🇦🇪 United Arab Emirates IVD regulatory data

Explore the IVD requirements currently available for United Arab Emirates, organized by topic and linked to official sources.

18 facts9 dimensions8 official sourceslatest verification 2026-10-21
Explore questions and sources →

Common questions about changing a registered IVD in United Arab Emirates. Each question separates published rules from details still awaiting verification.

Approval validity & renewal

What expires, what is retained periodically, and when must renewal start?

Compare countries

An expired MA does not itself permit ongoing activity

ivd-expired-ma-activity-limit
Fact-level verified

The marketing-right holder cannot continue the activity authorized by an ordinary MA after expiry until renewal unless EDE permits continued activity for particular medical products where necessary. A pending variation or renewal application is not itself that permission.

Conditions
  • Ordinary MA clause; conditional approvals have separate Art 10 rules.

Expired MA: retain the active-renewal exception

minor-variation-eligibility-pending-renewal-exception
Fact-level verified

The current amendment service rejects minor-variation requests after the MA expires unless a renewal application is active. Submit documents on the company’s official letterhead carrying its name/logo, and evidence of competent origin-country approval of the proposed changes where applicable. The current active-renewal exception qualifies the older guide’s unexpired-product rule; it does not independently authorize continued supply after MA expiry.

Site variation does not reset the site certificate term

site-variation-preserves-certificate-expiry
Fact-level verified

The manufacturer/site workflow guide states that the site registration certificate is valid for five years and a site variation does not change its expiry date. Keep renewal and variation separate; a site-data amendment does not itself extend a product MA.

Change control

Does the event require internal documentation, notice, prior approval, transfer, or a new application?

Compare countries

An IVD product MA transfer needs EDE approval

ivd-product-ma-transfer-and-separate-import-permit
Fact-level verified

EDE may approve transfer of a medical-product MA or conditional MA to another party under the conditions/cases set by the Executive Regulation. Import/export/re-export permissions themselves cannot be transferred; apply for a new permission when the marketing-right holder or importing/exporting establishment changes.

Conditions
  • A transfer to a different legal person requires EDE approval under the applicable Executive Regulation conditions and cases; an editable holder-name or manufacturer/site record does not itself grant that transfer.
  • The Article 17(2) transfer permission covers an ordinary MA or conditional MA and does not expressly extend to emergency-use approval.

Fundamental IVD changes may require a new MA

ivd-significant-change-new-ma-triggers
Fact-level verified

An existing MA/conditional MA ceases to support the product and a new application is required for the applicable Art 15 triggers, including fundamental medical-device design changes, substantial changes to manufacturing/production that may affect quality/safety/effectiveness, or applicable composition/classification/use-route changes. This is distinct from a same-entity administrative rename.

Conditions
  • Assess the actual IVD change and applicable statutory trigger; do not infer a fundamental design change from a new commercial REF alone.

Manufacturer/site master data is a separate amendment object

manufacturer-master-data-separate-amendment
Fact-level verified

If the registered manufacturer/company/site data changes, use the manufacturer-data amendment service for that object as applicable. Its site minor-variation guide requires an unexpired site and permits only certain open fields. Changing that record does not itself amend every product MA or transfer product MA ownership. A same-entity office/name change must first be located in the manufacturer master record and product approval particulars; do not assume a manufacturing-site move.

Dossier requirements

Which evidence, local deltas, formats, and signatures are required?

Compare countries

Product-amendment materials and conditional device scope

device-minor-variation-service-materials
Fact-level verified

The amendment service lists a company-signed/stamped minor-change application; the relevant change documents under the current applicable guide; the issued valid MA; a valid manufacturing-site registration; outer/inner packaging and leaflet; and, for certain changes, a sample and analysis certificate. Its medical-device branch is explicit. Select device-applicable materials and the actual change category; mixed pharmaceutical attachments such as active-substance or drug-PV materials are not a universal IVD/device checklist.

Conditions
  • The guide describes only two additional uploads in its minor-variation attachment screen; this screen limit does not remove otherwise applicable documentary requirements.
  • Read the current service requirement for a valid MA copy together with its exception accepting a minor-variation request while a renewal application is active; acceptance of the request does not itself authorize continued supply after expiry.

