🇿🇲 Zambia IVD regulatory data

Explore the IVD requirements currently available for Zambia, organized by topic and linked to official sources.

16 facts2 dimensions3 official sourceslatest verification 2026-09-16
Explore questions and sources →

Common questions about changing a registered IVD in Zambia. Each question separates published rules from details still awaiting verification.

Recall, FSCA & market surveillance

Which recall, FSCA, notice, and authority-coordination duties apply?

Compare countries

Class I: public notified within 24 hours, undistributed stock quarantined immediately, recall completed within 72 hours (s.6.2.4)

recall.class-i-24h-public-notification
Fact-level verified

For a Class I recall, the MAH, local responsible person, distributor and manufacturer must notify the public within 24 hours of the recall decision, quarantine undistributed stock immediately, and complete the recall within 72 hours. A problem that may lead to a Class I or II recall must be reported to ZAMRA by telephone within 24 hours (Appendix II note).

Conditions
  • Class I recalls only for the 24-hour public notification; the 24-hour telephone report applies to potential Class I or II recalls.

Effective from 2023-03-01

Reporting uses the Pharmaceutical Product Reporting Form (Part I) and Recall Notification Form (Part II) in Appendix II, also posted on the ZAMRA Forms page.

Recalls are classified I, II or III by ZAMRA; Classes I and II are urgent safety-related recalls

recall.classification-authority-decides
Fact-level verified

Recalls are classified by severity into Class I, II and III. Class I and II are 'Urgent Safety-Related Recalls' that must be reported to ZAMRA for evaluation; Class III are minimum-risk recalls that must nevertheless be reported. ZAMRA determines the class of each recall, and for international recalls it considers decisions of other national regulatory agencies.

Effective from 2023-03-01

Table 1 gives examples per class (Class I: serious/life-threatening defects; Class III: non-compliance with the Act's quality, safety or efficacy requirements).

Non-adherence to the Recall Guideline is prosecuted under s.46(4) MASA (s.9.0)

recall.guideline-non-adherence-penalty
Fact-level verified

Where there is non-adherence to the Recall Guideline, ZAMRA institutes regulatory action and prosecution in accordance with s.46(4) of the Medicines and Allied Substances Act (fine up to one million penalty units and/or up to three years' imprisonment).

Effective from 2023-03-01

The 2023 Recall Guideline covers allied substances, expressly including medical devices, and both voluntary and statutory recalls

recall.guideline-scope-includes-devices
Fact-level verified

ZAMRA's Guideline on Recall of Medicines and Allied Substances (March 2023) applies to all quality-defect reports and safety/efficacy incidents for medicines and allied substances, and to local responsible persons, MAHs, distributors, manufacturers, wholesalers, retailers, health facilities and the public; its definitions state that allied substances include medical and surgical sundries, medical devices and condoms.

Effective from 2023-03-01

Guideline is dated 'MARCH, 2023' on its cover; exact day not stated. The guideline also defines 'Correction' (repair, modification, adjustment, re-labelling or inspection without physical removal), the FSCA-type action.

Recall initiation and physical-recall timelines by class (Table 3)

recall.initiation-completion-timelines
Fact-level verified

Recalls must be initiated within 24 hours (Class I), 48 hours (Class II) or 72 hours (Class III), and physically completed within 72 hours (Class I), 14 days (Class II) or 30 days (Class III).

Effective from 2023-03-01

Section 1.4 (Table 2) separately defines recall levels A (consumer), B (retail/user facility) and C (wholesale).

Inspectors may enter, search, seize products and take samples under warrant (s.55(4) MASA)

recall.inspector-seizure-sampling-powers
Fact-level verified

A ZAMRA inspector may, on the authority of a warrant, enter premises or conveyances believed to be used contrary to the Act and, among other powers, seize any medicine or allied substance (and related records or devices) that bears on an inspection or investigation or is used contrary to the Act, and take samples for testing, examination or analysis.

Conditions
  • Entry requires a warrant and reasonable grounds (s.55(4)).

Effective from 2013-03-22

Market-surveillance enforcement power; s.55(5) requires a receipt for anything removed.

Interim report within 7 days, final report within 14 days, investigation/CAPA report within 30 days of the recall (s.6.2.6)

recall.interim-final-investigation-reports
Fact-level verified

The MAH, local responsible person, distributor or manufacturer must give ZAMRA an interim progress report within 7 calendar days of initiating the recall, a final report within 14 calendar days of commencing it (or obtain ZAMRA's approval for a later date), and a report on the investigation results and preventive action not more than 30 calendar days after the recall.

Effective from 2023-03-01

Final report content: circumstances, action taken, distribution extent in and outside Zambia, quantities returned/outstanding, confirmation of retailer returns, and method of destruction. Appendix IV provides the Post Recall Report Form.

Voluntary recall notices must be approved by ZAMRA before publication; statutory notices issue under s.46 (s.5.1)

recall.notice-approval-before-publication
Fact-level verified

A statutory recall notice is issued under s.46 of the Act. For a voluntary recall, the MAH, local responsible person, distributor or manufacturer must submit a draft notice with a factual statement of reasons and product identifiers to ZAMRA for approval before publication; the notice may not contain promotional material.

Effective from 2023-03-01

s.5.2 adds that for Class I/II recalls a media release (public warning) headed 'Urgent Medicine/Allied Substance Recall' is used.

