Changes after registration in Brunei

Source-linked rules for common changes to a registered IVD. Read the scope and conditions with each rule; unverified details are listed separately.

Does changing a device brand name or logo require an update?

A change to the registered brand name and a change to logo artwork can follow different rules. The legal manufacturer, device design and intended purpose remain the same.

A source-verified answer for this specific scenario is not yet published.

How do I update a manufacturer's name without changing the legal entity?

The same legal manufacturer changes its name, potentially alongside the device brand. A transfer to another legal entity is a separate question; permission to combine filings needs its own evidence.

A source-verified answer for this specific scenario is not yet published.

What changes when the manufacturer's office address changes?

An office address change for the same legal entity, potentially alongside name and brand changes. This does not cover moving a factory or manufacturing site, and a combined filing cannot be assumed.

A source-verified answer for this specific scenario is not yet published.

Does a catalogue or reference code change affect registration or UDI?

Check which identifier actually changes. A commercial catalogue, REF or order code is not automatically a device model, registered identifier, GTIN or UDI-DI. Rules for one identifier do not establish the route for another.

A source-verified answer for this specific scenario is not yet published.

Can several changes be filed together?

Changes to brand, manufacturer name, office address and reference codes may occur together, including with a real transfer to another legal manufacturer, registration holder or owner. Assess transfer with each single change, and transfer with brand plus reference, manufacturer name plus reference, or office address plus reference changes: seven transfer combinations in total. Evidence must establish the exact combination and the order of filings and approvals. These combinations and their sequence are not all verified by the rules below; a rule for one change does not prove that a single application, fee or review period covers the combination.

A source-verified answer for this specific scenario is not yet published.

Advertising, claims & promotion

Which claims, audiences, channels, approvals, and restrictions apply?

Compare countries

A medical advertisement (including for a device) may not claim to prevent, alleviate or cure any of the 22 Scheduled diseases, except in publications for health professionals

advertising.device-claims-for-scheduled-diseases-prohibited
Fact-level verified

Section 53(1)(a) of the Medicines Act prohibits publishing any medical advertisement that directly or indirectly claims, indicates or suggests that the article advertised will prevent, alleviate or cure any disease or condition in the Schedule, which lists 22 entries including AIDS, asthma, cancer, conception and pregnancy, deafness, diabetes, diseases of the eye, heart and kidney, epilepsy, hypertension, infertility, sexual function and tuberculosis. Advertisements distributed only to medical practitioners, dentists, pharmacists, nurses, midwives, veterinary surgeons and their trainees are exempt (s. 53(2)). The penalty is a fine of up to $5,000, up to 2 years' imprisonment or both (s. 53(3)).

Conditions
  • Because 'medical advertisement' covers devices used for a medicinal purpose, this prohibition is the one substantive statutory rule applying to device promotion in Brunei.
  • The Minister may exempt advertisements by Gazette order (s. 53(4)) and may amend the Schedule (s. 80).

Effective from 2010-07-01

The false-or-misleading advertising offence (s. 52) and the Minister's advertising-control orders (s. 54) are framed for medicinal products; no device-specific advertising order has been made

advertising.false-misleading-offence-medicinal-products-only
Fact-level verified

Section 52 of the Medicines Act makes it an offence to issue a false or misleading advertisement or representation 'relating to medicinal products', including an advertisement inconsistent with the information approved at registration under s. 11; it does not extend to devices. Section 54(1)(b) empowers the Minister by order to prohibit advertisements likely to lead to the use of 'any medicinal product, or any other substance or article' for treating or diagnosing specified diseases, and s. 54(2)(b) allows orders requiring prior approval of advertisements by the Authority; no such order reaching devices is listed among the Act's subsidiary legislation.

Conditions
  • Penalty under s. 52(7): fine up to $5,000, imprisonment up to 2 years or both.
  • Section 55 allows the Authority to require copies of advertisements.

Effective from 2010-07-01

The AGC's Medicines Act page lists only Rg 1, Rg 2 and Rg 3 as subsidiary legislation; an order under s. 54 would appear there or in the Gazette, and none was found.

