🇮🇩 Indonesia IVD regulatory data

Explore the IVD requirements currently available for Indonesia, organized by topic and linked to official sources.

12 facts7 dimensions5 official sourceslatest verification 2026-10-26
Explore questions and sources →

Common questions about changing a registered IVD in Indonesia. Each question separates published rules from details still awaiting verification.

Change control

Does the event require internal documentation, notice, prior approval, transfer, or a new application?

Compare countries

Administrative changes affecting all manufacturer labelling

change.all-product-administrative-notification-2025
Fact-level verified

For administrative changes affecting labelling of all products of a manufacturer, the marketing-authorization holder must notify the Ministry of Health.

Conditions
  • This clause alone does not establish that every brand, manufacturer-name/address or ownership change qualifies as administrative.
  • Apply the PMK11/2025 imported-device PB UMKU standard to its governing route; Article37 preserves specified pending/unverified/not-effective applications under the PMK14/2021 framework until the adjusted OSS system. Do not assume the same regime for every existing licence.

2021 guide: historical simultaneous-change submission rule

change.combination-highest-route-guide2021
Fact-level verified

The2021 guide says each simultaneous change is assessed separately, the changes may be submitted together with an explanation of the differences, and the registration type follows the highest route triggered by the changes. Its example combines a sterilization change(new application), package-layout variation and manufacturer-logo notification into one new application.

Conditions
  • One product with multiple changes; not blanket grouping across unrelated licences.
  • Legacy guide: verify compatibility with the current PMK11/2025 and PMK5/2026 framework.
  • The source describes the historical guide itself; current compatibility with later binding rules must be checked before relying on it as a current filing permission.

New IVD nomenclature and existing permits

change.ivd-nomenclature-circular2026
Fact-level verified

CircularFR.03.01/E.V/2204/2026 changes IVD category/subcategory/product-type/risk-class nomenclature per its annex. Already issued valid NIEs with the former nomenclature remain valid until expiry; holders may seek nomenclature adjustment through variation or renewal-with-variation.

Conditions
  • Source issued16October2026; portal posted20October2026.
  • Apply actual listed IVD product rows; do not infer that all IVD classes changed or that old approvals were invalidated.
  • Authority IVD nomenclature/classification adjustment only; not a rule classifying a commercial SMN/REF change or a substantive model/design change.

2021 guide: historical name, address, legal-entity and code examples

change.name-address-code-guide2021
Fact-level verified

The 2021 guide Section2.4.8 lists principal name/address, manufacturer legal entity, manufacturer address without physical relocation, product-owner name/address without changing manufacturer name/address, producer name without address or NIB change, manufacturer name without address or product change, product codes/types and specified distributor identity changes as variation examples where safety, quality, performance and specification are unaffected. The producer-name and distributor-name/address cases in letters(i),(k),(l) require a prior manual company-database request. Section2.4.9 separately lists product-name changes, distributor replacement/transfer and manufacturer/production-location replacement among new applications.

Conditions
  • Guidance predates PMK11/2025 and PMK5/2026; current binding rules prevail and any conflicting details need explicit reconciliation.
  • Do not conflate non-relocation address edits or legal-entity-name changes with acquisition/new manufacturer identity.
  • This is an attributed historical-guide finding. Read its Section2.4.8 examples with DiagramG and Section2.4.9; current case classification under later binding legislation is not certified here.

Stock depletion after labelling or agency transition

change.stock-depletion-2025
Fact-level verified

Old-label stock is given6months from approval of the new label; stock at the old distributor after an agency transition is given6months from issuance of the new authorization. Stock depletion requires Ministry approval of the operator application, and the application may be submitted only while the product authorization remains valid.

Conditions
  • Distinct from the separate import-to-domestic6month rule.
  • No automatic grace period without the stated approval.
  • Apply the PMK11/2025 imported-device PB UMKU standard to its governing route; Article37 preserves specified pending/unverified/not-effective applications under the PMK14/2021 framework until the adjusted OSS system. Do not assume the same regime for every existing licence.
  • Apply only to the actual approved labelling change or agency transition stated in Section5.g; a name, address or code edit does not by itself establish that this stock-depletion branch applies.

Imported-device technical-data changes

change.technical-data-2025
Fact-level verified

The technical annex to an imported-device permit records trade/brand name in relation to its written wording, type/size, packaging, producer/product owner, labelling, the accessory annex and other listed data. A technical-annex change is applied for through OSS and administratively/documentarily verified by the Ministry. Changes to business-operator/business-activity data follow the applicable legislation.

