🇸🇦 Saudi Arabia IVD regulatory data

Explore the IVD requirements currently available for Saudi Arabia, organized by topic and linked to official sources.

12 facts7 dimensions8 official sourceslatest verification 2026-10-21
Explore questions and sources →

Common questions about changing a registered IVD in Saudi Arabia. Each question separates published rules from details still awaiting verification.

Change control

Does the event require internal documentation, notice, prior approval, transfer, or a new application?

Compare countries

Combined changes and sequencing boundaries

combined.mdma.change.boundaries
Fact-level verified

Annex 1 records multiple affected change categories and the manufacturer must classify the complete change. This supports describing all linked brand/name/address/REF changes and their cumulative risk in the relevant MDMA update. It does not establish that a true legal-person transfer, AR/establishment change, Saudi-DI update, CFS or chemical permission can all be processed as one application with one fee or clock. Significant affected-device supply must await SFDA approval; coordinate other records with the accepted manufacturer and representation route.

Conditions
  • For a+d, b+d and c+d map the actual REF field separately.
  • For a+e, b+e, c+e, d+e, a+d+e, b+d+e and c+d+e resolve the successor route and each dependent record; no universal e-first statutory order is asserted.

Legal manufacturer and authorization identity

manufacturer.identity.and.succession
Fact-level verified

The manufacturer is responsible for obtaining MDMA, and the authorization is issued in the manufacturer’s name. Labels must identify the responsible legal manufacturer. A genuine transfer of legal-manufacturer responsibility requires assessment of the successor’s manufacturer identity, updated conformity/technical evidence, MDMA route and representation; a rename of the same company or a change of authorised representative does not by itself establish succession to another legal person’s MDMA.

Conditions
  • The exact amendment/new-application route for a particular foreign legal-person succession and its corporate evidence require SFDA acceptance of those facts; no automatic MDMA inheritance is established.

Annex 1 and change-specific evidence

mdma.change.form.materials
Fact-level verified

Use the Medical Device Change Form (Annex 1) with the affected MDMA number, manufacturer name/address, device name/model, national listing number and risk class. Select every affected category, explain and justify the change and risk assessment, state whether it is implemented or already supplied in Saudi Arabia, classify it, and attach relevant updated technical and label documents with a signed truth/completeness declaration.

Conditions
  • The supporting corporate, conformity and technical evidence must match the actual change; a tick-box form does not authorize every corporate transfer.

Significant and non-significant change deadlines

mdma.change.notification.and.approval
Fact-level verified

Report significant changes within 10 days of the change, update the relevant MDMA on GHAD with Annex 1 and relevant documents, and obtain SFDA approval before trading the affected device. Notify non-significant changes within 30 days: submit Annex 1 and documents by the designated MDMA email, then update GHAD when updating or renewing; alternatively send the prescribed short description first and complete the full notification information within 90 days of the change. These are manufacturer deadlines, not an SFDA decision SLA.

Conditions
  • The current Arabic wording controls; the 90-day alternative concerns completing notification information, not a universal GHAD deadline.

Registered identity changes versus artwork

mdma.change.triggers
Fact-level verified

For devices including IVDs already holding an MDMA, the manufacturer must evaluate, verify and validate each change under its QMS. The significant-change examples include device name/model and manufacturer name/address; artwork, colour, font and label layout alone are non-significant examples. Rebranding that changes the registered device name must therefore be distinguished from a purely cosmetic logo change.

Conditions
  • Confirm the exact approved name/model and actual safety/performance effects; a marketing description alone does not determine classification.

Dossier requirements

Which evidence, local deltas, formats, and signatures are required?

Compare countries

CFS and shipment documents after changes

cfs.and.shipment.change.information
Fact-level verified

Saudi CFS is a separate export service for a licensed manufacturer with MDMA: submit the Annex 6 request/declaration, manufacturer licence and MDMA verification. CFS validity is one year taking MDMA validity into account; Annex 6 requires informing SFDA within 10 calendar days of a change to submitted information. Shipment clearance instead uses Annex 8 and invoice/product/model/part/lot details. A change in a CFS field is not automatically a one-for-one replacement permission or an MDMA ownership transfer.

Conditions
  • CFS is conditional on Saudi manufacturing/export role and requested destination.
  • Chemical/IVD composition and permit-party changes require the separately held REQ5/Annex 2 and current chemical-form/list branch; no blanket exemption is inferred from conflicting English item8.

Economic-operator roles

Which manufacturer, representative, holder, applicant, importer, and distributor roles apply?

Compare countries

Foreign manufacturer representation at transition

authorised.representative.transition
Fact-level verified

A foreign manufacturer’s authorised-representative arrangement is a separate licensed role. On termination the manufacturer must appoint a replacement immediately, transfer relevant obligations and notify SFDA. The outgoing representative’s responsibilities continue until replacement or until the products are no longer marketed or in use, under the implementing regulation. This continuity rule does not transfer legal-manufacturer ownership of MDMA.

Conditions
  • Identify the affected device group, agreement, local licence and responsibilities; do not substitute an importer/account-access change for manufacturer succession.

Fees

What regulator fee applies to this class, route, event, and date?