'Medical advertisement' in the Medicines Act expressly covers any device, instrument, apparatus or contrivance used or represented for a medicinal purpose

advertising.medical-advertisement-definition-covers-devices
Fact-level verified

Part 8 of the Medicines Act defines 'medical advertisement' as an advertisement relating, or likely to cause a person to believe it relates, to any medicinal product 'or any device, instrument, apparatus or contrivance used or represented to be used for a medicinal purpose'. This is the only provision of Brunei law that expressly regulates medical devices, and it does so through advertising control rather than product approval.

Conditions
  • 'Advertisement' includes every form of advertising including internet and electronic transmission, but excludes spoken words other than recordings and broadcasts (s. 51(1)).
  • The identical definition appears in the Medicines Order, 2007 (S 79/2007), s. 51.

Effective from 2010-07-01

'Medicinal purpose' takes its meaning from s. 4(2): treating or preventing disease, diagnosing disease or ascertaining a physiological condition, contraception, inducing anaesthesia, or interfering with a physiological function.

Approval validity & renewal

What expires, what is retained periodically, and when must renewal start?

Compare countries

No device approval exists to expire or renew; validity rules apply to medicinal-product registrations (as specified in the product licence, else until cancelled) and one-year licences

validity.no-device-registration-validity-medicinal-rules
Fact-level verified

The only validity and renewal rules in Brunei health-product law are for medicinal products: the registration period is as specified in the product licence and, where none is specified, the registration is valid until cancelled (s. 11(6)-(8)); import, manufacturer's and wholesaler's licences are valid for one year and a clinical trial import licence for up to 3 years (s. 15(4)-(5)); licences are non-transferable (s. 15(6)). No validity period or renewal applies to a medical device because no device approval exists.

Conditions
  • Rg 2 Schedule 5 charges the product licence at no cost in the first year and $50 each subsequent year.

Effective from 2010-07-01

Verified negative across the whole Act.

Authority & regime

Which authority and binding regime govern this product, task, and date?

Compare countries

BDMCA's statutory functions are confined to medicinal products; no Brunei authority holds a medical-device mandate

authority.bdmca-mandate-medicinal-products-only
Fact-level verified

The Brunei Darussalam Medicines Control Authority established by s. 5 of the Medicines Act is the only product-regulatory authority created by Brunei statute, and s. 6 confines its functions to registration, licensing and listing of medicinal products and to the import, manufacture, distribution, labelling and disposal of medicinal products. The Act read in full confers no function over medical devices or in-vitro diagnostics on the Authority or on any other body.

Conditions
  • The Act's only sub-committee heads named in s. 6(e) are 'traditional or herbal medicines, cosmetics and the like'.
  • The Radiation Protection Act (Chapter 228) separately appoints an Authority over ionising-radiation apparatus (see authority.radiation-protection-authority-controlled-apparatus).

Effective from 2010-07-01

Verified negative: the whole Act was read; a case-insensitive search of its 106,966-character text layer finds 'device' once (s. 51 definition of 'medical advertisement') and 'medical device' never. The Authority's membership under s. 5(1) is the Director-General of Medical Services, the Director-General of Health Services, the Director of Pharmaceutical Services and at least five appointed members.

Change control

Does the event require internal documentation, notice, prior approval, transfer, or a new application?

Compare countries

No device variation procedure exists; change duties are the product licence holder's medicinal-product material-change notifications

change.no-device-variation-medicinal-material-change-duties
Fact-level verified

The only change-control rules in Brunei health-product law are for medicinal products: an applicant must notify changes to application information within 14 days (Act, s. 11(5)), and a product licence holder must forthwith inform the Authority of any material change in the specification, constituents, composition, manufacturing methods, quality procedures, storage arrangements, indications, label or leaflet of the medicinal product (Rg 2, Schedule 1, para. 2). The Ministry of Health's variation routes (MaV, MiV-PA, MiV-N) are for registered medicinal products. No variation or change-notification route exists for medical devices.

Conditions
  • Rg 2 Schedule 5 prices a major amendment of a product licence at $150 and a minor amendment at $50.