Conditions
  • Imported-device PB UMKU only.
  • Do not interpret correction of brand-name writing as blanket approval of a substantive brand or manufacturer replacement.
  • Apply the PMK11/2025 imported-device PB UMKU standard to its governing route; Article37 preserves specified pending/unverified/not-effective applications under the PMK14/2021 framework until the adjusted OSS system. Do not assume the same regime for every existing licence.

Dossier requirements

Which evidence, local deltas, formats, and signatures are required?

Compare countries

Documents for imported-device variations

change.dossier-2025
Fact-level verified

A variation application includes the old licence and annex, new labelling, a no-change-to-safety/benefit/quality statement, applicable distributor/CDB or repackaging CPB documents, agency appointment legalized by an Indonesian government representative or Apostille, licence/CFS documents, safety-event statement or event report, English and Indonesian manual/package insert/IFU, applicable nuclear permit, accessory/variant/type-code list and evidence specific to the proposed technical change.

Conditions
  • Imported-device variation; apply conditional documents only to their stated trigger.
  • Apostille is an express alternative here; do not retain an exclusive embassy-legalization rule from superseded standards.
  • Apply the PMK11/2025 imported-device PB UMKU standard to its governing route; Article37 preserves specified pending/unverified/not-effective applications under the PMK14/2021 framework until the adjusted OSS system. Do not assume the same regime for every existing licence.

Label, IFU & language

Which label and IFU elements, languages, symbols, and translation controls apply?

Compare countries

Relabelling after permit or distributor identity changes

change.restickering-2025
Fact-level verified

Relabelling/restickering is described for changes to the permit number and/or distributor name/address printed on packaging, and requires Ministry approval of the operator application.

Conditions
  • Does not establish an unrestricted right to relabel a foreign manufacturer identity change.
  • Apply the PMK11/2025 imported-device PB UMKU standard to its governing route; Article37 preserves specified pending/unverified/not-effective applications under the PMK14/2021 framework until the adjusted OSS system. Do not assume the same regime for every existing licence.
  • Only the changed permit number and/or distributor name/address on packaging triggers the cited branch; this is not unrestricted brand/foreign-manufacturer relabelling permission.

Registration pathways

Which registration, notification, listing, reliance, or conformity route applies?

Compare countries

Possibility of product grouping

change.grouping-2025
Fact-level verified

Some medical-device products can be registered in one product package/grouping.

Conditions
  • Grouping eligibility is product-specific; this clause does not permit consolidation of arbitrary independent licences.
  • Apply the PMK11/2025 imported-device PB UMKU standard to its governing route; Article37 preserves specified pending/unverified/not-effective applications under the PMK14/2021 framework until the adjusted OSS system. Do not assume the same regime for every existing licence.

Reliance & recognition

Which foreign decisions or evidence can be reused, with what effect and conditions?

Compare countries

Existing records include acceptance, restrictions and evidence reuse. The applicable product scope and conditions are preserved below; a reference does not by itself confer market access.

Indonesia approvals and evidence in other markets (0 records)

No linked record is currently maintained. This does not establish that a pathway is unavailable.

Prior approvals and evidence for Indonesia (8 records)

Regulatory reference

Indonesia import permit required; NMPA reliance eligibility not resolved by this clause

reliance.china-nmpa-no-recognition-route
Pending verification
Statement

PMK 5/2026 Pasal 58(4) requires the Indonesian izin edar before importing medical devices and PKRT, subject to the applicable exceptions. This establishes a local permit requirement. It does not determine whether reliance, evidence reuse or an abridged assessment may be available in obtaining that permit; the cited clause alone cannot support a country-wide finding that NMPA evidence has no benefit.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

indonesia

China Classes In Scope

Medical devices and PKRT within Pasal 58; local classification and application pathway require separate assessment

Edge Type

Eligibility Unresolved

Market Access Strength

undetermined

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Pasal 58(5) allows an import certificate in specified circumstances; Pasal 62 addresses special access.
  • Do not infer absence of reliance from a local permit requirement.
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Effective from 2026-05-04

Independent review of PMK 5/2026, Pasal 58(4)-(5) and 62, from the Ministry JDIH PDF; source-backed-partial reflects unresolved NMPA pathway eligibility.