Origin CFS/registration certificate if a new device MA is required

ivd-initial-device-ma-origin-free-sale-certificate
Fact-level verified

When an IVD is processed under the device initial-MA route, the service requires a valid origin-country competent-authority free-sale/registration certificate certified by the UAE Embassy. This initial-registration requirement does not prove that every same-entity minor variation needs a replacement CFS.

Conditions
  • Confirm EDE product classification and that the initial device-MA route applies.

Materials for manufacturer/company-data amendment

manufacturer-data-amendment-materials
Fact-level verified

The manufacturer-data service requires current origin-country authority approval of the requested changes and a signed/stamped letter comparing existing and proposed details. Its published list also includes certified company authorization, current GMP/manufacturing licence materials, product/site documentation and the current EDE registration certificate. Apply the actual device/site conditions; the mixed guide’s initial-registration attachment Y/N markers do not establish a blanket device exemption from this service list.

Conditions
  • This is a manufacturer/company-data service; do not present its documents as the complete true product MA-transfer dossier.

Fees

What regulator fee applies to this class, route, event, and date?

Compare countries

Fee for the separate manufacturer/company-data certificate

manufacturer-data-amendment-exact-fee
Fact-level verified

The fee for a certificate modifying registration data of a medical company or factory holding a marketing right is AED 2,000 per certificate under Cabinet 40/2025 Annex item 27, consistent with the current manufacturer-data service. This is a separate object from the product MA amendment certificate in item 28.

Conditions
  • Only if that company/factory-data certificate is actually required; not the total for all proposed product changes or true-MA transfer.

Label, IFU & language

Which label and IFU elements, languages, symbols, and translation controls apply?

Compare countries

IVD labels must match approved product information

ivd-label-information-matches-approval
Fact-level verified

The internal/external label and printed or electronic leaflet of a medical product must match the package information in the marketing-approval annex, or the origin-country approved particulars for an exempt product. EDE determines the required label/leaflet particulars.

Conditions
  • Applies to IVDs/reagents within the medical-device/medical-product definition. Arabic original prevails.

IVD leaflet languages

ivd-leaflet-arabic-english
Fact-level verified

Medical-product leaflets, printed or electronic, use at least Arabic and English unless the Chairperson or delegate issues an exception decision.

Conditions
  • Product-specific exceptions must be established; no blanket English-only IVD exception inferred.

Official forms & portals

Which current form, portal, account, and submission package must be used?

Compare countries

Product amendment submission workflow

minor-variation-submission-and-certificate
Fact-level verified

Search/select the registered product, modify the fields actually open for the product/pack, attach the applicable supporting documents, review/accept terms, proceed through checkout and submit for EDE processing. The resulting minor-variation certificate records approved change details; its multiple-section layout is not permission to group any proposed combination or a new juridical owner into one filing.

Conditions
  • Follow the current service sequence of completing the application, attaching the applicable documents and paying before submission. The guide illustrates product/pack selection, open fields, attachments, terms, checkout and the resulting certificate.

Reliance & recognition

Which foreign decisions or evidence can be reused, with what effect and conditions?

Compare countries

Existing records include acceptance, restrictions and evidence reuse. The applicable product scope and conditions are preserved below; a reference does not by itself confer market access.

United Arab Emirates approvals and evidence in other markets (0 records)

No linked record is currently maintained. This does not establish that a pathway is unavailable.

Prior approvals and evidence for United Arab Emirates (4 records)

United Arab EmiratesAustralia TGA · Health Canada

Prior approval / evidence record

Health Canada or Australia TGA

Authority

Health Canada or Australia TGA

Accepted

Yes

Benefit

A Canadian or Australian registration can serve as the free sale or registration certificate from the country of origin, but EDE names EC, 510(k) and PMA as the conformity certificate types.

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Prior approval / evidence record

FDA 510(k)/PMA

Authority

FDA 510(k)/PMA

Accepted

Yes

Benefit

Required, not optional. EDE lists a certificate of quality conformity or marketing authorisation such as EC, 510(k) or PMA, matched to the device class, among the documents that must accompany the application, alongside a free sale or registration certificate from the country of origin attested by the UAE Embassy. There is no separate recognition route and no second, shorter service.

