conditions
  • Applies to NNGYK device/IVD procedures handled as administrative cases.

Submission portal referenced on the page: https://ogyei.gov.hu/eugyintezes.

Recall, FSCA & market surveillance

Which recall, FSCA, notice, and authority-coordination duties apply?

Compare countries

Registration pathways

Which registration, notification, listing, reliance, or conformity route applies?

Compare countries

Distributor NOR notification before distribution

registration.distributor-nor-notification
Fact-level verified

A distributor established in Hungary must notify the authority (NNGYK) before starting to distribute devices on the Hungarian market; distributors are registered in the national register NOR, not in EUDAMED.

Conditions
  • Applies to distributors with a Hungarian seat; covers devices and, per 4/2009 EüM rendelet 17. § (7), nursing technical devices.

Mandatory EUDAMED Actor registration

registration.eudamed-actor-mandatory
Fact-level verified

Until NNGYK officially declares each EUDAMED module functional, use of the EUDAMED Actor registration module is mandatory in Hungary for the registration of economic operators, fulfilling 4/2009 EüM rendelet 17. § (1), MDR Article 31, 8/2003 ESzCsM rendelet 7/A. § (1) and IVDR Article 27.

Conditions
  • Operators acting in several roles must register separately for each role and receive a different SRN for each.

SRN example format on the page: HU-MF-123456789; registration via the EUDAMED webgate with EU Login; NNGYK issues an official certificate (hatósági bizonyítvány) via cégkapu.

IVD notification and attachments

registration.ivd-notification-legacy
Fact-level verified

IVD notification/registration under 8/2003. (III. 13.) ESzCsM rendelet 7/A. § uses an official notification form with attachments that vary by IVD tier; for general IVDs the manufacturer's declaration of conformity per Annex 3 of the decree must be attached.

Conditions
  • Self-test IVDs (except blood glucose) and List B/List A devices attach designated-body certificates (design documentation, full QA or type-examination variants); the fee must be paid when submitting.

Changes to notified data are submitted on the same form.

Reliance & recognition

Which foreign decisions or evidence can be reused, with what effect and conditions?

Compare countries

Existing records include acceptance, restrictions and evidence reuse. The applicable product scope and conditions are preserved below; a reference does not by itself confer market access.

Hungary approvals and evidence in other markets (2 records)

ParaguayBrazil ANVISA · US FDA · Health Canada · Austria · Belgium · Denmark · Finland · France · Germany · Hungary · Ireland · Italy · Netherlands · New Zealand · Norway · Spain · Sweden · Japan MHLW / PMDA · UK MHRA · Swissmedic · Australia TGA · WHO

Prior approval / evidence record

IVD Annex I reference-country list drives the 30/45/120-business-day clocks

reliance.ivd-anexo-i-reference-country-list
Fact-level verified
Statement

For IVDs, DINAVISA decides within 30 business days for products manufactured nationally or in a country of Annex I, within 45 business days for products manufactured outside Annex I but registered and commercialised in an Annex I country (which must additionally file the foreign registration certificate and free-sale certificate), and within 120 business days for other imports. Annex I names high-surveillance and adequate-surveillance authorities and the full MERCOSUR members; the United States and Japan are on it and Korea is not.

Source Approval

Sanitary registration and free-sale certificate in force in an Annex I country — high-surveillance authorities (Germany, Austria, Belgium, Canada, Denmark, Spain, United States, France, Israel, Italy, Japan, Netherlands, United Kingdom, Sweden, Switzerland) or adequate-surveillance authorities (Australia, Chile, Cuba, Finland, Hungary, Ireland, Luxembourg, Mexico, Norway, New Zealand); Annex I also lists full MERCOSUR members (Argentina, Brazil, Paraguay, Uruguay) and PAHO/WHO reference authorities

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Premarket Review

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • The Annex I list derives from Ley N° 3283/2007 Art. 11 (high and adequate sanitary surveillance).
  • A product made outside Annex I is admitted on a 45-day clock only if it is registered and commercialised in an Annex I country.
  • Korea (MFDS) does not appear in Annex I, so a Korean-only IVD authorisation does not by itself reach the 30/45-day clocks.
  • A Paraguayan sanitary registration is still required in every case.