Compare countries

MDMA change-fee categories

mdma.variation.fee.categories
Fact-level verified

The MDMA fee table’s reduced SAR 5,000 category covers renewal, adding devices or models, changing brand name, updating the technical file, a transition-period letter and minor manufacturer name/address changes. The administrative SAR 1,100 category covers only label-only, IFU-only and advertising-material-only updates. A change of legal manufacturer name/address or risk classification attracts the full risk-class fee: A SAR 15,000, B SAR 19,000, C SAR 21,000 and D SAR 23,000. Classify the actual SFDA request; a same-company description does not by itself decide whether the minor or legal-manufacturer fee category applies.

Conditions
  • The SAR 5,000 reduced merged cell spans Renew through Minor change in manufacturer name/address. The SAR 1,100 administrative merged cell begins at Update label(s) only and ends at Update advertising materials only.
  • The35–60 working-day heading is confined to new applications; this table does not publish a current universal variation-review duration.
  • The URL upload month is not an effective-date clause. Apply the accepted request category at filing; no single combined-case charge is established.

Label, IFU & language

Which label and IFU elements, languages, symbols, and translation controls apply?

Compare countries

Manufacturer identity on labels and import records

manufacturer.label.and.import.consistency
Fact-level verified

The responsible legal manufacturer’s name and address must remain consistent with the authorization and label evidence. Import shipment records and the shipment declaration must identify the product and manufacturer consistently with the approved information; imported labels must correspond to the labels submitted to SFDA. Evaluate a brand, name, address, REF or successor change across those linked documents before affected supply.

Conditions
  • Manufactured-for wording alone does not establish assumption of legal-manufacturer responsibility.
  • Do not infer a general right to relabel or sell through old stock.

Reliance & recognition

Which foreign decisions or evidence can be reused, with what effect and conditions?

Compare countries

Existing records include acceptance, restrictions and evidence reuse. The applicable product scope and conditions are preserved below; a reference does not by itself confer market access.

Saudi Arabia approvals and evidence in other markets (4 records)

BahrainSaudi Arabia · US FDA · Australia TGA · Japan MHLW / PMDA · UK MHRA · Ireland · Swissmedic · France · Germany

Prior approval / evidence record

Registration certificate from SFDA (Saudi Arabia), US FDA, TGA (Australia), PMDA (Japan), MHRA (United Kingdom), HPRA (Ireland), Swissmedic (Switzerland), ANSM (France) or BfArM (Germany)

Authority

Registration certificate from SFDA (Saudi Arabia), US FDA, TGA (Australia), PMDA (Japan), MHRA (United Kingdom), HPRA (Ireland), Swissmedic (Switzerland), ANSM (France) or BfArM (Germany)

Accepted

Yes

Benefit

Article (7) bis of Resolution No. (48) of 2020, inserted by Resolution No. (69) of 2024, provides by way of exception to the ordinary Article (7) prohibition that medical devices and products registered in one of the nine listed reference countries 'shall also be registered in the Kingdom of Bahrain' on the basis of that country's registration certificate, on payment of the applicable fee. This is a documentary registration route rather than a full de novo technical assessment, but the application is still filed by a licensed Bahraini Authorized Representative and NHRA retains the decision.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

BahrainUS FDA · Japan MHLW / PMDA · Saudi Arabia · Australia TGA · UK MHRA · Ireland · Swissmedic · France · Germany

Prior approval / evidence record

Registration certificate from a listed reference country is an alternative basis for Bahraini registration (Art. 7 bis)

reliance.reference-country-registration-certificate-route
Fact-level verified
Statement

Article (7) bis of Resolution No. (48) of 2020, inserted by Resolution No. (69) of 2024, provides that by way of exception to the ordinary Article (7) prohibition, medical devices and products registered in one of nine listed reference countries 'shall also be registered in the Kingdom of Bahrain' on the basis of the registration certificate issued by that country, following payment of the applicable fee. The listed reference authorities are SFDA (Saudi Arabia), FDA (United States), TGA (Australia), PMDA (Japan), MHRA (United Kingdom), HPRA (Ireland), Swissmedic (Switzerland), ANSM (France) and BfArM (Germany). A holder of a US FDA or Japan PMDA registration therefore has a documentary registration route in Bahrain rather than a full de novo technical assessment.

Source Approval

US FDA registration/clearance; Japan PMDA approval; Saudi SFDA registration; Australia TGA registration; UK MHRA registration; Ireland HPRA registration; Swissmedic (Switzerland) registration; France ANSM registration; Germany BfArM registration

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Only the nine authorities named in Article (7) bis are listed; Korea MFDS, Health Canada and China NMPA are not among them
  • The EU CE mark as such is not listed; the Union route appears only through three named national authorities (HPRA, ANSM, BfArM)
  • The application is still made by a Bahraini licensed Authorized Representative and the applicable fee must be paid
  • In force from the day following publication in the Official Gazette (No. 3781, 21 November 2024)
  • Resolution No. (69) of 2024 replaced Article (10) of Resolution No. (48) of 2020 so that marketing authorisation runs for three years