Effective from 2010-07-01

Verified negative: Rg 2 read in full; every standard provision is expressed for 'medicinal product'.

Classification rules

What local class follows from which rules and decisive product facts?

Compare countries

No device risk classification exists; the only product classes in Brunei law are the medicinal-product supply categories

classification.no-device-risk-class-axis
Fact-level verified

Brunei has no medical-device risk-class axis. The only classification in the Medicines Act is the supply classification of medicinal products into general-sale-list products (s. 24), products sold only from a registered pharmacy under a pharmacist's supervision (s. 25) and prescription-only products (s. 30); the Act and its regulations contain no class A to D, class I to IV or other device grading.

Conditions
  • Brunei signed but has not ratified the ASEAN Agreement on Medical Device Directive, so its class A to D rules do not apply (see the registry's reliance.asean-amdd-signed-not-ratified).
  • AITI grades radio equipment into three interference-risk levels (see classification.aiti-radio-equipment-interference-levels), which is not a device class.

Effective from 2010-07-01

Verified negative: the whole Act and Rg 1 to Rg 3 were read; no provision assigns a risk class to any instrument, apparatus or appliance.

Clinical investigation & performance

Which clinical or IVD performance evidence or study is required and accepted?

Compare countries

No clinical trial of 'any product' may be conducted without the BDMCA's prior written approval; the Act does not confine the definition to medicinal products

clinical.any-clinical-trial-requires-bdmca-written-approval
Fact-level verified

Section 23(1) of the Medicines Act prohibits conducting any clinical trial in Brunei Darussalam without prior written approval from the Brunei Darussalam Medicines Control Authority. The Act's s. 2(1) defines a 'clinical trial' as investigations on persons to determine 'the safety, effectiveness and other effects of any product', and 'product' is not defined, so the approval requirement is not on its face limited to medicinal products. No rules under s. 23(2) prescribing trial conditions, and no device-specific investigation guidance, have been published.

Conditions
  • The Act's long title and every other operative provision concern medicinal and cosmetic products; whether the Authority treats a device investigation as a 'clinical trial' is not addressed in any published guidance.
  • A clinical trial import licence under s. 15(1)(d) authorises import of 'one product' for trial purposes notwithstanding non-registration, valid up to 3 years (s. 15(5)); Rg 3 reg. 4 exempts trial products from certain label particulars.

Effective from 2010-07-01

Recorded as the text stands; the AGC's subsidiary-legislation page for the Act lists no clinical-trial rules. No performance-study or IVD-specific provision exists anywhere in the Act.

Device definition & scope

Is the product a device, IVD, accessory, software, or borderline product locally?

Compare countries

Brunei law defines no 'medical device'; the Medicines Act expressly excludes instruments, apparatus and appliances from 'medicinal product'

definition.no-statutory-device-definition-instrument-apparatus-excluded
Fact-level verified

No Brunei statute or regulation defines 'medical device' or 'in-vitro diagnostic'. The Medicines Act's s. 4(1) definition of 'medicinal product' excludes 'an instrument, apparatus or appliance' and requires that the substance or article be for use by being administered to human beings or animals, or as an ingredient of something so administered, so neither devices nor reagents used outside the body fall within the Act's registration regime.

Conditions
  • The identical exclusion appears in s. 4(1) of the Medicines Order, 2007 (S 79/2007).
  • The only positive statutory definition capturing any medical equipment is 'controlled apparatus' in the Radiation Protection Act (Chapter 228), s. 2(1).

Effective from 2010-07-01

Verified negative across the Act and all three of its regulations (Rg 1 Cosmetic Products, Rg 2 Licensing, Standard Provisions and Fees, Rg 3 Labelling), each read in full with zero occurrences of 'device'. Although s. 4(2)(b) names diagnosis as a medicinal purpose, s. 4(1) still requires administration to the body, which an in-vitro reagent does not satisfy; the Act publishes no guidance on that point.

Dossier requirements

Which evidence, local deltas, formats, and signatures are required?