IndonesiaEU conformity assessment

Prior approval / evidence record

EU CE Mark

Authority

EU CE Mark

Accepted

Yes

Benefit

Supports the Certificate of Free Sale and the technical evidence in the CSDT, but Kemenkes publishes no shorter period for it

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndonesiaUS FDA

Prior approval / evidence record

FDA 510(k)

Authority

FDA 510(k)

Accepted

Yes

Benefit

Supports the Certificate of Free Sale and the technical evidence in the CSDT, but Kemenkes publishes no shorter period for it

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndonesiaUS FDA · EU conformity assessment · ASEAN

Prior approval / evidence record

Class C

Id

class-c

Class Name

Class C

Risk Level

high

Pathway

Full Registration

Official Timeline
Min

30

Max

30

Unit

working-days

Verification Status

official-verified

Official Source
Title

Kemkes Regalkes — evaluation guideline Annex 8 service-standard clocks

Url

https://regalkes.kemkes.go.id/informasi_alkes/Indonesia%20Guideline%20for%20Evaluation%20of%20Medical%20Device%20and%20IVD%202020.pdf

Accessed Date

2026-08-25

Realistic Timeline
Min

4

Max

6

Unit

months

Prior Approval Benefits
  • ASEAN MRA countries
  • FDA
  • EU CE Mark
Notes
  • Official service-standard clock for imported Class C izin edar (new application): 30 working days of evaluation (Annex 8)
  • Clinical data may be required
Clinical Trial Required

No

Local Testing Required

Yes

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Prior approval / evidence record

ASEAN member-state registration

Authority

ASEAN member-state registration

Accepted

No

Benefit

There is no ASEAN mutual recognition agreement for device registration — the AMDD harmonises the dossier, not the decision

Timeline Savings

None

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Prior approval / evidence record

ASEAN Medical Device Directive (AMDD) harmonisation

Name

ASEAN Medical Device Directive (AMDD) harmonisation

Description

The AMDD gives the ten member states one device definition, one set of classification rules, one set of essential principles and one dossier structure, the Common Submission Dossier Template. It is not a reliance route. The ACCSQ Medical Device Product Working Group states in terms that the Directive is "not intended to confer mutual recognition" and that each member state retains the sovereign right to make its own registration decisions, so a registration in one ASEAN market authorises nothing in another and shortens no national review period.

Timeline Reduction

None — the saving is in dossier preparation, since one CSDT core can be reused across member states with national addenda

Eligibility
  • Any device filed in an ASEAN member state that has transposed the AMDD into national law

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndonesiaUS FDA · EU conformity assessment · ASEAN

Prior approval / evidence record

Class D

Id

class-d

Class Name

Class D

Risk Level

highest

Pathway

Full Registration

Official Timeline
Min

45

Max

45

Unit

working-days

Verification Status

official-verified

Official Source
Title

Kemkes Regalkes — evaluation guideline Annex 8 service-standard clocks

Url

https://regalkes.kemkes.go.id/informasi_alkes/Indonesia%20Guideline%20for%20Evaluation%20of%20Medical%20Device%20and%20IVD%202020.pdf

Accessed Date

2026-08-25

Realistic Timeline
Min

5

Max

7

Unit

months

Prior Approval Benefits
  • ASEAN MRA countries
  • FDA
  • EU CE Mark
Notes
  • Official service-standard clock for imported Class D izin edar (new application): 45 working days of evaluation (Annex 8)
  • Clinical evidence required
Clinical Trial Required

Yes

Local Testing Required

Yes

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndonesiaUS FDA · EU conformity assessment · ASEAN

Prior approval / evidence record

Class B

Id

class-b

Class Name

Class B

Risk Level

medium

Pathway

Registration

Official Timeline
Min

30

Max

30

Unit

working-days

Verification Status

official-verified

Official Source
Title

Kemkes Regalkes — evaluation guideline Annex 8 service-standard clocks

Url

https://regalkes.kemkes.go.id/informasi_alkes/Indonesia%20Guideline%20for%20Evaluation%20of%20Medical%20Device%20and%20IVD%202020.pdf

Accessed Date

2026-08-25

Realistic Timeline
Min

4

Max

6

Unit

months

Prior Approval Benefits
  • ASEAN MRA countries
  • FDA
  • EU CE Mark
Notes
  • Official service-standard clock for imported Class B izin edar (new application): 30 working days of evaluation (Annex 8)
  • ASEAN harmonization applies
Clinical Trial Required

No

Local Testing Required

Yes

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Review clocks

What review, validation, and applicant stop-clock rules apply?