Timeline Savings

None published — 45 working days is the only figure EDE publishes, and it is the same for every class

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Prior approval / evidence record

NMPA evidence requires product-specific EDE eligibility assessment

reliance.china-nmpa-origin-evidence
Pending verification
Statement

Article 6(1)(c) provides for EDE assessment of reference-country marketing approvals or research information demonstrating effectiveness, safety and quality. It does not name NMPA, automatically accept an NMPA certificate or Class I filing record, or create an NMPA-specific abridged route. Any Chinese origin-country evidence must meet the applicable EDE documentary requirements and remains subject to EDE assessment.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

uae

China Classes In Scope

No NMPA class or filing record is expressly accepted by the cited provision; product-specific EDE requirements must be met.

Edge Type

Eligibility Unresolved

Market Access Strength

undetermined

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • This general evidentiary route is not confirmation that any particular NMPA document is sufficient.

Effective from 2025-01-02

Official English expression independently compared in full with the UAE legislation portal; Arabic original prevails. R217 review, 18 September 2026.

United Arab EmiratesEU conformity assessment

Prior approval / evidence record

EU CE Mark

Authority

EU CE Mark

Accepted

Yes

Benefit

An EC certificate is one of the accepted conformity certificates for the same mandatory document slot, with an EC declaration of conformity acknowledgement required in addition.

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Review clocks

What review, validation, and applicant stop-clock rules apply?

Compare countries

Published manufacturer-data amendment duration

manufacturer-data-amendment-service-duration
Fact-level verified

EDE publishes 10 working days for the manufacturer/company-registration-data amendment service and describes this in its FAQ as the average duration. The workflow guide gives the applicant 60 days to provide corrections; that response deadline is not a 60-day review SLA.

Conditions
  • Object-specific published service duration, not a guaranteed complete product MA/transfer project duration.

Published duration for the applicable amendment service

minor-variation-published-service-duration
Fact-level verified

The current EDE medical-product amendment service publishes 22 working days. Use this only once the proposed change qualifies for that service; it is not the duration of manufacturer-master-data amendment, a new MA, true juridical transfer or the whole preparation/foreign-approval/combined-change project.

Conditions
  • Published duration for the eligible product-MA amendment service only; preparation, foreign approvals, corrections and other service objects are separate.
  • No statutory stop-clock or guaranteed end-to-end completion promise is specified by this service page.

UDI, listing & traceability

Which UDI, device listing, database, and traceability duties apply?

Compare countries

Trade name, MAH, labels and GTIN are separate fields

minor-variation-open-fields-and-identifiers
Fact-level verified

The mixed device-applicable guide lists trade name, MAH/contact/address, manufacturer-related manufacturing-stage/site data, label/artwork attachments and a pack-size GTIN among fields that may be open for minor variation. Only fields actually open may be modified. The catalogue does not equate an internal orderable SMN/REF with model or GTIN/UDI, or establish that editing the MAH name field transfers the MA to a new legal person.

Conditions
  • Confirm the existing registered product/pack identifiers and proposed old-to-new mapping.
  • For an IVD, confirm its EDE device classification and the actual open fields before mapping an orderable REF to a registered product, model or pack identifier.

Related structured datasets

Classification systems · 1 record
Open United Arab Emirates IVD risk classification →
Id

uae:ivd

Jurisdiction Id

uae

Domains
  • ivd
Name

United Arab Emirates IVD risk classification

Classes
  • Id

    a

    Label

    IVD Class A

    Normalized Risk Level

    low

  • Id

    b

    Label

    IVD Class B

    Normalized Risk Level

    medium

  • Id

    c

    Label

    IVD Class C

    Normalized Risk Level

    high

  • Id

    d

    Label

    IVD Class D

    Normalized Risk Level

    highest

Regulator

EDE

Official Domains
  • ede.gov.ae
  • uaelegislation.gov.ae
Source Status

official-source-verified

Source Title

UAE government product-classification guidelines

Source Url

https://mohap.gov.ae/documents/20117/1212145/Classification+of+a+Product-+Guidelines-277.pdf/dfa0c036-89f5-b5da-18f5-88f8ce3a88be

Verified On

2026-08-13

Source Supports
  • ivd-class-axis
Registration pathways, prior approvals and planning · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open United Arab Emirates registration →
Country