Source language: Spanish. Translation of the quoted Annex I heading and entries: 'Countries with High and Adequate Sanitary Surveillance Authorities (Law 3283/07 Article 11)' / 'Countries with high sanitary surveillance regulatory authorities: Germany, Austria, Belgium, Canada, Denmark, Spain, United States, France, Israel, Italy, Japan, Netherlands, United Kingdom, Sweden and Switzerland.' / 'Full MERCOSUR members: Argentina, Brazil, Paraguay and Uruguay.' The excerpt is taken from the resolution's own Annex I; Art. 6 supplies the 30- and 45-business-day clocks.

PeruUS FDA · UK MHRA · Health Canada · Japan MHLW / PMDA · Australia TGA · Swissmedic · South Korea MFDS · Germany · France · Netherlands · Spain · Italy · Belgium · Portugal · Ireland · Austria · Hungary · Denmark · Norway · Sweden

Prior approval / evidence record

País de Alta Vigilancia Sanitaria (art. 9 D.S. 016-2011-SA): United States, Canada, Japan, Australia, Switzerland, South Korea, United Kingdom, Germany, France, Netherlands, Spain, Italy, Belgium, Portugal, Ireland, Austria, Hungary, Denmark, Norway, Sweden

Authority

País de Alta Vigilancia Sanitaria (art. 9 D.S. 016-2011-SA): United States, Canada, Japan, Australia, Switzerland, South Korea, United Kingdom, Germany, France, Netherlands, Spain, Italy, Belgium, Portugal, Ireland, Austria, Hungary, Denmark, Norway, Sweden

Accepted

Yes

Benefit

The Peruvian route to relying on a foreign approval is documentary, not procedural. A certificate of free sale or a GMP certificate issued by a high-surveillance authority is accepted directly, and if the issuing authority appears on the Resolución Directoral 169-2014-DIGEMID list the document does not need an apostille. Everything else still goes through DIGEMID certification.

Timeline Savings

None published — the TUPA period is set by class, not by origin

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Prior approvals and evidence for Hungary (2 records)

HungaryUS FDA

Prior approval / evidence record

FDA 510(k)/PMA

Authority

FDA 510(k)/PMA

Accepted

No

Benefit

No substitute for EU MDR conformity assessment; useful as predicate documentation in the technical file

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

Regulatory reference

NMPA certificate does not replace EU conformity assessment

reliance.china-nmpa-no-recognition-route
Fact-level verified
Statement

For ordinary market placement under MDR/IVDR, a Chinese approval does not replace conformity assessment under MDR Article 52 or IVDR Article 48. The applicable EU class and procedure must be established independently; lower-risk devices may self-declare, while other devices require notified-body assessment. This is a finding about the ordinary conformity route, not a ban on Chinese devices or their technical evidence.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

hungary

China Classes In Scope

All NMPA classes (I filing, II, III) -> all EU MDR/IVDR classes

Edge Type

No Reliance Route

Market Access Strength

none

Regulatory Layer

Product Marketing

Local Application Required

null

Target Decision Retained

null

Conditions
  • Determine the local class independently; NMPA Class III is not an automatic EU class assignment.
  • A local regulatory application/decision is not universal: MDR Article 52(7) and IVDR Article 48(10) permit specified self-declaration routes.
  • Clinical, performance and technical evidence must meet the applicable EU requirements; the certificate alone does not replace them.
  • Special-purpose, investigational, custom-made and derogation regimes require separate assessment.

Independent review of current EUR-Lex consolidations supports this bounded ordinary-route conclusion. It does not establish the absence of every special route.

Special access & derogation

Is an emergency, compassionate, custom, humanitarian, or other exceptional path available?

Compare countries

IVD deviation permit (IVDR Art. 54)

derogation.ivd-deviation-permit
Fact-level verified

Where the conditions of IVDR Article 54(1) exist, NNGYK may upon application permit that the device be placed on the market without the conformity assessment procedures of IVDR Article 48 having been carried out (deviation permit under 8/2003. (III. 13.) ESzCsM rendelet 4. § (1)).