Compare countries

Published imported-device service windows

change.service-times-2025
Fact-level verified

The imported-device standard lists maximum issuance windows: new class A 21, B 36, C 36 and D 55 working days; renewal 13 working days; variation and renewal-with-variation 16 working days. Notification-permit services have a separate 12-working-day window. Applicant corrections may be requested up to twice, each at most 10 working days, or 15 for class D. Hari means the working days prescribed by the central government.

Conditions
  • Published regulatory service windows, not empirical best/expected/worst predictions or a guaranteed all-in elapsed project duration.
  • Do not count preparation, prerequisite certificates or applicant corrections as automatically included or invent a start trigger not stated in the cited clause.
  • Apply the PMK11/2025 imported-device PB UMKU standard to its governing route; Article37 preserves specified pending/unverified/not-effective applications under the PMK14/2021 framework until the adjusted OSS system. Do not assume the same regime for every existing licence.
  • Use the actual eligible registration type and class. A company/holder transfer or substantive identity change is not automatically a 16-day variation. These are issuance service windows, not applicant submission deadlines or guaranteed total elapsed project duration.

UDI, listing & traceability

Which UDI, device listing, database, and traceability duties apply?

Compare countries

KFA reference-identifier document linkage

change.kfa-reference-identifiers2026
Fact-level verified

CircularFR.03.01/E.V/1582/2026 requires NIE holders to validate KFA reference URLs to licensing documents, including NIE and applicable IDAK/CDB/CPB, using OSS-RBA downloaded PDFs. Invalid or inaccessible links require obtaining the correct documents. Failure can return valid KFA codes to draft and prevent synchronization to the sector e-catalog.

Conditions
  • KFA document-reference validation, not proof that manufacturer SMN/REF changes are themselves exempt or that KFA equals a commercial product code.

Related structured datasets

Classification systems · 1 record
Open Risk classes A through D. →
Id

indonesia:medical-device

Jurisdiction Id

indonesia

Domains
  • medical-device
  • ivd
Name

Risk classes A through D.

Classes
  • Id

    a

    Label

    Class A

    Normalized Risk Level

    low

  • Id

    b

    Label

    Class B

    Normalized Risk Level

    medium

  • Id

    c

    Label

    Class C

    Normalized Risk Level

    high

  • Id

    d

    Label

    Class D

    Normalized Risk Level

    highest

Regulator

Ministry of Health

Official Domains
  • kemkes.go.id
Source Status

official-source-verified

Source Title

Peraturan Menteri Kesehatan No. 5 Tahun 2026, Pasal 55 (klasifikasi perbekalan kesehatan)

Source Url

https://farmalkes.kemkes.go.id/unduh/peraturan-menteri-kesehatan-nomor-5-tahun-2026-tentang-perbekalan-kesehatan/

Verified On

2026-08-25

Source Supports
  • class-axis
Notes
  • PMK 5/2026 Pasal 55 (previously PMK 62/2017): class A low risk, B low-to-moderate, C moderate-to-high, D high risk; additionally classified by intended use into categories and subcategories per ministerial technical guideline.
  • PMK 14/2021 product standards: IVD classes A-D by risk to individual/public (A low/low; B moderate/low; C high/moderate; D high/high); a 'class A tertentu' subset (home-use, no special skill, non-sterile, non-IVD) is eligible for notification.
Registration pathways, prior approvals and planning · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Indonesia registration →
Country

Indonesia

Jurisdiction Id

indonesia

Country Code

ID

Flag

🇮🇩

Regulator

Ministry of Health

Regulator Abbrev

Ministry of Health

Region

asia-pacific

Tier

2

Risk Classification

4-class (A, B, C, D)

Classes
  • Id

    class-a

    Class Name

    Class A

    Risk Level

    low

    Pathway

    Notification

    Official Timeline
    Min

    15

    Max

    15

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    Kemkes Regalkes — evaluation guideline Annex 8 service-standard clocks

    Url

    https://regalkes.kemkes.go.id/informasi_alkes/Indonesia%20Guideline%20for%20Evaluation%20of%20Medical%20Device%20and%20IVD%202020.pdf

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    2

    Max

    4

    Unit

    months

    Notes
    • Official service-standard clock for imported Class A izin edar (new application): 15 working days of evaluation (evaluation guideline Annex 8)
    • Simplified notification
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    No