Recall/withdrawal may be regulated by statutory instrument, and ZAMRA guidelines are binding once published (ss.68-69 MASA)

recall.regulation-and-guideline-powers
Fact-level verified

The Minister may by statutory instrument make regulations on the recall or withdrawal of medicines and allied substances that do not meet prescribed quality, efficacy and safety standards (s.69(2)(g)); separately, guidelines made by ZAMRA under s.68 become binding on all regulated persons upon publication in a daily newspaper of general circulation.

Effective from 2013-03-22

The 2023 Recall Guideline cites s.46 and s.68 as its legal basis (Guideline s.1.0).

Selling a product subject to a recall notice is an offence (s.46(3)-(4) MASA)

recall.sale-of-recalled-product-offence
Fact-level verified

No person may sell a medicine or allied substance that is the subject of a s.46(1) recall notice; contravention is an offence punishable on conviction by a fine of up to one million penalty units, imprisonment of up to three years, or both.

Effective from 2013-03-22

The Recall Guideline (s.9.0) states that non-adherence to the guideline is prosecuted under s.46(4).

ZAMRA may direct the return (recall) of a medicine or allied substance by written notice or Gazette (s.46(1)-(2) MASA)

recall.statutory-recall-power
Fact-level verified

Where ZAMRA determines it is not in the public interest that a medicine or allied substance be made available to the public, it may by written notice served on any person, or by Gazette notice, direct that person to return the product to the manufacturer, to the importer (for imported products) or to the Authority or its designee, and may direct the manufacturer or importer how to deal with or dispose of the returned stock.

Conditions
  • Applies to any medicine or allied substance, which includes medical devices (s.2).

Effective from 2013-03-22

Marginal note reads 'Recall of medicine, etc. from circulation'.

Circumstances in which ZAMRA triggers a statutory (mandatory) recall (s.2.2)

recall.statutory-recall-triggers
Fact-level verified

ZAMRA triggers a statutory recall where a medicine or allied substance does not meet the set standard of quality, safety and efficacy; where continued supply violates the Act or another law; or where the Authority determines that continued availability is not in the public interest. Voluntary recalls (s.2.1) are triggered by the MAH/manufacturer on quality, complaint, stability or serious-safety-risk findings.

Effective from 2023-03-01

Definition on p.5: 'Mandatory (non-voluntary) recall: Is when the Authority requests/orders a product recall due to non-compliance to regulatory requirements.'

Special access & derogation

Is an emergency, compassionate, custom, humanitarian, or other exceptional path available?

Compare countries

Emergency authorisation to manufacture or import a generic of a patented medicine or allied substance (s.57 MASA)

special.emergency-generic-authorisation
Fact-level verified

Where the Minister declares a national health emergency or state of extreme urgency, where the public interest so requires, or where a product is needed for public non-commercial use, ZAMRA may authorise local manufacture or importation of a generic formulation of a medicine or allied substance notwithstanding a Zambian patent; the authorisation cannot exclude the patent holder and is subject to court review.

Conditions
  • Triggers: declared national health emergency / extreme urgency, public interest, or public non-commercial use.
  • Authorisation is subject to review by a court (s.57(3)).

Effective from 2013-03-22

This is a patent-override (compulsory-use) power triggered by emergencies rather than a product-approval waiver; the marketing-authorisation waiver for a declared health emergency sits in s.39(4)(e).

Statutory exemptions from the marketing-authorisation requirement (s.39(4) MASA)

special.marketing-authorisation-exemptions
Fact-level verified

Section 39(4) of the Medicines and Allied Substances Act 2013 lists the cases to which the marketing-authorisation requirement of s.39(1) does not apply: donated medicines; personal or relative's use on prescription; import by a physician, dentist or veterinary surgeon on prescription; pharmacist-compounded medicine; medicine imported or exported in response to a declared health emergency; medicine imported by an authorised institution for its patients; and any medicine or allied substance used for a clinical trial.

Conditions
  • Limbs (a)-(f) are drafted in terms of 'medicine'; only limb (g) (clinical trial) expressly says 'medicine or allied substance'. Medical devices are 'allied substances' under s.2, so the application of limbs (a)-(f) to devices is not explicit on the face of the Act.

Effective from 2013-03-22

s.2 defines 'allied substances' to include 'medical and surgical sundries, medical devices and condoms'. s.39(1) itself applies to 'any medicine or allied substance'.

Personal-use import permit for a medicine or allied substance (SI 57 of 2017, reg. 4)

special.personal-use-import-permit
Fact-level verified

A person who intends to import any medicine or allied substance for personal use must apply to ZAMRA for a permit in Form II of the Schedule to SI 57 of 2017 and pay the prescribed fee. 'Personal use' includes use by the person's relative or animal.

Conditions
  • Applies to importation for personal use only; commercial imports use the reg. 3 / Form I permit.

Effective from 2017-07-28

Read from the gazette scan via OCR; reg. 4(2) separately provides for the Authority to notify its decision, and reg. 7(2) for issue of the permit in Form VI.

Travellers carrying prescribed medicine or allied substance for own use are exempt from the permit regime (SI 57 of 2017, reg. 13)

special.traveller-exemption
Fact-level verified

The Importation and Exportation Regulations do not apply to a traveller entering Zambia with a medicine or allied substance for the traveller's own use, provided the traveller furnishes the port-of-entry inspector with written confirmation from the competent authority of the prescribing country that the product was prescribed for the traveller.