United Arab Emirates

Jurisdiction Id

uae

Country Code

AE

Flag

🇦🇪

Regulator

Emirates Drug Establishment

Regulator Abbrev

EDE

Region

middle-east

Tier

2

Risk Classification

Medical devices: Class I, II, III, IV; IVDs: Class A, B, C, D

Classes
  • Id

    class-i

    Class Name

    Class I

    Risk Level

    low

    Pathway

    Registration

    Official Timeline
    Min

    45

    Max

    45

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    UAE EDE — medical device marketing authorization (45 working days)

    Url

    https://www.ede.gov.ae/en/w/issuance-of-marketing-authorization-for-a-medical-device

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    2

    Max

    3

    Unit

    months

    Notes
    • EDE publishes one marketing authorisation service for medical devices with a completion duration of 45 working days, and it does not vary by class
    • A certificate of quality conformity or marketing authorisation — EC, 510(k) or PMA as appropriate to the class — is a required document, so a foreign approval is the price of entry rather than a shortcut
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    No

  • Id

    class-ii

    Class Name

    Class II

    Risk Level

    medium

    Pathway

    Registration

    Official Timeline
    Min

    45

    Max

    45

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    UAE EDE — medical device marketing authorization (45 working days)

    Url

    https://www.ede.gov.ae/en/w/issuance-of-marketing-authorization-for-a-medical-device

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    3

    Max

    6

    Unit

    months

    Notes
    • EDE publishes one marketing authorisation service for medical devices with a completion duration of 45 working days, and it does not vary by class
    • A certificate of quality conformity or marketing authorisation — EC, 510(k) or PMA as appropriate to the class — is a required document, so a foreign approval is the price of entry rather than a shortcut
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

  • Id

    class-iii

    Class Name

    Class III

    Risk Level

    high

    Pathway

    Registration

    Official Timeline
    Min

    45

    Max

    45

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    UAE EDE — medical device marketing authorization (45 working days)

    Url

    https://www.ede.gov.ae/en/w/issuance-of-marketing-authorization-for-a-medical-device

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    4

    Max

    8

    Unit

    months

    Notes
    • EDE publishes one marketing authorisation service for medical devices with a completion duration of 45 working days, and it does not vary by class
    • Safety and efficacy data are a required document at Class III and IV
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

  • Id

    class-iv

    Class Name

    Class IV

    Risk Level

    highest

    Pathway

    Full Registration

    Official Timeline
    Min

    45

    Max

    45

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    UAE EDE — medical device marketing authorization (45 working days)

    Url

    https://www.ede.gov.ae/en/w/issuance-of-marketing-authorization-for-a-medical-device

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    6

    Max

    12

    Unit

    months

    Notes
    • EDE publishes one marketing authorisation service for medical devices with a completion duration of 45 working days, and it does not vary by class
    • Safety and efficacy data are a required document at Class III and IV
    Clinical Trial Required

    Yes

    Local Testing Required

    No

    Gmp Required

    Yes

Prior Approvals
  • Authority

    FDA 510(k)/PMA

    Accepted

    Yes

    Benefit

    Required, not optional. EDE lists a certificate of quality conformity or marketing authorisation such as EC, 510(k) or PMA, matched to the device class, among the documents that must accompany the application, alongside a free sale or registration certificate from the country of origin attested by the UAE Embassy. There is no separate recognition route and no second, shorter service.

    Timeline Savings

    None published — 45 working days is the only figure EDE publishes, and it is the same for every class

  • Authority

    EU CE Mark

    Accepted

    Yes

    Benefit

    An EC certificate is one of the accepted conformity certificates for the same mandatory document slot, with an EC declaration of conformity acknowledgement required in addition.

    Timeline Savings

    None published

  • Authority

    Health Canada or Australia TGA

    Accepted

    Yes

    Benefit

    A Canadian or Australian registration can serve as the free sale or registration certificate from the country of origin, but EDE names EC, 510(k) and PMA as the conformity certificate types.

    Timeline Savings

    None published

Key Requirements
  • UAE Authorized Representative
  • Arabic/English labeling
  • ISO 13485 certification
  • Free Sale Certificate
  • GCC conformity assessment
  • Related EDE clocks: MA renewal 15 working days; classification letter 10 working days; minor-variation certificate 22 working days
  • Import permits: initial import requests 3 working days, shipment clearance 2 working days; advertisement licence 3 working days; manufacturer (site) registration 2-4 weeks
Local Rep Required

Yes

Local Rep Name

UAE Authorized Representative

Mdsap Accepted

No

Mdsap Benefit

Not recognised. EDE names EC, 510(k) and PMA as the conformity certificate types, and requires ISO 13485 separately.