  • Id

    class-b

    Class Name

    Class B

    Risk Level

    medium

    Pathway

    Registration

    Official Timeline
    Min

    30

    Max

    30

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    Kemkes Regalkes — evaluation guideline Annex 8 service-standard clocks

    Url

    https://regalkes.kemkes.go.id/informasi_alkes/Indonesia%20Guideline%20for%20Evaluation%20of%20Medical%20Device%20and%20IVD%202020.pdf

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    4

    Max

    6

    Unit

    months

    Prior Approval Benefits
    • ASEAN MRA countries
    • FDA
    • EU CE Mark
    Notes
    • Official service-standard clock for imported Class B izin edar (new application): 30 working days of evaluation (Annex 8)
    • ASEAN harmonization applies
    Clinical Trial Required

    No

    Local Testing Required

    Yes

    Gmp Required

    Yes

  • Id

    class-c

    Class Name

    Class C

    Risk Level

    high

    Pathway

    Full Registration

    Official Timeline
    Min

    30

    Max

    30

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    Kemkes Regalkes — evaluation guideline Annex 8 service-standard clocks

    Url

    https://regalkes.kemkes.go.id/informasi_alkes/Indonesia%20Guideline%20for%20Evaluation%20of%20Medical%20Device%20and%20IVD%202020.pdf

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    4

    Max

    6

    Unit

    months

    Prior Approval Benefits
    • ASEAN MRA countries
    • FDA
    • EU CE Mark
    Notes
    • Official service-standard clock for imported Class C izin edar (new application): 30 working days of evaluation (Annex 8)
    • Clinical data may be required
    Clinical Trial Required

    No

    Local Testing Required

    Yes

    Gmp Required

    Yes

  • Id

    class-d

    Class Name

    Class D

    Risk Level

    highest

    Pathway

    Full Registration

    Official Timeline
    Min

    45

    Max

    45

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    Kemkes Regalkes — evaluation guideline Annex 8 service-standard clocks

    Url

    https://regalkes.kemkes.go.id/informasi_alkes/Indonesia%20Guideline%20for%20Evaluation%20of%20Medical%20Device%20and%20IVD%202020.pdf

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    5

    Max

    7

    Unit

    months

    Prior Approval Benefits
    • ASEAN MRA countries
    • FDA
    • EU CE Mark
    Notes
    • Official service-standard clock for imported Class D izin edar (new application): 45 working days of evaluation (Annex 8)
    • Clinical evidence required
    Clinical Trial Required

    Yes

    Local Testing Required

    Yes

    Gmp Required

    Yes

Expedited Pathways
  • Name

    ASEAN Medical Device Directive (AMDD) harmonisation

    Description

    The AMDD gives the ten member states one device definition, one set of classification rules, one set of essential principles and one dossier structure, the Common Submission Dossier Template. It is not a reliance route. The ACCSQ Medical Device Product Working Group states in terms that the Directive is "not intended to confer mutual recognition" and that each member state retains the sovereign right to make its own registration decisions, so a registration in one ASEAN market authorises nothing in another and shortens no national review period.

    Timeline Reduction

    None — the saving is in dossier preparation, since one CSDT core can be reused across member states with national addenda

    Eligibility
    • Any device filed in an ASEAN member state that has transposed the AMDD into national law
Prior Approvals
  • Authority

    ASEAN member-state registration

    Accepted

    No

    Benefit

    There is no ASEAN mutual recognition agreement for device registration — the AMDD harmonises the dossier, not the decision

    Timeline Savings

    None

  • Authority

    FDA 510(k)

    Accepted

    Yes

    Benefit

    Supports the Certificate of Free Sale and the technical evidence in the CSDT, but Kemenkes publishes no shorter period for it

    Timeline Savings

    None published

  • Authority

    EU CE Mark

    Accepted

    Yes

    Benefit

    Supports the Certificate of Free Sale and the technical evidence in the CSDT, but Kemenkes publishes no shorter period for it

    Timeline Savings

    None published

Key Requirements
  • Indonesia Authorized Representative
  • Bahasa Indonesia labeling
  • Local testing from accredited labs
  • ISO 13485 certification
  • Free Sale Certificate
  • Annex 8 additional clocks: additional-data evaluation 10 working days (Class A/B) or 15 (Class C/D); renewal 7 working days; variation (perubahan) 10 working days
Local Rep Required