Conditions
  • Own use only; written confirmation from the foreign competent authority must be produced at the port of entry.

Effective from 2017-07-28

'competent authority' is defined in reg. 2 as the medicines and allied substances regulatory authority of the country where the product was prescribed.

Related structured datasets

Classification systems · 1 record
Open ZAMRA Classes A, B, C and D — in-vitro diagnostic devices →
Id

zambia:ivd

Jurisdiction Id

zambia

Domains
  • ivd
Name

ZAMRA Classes A, B, C and D — in-vitro diagnostic devices

Description

Zambia classifies IVDs on a parallel four-class system based on individual and public-health risk, published in the ZAMRA guideline for Grant of Marketing Authorisation of In-Vitro Diagnostic Devices for Human Use (ZAMRA/MA/GUD/0003): Class A low individual and low public-health risk, Class B moderate individual and/or low public-health risk, Class C high individual and/or moderate public-health risk, Class D high individual and high public-health risk. The IVD guideline excludes in-vitro diagnostics for veterinary use.

Classes
  • Id

    a

    Label

    Class A — low individual and low public-health risk

    Normalized Risk Level

    low

  • Id

    b

    Label

    Class B — moderate individual and/or low public-health risk

    Normalized Risk Level

    medium

  • Id

    c

    Label

    Class C — high individual and/or moderate public-health risk

    Normalized Risk Level

    high

  • Id

    d

    Label

    Class D — high individual and high public-health risk

    Normalized Risk Level

    highest

Regulator

Zambia Medicines Regulatory Authority

Official Domains
  • zamra.co.zm
Source Status

official-source-verified

Reference Sources
  • Title

    ZAMRA IVD guideline section 1.14 and the IVD risk-band table

    Url

    https://www.zamra.co.zm/wp-content/uploads/2026/06/Guidelines-for-grant-of-Marketing-Authorisation-of-in-vitro-diagnostic-devices-for-Human-use.pdf

    Reviewed On

    2026-09-16

    Supports

    The IVD class axis is the Authority's own published risk-band table rather than an imported GHTF document, and IVD applications are filed physically with the Authority — the two device pathways differ in both classification wording and filing channel.

Source Title

ZAMRA, Guidelines for Grant of Marketing Authorisation of In-Vitro Diagnostic Devices for Human Use (ZAMRA/MA/GUD/0003, version 1.0)

Source Url

https://www.zamra.co.zm/wp-content/uploads/2026/06/Guidelines-for-grant-of-Marketing-Authorisation-of-in-vitro-diagnostic-devices-for-Human-use.pdf

Verified On

2026-09-16

Source Supports
  • class-axis
  • classification-rules
Registration pathways, prior approvals and planning · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Zambia registration →
Country

Zambia

Jurisdiction Id

zambia

Country Code

ZM

Flag

🇿🇲

Regulator

Zambia Medicines Regulatory Authority

Regulator Abbrev

ZAMRA

Region

africa

Tier

3

Risk Classification

4-class (A, B, C and D) — GHTF/IMDRF-derived rules in ZAMRA/MA/GUD/0002 for non-IVDs and a parallel four-class individual/public-health-risk table for IVDs in ZAMRA/MA/GUD/0003

Classes
  • Id

    ivd-marketing-authorisation

    Class Name

    Class A to D in-vitro diagnostic device — marketing authorisation

    Risk Level

    medium

    Pathway

    Marketing authorisation application to ZAMRA under s.39 of the Medicines and Allied Substances Act No. 3 of 2013, filed physically in duplicate with one hard and one soft copy of the full product dossier under ZAMRA/MA/GUD/0003

    Official Timeline
    Min

    180

    Max

    180

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    ZAMRA, Guidelines for Grant of Marketing Authorisation of In-Vitro Diagnostic Devices for Human Use (ZAMRA/MA/GUD/0003, version 1.0) — sections 1.9.3 and 1.10.1-1.10.2

    Url

    https://www.zamra.co.zm/wp-content/uploads/2026/06/Guidelines-for-grant-of-Marketing-Authorisation-of-in-vitro-diagnostic-devices-for-Human-use.pdf

    Accessed Date

    2026-09-16

    Realistic Timeline
    Min

    9

    Max

    14

    Notes
    • The published clock is the IVD guideline's: 'All applications shall be evaluated within one hundred and eighty (180) working days from the date of receipt in the order of submission.' It is a queue-inclusive figure, not a pure assessment time, because it runs in order of submission.
    • The clock is not suspended quietly: where the Authority requests additional data the applicant must supply it within 90 working days of the request, and any extension must be requested in writing within 14 days of receipt of the request and be approved by the Authority. If the Technical Committee defers an application for unresolved safety or quality issues and the data is not provided within 90 days, the application is refused; if the applicant misses the response deadline the application is closed and must be refiled as a new application.
    • The realistic range adds the queue and the two-cycle query pattern to the 180 working days, which is roughly 8.3 working months of assessment on its own.
    Clinical Trial Required

    No

    Local Testing Required

    Yes

    Gmp Required

    Yes

  • Id

    non-ivd-expedited-evaluation

    Class Name

    Class A to D non-IVD medical device — expedited evaluation (fast track)

    Risk Level

    medium

    Pathway

    Expedited evaluation authorised by the Authority, charged at double the applicable marketing authorisation fee