Compare countries

No device dossier exists; the only application-content rule is the Authority's discretion over medicinal-product applications

dossier.no-device-dossier-medicinal-application-contents
Fact-level verified

The Medicines Act prescribes no dossier for any product beyond s. 11(2), under which an application to register a medicinal product must be accompanied by the prescribed processing fee and 'such documents, items, samples, particulars and other information as the Authority may require', and s. 13, under which the Authority may require a manufacturer's declaration of compliance with the laws of the country of manufacture. No technical-file, CSDT, STED or other device dossier requirement exists in Brunei law or Ministry of Health guidance.

Conditions
  • The ASEAN Common Submission Dossier Template does not apply because Brunei has not ratified the AMDD.

Effective from 2010-07-01

Verified negative across the whole Act; s. 11(3) allows the Authority to charge the cost of any laboratory investigation before registration of a medicinal product.

Economic-operator roles

Which manufacturer, representative, holder, applicant, importer, and distributor roles apply?

Compare countries

Brunei creates no manufacturer, authorised-representative, importer or distributor role for devices; the licensed roles exist only for medicinal products

roles.no-device-operator-roles
Fact-level verified

The only economic-operator roles in Brunei health-product law are those created by s. 15(1) of the Medicines Act: the holders of an import licence, a manufacturer's licence, a wholesaler's licence and a clinical trial import licence, each defined by reference to 'registered products', together with the product licence holder under s. 11(8). No Brunei instrument creates a local authorised representative, responsible person or registrant role for medical devices.

Conditions
  • The Radiation Protection Act imposes duties on 'licensees' holding a controlled-apparatus licence and notice duties on sellers and purchasers of controlled apparatus (see udi.radiation-controlled-apparatus-sale-purchase-notices).
  • AITI's dealer regime requires a Dealer Licence for commercial importers of radio equipment (pathways.aiti-type-approval-before-import).

Effective from 2010-07-01

Verified negative: the whole Act and Rg 2 (whose Schedules 1 to 4 set the standard provisions for product, import, wholesale dealer's and manufacturer's licences, all for medicinal products) were read.

Establishment, import & distribution

Which establishment licences, importer permissions, and distribution controls apply?

Compare countries

The customs controlled-goods table has no medical-device category; the Ministry of Health's permit category is 'Medicines, Herbal, Cosmetics'

establishment.customs-controlled-goods-no-device-category
Fact-level verified

The Royal Customs and Excise Department's table of goods requiring an import permit from a responsible agency contains no row for medical devices, medical equipment, surgical instruments or diagnostic kits. The only Ministry of Health Pharmaceutical Services category is 'Medicines, Herbal, Cosmetics', whose tariff list (HSC-K) covers chapters 27 to 38 only and contains no heading of chapter 90 or headings 3005, 3006 or 4015. General (non-radiating, non-radio) medical devices are therefore imported on an ordinary customs declaration without a health-agency permit.

Conditions
  • Two device sub-classes are controlled under other rows: X-ray and CT apparatus of heading 9022 (Radiation Safety and Quality Unit, list HSC-N) and radio or communications equipment (AITI, list HSC-G).
  • Legal basis stated on the page: 'All goods may be imported or exported except for prohibited goods under Section 31 of the Customs Order, 2006.'

Verified negative: all 30 rows of the table were read on 2026-09-16, and the linked HSC-K PDF (r210-bn-customs-hsc-k-medicine) was read in full: its tariff lines fall only in headings 2711, 2801-2853, 2901-2942, 3001-3004, 3301-3306 and 3824. The HSC-G telecommunications list covers headings 8471, 8517-8531, 8543, 8802, 9015, 9102 and 9503, none of which is a medical heading.

The Ministry of Health import and wholesaler's licence is required for medicinal products only

establishment.medicinal-import-licence-not-devices
Fact-level verified

The Department of Pharmaceutical Services' licensing guideline states that all medicinal products are regulated under the Medicines Order, 2007, that the Import Licence and Wholesaler's Licence are required for the importation, storage and supply of medicinal products, and that the licence authorises the import, storage and wholesale of registered medicinal products. No establishment licence, import licence or wholesale licence is required for a medical-device importer or distributor.