Yes

Local Rep Name

Indonesia Authorized Representative

Mdsap Accepted

No

Mdsap Benefit

Not accepted but ISO 13485 required

Registration Validity

5 years

Verification Status

official-verified

Official Sources
  • Title

    Kemkes Regalkes — evaluation guideline Annex 8 service-standard clocks

    Url

    https://regalkes.kemkes.go.id/informasi_alkes/Indonesia%20Guideline%20for%20Evaluation%20of%20Medical%20Device%20and%20IVD%202020.pdf

    Accessed Date

    2026-08-25

  • Title

    ACCSQ Medical Device Product Working Group — Implementing the ASEAN Medical Device Directive: "Not intended to confer mutual recognition; each member state retains sovereign right for registration decisions"

    Url

    https://www.ahwp.info/sites/default/files/6%20Implementing%20the%20ASEAN%20Medical%20Device%20Directive%201.0.pdf

    Accessed Date

    2026-08-30

Reviewed On

2026-08-30

Last Updated

2026-08-30

Process
Template Id

default

Status

planning-template

Usage Rule

Maintained planning steps and estimates, not verified regulator deadlines.

Steps
  • Step

    1

    Title

    Pre-Submission Preparation

    Description

    Gather technical documentation, quality management system (QMS) records, and clinical data required by Ministry of Health. Appoint a Indonesia Authorized Representative.

    Duration

    1-3 months

    Icon

    📋

  • Step

    2

    Title

    Application Submission

    Description

    Submit complete registration dossier to Ministry of Health including device description, risk analysis, and testing reports.

    Duration

    1-2 weeks

    Icon

    📤

  • Step

    3

    Title

    Administrative Review

    Description

    Ministry of Health reviews application for completeness. May request additional documents or clarifications.

    Duration

    2-4 weeks

    Icon

    🔍

  • Step

    4

    Title

    Technical Evaluation

    Description

    In-depth technical assessment of device safety and performance. Ministry of Health may issue questions.

    Duration

    15 working days

    Icon

    ⚙️

  • Step

    5

    Title

    Registration Approval

    Description

    Upon successful review, Ministry of Health issues Indonesia market authorization certificate.

    Duration

    1-2 weeks

    Icon

    ✅

Government fee schedules · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Indonesia fees →
Country

Indonesia

Jurisdiction Id

indonesia

Country Code

ID

Authority

Indonesia MoH

Low Risk Local
Min

1500000

Max

3000000

Currency

IDR

Per Product

Yes

High Risk Local
Min

3000000

Max

5000000

Currency

IDR

Per Product

Yes

Details
  • Id

    registration-class-a

    Item

    New registration - Class A

    Currency

    IDR

    Amount Local

    1500000

    Amount Local Formatted

    IDR 1,500,000

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://regalkes.kemkes.go.id/informasi_alkes/011.%20Modul%20Perizinan%20Alat%20Kesehatan.pdf

    Source Title

    Indonesia Ministry of Health medical-device licensing module

    Verified On

    2026-08-25

  • Id

    registration-class-b

    Item

    New registration - Class B

    Currency

    IDR

    Amount Local

    3000000

    Amount Local Formatted

    IDR 3,000,000

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://regalkes.kemkes.go.id/informasi_alkes/011.%20Modul%20Perizinan%20Alat%20Kesehatan.pdf

    Source Title

    Indonesia Ministry of Health medical-device licensing module

    Verified On

    2026-08-25

  • Id

    registration-class-c

    Item

    New registration - Class C

    Currency

    IDR

    Amount Local

    3000000

    Amount Local Formatted

    IDR 3,000,000

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://regalkes.kemkes.go.id/informasi_alkes/011.%20Modul%20Perizinan%20Alat%20Kesehatan.pdf

    Source Title

    Indonesia Ministry of Health medical-device licensing module

    Verified On

    2026-08-25

  • Id

    registration-class-d

    Item

    New registration - Class D

    Currency

    IDR

    Amount Local

    5000000

    Amount Local Formatted

    IDR 5,000,000

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://regalkes.kemkes.go.id/informasi_alkes/011.%20Modul%20Perizinan%20Alat%20Kesehatan.pdf

    Source Title

    Indonesia Ministry of Health medical-device licensing module

    Verified On

    2026-08-25

Notes
  • The official module distinguishes new registration from renewal/change. Renewal/change fees are lower: IDR 500,000 for Class A and IDR 1,000,000 for Classes B-D.
  • Domestically manufactured PKRT (perbekalan kesehatan rumah tangga) follow a separate PNBP table: IDR 1,000,000 / 2,000,000 / 3,000,000 for classes 1/2/3.
  • The module says the payable PNBP follows the regulations in force, so confirm the portal assessment before payment.
Last Updated