    Official Timeline
    Min

    90

    Max

    90

    Unit

    days

    Verification Status

    official-verified

    Official Source
    Title

    ZAMRA, Guidelines on Application for Grant of Marketing Authorisation of Non In-Vitro Diagnostic Medical Devices for Human Use (version 2, implemented 21 June 2024) — section 1(h) and the definition of 'expedited evaluation'

    Url

    https://www.zamra.co.zm/wp-content/uploads/2026/06/GUIDELINES-ON-APPLICATION-FOR-GRANT-OF-MARKETING-AUTHORISATION-OF-NON-IVDS.pdf

    Accessed Date

    2026-09-16

    Realistic Timeline
    Min

    4

    Max

    8

    Notes
    • The non-IVD guideline defines 'expedited evaluation' as 'the express evaluation of a product application within a 90 - day period'. That definition is the published clock this row carries.
    • The Authority authorises expedited evaluation and the guideline states no criteria for granting it, so holding a WHO prequalification, an FDA approval, a CE mark or an approval in an IMDRF member state does not by itself trigger it — those are disclosure items on the Annex I form. The published commercial signal is the fee schedule's 'Fast track fees' head at double the applicable marketing authorisation fee.
    • The IVD guideline defines the equivalent as a 'Fast Track process' without a period of its own.
    Clinical Trial Required

    No

    Local Testing Required

    Yes

    Gmp Required

    Yes

  • Id

    condom-marketing-authorisation

    Class Name

    Condoms — marketing authorisation under the dedicated condom guideline

    Pathway

    Application to ZAMRA for marketing authorisation of a condom under s.39 of the Medicines and Allied Substances Act No. 3 of 2013, evaluated against the condom guideline's dossier requirements with laboratory analysis by the National Drug Quality Control Laboratory against ISO 4074:2015 (male) or ISO 25841:2014 (female), and tabled before the Technical Committee for approval

    Official Timeline
    Min

    365

    Max

    365

    Unit

    days

    Verification Status

    official-verified

    Official Source
    Title

    Guidelines on the Application for Marketing Authorisation of Condoms — ZAMRA guidance for the preparation and submission of dossiers, §1.12.1-1.12.2 Time Frames (acknowledgement within 14 days; review of a new application within 365 days on a first-come basis)

    Url

    https://www.zamra.co.zm/wp-content/uploads/2026/06/Application-for-Marketing-Authorisation-Condoms.pdf

    Accessed Date

    2026-09-16

    Realistic Timeline
    Min

    12

    Max

    18

    Notes
    • Section 1.12.2 (Time Frames — Evaluation of new applications) of the ZAMRA condom guideline: 'All applications shall be reviewed within 365 days from the date of receipt on a first come basis.' The period is queue-inclusive rather than pure assessment time, because it runs on a first-come basis; the same section separately commits the Authority to acknowledge receipt within 14 days (s.1.12.1), which also binds the Authority but is not the review clock.
    • The applicant-facing side of the same section is not part of the clock: 'The applicant shall be requested to provide additional data when required within 120 days of such a request. Should additional time be required, a formal request must be made and approved by the Authority.'
    • The instrument is a device instrument: condoms are an allied substance under the Medicines and Allied Substances Act No. 3 of 2013, and ZAMRA lists this guideline in the Medical Devices section of its guidelines page.
    • CAVEAT 1 — INSTRUMENT CURRENCY, stated rather than resolved: the guideline's own version table carries 'Version 1 draft March 2016', 'Version 1 released for comment April 2016' and 'Deadline for comment May 2016', then leaves both 'Version 2 published for implementation' and 'Date of implementation' blank, and the PDF's metadata Subject field reads 'DRAFT'; the file's embedded metadata gives CreationDate 6 September 2016 and ModDate 24 August 2023, and it is the version the Authority currently hosts in the Medical Devices section of its guidelines page.
    • CAVEAT 2 — INSTRUMENT CONFLICT, stated rather than resolved: the Authority's later general guideline, the Guidelines on Application for Grant of Marketing Authorisation of Non In-Vitro Diagnostic Medical Devices Version 2 published for implementation on 21 June 2024, governs non-IVD medical devices generally, states only that 'An application shall be evaluated on a first come first serve basis unless expedited evaluation has been authorised by the Authority', and fixes no period; condoms are a non-IVD device, so on one reading the 2024 general guideline governs and this 365-day clock is superseded, and on another reading the condom-specific guideline still governs because ZAMRA continues to publish it as the condom route and neither document says it replaces the other. The 365 days is published because it is a period the Authority fixed in a device instrument; the two readings are stated rather than reconciled.
    • The row is route-wide rather than class-specific because the condom guideline fixes one period for the product class it governs, and condoms do not sit on the A-D device axis.
    Clinical Trial Required

    No

    Local Testing Required

    Yes

    Gmp Required

    Yes

Prior Approvals
  • Authority

    WHO Prequalification; US FDA approval; EU CE marking; approval in any IMDRF member state

    Accepted

    No

    Benefit

    Disclosure only. The Annex I application form requires the applicant to state whether the device is WHO prequalified, US FDA approved, CE marked or approved in any IMDRF member state and to attach valid documentation, but neither the non-IVD nor the IVD guideline attaches any procedural consequence — no shortened clock, no abridged dossier and no substitute for ZAMRA's own evaluation.

  • Authority

    Marketing approval in a SADC state

    Accepted

    No

    Benefit

    Disclosure only. The application must list the SADC states where the device has obtained marketing approval and the dossier must include documentary proof of registration in other countries, specifically those in the SADC region. No SADC approval abridges the Zambian evaluation; there is no SADC device approval to rely on.