Conditions
  • Rg 2 reg. 3 makes an import licence granted to a person not authorised by the product licence holder a per-consignment licence.

Verified negative: the guideline (27,574 characters) was read in full and contains no occurrence of 'device'.

Fees

What regulator fee applies to this class, route, event, and date?

Compare countries

No medical-device fee exists; the only health-product tariff (Rg 2, Schedule 5) prices medicinal-product licences and certificates

fees.no-device-fee-medicines-tariff-heads
Fact-level verified

Schedule 5 of the Medicines (Licensing, Standard Provisions and Fees) Regulations is Brunei's only fee tariff for health products. Its five heads are product licence (processing fee $200; first year no charge; $50 each subsequent year; major amendment $150, minor $50), import licence, wholesale dealer's licence ($100 per year), manufacturer's licence and certificates (certificate to export a medicinal product $50; statement of licensing status $25). No head prices any medical-device application, because none exists.

Conditions
  • Fees are non-refundable (reg. 4(2)); the cost of any assessment or evaluation of a medicinal product is borne by the applicant unless the Authority requires otherwise (reg. 5).
  • Amounts are in Brunei dollars.

Effective from 2010-07-01

Verified negative: the Regulations were read in full and contain no occurrence of 'device'.

Label, IFU & language

Which label and IFU elements, languages, symbols, and translation controls apply?

Compare countries

No device labelling or IFU rule exists; the only labelling instrument covers medicinal products and its sole language rule is an English statement of certain substances

label.no-device-labelling-rule-medicinal-english-only
Fact-level verified

Brunei's only product-labelling instrument is the Medicines (Labelling) Regulations, made under s. 81(1) of the Medicines Act, whose rules cover dispensed medicinal products, clinical-trial products, products containing Schedule 1 substances (which must carry a statement in English declaring the substance) and date-stamping of Schedule 2 products. Sections 46 and 47 of the Act confine the labelling and leaflet rule-making powers to medicinal products. No labelling, instructions-for-use or language requirement exists for medical devices or IVDs.

Conditions
  • The Act's rule-making power in s. 81(2) could extend regulations to non-medicinal articles used for a medicinal purpose, but no such regulation has been made (the AGC lists only Rg 1 to Rg 3 under the Act).

Effective from 2010-07-01

Verified negative: the Regulations (13,205 characters) and the Act's Part 7 (ss. 46 to 50) were read in full; every provision is expressed in terms of 'medicinal product'.

Manufacturer PMS

Which PMS plan, report, PMCF or PMPF, trend, and lifecycle-update duties apply?

Compare countries

No post-market surveillance duty applies to device manufacturers; the licence-holder duties in Rg 2 attach to medicinal products

pms.no-device-pms-duty-medicinal-licence-holder-duties
Fact-level verified

The only post-market duties on manufacturers in Brunei health-product law are the standard provisions of the product licence in Rg 2 Schedule 1: to inform the Authority forthwith of information casting doubt on the safety, quality or efficacy data submitted (para. 3), to report adverse-effect reports within 7 days (para. 4), to withhold or withdraw batches found harmful or non-conforming if directed (para. 6) and to notify any withdrawal decision (para. 7), each expressed for 'any medicinal product to which the licence relates'. No manufacturer post-market surveillance, periodic safety update or trend-reporting duty exists for medical devices or IVDs.

Conditions
  • The Pharmacovigilance Guidelines (Part 2) elaborate these duties for licensed manufacturers, wholesalers, importers and product licence holders of medicinal products.
  • Radiation Protection Act licensees owe continuing safety duties to employees and third parties (ss. 10 and 11).

Effective from 2010-07-01

Verified negative: Rg 2 read in full.

QMS, audit & inspection

Which QMS, audit, inspection, and certificate scope is accepted or required?