Registration Validity

The marketing authorisation certificate is valid for five years from the date of the registration committee approval, not from the date of application.

Verification Status

official-verified

Official Sources
  • Title

    UAE EDE — medical device marketing authorization (45 working days)

    Url

    https://www.ede.gov.ae/en/w/issuance-of-marketing-authorization-for-a-medical-device

    Accessed Date

    2026-08-25

Reviewed On

2026-08-30

Last Updated

2026-08-30

Process
Template Id

default

Status

planning-template

Usage Rule

Maintained planning steps and estimates, not verified regulator deadlines.

Steps
  • Step

    1

    Title

    Pre-Submission Preparation

    Description

    Gather technical documentation, quality management system (QMS) records, and clinical data required by EDE. Appoint a UAE Authorized Representative.

    Duration

    1-3 months

    Icon

    📋

  • Step

    2

    Title

    Application Submission

    Description

    Submit complete registration dossier to EDE including device description, risk analysis, and testing reports.

    Duration

    1-2 weeks

    Icon

    📤

  • Step

    3

    Title

    Administrative Review

    Description

    EDE reviews application for completeness. May request additional documents or clarifications.

    Duration

    2-4 weeks

    Icon

    🔍

  • Step

    4

    Title

    Technical Evaluation

    Description

    In-depth technical assessment of device safety and performance. EDE may issue questions.

    Duration

    45 working days

    Icon

    ⚙️

  • Step

    5

    Title

    Registration Approval

    Description

    Upon successful review, EDE issues United Arab Emirates market authorization certificate.

    Duration

    1-2 weeks

    Icon

    ✅

Government fee schedules · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open United Arab Emirates fees →
Country

United Arab Emirates

Jurisdiction Id

uae

Country Code

AE

Authority

EDE

Low Risk Local
Min

5100

Max

5100

Currency

AED

Per Product

Yes

High Risk Local
Min

5100

Max

5100

Currency

AED

Per Product

Yes

Details
  • Id

    application-fee

    Item

    Marketing-authorization application fee

    Currency

    AED

    Amount Local

    100

    Amount Local Formatted

    AED 100

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.ede.gov.ae/en/w/issuance-of-marketing-authorization-for-a-medical-device

    Source Title

    EDE marketing authorization for a medical device

    Verified On

    2026-08-25

  • Id

    device-registration

    Item

    Device Registration

    Currency

    AED

    Amount Local

    5000

    Amount Local Formatted

    AED 5,000

    Frequency

    Per device

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.ede.gov.ae/en/w/issuance-of-marketing-authorization-for-a-medical-device

    Source Title

    EDE marketing authorization for a medical device

    Verified On

    2026-08-25

  • Id

    renewal-application-fee

    Item

    Marketing-authorization renewal application fee

    Currency

    AED

    Amount Local

    100

    Amount Local Formatted

    AED 100

    Frequency

    Per renewal application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.ede.gov.ae/en/w/renewal-of-marketing-authorization-for-a-medical-device

    Source Title

    EDE renewal of marketing authorization for a medical device

    Verified On

    2026-08-25

  • Id

    renewal-registration-fee

    Item

    Marketing-authorization renewal fee

    Currency

    AED

    Amount Local

    2500

    Amount Local Formatted

    AED 2,500

    Frequency

    Per renewal

    Notes

    The renewal service card prints the fee line as "Renewal of the registration of pharmaceutical product for general sale: AED 2500" — label wording appears generic, amount as published.

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.ede.gov.ae/en/w/renewal-of-marketing-authorization-for-a-medical-device

    Source Title

    EDE renewal of marketing authorization for a medical device

    Verified On

    2026-08-25

  • Id

    classification-letter

    Item

    Classification-of-a-product letter

    Currency

    AED

    Amount Local

    500

    Amount Local Formatted

    AED 500

    Frequency

    Per letter

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.ede.gov.ae/en/w/classification-of-a-product

    Source Title

    EDE classification of a product letter

    Verified On

    2026-08-25

  • Id

    import-permit-application

    Item

    Import permit - application fee

    Currency

    AED

    Amount Local

    100

    Amount Local Formatted

    AED 100

    Frequency

    Per application

    Notes

    Chemical-precursor import permits are 1% of invoice value, minimum AED 200 per permit.