  • Authority

    ZAMRA or another recognised competent authority GMP inspection report for the same manufacturing site

    Accepted

    Yes

    Benefit

    Abridges the site-inspection step rather than the dossier review: for IVDs the Authority 'shall consider' and for non-IVDs it 'may consider' a previous satisfactory inspection report of the site, provided the inspection was conducted within 24 months preceding receipt of the application.

  • Authority

    China NMPA medical device registration certificate or Class I filing record

    Accepted

    No

    Benefit

    No route. The NMPA is not among the approvals the Annex I form asks the applicant to disclose, and no recognition procedure in the non-IVD or IVD guideline names it. An NMPA certificate may be filed as documentary proof of registration in other countries, which carries no procedural benefit, and it cannot substitute for the Free Sale Certificate, which must be addressed to Zambia from the manufacturer's own authority.

Key Requirements
  • An applicant not resident in Zambia must appoint a local responsible person with a power of attorney, defined as a natural or legal person resident in Zambia appointed by a foreign-based marketing authorisation holder to be responsible for all regulatory matters in respect of products granted marketing authorisation; the Annex I form separately requires the authorised local distributor's details and letter of appointment.
  • An original Free Sale Certificate addressed specifically to Zambia from the manufacturer's home authority (notarized copies acceptable), the manufacturer's manufacturing licence, documentary proof of ISO 13485 compliance and documentary proof of registration in other countries, specifically those in the SADC region.
  • The dossier follows the guideline's Chapters 1-6 table of contents for Classes B to D, or the shorter Annex II package for Class A, with a dedicated Chapter 4 on Clinical Evidence and a Chapter 2 that must include global market history with incident and recall rates.
  • A complete application on filing: the guideline states an application is not accepted where any one of the six listed items is missing — cover letter, completed and signed Annex I form, proof of payment of the prescribed fee, at least two samples in the smallest commercial pack size labelled primarily in English, the completed Annex II checklist, and the product dossier.
  • Applications and supporting documents must be in English, in 12-point Times New Roman or Arial at 1.5 line spacing with sequentially numbered pages; the non-IVD dossier is submitted as an electronic text-selectable PDF via the Authority's online portal, while IVD applications are physically delivered to the Authority.
  • Samples may be analysed by the National Drug Quality Control Laboratory against the manufacturer's claimed specifications and against available local or internationally recognised standards, and the marketing authorisation fees exclude laboratory testing and site audit fees, which are charged separately.
  • An applicant must facilitate site audits of its manufacturing site to demonstrate ISO 13485 compliance; where the Authority cannot inspect it shall (IVDs) or may (non-IVDs) consider previous satisfactory inspection reports from ZAMRA or other recognised competent authorities, provided the inspection was conducted within 24 months preceding receipt of the application.
  • A marketing authorisation is valid for five years from the date of grant subject to terms and conditions the Authority may specify, with an annual retention fee payable by the marketing authorisation holder; suspension or revocation may follow non-payment or an unforeseen high public or personal health risk.
  • Changes are made by amendment application, with the fee schedule distinguishing a minor amendment (US$ 100) from a major amendment (US$ 500) for an imported allied substance; the applicant remains responsible for the product information supporting the application and its variations.
  • Pre-market disclosure obligations: the Annex I form requires the applicant to state whether the device has had previous recalls, reportable adverse incidents, banning in other countries or post-market surveillance studies, and to attach the relevant documentation.
Local Rep Required

Yes

Local Rep Name

Local responsible person (resident in Zambia, appointed by the foreign marketing authorisation holder with a power of attorney); an authorised local distributor is named separately

Mdsap Accepted

No

Mdsap Benefit

MDSAP is not named in ZAMRA's device instruments. The recognised QMS evidence is an ISO 13485 certificate plus a site audit, or a ZAMRA or other recognised competent authority inspection report no more than 24 months old.

Registration Validity

5 years from the date of grant, subject to an annual retention fee

Last Updated

2026-09-16

Verification Status

partially-verified

Official Sources
  • Title

    ZAMRA, Guidelines on Application for Grant of Marketing Authorisation of Non In-Vitro Diagnostic Medical Devices for Human Use (version 2, implemented 21 June 2024) — section 1(h) and the definition of 'expedited evaluation'

    Url

    https://www.zamra.co.zm/wp-content/uploads/2026/06/GUIDELINES-ON-APPLICATION-FOR-GRANT-OF-MARKETING-AUTHORISATION-OF-NON-IVDS.pdf

    Accessed Date

    2026-09-16

  • Title

    ZAMRA, Guidelines for Grant of Marketing Authorisation of In-Vitro Diagnostic Devices for Human Use (ZAMRA/MA/GUD/0003, version 1.0) — sections 1.9.3 and 1.10.1-1.10.2

    Url

    https://www.zamra.co.zm/wp-content/uploads/2026/06/Guidelines-for-grant-of-Marketing-Authorisation-of-in-vitro-diagnostic-devices-for-Human-use.pdf

    Accessed Date

    2026-09-16

  • Title

    Guidelines on the Application for Marketing Authorisation of Condoms — ZAMRA guidance for the preparation and submission of dossiers, §1.12.1-1.12.2 Time Frames (acknowledgement within 14 days; review of a new application within 365 days on a first-come basis)