Conditions
  • National implementation of the IVDR Article 54 derogation.

Submission communications & appeal

How are deficiencies answered, clocks handled, and decisions escalated or appealed?

Compare countries

Ákr. treatment of Actor data upload

procedure.akr-treatment-eudamed-actor
Fact-level verified

NNGYK treats the EUDAMED Actor data upload as an application (kérelem) under the general administrative procedure code (2016. évi CL. törvény, Ákr.); deficiencies are communicated by decision (végzés) via cégkapu and are remedied by editing the original APP request in EUDAMED.

Conditions
  • Applies to EUDAMED Actor registration handled by NNGYK.

Vigilance & incident reporting

What is reportable, by whom, where, in what format, and by what deadline?

Compare countries

Incident reporting channel and forms

vigilance.reporting-channel-forms
Fact-level verified

Until the EUDAMED vigilance modules become functional, incidents and accidents (including those arising in clinical investigations) are reported to the NNGYK mailbox amd.vig@nngyk.gov.hu stating the case number (iktatószám) and CIV-ID; NNGYK publishes separate reporting forms for manufacturer, distributor and healthcare-provider reporters.

Conditions
  • Applies to national incident reporting to NNGYK.

The separate per-role forms are published on the varatlan_esemeny_jelentes_formanyomtatvanyok page.

Related structured datasets

Classification systems · 0 records

No maintained record.

Registration pathways, prior approvals and planning · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Hungary registration →
Country

Hungary

Jurisdiction Id

hungary

Country Code

HU

Flag

🇭🇺

Regulator

National Public Health and Medical Officer Service

Regulator Abbrev

NNGYK

Region

europe

Tier

3

Risk Classification

4-class (I, IIa, IIb, III) — EU MDR

Classes
  • Id

    class-i

    Class Name

    Class I

    Risk Level

    low

    Pathway

    Self-declaration (no notified body)

    Official Timeline
    Min

    0

    Max

    0

    Unit

    days

    Verification Status

    official-verified

    Official Source
    Title

    EUR-Lex Regulation (EU) 2017/745 (MDR)

    Url

    https://eur-lex.europa.eu/eli/reg/2017/745/oj

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    6

    Max

    9

    Unit

    months

    Notes
    • MDR Art. 52(7): non-sterile, non-measuring, non-reusable-surgical Class I devices are self-declared, so no authority or notified body reviews the file and there is no review clock
    • Elapsed time is Eudamed actor and device registration, which has no guaranteed processing time, plus any national registration duty
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

Key Requirements
  • EU Authorized Representative (for non-EU manufacturers)
  • National registration/notification deadlines follow the general administrative procedure code (Ákr. 2016. CL.)
  • Clinical investigation validation: formal validity reviewed within 10 days (+5 extension), cure 10 days (+20); the procedure lapses without a response; substantive assessment 45 days (+20 with expert body)
  • PMCF studies notified at least 30 days before start; substantial modifications start earliest 38 days after notification (+7)
  • The IVD page still cites the repealed Ket. — treat as outdated; Art. 78 multicentre single-application procedures are not available in Hungary before 2027
Prior Approvals
  • Authority

    FDA 510(k)/PMA

    Accepted

    No

    Benefit

    No substitute for EU MDR conformity assessment; useful as predicate documentation in the technical file

Local Rep Required

Yes

Local Rep Name

EU Authorized Representative

Mdsap Accepted

No

Mdsap Benefit

ISO 13485 accepted as QMS evidence

Registration Validity

5 years (CE certificate)

Last Updated

2026-08-29

Verification Status

partially-verified

Official Sources
  • Title

    Commission Implementing Regulation (EU) 2026/977, Art. 2 — maximum notified-body conformity assessment timelines (30 + 120/90 parallel + 20 days)

    Url

    https://eur-lex.europa.eu/eli/reg_impl/2026/977/oj

    Effective Date

    2027-02-25

    Accessed Date

    2026-08-29

  • Title

    EUR-Lex Regulation (EU) 2017/745 (MDR)

    Url

    https://eur-lex.europa.eu/eli/reg/2017/745/oj

    Accessed Date

    2026-08-25

  • Title

    EUR-Lex MDR consolidated text — Annex IX §5.1 (expert-panel clinical evaluation consultation: 21/60 days) and §5.2(d) (ancillary medicinal-substance opinion: 210 days)

    Url

    https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:02017R0745-20250110

    Accessed Date

    2026-08-29

Reviewed On

2026-08-29

Process
Template Id

default

Status

planning-template

Usage Rule

Maintained planning steps and estimates, not verified regulator deadlines.