Effective from 2024-11-22

Verbatim source passage: "By way of exception to Article (7) of this Resolution, medical devices and products registered in one of the following reference countries shall also be registered in the Kingdom of Bahrain: 1. Kingdom of Saudi Arabia - Represented by the Saudi Food and Drug Authority (SFDA). 2. United States of America - Represented by the Food and Drug Administration (FDA). 3. Australia - Represented by the Therapeutic Goods Administration (TGA). 4. Japan - Represented by the Pharmaceuticals and Medical Devices Agency (PMDA). 5. United Kingdom - Represented by the Medicines and Healthcare products Regulatory Agency (MHRA). 6. Ireland - Represented by the Health Products Regulatory Authority (HPRA). 7. Switzerland - Represented by Swissmedic(Swissmedic). 8. France - Represented by the French National Agency for Medicines and Health Products Safety (ANSM). 9. Germany - Represented by the Federal Institute for Drugs and Medical Devices (BfArM). Based on the registration certificate issued by one of these countries, following the payment of the applicable fee."

BahrainSaudi Arabia · US FDA · UK MHRA · Australia TGA · Health Canada · Japan MHLW / PMDA · Swissmedic · Ireland · Denmark · New Zealand · France · Netherlands · Belgium

Prior approval / evidence record

Free Sale Certificate may be issued by the country of origin or by a reference country

reliance.free-sale-certificate-origin-or-reference-country
Fact-level verified
Statement

At renewal, Bahrain requires a Free Sale Certificate (FSC) or Certificate to Foreign Government issued by the regulatory authority of the country of origin or of a reference country (Registration Guideline 9.1, section 10 item 4). The guideline's glossary (item 14) defines reference countries as Saudi Arabia, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland and Belgium, so a certificate from one of those authorities substitutes for an origin-country free-sale certificate. A Free Sale Certificate is also required on manufacturer acquisition (section 12).

Source Approval

Free Sale Certificate or Certificate to Foreign Government from the country of origin or from a reference country (Saudi Arabia, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland, Belgium)

Edge Type

Origin Country Evidence

Market Access Strength

Evidence Only

Regulatory Layer

Premarket Review

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • The classification guideline separately accepts a Free Sale Certificate OR Registration Certificate from one of the reference countries as a classification requirement
  • The reference-country list in the registration guideline glossary includes Canada, Denmark, the Netherlands ('Holland') and New Zealand, which are not in the Article (7) bis list of Resolution No. (69) of 2024, and omits Germany
  • Free Sale Certificate is also required on manufacturer acquisition (Section 12)

Verbatim source passages: "Free Sale Certificate (FSC) or Certificate to foreign government issued by the regulatory authority of the country of origin or a reference country." and "Reference countries: Saudi Arabia, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland, Belgium."

BahrainSaudi Arabia · US FDA · UK MHRA · Australia TGA · Health Canada · Japan MHLW / PMDA · Swissmedic · Ireland · Denmark · New Zealand · France · Netherlands · Belgium

Prior approval / evidence record

Classification of a product as a medical device may be evidenced by a registration certificate from a reference country

reliance.reference-country-certificate-for-classification
Fact-level verified
Statement

To obtain an NHRA classification letter, the applicant must supply a Free Sale Certificate or Registration Certificate issued by the competent authority of the country of origin or by one of the reference countries listed in the classification guideline (SFDA, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland, Belgium), together with a recognised quality assurance certificate such as ISO 13485, CE mark or FDA.

Source Approval

Registration certificate or Free Sale Certificate from SFDA (Saudi Arabia), US FDA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland or Belgium

Edge Type

Origin Country Evidence

Market Access Strength

Evidence Only

Regulatory Layer

classification

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • A classification letter is a separate, fee-bearing NHRA service that precedes importation and registration
  • The reference-country list here is broader than the Article (7) bis list in Resolution No. (69) of 2024 and omits Germany and includes Canada, Denmark, the Netherlands and New Zealand
  • For class I products without CE marking a manufacturer's Declaration of Conformity may be provided

Verbatim source passage: "1. Free Sale Certificate OR Registration Certificate issued from competent authority in country of origin classifies the product as a medical device. 2. Free Sale Certificate OR Registration Certificate issued from one of the following reference countries: (SFDA, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, Newzland, France, and Holland, Belgium.)."

Prior approvals and evidence for Saudi Arabia (4 records)

Prior approval / evidence record

SFDA-NMPA regulatory cooperation MoU (2024): no recognition effect

reliance.china-nmpa-bilateral-programme
Pending verification
Statement

SFDA and China's NMPA signed a Memorandum of Understanding on regulatory cooperation in medicines, medical devices and cosmetics at NMPA headquarters in Beijing on 1 November 2024 (SFDA CEO Hisham Aljadhey and NMPA Commissioner Li Li). SFDA describes its scope as exchange of regulatory information on laws and regulations and sharing of regulatory experience and best practices; the Saudi Council of Ministers approved the MoU on 12 May 2025. No implementing measure giving NMPA certificates any reliance or recognition effect in the MDMA process has been published by SFDA as of the check date, so the MoU is a cooperation framework, not a recognition route.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

saudi-arabia

China Classes In Scope

Not class-specific; regulator-to-regulator cooperation covering drugs, medical devices and cosmetics