    Url

    https://www.zamra.co.zm/wp-content/uploads/2026/06/Application-for-Marketing-Authorisation-Condoms.pdf

    Accessed Date

    2026-09-16

  • Title

    ZAMRA Notice to all Manufacturers, Wholesalers, Retailers and Importers of Non-IVD Medical Devices in Zambia, 10 September 2024 — non-IVD device registration in effect on a phased basis from Class A and Class B, with import controls only for Class C and Class D until further notice

    Url

    https://www.zamra.co.zm/wp-content/uploads/2026/06/NOTICE-TO-ALL-IMPORTERS-OF-NON-IVDS.pdf

    Accessed Date

    2026-09-16

Reviewed On

2026-09-16

Process
Template Id

default

Status

planning-template

Usage Rule

Maintained planning steps and estimates, not verified regulator deadlines.

Steps
  • Step

    1

    Title

    Pre-Submission Preparation

    Description

    Gather technical documentation, quality management system (QMS) records, and clinical data required by ZAMRA. Appoint a Local responsible person (resident in Zambia, appointed by the foreign marketing authorisation holder with a power of attorney); an authorised local distributor is named separately.

    Duration

    1-3 months

    Icon

    📋

  • Step

    2

    Title

    Application Submission

    Description

    Submit complete registration dossier to ZAMRA including device description, risk analysis, and testing reports.

    Duration

    1-2 weeks

    Icon

    📤

  • Step

    3

    Title

    Administrative Review

    Description

    ZAMRA reviews application for completeness. May request additional documents or clarifications.

    Duration

    2-4 weeks

    Icon

    🔍

  • Step

    4

    Title

    Technical Evaluation

    Description

    In-depth technical assessment of device safety and performance. ZAMRA may issue questions.

    Duration

    180 working days

    Icon

    ⚙️

  • Step

    5

    Title

    Registration Approval

    Description

    Upon successful review, ZAMRA issues Zambia market authorization certificate.

    Duration

    1-2 weeks

    Icon

    ✅

Government fee schedules · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Zambia fees →
Jurisdiction Id

zambia

Country

Zambia

Country Code

ZM

Authority

ZAMRA

Low Risk Local
Min

500

Max

500

Currency

USD

Per Product

Yes

High Risk Local
Min

500

Max

500

Currency

USD

Per Product

Yes

Details
  • Id

    marketing-authorisation-application

    Item

    Application for marketing authorisation — allied substances imported as finished products

    Currency

    USD

    Amount Local

    500

    Amount Local Formatted

    US$ 500.00

    Frequency

    Per product application

    Notes

    Fee schedule item 3(a), imported-products table. Medical devices are 'allied substances' under the Medicines and Allied Substances Act No. 3 of 2013, so this allied-substance head is the operative device application fee; ZAMRA publishes no separate medical-device fee head. The fee covers evaluation of the initial submission only — laboratory testing and site audits are charged separately.

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.zamra.co.zm/wp-content/uploads/2026/07/FEES-FOR-MARKETING-AUTHORISATION-2.pdf

    Source Title

    ZAMRA, Fees for Marketing Authorisation — 'Application for Marketing Authorisation: Allied Substances Imported as Finished Products' and the other imported-products heads, reproducing the Medicines and Allied Substances (Fees) Regulations (S.I. No. 38 of 2016)

    Verified On

    2026-09-16

  • Id

    annual-retention

    Item

    Annual retention fee — allied substances imported as finished products

    Currency

    USD

    Amount Local

    200

    Amount Local Formatted

    US$ 200.00

    Frequency

    Annually, for the life of the marketing authorisation

    Notes

    Fee schedule item 5(e). Retention is payable annually even though the marketing authorisation runs five years; non-payment is a ground on which the Authority may suspend or revoke the authorisation and remove the device from the register.

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.zamra.co.zm/wp-content/uploads/2026/07/FEES-FOR-MARKETING-AUTHORISATION-2.pdf

    Source Title

    ZAMRA, Fees for Marketing Authorisation — 'Application for Marketing Authorisation: Allied Substances Imported as Finished Products' and the other imported-products heads, reproducing the Medicines and Allied Substances (Fees) Regulations (S.I. No. 38 of 2016)

    Verified On

    2026-09-16

  • Id

    marketing-authorisation-renewal

    Item

    Renewal of marketing authorisation — allied substances imported as finished products

    Currency

    USD

    Amount Local

    350

    Amount Local Formatted

    US$ 350.00

    Frequency

    Every five years

    Notes

    Fee schedule item 6(e).

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.zamra.co.zm/wp-content/uploads/2026/07/FEES-FOR-MARKETING-AUTHORISATION-2.pdf

    Source Title

    ZAMRA, Fees for Marketing Authorisation — 'Application for Marketing Authorisation: Allied Substances Imported as Finished Products' and the other imported-products heads, reproducing the Medicines and Allied Substances (Fees) Regulations (S.I. No. 38 of 2016)

    Verified On

    2026-09-16

  • Id

    late-renewal-penalty

    Item

    Late submission of an application for renewal of marketing authorisation

    Currency

    USD

    Amount Local

    5

    Amount Local Formatted

    US$ 5.00

    Frequency

    For each day the application is late

    Notes

    Fee schedule item 6(f), imported-products table.