Steps
  • Step

    1

    Title

    Pre-Submission Preparation

    Description

    Gather technical documentation, quality management system (QMS) records, and clinical data required by NNGYK. Appoint a EU Authorized Representative.

    Duration

    1-3 months

    Icon

    📋

  • Step

    2

    Title

    Application Submission

    Description

    Submit complete registration dossier to NNGYK including device description, risk analysis, and testing reports.

    Duration

    1-2 weeks

    Icon

    📤

  • Step

    3

    Title

    Administrative Review

    Description

    NNGYK reviews application for completeness. May request additional documents or clarifications.

    Duration

    2-4 weeks

    Icon

    🔍

  • Step

    4

    Title

    Technical Evaluation

    Description

    In-depth technical assessment of device safety and performance. NNGYK may issue questions.

    Duration

    0 days

    Icon

    ⚙️

  • Step

    5

    Title

    Registration Approval

    Description

    Upon successful review, NNGYK issues Hungary market authorization certificate.

    Duration

    1-2 weeks

    Icon

    ✅

Government fee schedules · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Hungary fees →
Jurisdiction Id

hungary

Country

Hungary

Country Code

HU

Authority

NNGYK

Low Risk Local
Min

8100

Max

8100

Currency

HUF

Per Product

No

High Risk Local
Min

97065

Max

97065

Currency

HUF

Per Product

No

Details
  • Id

    eudamed-actor-registration

    Item

    EUDAMED actor registration

    Currency

    HUF

    Amount Local

    97065

    Amount Local Formatted

    HUF 97,065

    Frequency

    Per registration

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://ogyei.gov.hu/eljarasi_dijak

    Source Title

    Hungary OGYEI procedural fee table for medical devices

    Verified On

    2026-08-25

  • Id

    custom-made-manufacturer-registration

    Item

    Custom-made-device manufacturer registration

    Currency

    HUF

    Amount Local

    8100

    Amount Local Formatted

    HUF 8,100

    Frequency

    Per registration

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://ogyei.gov.hu/eljarasi_dijak

    Source Title

    Hungary OGYEI procedural fee table for medical devices

    Verified On

    2026-08-25

  • Id

    cab-designation

    Item

    Conformity-assessment-body designation

    Currency

    HUF

    Amount Local

    167000

    Amount Local Formatted

    HUF 167,000

    Frequency

    Per designation

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://ogyei.gov.hu/eljarasi_dijak

    Source Title

    Hungary OGYEI procedural fee table for medical devices

    Verified On

    2026-08-25

  • Id

    individual-measure-aids-certificate

    Item

    Certificate for aids for individual measures

    Currency

    HUF

    Amount Local

    189000

    Amount Local Min

    189000

    Amount Local Max

    508950

    Amount Local Formatted

    HUF 189,000–508,950

    Frequency

    Per certificate

    Notes

    Tiered by headcount: 189,000 / 276,000 / 508,950.

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://ogyei.gov.hu/eljarasi_dijak

    Source Title

    Hungary OGYEI procedural fee table for medical devices

    Verified On

    2026-08-25

  • Id

    change-notification

    Item

    Notification of changes

    Currency

    HUF

    Amount Local

    21600

    Amount Local Formatted

    HUF 21,600

    Frequency

    Per notification

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://ogyei.gov.hu/eljarasi_dijak

    Source Title

    Hungary OGYEI procedural fee table for medical devices

    Verified On

    2026-08-25

  • Id

    english-language-certificate

    Item

    English-language certificate

    Currency

    HUF

    Amount Local

    47250

    Amount Local Formatted

    HUF 47,250

    Frequency

    Per 60 products

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://ogyei.gov.hu/eljarasi_dijak