Edge Type

Cooperation Mou

Market Access Strength

none

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Scope per SFDA: exchange of regulatory information on laws and regulations; sharing of regulatory experiences and best practices (medicines, medical devices, cosmetics)
  • Council of Ministers approval (14/11/1446 AH = 12 May 2025) on the Saudi side
  • No SFDA reliance, abridged-review or inspection-recognition mechanism for NMPA-approved devices found on sfda.gov.sa (EN/AR searches through 2026-09-15)
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Effective from 2024-11-01

Source passage: وقّعت الهيئة العامة للغذاء والدواء مذكرة تفاهم لتبادل الخبرات والمعلومات التنظيمية مع الإدارة الوطنية للمنتجات الطبية (NMPA) في جمهورية الصين الشعبية، وذلك في مقر الأخيرة بالعاصمة بكين. || Areas of cooperation outlined in the MoU regarding medicines, medical devices, and cosmetics include exchanging regulatory information pertaining to laws and regulations, as well as sharing regulatory experiences and best practices.

Regulatory reference

China NMPA approval: no recognition regime

reliance.china-nmpa-no-recognition-route
Pending verification
Statement

Under the Medical Devices Law (Royal Decree M/54) framework, the SFDA Medical Devices Marketing Authorization (MDMA) is a standalone technical-file assessment: MDS-REQ 1 v6.0 requires conformity with the Essential Principles (Annexes 1-2), technical documentation (Annexes 3-4) and an ISO 13485:2016 QMS, and neither lists reference regulators nor accepts any foreign marketing approval as a basis for authorization. The GHTF-founding-member basis (Australia, Canada, Japan, USA, EU/EFTA) of the pre-2021 Interim Regulation guidance MDS-G5 no longer appears in the current requirement. A China NMPA registration certificate therefore confers no abridged, verification or recognition route, and China was never a listed reference jurisdiction.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

saudi-arabia

China Classes In Scope

All NMPA classes (Class I filing, Class II, Class III); SFDA MDMA covers Saudi Classes A-D and IVDs

Edge Type

No Reliance Route

Market Access Strength

none

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Applies to the MDMA route for all device classes (A, B, C, D) and IVDs
  • Applicant must be a licensed local manufacturer or licensed Authorized Representative of the overseas manufacturer
  • Technical documentation per Annex 3/4 and ISO 13485:2016 QMS required; SFDA may request full technical documentation
  • The SFDA 'Verification and Abridged' pathway (reliance on reference agencies) is a Drug Sector process for medicines only, not medical devices
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Effective from 2021-12-19

Source passage: Medical device may be made available in return for payment or free of charge, with a view to distribution and/or use within the KSA only if it is registered at SFDA and obtaining medical devices marketing authorization (MDMA). 1 General Medical device shall comply with the "Essential Principles of Safety and Performance" specified in Annex (1) and Annex (2). 2 Medical device manufacture shall: - Prepare, hold and update the "Medical Device Technical Documentation" and/or "IVD Technical Documentation" that confirm to "Essential Principles of Safety and Performance" specified in Annex (3) and Annex (4) - Establish, document and maintain an effective quality management system (QMS) according to the international ISO standard (ISO 13485:2016) or any identical adopted standard for the same issue/version.

Prior approval / evidence record

FDA 510(k)

Authority

FDA 510(k)

Accepted

No

Benefit

GHTF route canceled Jan 2022 - data helps but no expedited path

Timeline Savings

None

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Saudi ArabiaEU conformity assessment

Prior approval / evidence record

EU CE Mark

Authority

EU CE Mark

Accepted

No

Benefit

GHTF route canceled Jan 2022

Timeline Savings

None

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

UDI, listing & traceability

Which UDI, device listing, database, and traceability duties apply?

Compare countries

UDI versus catalogue/REF changes

udi.brand.model.and.catalogue
Fact-level verified

A new UDI-DI is required for a change that can cause misidentification or ambiguity in traceability, including the listed brand/trade-name and version/model changes. Catalogue number, version/model and primary UDI-DI are separate Saudi-DI data fields. An orderable REF/SMN change is therefore assessed by its actual field and device effect; it is not automatically a model or UDI-DI change. For changes not requiring a new UDI-DI, update the relevant Saudi-DI record within 10 working days.

Conditions
  • Apply the UDI-DI packaging and identity rules in addition to MDMA change control.
  • No universal old-stock grace period follows from these field-update rules.

Related structured datasets

Classification systems · 1 record
Open Risk classes A through D, with device-type-specific handling. →
Id

saudi-arabia:medical-device

Jurisdiction Id

saudi-arabia

Domains
  • medical-device
  • ivd
Name

Risk classes A through D, with device-type-specific handling.