Compare countries

No ISO 13485 or GMP requirement applies to devices; the Act's pre-licence inspections and inspection powers run to medicinal products

qms.no-device-qms-inspection-medicinal-licence-inspection
Fact-level verified

The only quality-system control in the Medicines Act is that import, manufacturer's and wholesaler's licences for medicinal products 'shall only be issued after an inspection' (s. 15(1)), and the inspection power in s. 60(1) is limited to substances or articles appearing to be medicinal products, their containers, labels and leaflets, and plant used to manufacture medicinal products. No quality-management-system, ISO 13485, GMP certificate or audit requirement applies to a medical-device manufacturer, importer or distributor.

Conditions
  • AITI may direct an importer of radio equipment to produce it for testing or inspection (qms.aiti-testing-and-inspection-direction).

Effective from 2010-07-01

Verified negative across the whole Act; s. 36(2)(c) allows rules on 'any apparatus, equipment, furnishings or utensils used in any such premises' where medicinal products are made or stored, which is a premises rule, not a device rule.

Recall, FSCA & market surveillance

Which recall, FSCA, notice, and authority-coordination duties apply?

Compare countries

The Authority's statutory recall and disposal directions (s. 75) relate to registered medicinal products; no device recall or FSCA power exists

recall.act-recall-directions-registered-medicinal-products-only
Fact-level verified

Section 75(1) of the Medicines Act empowers the Authority to issue directions, in particular relating to the recall from the market and the disposal of any registered medicinal product, and s. 75(2) makes contravention an offence. The Act contains no recall, field safety corrective action or market-withdrawal power over medical devices, and its seizure and forfeiture powers (ss. 60(4), 62 to 64) operate on articles connected with offences under the Act, which a device cannot be.

Conditions
  • AITI may direct a person to cease operating radio equipment (qms.aiti-testing-and-inspection-direction).
  • Radiation Protection Act Part 8 gives authorised officers remedial-order and stop-work-order powers over licensees (ss. 21 to 22).

Effective from 2010-07-01

Verified negative across the whole Act.

Brunei's recall guidance (February 2022) applies to registered and specially-imported medicinal products only

recall.guidance-2022-scope-medicinal-products-only
Fact-level verified

The Ministry of Health's Guidance for Medicinal Product Defect, Quarantine and Recall states that it applies to licensed manufacturers, wholesalers, importers and product licence holders for registered medicinal products and unregistered medicinal products imported via import permit on special approval, and reproduces the Medicines Order's exclusion of instruments, apparatus and appliances. Its recall classes (Class 1 to 3), recall levels and defect-reporting forms (Annex II to IV) therefore do not apply to medical devices.

Conditions
  • Defect reports for medicinal products go to the Pharmacovigilance Section at productdefect.pharmacy@moh.gov.bn.

Effective from 2022-02-01

Verified negative: the guidance (97,930 characters) was read in full with zero occurrences of 'device'.

Registration pathways

Which registration, notification, listing, reliance, or conformity route applies?

Compare countries

The only pre-market authorisation duty in Brunei law attaches to medicinal products; devices have no registration, notification or licence route

pathways.no-device-premarket-authorisation
Fact-level verified

Section 10(1) of the Medicines Act prohibits the manufacture, sale, supply or import of an unregistered 'medicinal product', and s. 11(1) empowers the Authority to register only 'any medicinal product'. Because s. 4(1) excludes instruments, apparatus and appliances from that term, no pre-market registration, notification or product licence exists or can be applied for in respect of a medical device or IVD.

Conditions
  • Two sector regimes gate specific equipment at the border rather than by product registration: the Radiation Protection Act licence for controlled apparatus (pathways.radiation-licence-for-controlled-apparatus) and AITI type approval for radio equipment (pathways.aiti-type-approval-before-import).

Effective from 2010-07-01

Verified negative: the whole Act was read. Sections 56 and 81(2) allow the Minister by order or regulation to extend provisions of the Act to 'articles or substances which are not medicinal products' used for a medicinal purpose; the AGC's subsidiary-legislation page for the Act lists only Rg 1 (Cosmetic Products), Rg 2 and Rg 3, none of which extends the Act to devices.

Review clocks

What review, validation, and applicant stop-clock rules apply?