    Source Title

    Hungary OGYEI procedural fee table for medical devices

    Verified On

    2026-08-25

  • Id

    deviation-permit

    Item

    Deviation permit

    Currency

    HUF

    Amount Local

    81810

    Amount Local Formatted

    HUF 81,810

    Frequency

    Per permit

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://ogyei.gov.hu/eljarasi_dijak

    Source Title

    Hungary OGYEI procedural fee table for medical devices

    Verified On

    2026-08-25

  • Id

    clinical-performance-application

    Item

    Clinical-investigation / performance-study application

    Currency

    HUF

    Amount Local

    243000

    Amount Local Formatted

    HUF 243,000

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://ogyei.gov.hu/eljarasi_dijak

    Source Title

    Hungary OGYEI procedural fee table for medical devices

    Verified On

    2026-08-25

  • Id

    ionizing-radiation-investigation

    Item

    Clinical investigation of devices with ionizing radiation

    Currency

    HUF

    Amount Local

    397845

    Amount Local Formatted

    HUF 397,845

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://ogyei.gov.hu/eljarasi_dijak

    Source Title

    Hungary OGYEI procedural fee table for medical devices

    Verified On

    2026-08-25

  • Id

    distributor-notification

    Item

    Distributor notification

    Currency

    HUF

    Amount Local

    89100

    Amount Local Formatted

    HUF 89,100

    Frequency

    Per notification

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://ogyei.gov.hu/eljarasi_dijak

    Source Title

    Hungary OGYEI procedural fee table for medical devices

    Verified On

    2026-08-25

  • Id

    article-74-1-notification

    Item

    Notification under MDR Art. 74(1)

    Currency

    HUF

    Amount Local

    99000

    Amount Local Formatted

    HUF 99,000

    Frequency

    Per notification

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://ogyei.gov.hu/eljarasi_dijak

    Source Title

    Hungary OGYEI procedural fee table for medical devices

    Verified On

    2026-08-25

Notes
  • Amounts follow the NNGYK fee schedule as published at the 2026 research date.
Last Updated

2026-08-25

Verification Status

official-verified

Official Sources
  • Title

    Hungary OGYEI procedural fee table for medical devices

    Url

    https://ogyei.gov.hu/eljarasi_dijak

    Accessed Date

    2026-08-25

Reviewed On

2026-08-25

Amount Status

exact-current

Range Available

Yes

Additional certifications and licences · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Hungary certifications →
Slug

hungary

Market

Hungary

Confidence

high

Schemes
  • Id

    lvd-medical-exclusion-mdr-electrical-safety

    Name

    MDR electrical safety in place of the Low Voltage Directive

    Category

    Medical Electrical Safety

    Status

    Included In Medical Registration

    Object

    Finished Device

    Note

    Annex II of Directive 2014/35/EU places electrical equipment for radiology and medical purposes outside the Low Voltage Directive, so there is no separate national electrical-safety certificate for a medical device. Electrical safety is assessed inside the MDR conformity assessment against MDR Annex I, with the harmonised IEC/EN 60601 series carrying the presumption of conformity.

    Source Urls
    • https://eur-lex.europa.eu/eli/dir/2014/35/oj
    Source Binding

    candidate

  • Id

    emc-directive-disapplied-by-mdr

    Name

    MDR electromagnetic compatibility in place of the EMC Directive

    Category

    emc

    Status

    Included In Medical Registration

    Object

    Finished Device

    Note

    Article 2(3) of Directive 2014/30/EU switches the EMC Directive off where other Union legislation lays the same essential requirements down more specifically, which MDR Annex I sections 18.5 and 18.6 do for medical devices. EMC is therefore dossier evidence inside the CE route and not a separate national EMC conformity step.