Classes
  • Id

    a

    Label

    Class A

    Normalized Risk Level

    low

  • Id

    b

    Label

    Class B

    Normalized Risk Level

    medium

  • Id

    c

    Label

    Class C

    Normalized Risk Level

    high

  • Id

    d

    Label

    Class D

    Normalized Risk Level

    highest

Regulator

SFDA

Official Domains
  • sfda.gov.sa
Source Status

official-source-verified

Source Title

SFDA Medical Device System (MDSysExcE), Art. 21

Source Url

https://www.sfda.gov.sa/sites/default/files/2023-10/MDSysExcE.pdf

Verified On

2026-08-25

Source Supports
  • class-axis
  • classification-rules
Notes
  • Art. 21.1: Class A low, B low-to-medium, C medium-to-high, D high; Art. 21.5: the manufacturer classifies per SFDA-published rules and SFDA may refuse an unjustified classification claim.
  • 22 MDR-harmonized classification rules in MDS-REQ1 Annex (5) (non-invasive, invasive, active, special): Rule 1 defaults non-invasive devices to class A; Rule 6 classifies surgically invasive transient-use devices as class B subject to exceptions.
  • IVDs are classified under 7 separate rules in Annex (5); the default/catch-all is class B (Rule 6) and controls without assigned value are class B (Rule 7); Annex 5 Table 1 lists 6 categories of products without a medical purpose treated as devices.
  • Mechanics (Art. 21.2-21.3): highest class when multiple rules apply; devices and accessories classified separately; software classified by the intended purpose of the device it influences or standalone by the rules; procedure packs take the highest classification of their components.
Registration pathways, prior approvals and planning · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Saudi Arabia registration →
Country

Saudi Arabia

Jurisdiction Id

saudi-arabia

Country Code

SA

Flag

🇸🇦

Regulator

Saudi Food and Drug Authority

Regulator Abbrev

SFDA

Region

middle-east

Tier

2

Risk Classification

4-class (A, B, C, D)

Classes
  • Id

    class-a

    Class Name

    Class A

    Risk Level

    low

    Pathway

    MDMA Registration (Technical File Assessment)

    Official Timeline
    Min

    35

    Max

    35

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    SFDA FAQ — MDMA issued within 35 business days

    Url

    https://www.sfda.gov.sa/en/faq/what-are-fees-marketing-authorization-application-and-how-long-does-it-take-issue

    Accessed Date

    2026-08-30

    Realistic Timeline
    Min

    1.5

    Max

    3

    Unit

    months

    Notes
    • SFDA publishes one lead time for every fee group in its MDMA evaluation fees table: 35 working days, from the lowest group (all Class I and general IVDs) to the highest. The FAQ repeats the same 35-business-day maximum without distinguishing risk class
    • There is no separate low-risk listing route any more. The Medical Device National Registry, which let non-sterile non-measuring Class A devices be listed without a technical file or an Authorized Representative, was cancelled in September 2022
    • What Class A still gets is a lighter dossier, not a lighter process: MDS-REQ 1 item 4 cuts the submission for low-risk devices that are not IVDs, sterile, measuring, reusable surgical instruments or novel down to six elements — device description and specification, information provided by the manufacturer, essential principles checklist, evidence of compliance, risk management file, and post-market surveillance plan and report
    • The abbreviated dossier is conditional: SFDA may demand the full technical documentation at any point, including after the MDMA has been granted, and it must be supplied within 10 days
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    No

  • Id

    class-b

    Class Name

    Class B

    Risk Level

    medium

    Pathway

    MDMA Registration

    Official Timeline
    Min

    35

    Max

    35

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    SFDA FAQ — MDMA issued within 35 business days

    Url

    https://www.sfda.gov.sa/en/faq/what-are-fees-marketing-authorization-application-and-how-long-does-it-take-issue

    Accessed Date

    2026-08-30

    Realistic Timeline
    Min

    1.5

    Max

    3

    Unit

    months

    Notes
    • SFDA FAQ: the marketing authorization (MDMA) is issued within 35 business days as a maximum if all requirements are fulfilled
    • Full technical file required since Jan 2022
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

  • Id

    class-c

    Class Name

    Class C

    Risk Level

    high

    Pathway

    MDMA Registration

    Official Timeline
    Min

    35

    Max

    35

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    SFDA FAQ — MDMA issued within 35 business days

    Url

    https://www.sfda.gov.sa/en/faq/what-are-fees-marketing-authorization-application-and-how-long-does-it-take-issue

    Accessed Date

    2026-08-30

    Realistic Timeline
    Min

    3

    Max

    5

    Unit

    months

    Notes
    • SFDA FAQ: the marketing authorization (MDMA) is issued within 35 business days as a maximum if all requirements are fulfilled
    • Clinical data may be required
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

  • Id

    class-d

    Class Name

    Class D

    Risk Level

    highest

    Pathway

    MDMA Registration

    Official Timeline
    Min

    35

    Max

    35

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    SFDA FAQ — MDMA issued within 35 business days

    Url

    https://www.sfda.gov.sa/en/faq/what-are-fees-marketing-authorization-application-and-how-long-does-it-take-issue

    Accessed Date

    2026-08-30

    Realistic Timeline
    Min

    4

    Max

    6

    Unit

    months

    Notes
    • SFDA FAQ: the marketing authorization (MDMA) is issued within 35 business days as a maximum if all requirements are fulfilled
    • Extensive clinical data required
    Clinical Trial Required

    Yes

    Local Testing Required

    No

    Gmp Required

    Yes

Prior Approvals
  • Authority

    FDA 510(k)