2026-08-25

Verification Status

official-verified

Official Sources
  • Title

    Indonesia Ministry of Health medical-device licensing module

    Url

    https://regalkes.kemkes.go.id/informasi_alkes/011.%20Modul%20Perizinan%20Alat%20Kesehatan.pdf

    Accessed Date

    2026-08-25

  • Title

    Indonesia regalkes PNBP tariff: izin edar certificate fees by class

    Url

    https://regalkes.kemkes.go.id/index.php/home/fileDownload/BERITA_FILE.pdf/1000786

    Accessed Date

    2026-08-25

Reviewed On

2026-08-25

Amount Status

exact-current

Range Available

Yes

Additional certifications and licences · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Indonesia certifications →
Slug

indonesia

Market

Indonesia

Confidence

low

Schemes
  • Id

    medical-electrical-evidence

    Name

    Medical electrical evidence

    Category

    Medical Electrical Safety

    Status

    Included In Medical Registration

    Object

    Finished Device

    Note

    Published Ditjen Farmalkes assessment guide (Pedoman Penilaian Alat Kesehatan dan PKRT): function/performance test evidence or a QC Certificate of Analysis for the finished device; electrically powered devices need electrical-safety testing per the IEC 60601 series (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance), with the testing laboratory's ISO/IEC 17025 certificate; product-specific evidence quoted by the guide includes condoms (uji daya letup dan kebocoran) and sanitary pads (uji fluoresensi dan uji daya serap). PMK 14/2021 (as amended by PMK 8/2022 and 17/2024) was repealed in part by Permenkes 11/2025 (set 3 October 2025), which now carries the OSS health-sector business/product standards under PP 28/2025. PMK 54/2015 (device testing and calibration) was revoked by PMK 5/2026. Read the published assessment guidance subject to the current framework and the consistency condition for earlier implementing provisions in Permenkes 5/2026 Article 124.

    Source Urls
    • https://regalkes.kemkes.go.id/informasi_alkes/PedomanPenilaianAlatKesehatandanPKRT.pdf
    • https://jdih.kemkes.go.id/documents/peraturan-menteri-kesehatan-nomor-11-tahun-2025
    • https://jdih.kemkes.go.id/documents/peraturan-menteri-kesehatan-nomor-5-tahun-2026
    Source Binding

    reviewed

  • Id

    komdigi-telecommunications-device-certification

    Name

    Komdigi telecommunications-device certification

    Category

    Radio Spectrum

    Status

    Mandatory If Feature

    Object

    Covered Telecommunications Device

    Source Urls
    • https://layanan.komdigi.go.id/layanan/detail/?id=9
    Source Binding

    candidate

  • Id

    sni-compulsory-certification

    Name

    SNI compulsory certification

    Category

    General Electrical Components

    Status

    Mandatory If Feature

    Object

    Listed Product Or Component

    Source Binding

    unresolved

  • Id

    medical-device-halal-class-a

    Name

    Medical-device halal Class A

    Category

    Halal And Material Origin

    Status

    Transition Not Yet Mandatory

    Object

    In Scope Class A Device

    Effective Date

    2026-10-18

    Post Effective Status

    Mandatory Pre Market

    Source Urls
    • https://peraturan.bpk.go.id/Details/291786
    Source Binding

    candidate

  • Id

    medical-device-halal-class-b

    Name

    Medical-device halal Class B

    Category

    Halal And Material Origin

    Status

    Transition Not Yet Mandatory

    Object

    In Scope Class B Device

    Effective Year

    2029

    Source Urls
    • https://peraturan.bpk.go.id/Details/291786
    Source Binding

    candidate

  • Id

    medical-device-halal-class-c

    Name

    Medical-device halal Class C

    Category

    Halal And Material Origin

    Status

    Transition Not Yet Mandatory

    Object

    In Scope Class C Device

    Effective Year

    2034

    Note

    Class D timing is governed separately and must remain authority-confirmation-needed until the applicable presidential rule and availability conditions are resolved.