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.zamra.co.zm/wp-content/uploads/2026/07/FEES-FOR-MARKETING-AUTHORISATION-2.pdf

    Source Title

    ZAMRA, Fees for Marketing Authorisation — 'Application for Marketing Authorisation: Allied Substances Imported as Finished Products' and the other imported-products heads, reproducing the Medicines and Allied Substances (Fees) Regulations (S.I. No. 38 of 2016)

    Verified On

    2026-09-16

  • Id

    evaluation-additional-information

    Item

    Evaluating additional information for an application — inadequate technical information (quality, safety or efficacy)

    Currency

    USD

    Amount Local

    400

    Amount Local Formatted

    US$ 400.00

    Frequency

    Per occurrence

    Notes

    Fee schedule item 4(a), imported-products table.

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.zamra.co.zm/wp-content/uploads/2026/07/FEES-FOR-MARKETING-AUTHORISATION-2.pdf

    Source Title

    ZAMRA, Fees for Marketing Authorisation — 'Application for Marketing Authorisation: Allied Substances Imported as Finished Products' and the other imported-products heads, reproducing the Medicines and Allied Substances (Fees) Regulations (S.I. No. 38 of 2016)

    Verified On

    2026-09-16

  • Id

    minor-amendment

    Item

    Minor amendment to a marketing authorisation

    Currency

    USD

    Amount Local

    100

    Amount Local Formatted

    US$ 100.00

    Frequency

    Per amendment

    Notes

    Fee schedule item 7(a), imported-products table. The criteria distinguishing a minor from a major amendment sit in ZAMRA's separate amendment guideline, which was not retrieved.

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.zamra.co.zm/wp-content/uploads/2026/07/FEES-FOR-MARKETING-AUTHORISATION-2.pdf

    Source Title

    ZAMRA, Fees for Marketing Authorisation — 'Application for Marketing Authorisation: Allied Substances Imported as Finished Products' and the other imported-products heads, reproducing the Medicines and Allied Substances (Fees) Regulations (S.I. No. 38 of 2016)

    Verified On

    2026-09-16

  • Id

    major-amendment

    Item

    Major amendment to a marketing authorisation

    Currency

    USD

    Amount Local

    500

    Amount Local Formatted

    US$ 500.00

    Frequency

    Per amendment

    Notes

    Fee schedule item 7(b), imported-products table.

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.zamra.co.zm/wp-content/uploads/2026/07/FEES-FOR-MARKETING-AUTHORISATION-2.pdf

    Source Title

    ZAMRA, Fees for Marketing Authorisation — 'Application for Marketing Authorisation: Allied Substances Imported as Finished Products' and the other imported-products heads, reproducing the Medicines and Allied Substances (Fees) Regulations (S.I. No. 38 of 2016)

    Verified On

    2026-09-16

Notes
  • Medical devices are regulated as 'allied substances' under the Medicines and Allied Substances Act No. 3 of 2013, so the allied-substance lines in ZAMRA's fee schedule are the device fee lines. ZAMRA publishes no separate medical-device fee head, and no class-differentiated fee: the application fee is the same for Class A and Class D.
  • The schedule runs two tables. Locally manufactured or packaged allied substances pay in Zambian Kwacha (ZMW 1,500 marketing authorisation, ZMW 1,000 annual retention, ZMW 1,200 renewal, ZMW 400 minor amendment, ZMW 1,950 major amendment, all expressed as fee units at 30 Ngwee per unit). Imported finished products — the table that applies to a foreign device manufacturer — are charged in United States dollars. This record publishes the imported-products table, so its currency is USD; the two tables cannot be mixed inside one record and the importer requires one currency throughout.
  • Fast tracking is charged at double the applicable marketing authorisation fee (fee schedule 'Other fees', item l). That is a published formula rather than a fixed head, so no exact row is asserted here: for an imported allied substance it resolves to US$ 1,000.00. The non-IVD guideline defines 'expedited evaluation' as express evaluation within a 90-day period and the Authority authorises it without published criteria.
  • Good-manufacturing-practice inspection fees for foreign-based manufacturers are set by region and are charged on top of the marketing authorisation fee: full site in Southern Africa US$ 3,500.00, rest of Africa US$ 5,000.00, Far East/Asia US$ 6,500.00, Europe/America/Australia US$ 7,500.00, each additional production line US$ 1,500.00, and GMP document evaluation (desk audit) US$ 3,500.00 per manufacturing site.
  • The 2016 instrument (S.I. No. 38 of 2016) was read by OCR from a scanned gazette PDF in the spine round and its US-dollar amounts were cross-checked against this current published schedule, where they agree unchanged.
  • Advertising and promotion of allied substances carries separate heads in the Kwacha table (ZMW 5,000 advertising to the general public, ZMW 1,000 promotion to health-care professionals, ZMW 2,000 exhibition at a public event). They are worded for medicines and are not published in the imported-products USD table, so they are not asserted as USD amounts here.
Last Updated

2026-09-16

Verification Status

official-verified

Official Sources
  • Title

    ZAMRA, Fees for Marketing Authorisation — 'Application for Marketing Authorisation: Allied Substances Imported as Finished Products' and the other imported-products heads, reproducing the Medicines and Allied Substances (Fees) Regulations (S.I. No. 38 of 2016)

    Url

    https://www.zamra.co.zm/wp-content/uploads/2026/07/FEES-FOR-MARKETING-AUTHORISATION-2.pdf