    Accepted

    No

    Benefit

    GHTF route canceled Jan 2022 - data helps but no expedited path

    Timeline Savings

    None

  • Authority

    EU CE Mark

    Accepted

    No

    Benefit

    GHTF route canceled Jan 2022

    Timeline Savings

    None

Key Requirements
  • Saudi Authorized Representative
  • Complete Technical File (since Jan 2022)
  • Arabic labeling
  • ISO 13485 certification
  • One route for everything: MDMA-2, the Technical File Assessment, has been the only live pathway since January 2022, after the GHTF reliance route closed at the end of 2021 and the low-risk national registry closed in September 2022
  • Filed through the GHAD unified electronic system by a licensed local manufacturer or by the Authorized Representative, who must hold their own GHAD licence before the MDMA application can be lodged (MDS-REQ 1 item 4)
  • Clinical trial approval granted no later than 60 days when requirements are met (MDS-ExcE Art. 7.5(1))
  • Advertising review: 1st Track (within the MDMA application) 35 days; 2nd Track (separate) 10 days; missing documents must be provided within 90 days of the incompleteness notification or the application is VOID
  • Post-approval reporting clocks: significant changes must reach SFDA through the electronic system within 10 days, non-significant changes within 30 days (MDS-REQ 1 item 5)
  • Renewal is filed before expiry and the request can be submitted from 90 days before the expiry date (MDS-REQ 1 item 6)
Local Rep Required

Yes

Local Rep Name

Saudi Authorized Representative

Mdsap Accepted

No

Mdsap Benefit

Not accepted - GHTF route was canceled January 2022

Registration Validity

3 years for Class A non-sterile non-measuring devices; for all other classes the MDMA runs to the remaining validity of the underlying reference-market certificate, or 3 years where that certificate has no defined validity

Verification Status

official-verified

Official Sources
  • Title

    SFDA FAQ — MDMA issued within 35 business days

    Url

    https://www.sfda.gov.sa/en/faq/what-are-fees-marketing-authorization-application-and-how-long-does-it-take-issue

    Accessed Date

    2026-08-30

  • Title

    SFDA MDMA evaluation fees table — 35 working day lead time for every fee group, MDMA validity 3 years

    Url

    https://www.sfda.gov.sa/sites/default/files/2019-10/Fees-Rev-Times-Mark-Aut-App-ar%2Cen_9.pdf

    Accessed Date

    2026-08-30

  • Title

    SFDA MDS-REQ 1 V6 — Requirements for Medical Devices Marketing Authorization

    Url

    https://www.sfda.gov.sa/sites/default/files/2021-12/REQ1En_0.pdf

    Accessed Date

    2026-08-30

  • Title

    SFDA MDS-ExcE — clinical trial approval no later than 60 days (Art. 7.5(1))

    Url

    https://www.sfda.gov.sa/sites/default/files/2023-10/MDSysExcE.pdf

    Accessed Date

    2026-08-25

Reviewed On

2026-08-30

Last Updated

2026-08-30

Process
Template Id

default

Status

planning-template

Usage Rule

Maintained planning steps and estimates, not verified regulator deadlines.

Steps
  • Step

    1

    Title

    Pre-Submission Preparation

    Description

    Gather technical documentation, quality management system (QMS) records, and clinical data required by SFDA. Appoint a Saudi Authorized Representative.

    Duration

    1-3 months

    Icon

    📋

  • Step

    2

    Title

    Application Submission

    Description

    Submit complete registration dossier to SFDA including device description, risk analysis, and testing reports.

    Duration

    1-2 weeks

    Icon

    📤

  • Step

    3

    Title

    Administrative Review

    Description

    SFDA reviews application for completeness. May request additional documents or clarifications.

    Duration

    2-4 weeks

    Icon

    🔍

  • Step

    4

    Title

    Technical Evaluation

    Description

    In-depth technical assessment of device safety and performance. SFDA may issue questions.

    Duration

    35 working days

    Icon

    ⚙️

  • Step

    5

    Title

    Registration Approval

    Description

    Upon successful review, SFDA issues Saudi Arabia market authorization certificate.

    Duration

    1-2 weeks

    Icon

    ✅

Government fee schedules · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Saudi Arabia fees →
Country

Saudi Arabia

Jurisdiction Id

saudi-arabia

Country Code

SA

Authority

SFDA

Low Risk Local
Min

15000

Max

15000

Currency

SAR

Per Product

Yes

High Risk Local
Min

19000

Max

23000

Currency

SAR

Per Product

Yes

Details
  • Id

    application-assessment-class-a

    Item

    Application Assessment - Class A

    Currency

    SAR

    Amount Local

    15000

    Amount Local Formatted

    SAR 15,000

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sfda.gov.sa/sites/default/files/2024-12/RISK%20CLASSES_0.pdf

    Source Title

    SFDA medical-device marketing-authorization evaluation fees by risk class

    Verified On

    2026-08-25

  • Id

    application-assessment-class-b

    Item

    Application Assessment - Class B

    Currency

    SAR

    Amount Local

    19000

    Amount Local Formatted

    SAR 19,000

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sfda.gov.sa/sites/default/files/2024-12/RISK%20CLASSES_0.pdf

    Source Title

    SFDA medical-device marketing-authorization evaluation fees by risk class

    Verified On

    2026-08-25

  • Id

    application-assessment-class-c

    Item

    Application Assessment - Class C

    Currency

    SAR

    Amount Local

    21000

    Amount Local Formatted

    SAR 21,000

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sfda.gov.sa/sites/default/files/2024-12/RISK%20CLASSES_0.pdf