Compare countries

The Medicines Act sets no review period for any application and lets the Authority reject without reasons; no device review clock exists

clock.no-statutory-review-period-any-product
Fact-level verified

The Medicines Act contains no time limit within which the Authority must decide an application for registration or a licence; the only periods it fixes are the applicant's 14-day duty to notify changes to application information (s. 11(5)) and the 14-day appeal window (s. 21). Section 14 allows the Authority to reject a registration application without assigning any reason. No published review timeline exists for medical devices because no device application exists.

Conditions
  • AITI publishes processing times for radio-equipment type approval (clock.aiti-type-approval-processing-times).

Effective from 2010-07-01

Verified negative: every section of the Act was read for a decision deadline; none exists. The Ministry of Health publishes no target timelines for medicinal-product registration either.

Special access & derogation

Is an emergency, compassionate, custom, humanitarian, or other exceptional path available?

Compare countries

No special-access, emergency-use or derogation route exists for devices; the Act's exemptions attach to medicinal products

special.no-device-special-access-medicinal-exemptions
Fact-level verified

The exemptions in s. 18 of the Medicines Act (research or registration samples, pharmacy dispensing, teaching institutions, clinical-trial or registration samples, and manufacture or import 'solely for the purpose of treatment of any person suffering from a life-threatening illness') and the Ministry of Health's Import Permit with Special Approval all relieve the s. 10(1) duty for medicinal products. Because devices are outside s. 10(1) altogether, no special-access, named-patient or emergency-use derogation exists or is needed for a medical device.

Conditions
  • Section 76 lets the Minister exempt any person or class of persons from any provision of the Act by Gazette notification.
  • AITI's Personal Type Approval is a single-use import route for personally used radio equipment (§4.7).

Effective from 2010-07-01

Verified negative across the whole Act. Section 10(3) additionally exempts personal-luggage imports of up to one month's supply, public officers on duty and consented transit.

Standards, testing & ancillary approvals

Which standards, tests, and radio, electrical, metrology, or other overlays are triggered?

Compare countries

Brunei recognises no device standards; the Act's only standards clause enforces pharmacopoeial monographs for medicinal products

standards.no-device-standards-pharmacopoeial-only
Fact-level verified

The only standards provision in Brunei health-product law is s. 35 of the Medicines Act, which requires a medicinal product sold under a monograph name to comply with the monograph in the British, European or United States Pharmacopoeia or other named publications. No Brunei instrument recognises ISO, IEC or other standards for medical devices, and no device testing or certificate is required by the Ministry of Health.

Conditions
  • Sector regimes name standards for their own objects: AITI requires conformity evidence to 'international standards ... recognised by the Authority' for Level 2 and 3 radio equipment, and AENBD's guideline lists accepted foreign electrical safety marks (standards.aenbd-accepted-electrical-safety-marks).
  • Ministry of Health tenders may specify standards as procurement conditions (registry fact reliance.procurement-international-safety-standard).

Effective from 2010-07-01

Verified negative across the whole Act and Rg 1 to Rg 3. The s. 35(4) list continues beyond the excerpt with further named compendia.

Submission communications & appeal

How are deficiencies answered, clocks handled, and decisions escalated or appealed?

Compare countries

The only appeal in the Medicines Act is a 14-day written appeal to the Minister against suspension, cancellation or revocation decisions on medicinal products; no device submission or appeal route exists

submission.no-device-appeal-medicinal-14-day-appeal-to-minister
Fact-level verified

Section 21 of the Medicines Act gives a person aggrieved by an Authority decision under s. 20 (suspension or cancellation of a medicinal-product registration, or suspension, revocation or variation of a licence) a written appeal to the Minister within 14 days of the decision being made known, the Minister's decision being final. No appeal lies against a rejection under s. 14, which may be made without reasons. Because no device application, deficiency or decision process exists, no device communication or appeal route exists.

Conditions
  • Medicinal-product submissions are lodged by appointment at the Drug Registration Unit on fixed weekdays (forms.moh-forms-medicinal-cosmetic-poisons-only).
  • Radiation Protection Act decisions carry a separate 30-day appeal (submission.radiation-licence-appeal-30-days).