    Source Urls
    • https://eur-lex.europa.eu/eli/dir/2014/30/oj
    Source Binding

    candidate

  • Id

    nmhh-2-2017-radio-conformity-and-hungarian-language

    Name

    2/2017. (I. 17.) NMHH rendelet on radio equipment, with Hungarian-language duties

    Category

    Radio Spectrum

    Status

    Mandatory If Feature

    Object

    Radio Equipment

    Note

    The Hungarian transposition of Directive 2014/53/EU, consolidated on the Nemzeti Jogszabálytár. 3. § (1) a) imports the safety requirements of the NGM decree on low-voltage electrical products expressly without applying its voltage limit, 3. § (1) b) imports the essential requirements of the NMHH EMC decree, and the same subsection ties radio-equipment design back to the national RoHS and WEEE decrees, so those obligations are not separable from the radio assessment. There is no Hungarian type approval. What is genuinely national is the language layer, and it is the part most often missed: 7. § (8) requires the manufacturer's name and contact address to appear in Hungarian on the equipment, its packaging or the accompanying documentation; 7. § (9) requires the operating instructions and safety information to accompany the equipment in both English and Hungarian; 15. § (3) and (5) require the EU declaration of conformity, and the simplified declaration, to be drawn up in Hungarian; and 7. § (14) lets the market-surveillance authority demand the whole compliance file in Hungarian. The market-surveillance authority is the Hivatal of the Nemzeti Média- és Hírközlési Hatóság, and 10. § (1) makes the distributor verify the Hungarian-language documentation before placing the equipment on the market.

    Source Urls
    • https://njt.jog.gov.hu/jogszabaly/2017-2-20-3H
    Source Binding

    candidate

  • Id

    korm-374-2012-rohs

    Name

    374/2012. (XII. 18.) Korm. rendelet restricting hazardous substances in electrical equipment

    Category

    Environmental And Batteries

    Status

    Evidence Only No Separate Certificate

    Object

    Electrical Electronic Medical Device

    Note

    Consolidated to 22 July 2026 on the Nemzeti Jogszabálytár. 1. § (2) h) excludes active implantable medical devices, and 2. § now defines that term by reference to Regulation (EU) 2017/745 rather than the repealed directives, so the decree is current. The Hungarian-specific dates sit in the transitional text: the DEHP, BBP, DBP and DIBP marketing restriction under 3. § (1) applies to medical devices including in vitro diagnostics, and to monitoring and control instruments including industrial ones, only from 22 July 2021, while the general date is 22 July 2019; and the restriction does not reach medical devices placed on the market up to 21 July 2014, in vitro diagnostics up to 21 July 2016 or industrial monitoring and control instruments up to 21 July 2017. Directive 2011/65/EU Annex I category 8 covers medical devices, Article 4(3) has bound medical devices since 22 July 2014 and in vitro diagnostic medical devices since 22 July 2016, and Article 2(4)(h) excludes active implantable medical devices.

    Source Urls
    • https://njt.jog.gov.hu/jogszabaly/2012-374-20-22
    Source Binding

    candidate

  • Id

    korm-197-2014-weee-scope-and-producer-takeback

    Name

    197/2014. (VIII. 1.) Korm. rendelet on waste electrical and electronic equipment

    Category

    Environmental And Batteries

    Status

    Mandatory If Feature

    Object

    Electrical Electronic Medical Device

    Note

    1. § (2) j) keeps out of scope medical devices and in vitro diagnostic medical devices expected to become infective before the end of their life cycle, together with active implantable medical devices; every other medical device inside the 1. melléklet categories is covered, with equipment defined in 2. § (1) 2. as working at up to 1000 V AC or 1500 V DC. The operative duty for a manufacturer is take-back rather than a certificate: under 4. § (1) the producer under individual compliance, or the concession company under collective compliance, must take back its own waste equipment and equipment of the same nature and function placed on the market by others, and 3. § lets the producer refuse a used device only where it presents a serious risk to the health or physical safety of the person taking it back, in which case the item is treated as waste. Directive 2012/19/EU Article 2(4)(g) excludes devices expected to be infective before end of life and active implantable devices; other medical devices are in scope, and Article 16 requires each Member State to keep a producer register.