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.ede.gov.ae/en/w/import-permit-for-medical-materials-and-products

    Source Title

    EDE import permit for medical materials and products

    Verified On

    2026-08-25

  • Id

    manufacturer-registration-application

    Item

    Manufacturer (site) registration - application fee

    Currency

    AED

    Amount Local

    100

    Amount Local Formatted

    AED 100

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.ede.gov.ae/en/w/registration-of-a-manufacturer-of-medical-products

    Source Title

    EDE registration of a manufacturer of medical products

    Verified On

    2026-08-25

  • Id

    manufacturer-registration

    Item

    Registration of a medical-products manufacturer

    Currency

    AED

    Amount Local

    10000

    Amount Local Formatted

    AED 10,000

    Frequency

    Per registration, renewed every 5 years

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.ede.gov.ae/en/w/registration-of-a-manufacturer-of-medical-products

    Source Title

    EDE registration of a manufacturer of medical products

    Verified On

    2026-08-25

  • Id

    facility-plans-approval

    Item

    Domestic manufacturing-facility licence - approval of architectural plans and drawings

    Currency

    AED

    Amount Local

    2000

    Amount Local Formatted

    AED 2,000

    Frequency

    Per approval

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.ede.gov.ae/en/w/issuance-of-a-license-for-a-medical-products-factory-and-contractual-companies-for-research-and-development

    Source Title

    EDE licensing of a medical-products factory: plans approval and final licensing

    Verified On

    2026-08-25

  • Id

    facility-final-licensing

    Item

    Domestic manufacturing-facility licence - final licensing of pharmaceutical and medical facilities

    Currency

    AED

    Amount Local

    50000

    Amount Local Formatted

    AED 50,000

    Frequency

    Per licence

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.ede.gov.ae/en/w/issuance-of-a-license-for-a-medical-products-factory-and-contractual-companies-for-research-and-development

    Source Title

    EDE licensing of a medical-products factory: plans approval and final licensing

    Verified On

    2026-08-25

  • Id

    advertisement-licence-application

    Item

    Advertisement licence - application fee

    Currency

    AED

    Amount Local

    100

    Amount Local Formatted

    AED 100

    Frequency

    Per application

    Notes

    The licence fee varies by advertising-facility category (clinic, healthcare centre, hospital) and advertisement type (print, audio, audiovisual).

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.ede.gov.ae/en/w/licensing-an-advertisement-for-a-medical-product

    Source Title

    EDE licensing an advertisement for a medical product

    Verified On

    2026-08-25

  • Id

    minor-variation-certificate

    Item

    Minor-variation (amendment) certificate for registered MA data

    Currency

    AED

    Amount Local

    1000

    Amount Local Formatted

    AED 1,000

    Frequency

    Per certificate

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.ede.gov.ae/en/w/issuance-of-a-certificate-of-amendment-to-the-registered-marketing-authorization-data-for-a-medical-product

    Source Title

    EDE certificate of amendment to registered marketing-authorization data

    Verified On

    2026-08-25

Notes
  • The official service supports AED 100 application plus AED 5,000 registration for device marketing authorization.
  • Manufacturer, facility, advertisement, variation, classification, and import-permit fees are separate EDE services with their own published amounts.
Last Updated

2026-08-25

Verification Status

official-verified

Official Sources
  • Title

    EDE marketing authorization for a medical device

    Url

    https://www.ede.gov.ae/en/w/issuance-of-marketing-authorization-for-a-medical-device

    Accessed Date

    2026-08-25

  • Title

    EDE renewal of marketing authorization for a medical device

    Url

    https://www.ede.gov.ae/en/w/renewal-of-marketing-authorization-for-a-medical-device