    Source Title

    SFDA medical-device marketing-authorization evaluation fees by risk class

    Verified On

    2026-08-25

  • Id

    application-assessment-class-d

    Item

    Application Assessment - Class D

    Currency

    SAR

    Amount Local

    23000

    Amount Local Formatted

    SAR 23,000

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sfda.gov.sa/sites/default/files/2024-12/RISK%20CLASSES_0.pdf

    Source Title

    SFDA medical-device marketing-authorization evaluation fees by risk class

    Verified On

    2026-08-25

  • Id

    establishment-licence-manufacturer

    Item

    Establishment licence - medical-device manufacturer

    Currency

    SAR

    Amount Local

    5000

    Amount Local Formatted

    SAR 5,000

    Frequency

    Per 5-year licence

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sfda.gov.sa/sites/default/files/2025-03/MDS-REQ%209_En_V2.pdf

    Source Title

    SFDA MDS-REQ9 v2.0 Annex 2: establishment licence periods and fees

    Verified On

    2026-08-25

  • Id

    establishment-licence-authorised-representative

    Item

    Establishment licence - authorised representative

    Currency

    SAR

    Amount Local

    2600

    Amount Local Formatted

    SAR 2,600

    Frequency

    Per year (1–10 years, per contract)

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sfda.gov.sa/sites/default/files/2025-03/MDS-REQ%209_En_V2.pdf

    Source Title

    SFDA MDS-REQ9 v2.0 Annex 2: establishment licence periods and fees

    Verified On

    2026-08-25

  • Id

    establishment-licence-distributor-class-a

    Item

    Establishment licence - distributor/importer Class A

    Currency

    SAR

    Amount Local

    25000

    Amount Local Formatted

    SAR 25,000

    Frequency

    Per year

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sfda.gov.sa/sites/default/files/2025-03/MDS-REQ%209_En_V2.pdf

    Source Title

    SFDA MDS-REQ9 v2.0 Annex 2: establishment licence periods and fees

    Verified On

    2026-08-25

  • Id

    establishment-licence-distributor-class-b

    Item

    Establishment licence - distributor/importer Class B

    Currency

    SAR

    Amount Local

    15000

    Amount Local Formatted

    SAR 15,000

    Frequency

    Per year

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sfda.gov.sa/sites/default/files/2025-03/MDS-REQ%209_En_V2.pdf

    Source Title

    SFDA MDS-REQ9 v2.0 Annex 2: establishment licence periods and fees

    Verified On

    2026-08-25

  • Id

    establishment-licence-distributor-class-c

    Item

    Establishment licence - distributor/importer Class C

    Currency

    SAR

    Amount Local

    8000

    Amount Local Formatted

    SAR 8,000

    Frequency

    Per year

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sfda.gov.sa/sites/default/files/2025-03/MDS-REQ%209_En_V2.pdf

    Source Title

    SFDA MDS-REQ9 v2.0 Annex 2: establishment licence periods and fees

    Verified On

    2026-08-25

  • Id

    establishment-licence-distributor-class-d

    Item

    Establishment licence - distributor/importer Class D

    Currency

    SAR

    Amount Local

    5000

    Amount Local Formatted

    SAR 5,000

    Frequency

    Per year

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sfda.gov.sa/sites/default/files/2025-03/MDS-REQ%209_En_V2.pdf

    Source Title

    SFDA MDS-REQ9 v2.0 Annex 2: establishment licence periods and fees

    Verified On

    2026-08-25

  • Id

    establishment-licence-warehouse

    Item

    Establishment licence - warehouse

    Currency

    SAR

    Amount Local

    800

    Amount Local Formatted

    SAR 800

    Frequency

    Per year (1–5 years; SAR 800 each additional year)

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sfda.gov.sa/sites/default/files/2025-03/MDS-REQ%209_En_V2.pdf

    Source Title

    SFDA MDS-REQ9 v2.0 Annex 2: establishment licence periods and fees

    Verified On

    2026-08-25

  • Id

    advertising-approval-lay-persons

    Item

    Advertising approval - directed to lay persons

    Currency

    SAR

    Amount Local

    3000

    Amount Local Formatted

    SAR 3,000

    Frequency

    Per approval

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sfda.gov.sa/sites/default/files/2026-07/MDS-REQ%208_En_V4.pdf

    Source Title

    SFDA MDS-REQ8 v4.0: advertising approval fees

    Source Effective Date

    2026-07-20

    Verified On

    2026-08-25

  • Id

    advertising-approval-practitioners

    Item

    Advertising approval - directed to health practitioners

    Currency

    SAR

    Amount Local

    6000

    Amount Local Formatted

    SAR 6,000

    Frequency

    Per approval

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.sfda.gov.sa/sites/default/files/2026-07/MDS-REQ%208_En_V4.pdf

    Source Title

    SFDA MDS-REQ8 v4.0: advertising approval fees

    Source Effective Date

    2026-07-20

    Verified On

    2026-08-25

Notes
  • SFDA marketing-authorization assessment fee by risk class (fee circular effective 28 February 2016, still linked by the current SFDA FAQ).
  • Establishment-licence periods and fees follow MDS-REQ9 v2.0 Annex 2; upgrading to a higher class costs the fee difference.
  • Advertising approvals follow MDS-REQ8 v4.0 (20 July 2026).
  • UDI, testing, and commercial service costs are excluded.
Last Updated