    Source Urls
    • https://njt.jog.gov.hu/jogszabaly/2014-197-20-22
    Source Binding

    candidate

  • Id

    korm-445-2012-batteries-regulation-implementation

    Name

    445/2012. (XII. 29.) Korm. rendelet implementing the Batteries Regulation

    Category

    Environmental And Batteries

    Status

    Mandatory If Feature

    Object

    Battery Producer Or Authorised Representative

    Note

    Worth contrasting with neighbours that have not moved: the Hungarian battery decree, consolidated to 18 August 2026, has already been rebuilt around Regulation (EU) 2023/1542 rather than left on the repealed Directive 2006/66/EC. 2. § (2) makes the Regulation's definitions directly applicable inside the decree, 5. § (2) sets the collection duty by reference to Articles 59(3) and 60(3) of the Regulation, 14. § sets the producer's record-keeping content by reference to Article 75, 24/A. § designates the országos hulladékgazdálkodási hatóság as the control and competent authority for Article 69(2)–(5), and 30. § (1a) states in terms that the decree lays down the provisions needed to implement the Regulation. 15. § (1) requires the annual data return to that authority by 20 February following the reference year. Article 11(2)(b) of Regulation (EU) 2023/1542 lets professional medical imaging and radiotherapy devices and in vitro diagnostic medical devices restrict portable-battery replacement to independent professionals, and Article 11(3) lifts the removability duty where continuity of supply is needed for safety or data integrity.

    Source Urls
    • https://njt.jog.gov.hu/jogszabaly/2012-445-20-22
    Source Binding

    candidate

  • Id

    epr-80-2023-registration-before-first-placing

    Name

    80/2023. (III. 14.) Korm. rendelet extended producer responsibility registration

    Category

    Environmental And Batteries

    Status

    Mandatory If Feature

    Object

    Producer Or Authorised Representative

    Note

    This is the single registration gate for a device that is electrical equipment or contains batteries, and its timing is stricter than the neighbouring one-month regimes. 23. § (1) requires the producer to apply to the országos hulladékgazdálkodási hatóság for entry in the register before starting activity with the circular product, declaring whether extended producer responsibility will be met collectively or individually under 8. §. 6. § a) adds a first step before that application: the producer must supply the 3. melléklet point 1 data to the concession company on its electronic platform, which is why practitioners describe the Hungarian entry as a double registration. 23. § (9) gives the authority 15 days to register a conforming application, 23. § (3) lets it register a producer as a collective performer of its own motion where no individual declaration or valid concession contract is produced, and 23. § (8) closes the obvious gap by making the extended producer responsibility fee payable on quantities placed on the market before registration was completed.

    Source Urls
    • https://njt.jog.gov.hu/jogszabaly/2023-80-20-22
    Source Binding

    reviewed

  • Id

    kornyezetvedelmi-termekdij-no-longer-covers-eee-or-batteries

    Name

    Környezetvédelmi termékdíj, which no longer reaches electrical equipment or batteries

    Category

    Environmental And Batteries

    Status

    Mandatory If Feature

    Object

    Producer Or Authorised Representative

    Note

    Recorded because the opposite is widely repeated and would send a manufacturer to the wrong authority. Hungary used to charge a környezetvédelmi termékdíj on electrical and electronic equipment and on batteries alongside the waste duties, but the tax authority's own current information booklet number 48, published 23 January 2026, lists only three categories of fee-liable product: egyéb kőolajtermék, egyéb műanyagtermék and egyéb vegyipari termék. Neither electrical and electronic equipment nor batteries appears anywhere in the fee-liable scope of that booklet. The residual product-fee mechanics still described there, such as the TKORNY notification to NAV within 15 days of starting the activity under Ktdt. 10. § (1), therefore bite only if the device business also places one of those three product groups on the Hungarian market; the waste-side obligation for the device itself now runs through the extended producer responsibility register and the concession company instead.

    Source Urls
    • https://nav.gov.hu/pfile/file?path=/ugyfeliranytu/nezzen-utana/inf_fuz/2026/48.-Kornyezetvedelmi-termekdij-2026.-01.-23
    Source Binding

    candidate

  • Id

    cmo-radiation-opinion-for-clinical-investigations

    Name

    Chief Medical Officer radiation opinion for clinical investigations

    Category

    Ionizing Radiation

    Status

    Mandatory If Feature

    Object

    Ionizing Radiation Device In Clinical Investigation

    Note

    A clinical investigation with a device emitting ionizing radiation additionally requires the radiation documentation and the opinion of the országos tisztifőorvos (Chief Medical Officer) under 235/2009. (X. 20.) Korm. rendelet 3/A. cím and 37/C. § (1)-(2); such applications also carry the distinct II.10 fee (HUF 397,845).