    Accessed Date

    2026-09-16

Reviewed On

2026-09-16

Amount Status

exact-current

Range Available

Yes

Additional certifications and licences · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Zambia certifications →
Slug

zambia

Market

Zambia

Confidence

low

Schemes
  • Id

    zamra-electrical-safety-emc-evidence

    Name

    ZAMRA electrical, radiation and standards evidence in the device dossier

    Category

    Medical Electrical Safety

    Status

    Included In Medical Registration

    Object

    Finished Device

    Note

    ZAMRA's Guideline on Application for Grant of Marketing Authorisation of Non In-Vitro Diagnostic Medical Devices, issued under s.68 of the Medicines and Allied Substances Act No. 3 of 2013, carries electrical and radiation safety as dossier evidence rather than as a separate certificate: the Chapter 3 Non-clinical Evidence block requires physical, chemical, electrical, radiation and software studies, and the Annex I application form requires the international or national standards with which the device complies. Zambia operates no low-voltage or EMC regime of its own for medical devices, so an applicant obtains no Zambian electrical-safety certificate — conformity is demonstrated inside the marketing authorisation, and it is the ZAMRA dossier that an applicant has to satisfy. The binding is recorded as candidate because the requirement is read from ZAMRA's own device guideline rather than from a dedicated electrical-safety instrument.

    Source Urls
    • https://www.zamra.co.zm/wp-content/uploads/2026/06/GUIDELINES-ON-APPLICATION-FOR-GRANT-OF-MARKETING-AUTHORISATION-OF-NON-IVDS.pdf
    Source Binding

    candidate

  • Id

    zicta-type-approval

    Name

    ZICTA equipment type approval

    Category

    Radio Spectrum

    Status

    Mandatory If Feature

    Object

    Covered Radio Or Terminal Equipment

    Note

    ZICTA states on its technical-regulation page that all telecommunications equipment and ICT devices must undergo type approval before being sold or used in Zambia, that equipment must comply with ZICTA's relevant technical standards, and that dealers must be registered with ZICTA; the department's mandate is described as overseeing the technical regulation aspects of the Information and Communications Technology Act No. 15 of 2009. A medical device with a radio or telecommunications function — a wirelessly connected patient monitor, a telemetry programmer or a connected IVD analyser — therefore needs a ZICTA type approval in addition to its ZAMRA marketing authorisation. ZAMRA's device guidelines do not mention ZICTA, so the two approvals are independent and neither substitutes for the other.

    Source Urls
    • https://www.zicta.zm/services/technical-regulation
    Source Binding

    reviewed

  • Id

    zcsa-compulsory-standards

    Name

    ZCSA compulsory standards

    Category

    Product Specific Conformity

    Status

    Mandatory At Import

    Object

    Product Under Compulsory Standard

    Note

    The Zambia Compulsory Standards Agency publishes the list of compulsory standards it administers. That list includes ZS ISO 4074: 2015 Natural Latex Rubber Condoms — Requirements and Test Methods, enforced under Statutory Instrument 78 of 2003, so a condom — a medical device class that ZAMRA also registers as an allied substance — must additionally satisfy the compulsory standard at import. The same list carries household electrical appliances, electric cables and batteries, but no medical electrical equipment standard such as the IEC 60601 series appears in it, so on the list as retrieved the agency's reach into devices is the condom entry rather than medical electrical equipment generally.

    Source Urls
    • https://www.zcsa.org.zm/index.php/list-of-compulsory-standards/
    Source Binding

    reviewed

  • Id

    rpa-ionising-radiation-licence

    Name

    Radiation Protection Authority ionising-radiation licence and permits

    Category

    Ionizing Radiation

    Status

    Mandatory For Use Site

    Object

    Radiation Product Importer Or Use Site

    Note

    The Radiation Protection Authority of Zambia licences and regulates facilities using radioactive materials and states that it issues authorisation — permits and licences — to deal with any activity relating to radiation sources, and that it is mandated by the Ionising Radiation Protection Act to maintain an inventory of radiation sources, devices and occupationally radiation exposed workers. The Authority's published renewal practice covers facilities using X-ray machines and radioactive sources, mines and industries using nuclear gauges and X-ray fluorescence analysers, baggage and cargo scanners, and exploration facilities. The obligation attaches to the importing or using facility rather than to the product's marketing authorisation, so a Zambian hospital, clinic or industrial site must hold a valid ionising-radiation licence before it operates the equipment. The binding is candidate because the Authority's licensing page names the Act without a section or a number, so no article is asserted.

    Source Urls
    • https://rpa.gov.zm/licensing/
    Source Binding

    candidate

Official Sources
  • Authority

    ZAMRA

    Topic

    marketing authorisation guideline for non-IVD medical devices — electrical, radiation and standards evidence in the dossier

    Url

    https://www.zamra.co.zm/wp-content/uploads/2026/06/GUIDELINES-ON-APPLICATION-FOR-GRANT-OF-MARKETING-AUTHORISATION-OF-NON-IVDS.pdf

  • Authority

    ZICTA

    Topic

    technical regulation and type approval of telecommunications and ICT equipment

    Url

    https://www.zicta.zm/services/technical-regulation

  • Authority

    ZCSA

    Topic

    list of Zambia national compulsory standards and their statutory instruments

    Url

    https://www.zcsa.org.zm/index.php/list-of-compulsory-standards/

  • Authority

    RPA

    Topic

    licensing of activities involving ionising radiation sources and devices

    Url

    https://rpa.gov.zm/licensing/