2026-08-25

Verification Status

official-verified

Official Sources
  • Title

    SFDA medical-device marketing-authorization evaluation fees by risk class

    Url

    https://www.sfda.gov.sa/sites/default/files/2024-12/RISK%20CLASSES_0.pdf

    Accessed Date

    2026-08-25

  • Title

    SFDA MDS-REQ9 v2.0 Annex 2: establishment licence periods and fees

    Url

    https://www.sfda.gov.sa/sites/default/files/2025-03/MDS-REQ%209_En_V2.pdf

    Accessed Date

    2026-08-25

  • Title

    SFDA MDS-REQ8 v4.0: advertising approval fees

    Url

    https://www.sfda.gov.sa/sites/default/files/2026-07/MDS-REQ%208_En_V4.pdf

    Effective Date

    2026-07-20

    Accessed Date

    2026-08-25

Reviewed On

2026-08-25

Amount Status

exact-current

Range Available

Yes

Additional certifications and licences · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Saudi Arabia certifications →
Slug

saudi-arabia

Market

Saudi Arabia

Confidence

low

Schemes
  • Id

    sfda-electrical-emc-evidence

    Name

    SFDA electrical/EMC evidence

    Category

    Medical Electrical Safety

    Status

    Included In Medical Registration

    Object

    Finished Device

    Note

    MDMA prerequisite: QMS per ISO 13485:2016 or an identical adopted standard of the same issue (REQ1); risk management per SFDA.MD/ISO 14971 (Essential Principle 3(b)); labelling symbols per SFDA.MD/ISO 15223-1 with unrecognised symbols explained in the IFU; establishment QMS certificates per Saudi Standard SFDA.MD/GSO ISO 13485:2017 from SFDA-designated Conformity Assessment Bodies (Implementing Regulation Arts. 21-23) or an SFDA inspection report; licensed clinical-trial establishments comply with SFDA.MD/ISO 14155 and SFDA.MD/ISO 20916.

    Source Urls
    • https://www.sfda.gov.sa/sites/default/files/2021-12/REQ1En_0.pdf
    • https://www.sfda.gov.sa/sites/default/files/2023-10/MDSysExcE.pdf
    Source Binding

    candidate

  • Id

    cst-equipment-approval-licensing-and-coc

    Name

    CST equipment approval/licensing and CoC

    Category

    Radio Spectrum

    Status

    Mandatory If Feature

    Object

    Covered Telecommunications Equipment

    Source Binding

    unresolved

  • Id

    saber-product-certificate-of-conformity

    Name

    SABER Product Certificate of Conformity

    Category

    Import Conformity

    Status

    Mandatory If Feature

    Object

    Regulated Product Or Component

    Source Urls
    • https://saber.sa/Home/index/
    Source Binding

    candidate

  • Id

    saber-shipment-certificate-of-conformity

    Name

    SABER Shipment Certificate of Conformity

    Category

    Import Conformity

    Status

    Mandatory At Import

    Object

    Regulated Shipment

    Source Urls
    • https://saber.sa/Home/index/
    Source Binding

    candidate

  • Id

    saso-iecee-recognition

    Name

    SASO IECEE Recognition

    Category

    General Electrical Components

    Status

    Mandatory If Feature

    Object

    Listed Electrotechnical Product

    Source Binding

    unresolved

  • Id

    halal-certification

    Name

    Halal certification

    Category

    Halal And Material Origin

    Status

    Voluntary Or Procurement

    Object

    Eligible Product Or Claim

    Source Binding

    unresolved

Official Sources
  • Authority

    CST

    Topic

    communications equipment licensing

    Url

    https://www.cst.gov.sa/services/wireless-equipment-licensing

  • Authority

    SASO

    Topic

    SABER product conformity

    Url

    https://saso.gov.sa/en/eservices/pages/maineservicesdetails.aspx?serviceid=396

  • Authority

    SABER

    Topic

    product and shipment conformity portal

    Url

    https://saber.sa/Home/index/

  • Authority

    SFDA

    Topic

    MDMA requirements ISO 13485 QMS ISO 14971 risk management and ISO 15223-1 symbols

    Url

    https://www.sfda.gov.sa/sites/default/files/2021-12/REQ1En_0.pdf

  • Authority

    SFDA

    Topic

    establishment licensing SFDA.MD/GSO ISO 13485 CAB certificates

    Url

    https://www.sfda.gov.sa/sites/default/files/2023-10/MDSysExcE.pdf

  • Authority

    SFDA

    Topic

    clinical trial verification establishments ISO 14155 and ISO 20916

    Url

    https://www.sfda.gov.sa/sites/default/files/2025-03/MDS-REQ%209_En_V2.pdf

Keep researching this market
View the Saudi Arabia resource hub →
Compare regulatory data across all countries →
DJ Fang
Need to apply Saudi Arabia's requirements to a product?
Confirm the current SFDA pathway, evidence, and local-role details for the actual device before filing.