    Source Urls
    • https://ogyei.gov.hu/utmutato_az_mdr_szerinti_klinikai_vizsgalathoz
    Source Binding

    candidate

Official Sources
  • Authority

    EUR-Lex

    Topic

    Low Voltage Directive 2014/35/EU Annex II — electrical equipment for radiology and medical purposes outside scope

    Url

    https://eur-lex.europa.eu/eli/dir/2014/35/oj

  • Authority

    EUR-Lex

    Topic

    EMC Directive 2014/30/EU Article 2(3) — more specific Union legislation displaces the directive

    Url

    https://eur-lex.europa.eu/eli/dir/2014/30/oj

  • Authority

    EUR-Lex

    Topic

    RoHS Directive 2011/65/EU Annex I category 8 and Article 4(3) medical-device application dates

    Url

    https://eur-lex.europa.eu/eli/dir/2011/65/oj

  • Authority

    EUR-Lex

    Topic

    WEEE Directive 2012/19/EU Article 2(4)(g) medical-device exclusions and Article 16 national producer register

    Url

    https://eur-lex.europa.eu/eli/dir/2012/19/oj

  • Authority

    EUR-Lex

    Topic

    Batteries Regulation (EU) 2023/1542 Articles 11 and 55 — medical-device removability derogation and producer register

    Url

    https://eur-lex.europa.eu/eli/reg/2023/1542/oj

  • Authority

    Nemzeti Jogszabálytár (njt.jog.gov.hu)

    Topic

    2/2017. (I. 17.) NMHH rendelet a rádióberendezésekről — 3. §, 7. §, 10. § and 15. §

    Url

    https://njt.jog.gov.hu/jogszabaly/2017-2-20-3H

  • Authority

    Nemzeti Jogszabálytár (njt.jog.gov.hu)

    Topic

    374/2012. (XII. 18.) Korm. rendelet — scope exclusions and the phthalate application dates for medical devices

    Url

    https://njt.jog.gov.hu/jogszabaly/2012-374-20-22

  • Authority

    Nemzeti Jogszabálytár (njt.jog.gov.hu)

    Topic

    197/2014. (VIII. 1.) Korm. rendelet — 1. § (2) j) medical-device exclusions, 2. § definitions and 3.–4. § take-back

    Url

    https://njt.jog.gov.hu/jogszabaly/2014-197-20-22

  • Authority

    Nemzeti Jogszabálytár (njt.jog.gov.hu)

    Topic

    445/2012. (XII. 29.) Korm. rendelet — 2. §, 5. §, 14.–15. §, 24/A. § and 30. § (1a) implementing Regulation (EU) 2023/1542

    Url

    https://njt.jog.gov.hu/jogszabaly/2012-445-20-22

  • Authority

    Nemzeti Jogszabálytár (njt.jog.gov.hu)

    Topic

    80/2023. (III. 14.) Korm. rendelet — 6. § and 23. § extended producer responsibility registration

    Url

    https://njt.jog.gov.hu/jogszabaly/2023-80-20-22

  • Authority

    Nemzeti Adó- és Vámhivatal (NAV)

    Topic

    Információs füzet 48. Környezetvédelmi termékdíj, 23 January 2026 — the fee-liable product list and the TKORNY notification

    Url

    https://nav.gov.hu/pfile/file?path=/ugyfeliranytu/nezzen-utana/inf_fuz/2026/48.-Kornyezetvedelmi-termekdij-2026.-01.-23

  • Authority

    OGYÉI

    Topic

    MDR clinical-investigation guide — ionizing-radiation devices, CMO opinion

    Url

    https://ogyei.gov.hu/utmutato_az_mdr_szerinti_klinikai_vizsgalathoz

Keep researching this market
View the Hungary resource hub →