Effective from 2010-07-01

Verified negative across the whole Act; s. 44 provides a separate appeal on pharmacy disqualification.

UDI, listing & traceability

Which UDI, device listing, database, and traceability duties apply?

Compare countries

No UDI, device listing or traceability duty exists; the statutory register covers medicinal and cosmetic products and the only identifier is the licence number on medicinal-product labels

udi.no-device-udi-or-listing-medicinal-register
Fact-level verified

Section 12 of the Medicines Act requires the Authority to keep a register of medicinal products and cosmetic products, and Rg 2 Schedule 1 para. 8 requires the product licence holder to state the licence number on the label and package. No unique device identifier, device listing, registration number or traceability requirement applies to medical devices or IVDs in any Brunei instrument.

Conditions
  • Rg 2 Schedules 1 to 3 require licence holders to keep importation, sale and supply records of medicinal products for 2 years for inspection.
  • Traceability duties for ionising-radiation apparatus exist under the Radiation Protection Act (udi.radiation-controlled-apparatus-sale-purchase-notices).

Effective from 2010-07-01

Verified negative across the whole Act and Rg 1 to Rg 3; the Ministry of Health publishes lists of registered medicinal products, notified cosmetic products and approved traditional medicines and health supplements, and no device list.

Vigilance & incident reporting

What is reportable, by whom, where, in what format, and by what deadline?

Compare countries

The statutory adverse-reaction reporting duty (s. 74) attaches to registered medicinal products; no device incident-reporting duty exists

vigilance.act-adverse-reaction-duty-registered-medicinal-products-only
Fact-level verified

Section 74 of the Medicines Act requires a licensed manufacturer, wholesaler, importer or product licence holder to notify the Authority immediately of any adverse reactions arising from the use of the registered medicinal product. The Act read in full contains no adverse-event, incident or malfunction reporting duty for medical devices or IVDs, and no timeline or form for such reports exists.

Conditions
  • Rg 2 Schedule 1 para. 4 fixes a 7-day window for product licence holders to pass on adverse-effect reports for medicinal products.
  • The Pharmacovigilance Guidelines set 7-calendar-day and 8-day follow-up timelines for serious ADRs to medicinal products (vigilance.pv-guidelines-scope-and-timelines-medicinal).

Effective from 2010-07-01

Verified negative: whole Act read; this upgrades the registry's earlier partial fact vigilance.no-device-adverse-event-reporting to a statute-level verified negative.

The Pharmacovigilance Guidelines define their subject by the Medicines Order's 'medicinal product' definition and set 7-calendar-day reporting for serious ADRs; no device vigilance guidance exists

vigilance.pv-guidelines-scope-and-timelines-medicinal
Fact-level verified

Brunei's Pharmacovigilance Guidelines adopt the Medicines Order, 2007 s. 4 definition of 'medicinal product', which excludes 'an instrument, apparatus or appliance', and require licence holders to report serious adverse drug reactions to the National Adverse Drug Reaction Monitoring Centre no later than 7 calendar days after first awareness, with follow-up within 8 additional calendar days. The Guidelines contain no device incident-reporting provision.

Conditions
  • Reporting contact stated in the Guidelines: nadrmc.dps@moh.gov.bn; the Suspected ADR Report Form is published on the Ministry of Health site.

Effective from 2018-07-31

Verified negative: the Guidelines (56,648 characters) were read in full with zero occurrences of 'device'. The Table 1 cells are fragmented in the text layer; the timeline wording is repeated in the body: 'The relevant licence holders must write to NADRMC no later than 7 calendar days after first awareness' and 'within 8 additional calendar days from when the first report was submitted'.

Related structured datasets

Classification systems · 0 records

No maintained record.

Registration pathways, prior approvals and planning · 0 records

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

No maintained record.

Government fee schedules · 0 records

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

No maintained record.

Additional certifications and licences · 0 records

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

No maintained record.

DJ Fang
Need to apply Brunei's requirements to a product?
Confirm the current Ministry of Health pathway, evidence, and local-role details for the actual device before filing.
Pricing