    Source Urls
    • https://peraturan.bpk.go.id/Details/291786
    Source Binding

    candidate

  • Id

    bapeten-radiation-authorization

    Name

    BAPETEN radiation authorization

    Category

    Ionizing Radiation

    Status

    Mandatory If Feature

    Object

    Product And Use Site

    Note

    Current Ditjen Farmalkes assessment guide (Pedoman Penilaian Alat Kesehatan dan PKRT): devices containing ionizing radiation, such as X-ray units, require a Surat Izin Pemanfaatan Tenaga Nuklir issued by Badan Pengawas Tenaga Nuklir (BAPETEN) as part of the izin edar assessment. Locator quote: "alkes yang mengandung radiasi pengion seperti unit sinar X, dipersyaratkan Surat Izin Pemanfaatan Tenaga Nuklir yang dikeluarkan oleh Badan Pengawas Tenaga Nuklir (BAPETEN)".

    Source Binding

    unresolved

  • Id

    kemenkes-import-letters

    Name

    Kemenkes medical-device imports and special access

    Category

    Import Conformity

    Status

    Mandatory If Feature

    Object

    Commercial Or Special Circumstance Imports

    Note

    Permenkes 5/2026 (effective 4 May 2026) Article 58(4) requires an izin edar for medical-device and PKRT imports; Article 58(5) allows the Minister to issue an import letter in specified circumstances, with procedures to be set by the Minister under Article 58(7). Article 60 provides for customs/post-border checks. Article 62 governs special access without an izin edar, including purpose-specific approval and exceptions for personal use and other enumerated needs; assess the exact purpose rather than treating every special-access shipment as requiring the same permission. Article 125 repeals, among others, PMK 14/2016 (complementary goods, market testing and after-sales import recommendations), PMK 51/2014 and 7/2020 (special access), and PMK 60/2017 (import supervision). Earlier implementing provisions survive only insofar as consistent with the new regulation (Article 124). The older published export/import guide is historical operational context and does not by itself establish current letter names, sample re-testing requirements or an exemption. Business/distributor licensing also requires the applicable OSS standards under Permenkes 11/2025.

    Source Urls
    • https://jdih.kemkes.go.id/documents/peraturan-menteri-kesehatan-nomor-5-tahun-2026
    • https://jdih.kemkes.go.id/documents/peraturan-menteri-kesehatan-nomor-11-tahun-2025
    Source Binding

    reviewed

Official Sources
  • Authority

    Komdigi

    Topic

    telecommunications-device certification

    Url

    https://layanan.komdigi.go.id/layanan/detail/?id=9

  • Authority

    BPJPH

    Topic

    halal obligation beginning 18 October 2026

    Url

    https://bpjph.halal.go.id/read/tak-hanya-makanan-minuman-ini-jenis-produk-yang-wajib-bersertifikat-halal-mulai-18-oktober-2026

  • Authority

    Republic of Indonesia

    Topic

    medical-device halal implementation guideline

    Url

    https://peraturan.bpk.go.id/Details/291786

  • Authority

    Kementerian Kesehatan (JDIH)

    Topic

    Permenkes 11/2025 — current OSS health-sector business/product standards repealing PMK 14/2021 in part (set 3 October 2025, status Berlaku)

    Url

    https://jdih.kemkes.go.id/documents/peraturan-menteri-kesehatan-nomor-11-tahun-2025

  • Authority

    Direktorat Jenderal Farmasi dan Alat Kesehatan

    Topic

    Pedoman Penilaian Alat Kesehatan dan PKRT — IEC 60601 electrical-safety evidence and BAPETEN Surat Izin Pemanfaatan Tenaga Nuklir

    Url

    https://regalkes.kemkes.go.id/informasi_alkes/PedomanPenilaianAlatKesehatandanPKRT.pdf

  • Authority

    Direktorat Jenderal Farmasi dan Alat Kesehatan

    Topic

    Published export/import guide — historical import-letter and sample re-testing descriptions; assess continued applicability under Permenkes 5/2026

    Url

    https://regalkes.kemkes.go.id/informasi_alkes/Pedoman%20Ekspor%20Impor.pdf

  • Authority

    Kementerian Kesehatan (JDIH)

    Topic

    Permenkes 5/2026 — current medical-device import and special-access requirements (Articles 58, 60, 62), implementing-rule transition and repeals (Articles 124–126), effective 4 May 2026

    Url

    https://jdih.kemkes.go.id/documents/peraturan-menteri-kesehatan-nomor-5-tahun-2026

Keep researching this market
View the Indonesia resource hub →
Compare regulatory data across all countries →