    Accessed Date

    2026-08-25

  • Title

    EDE classification of a product letter

    Url

    https://www.ede.gov.ae/en/w/classification-of-a-product

    Accessed Date

    2026-08-25

  • Title

    EDE import permit for medical materials and products

    Url

    https://www.ede.gov.ae/en/w/import-permit-for-medical-materials-and-products

    Accessed Date

    2026-08-25

  • Title

    EDE registration of a manufacturer of medical products

    Url

    https://www.ede.gov.ae/en/w/registration-of-a-manufacturer-of-medical-products

    Accessed Date

    2026-08-25

  • Title

    EDE licensing of a medical-products factory: plans approval and final licensing

    Url

    https://www.ede.gov.ae/en/w/issuance-of-a-license-for-a-medical-products-factory-and-contractual-companies-for-research-and-development

    Accessed Date

    2026-08-25

  • Title

    EDE licensing an advertisement for a medical product

    Url

    https://www.ede.gov.ae/en/w/licensing-an-advertisement-for-a-medical-product

    Accessed Date

    2026-08-25

  • Title

    EDE certificate of amendment to registered marketing-authorization data

    Url

    https://www.ede.gov.ae/en/w/issuance-of-a-certificate-of-amendment-to-the-registered-marketing-authorization-data-for-a-medical-product

    Accessed Date

    2026-08-25

Reviewed On

2026-08-25

Amount Status

exact-current

Range Available

Yes

Additional certifications and licences · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open United Arab Emirates certifications →
Slug

uae

Market

United Arab Emirates

Confidence

low

Schemes
  • Id

    mohap-electrical-emc-evidence

    Name

    MOHAP electrical/EMC evidence

    Category

    Medical Electrical Safety

    Status

    Included In Medical Registration

    Object

    Finished Device

    Note

    EDE routes: manufacturer-site registration requires a valid legalized ISO 13485 certificate from the competent authority of the country of origin; device marketing authorization requires an EC Declaration of Conformity acknowledging conformity with the Medical Equipment Manual specifications, and animal-origin devices additionally need a certificate of conformity; import permits accept valid FDA registration, ISO 13485/EC quality certificates, a Free Sale Certificate or GMP.

    Source Binding

    unresolved

  • Id

    tdra-equipment-registration-level-1-2-3

    Name

    TDRA equipment registration Level 1/2/3

    Category

    Radio Spectrum

    Status

    Mandatory If Feature

    Object

    Covered Telecommunications Equipment

    Source Urls
    • https://tdra.gov.ae/en/services/equipment-registration
    Source Binding

    candidate

  • Id

    moiat-ecas

    Name

    MOIAT ECAS

    Category

    General Electrical Components

    Status

    Mandatory If Feature

    Object

    Regulated Product Or Component

    Source Urls
    • https://moiat.gov.ae/en/services/issue-conformity-certificates-for-regulated-products
    Source Binding

    candidate

  • Id

    fanr-radiation-authorization

    Name

    FANR radiation authorization

    Category

    Ionizing Radiation

    Status

    Mandatory If Feature

    Object

    Radiation Product Import Or Use Site

    Source Binding

    unresolved

  • Id

    halal-certification-mark

    Name

    Halal certification/mark

    Category

    Halal And Material Origin

    Status

    Voluntary Or Procurement

    Object

    Eligible Product Or Claim

    Source Binding

    unresolved

Official Sources
  • Authority

    TDRA

    Topic

    telecommunications equipment registration

    Url

    https://tdra.gov.ae/en/services/equipment-registration

  • Authority

    MOIAT

    Topic

    ECAS conformity certificate for regulated products

    Url

    https://moiat.gov.ae/en/services/issue-conformity-certificates-for-regulated-products

  • Authority

    EDE

    Topic

    manufacturer registration legalized ISO 13485 certificate

    Url

    https://www.ede.gov.ae/en/w/registration-of-a-manufacturer-of-medical-products

  • Authority

    EDE

    Topic

    marketing authorization EC declaration of conformity and animal-origin certificate

    Url

    https://www.ede.gov.ae/en/w/issuance-of-marketing-authorization-for-a-medical-device

  • Authority

    EDE

    Topic

    import permit accepted FDA ISO 13485 EC and free sale evidence

    Url

    https://www.ede.gov.ae/en/w/import-permit-for-medical-materials-and-products

Keep researching this market
View the United Arab Emirates resource hub →
Compare regulatory data across all countries →
DJ Fang
Need to apply United Arab Emirates's requirements to a product?
Confirm the current EDE pathway, evidence, and local-role details for the actual device before filing.