Roszdravnadzor · structured requirements

🇷🇺 Russia IVD regulatory data

Explore the IVD requirements currently available for Russia, organized by topic and linked to official sources.

14 facts5 dimensions4 official sourceslatest verification 2026-10-26
Explore questions and sources →

Common questions about changing a registered IVD in Russia. Each question separates published rules from details still awaiting verification.

Change control

Does the event require internal documentation, notice, prior approval, transfer, or a new application?

Compare countries

Trademark/name changes without expertise

brand.name.non.expertise
Fact-level verified

Clause 111(д) includes changing the device name in respect of a trademark or other means of individualization among dossier changes not requiring quality, efficacy and safety expertise. Clause 111(ж) separately includes label-design changes without changing the symbols used on the device marking.

Conditions
  • Apply only to a national Russian registration dossier under Decree1684.
  • A rule for one change does not establish a combined application, a single fee or a guaranteed total timeline.
  • The label-design condition does not establish that every logo/symbol change qualifies.

Applies to the stated national Russian dossier-change rule. The cited clause does not establish the EAEU procedure, a universal combined filing or an empirical turnaround forecast.

Dossier-change submission deadline

change.submission.deadline
Fact-level verified

For clause 111 non-expertise changes, clause 112 requires submission through the unified applicant cabinet no later than90 working days after the relevant data changed. Clause 120 provides the same filing deadline for clause 119 expertise changes.

Conditions
  • Apply only to a national Russian registration dossier under Decree1684.
  • A rule for one change does not establish a combined application, a single fee or a guaranteed total timeline.
  • This90workingday period is an applicant filing deadline; it is not an authority review duration or an old-label sell-through period.

Applies to the stated national Russian dossier-change rule. The cited clause does not establish the EAEU procedure, a universal combined filing or an empirical turnaround forecast.

Manufacturer name and office address changes

manufacturer.name.office.address.non.expertise
Fact-level verified

Clause 111(в) includes changes to the manufacturer legal-person name and address of location, and information about legal-person reorganization, among dossier changes not requiring quality, efficacy and safety expertise. It separately limits the production-site address case to administrative geographical/street/number/postcode changes.

Conditions
  • Apply only to a national Russian registration dossier under Decree1684.
  • A rule for one change does not establish a combined application, a single fee or a guaranteed total timeline.
  • Distinguish the same legal manufacturer changing its name/office address from a real manufacturing-site move or a different manufacturer.
  • This listing alone does not prove combined brand/name/address filing permission.

Applies to the stated national Russian dossier-change rule. The cited clause does not establish the EAEU procedure, a universal combined filing or an empirical turnaround forecast.

Removal of a registered model

model.removal.non.expertise
Fact-level verified

Clause 111(е) includes removing information about a medical-device model (make) from the register entry among dossier changes not requiring quality, efficacy and safety expertise.

Conditions
  • Apply only to a national Russian registration dossier under Decree1684.
  • A rule for one change does not establish a combined application, a single fee or a guaranteed total timeline.
  • Applies to removal of a registered model/make. Does not establish the treatment of a purely commercial REF/SMN code, addition of a model, or arbitrary renaming.
  • The clause changes registered model information; it does not itself require a particular label revision.

Applies to the stated national Russian dossier-change rule. The cited clause does not establish the EAEU procedure, a universal combined filing or an empirical turnaround forecast.

Changes requiring a new device registration

new.device.change.triggers
Fact-level verified

Clause 110 requires registration of a new medical device for a change to its intended purpose or principle of operation; inclusion of model/make information that does not satisfy the grouping criteria in clause 26; or simultaneous changes to manufacturer and manufacturing-site information, subject to the express exceptions for legal form, address of location, reorganization and other constituent-document information.

Conditions
  • Apply only to a national Russian registration dossier under Decree1684.
  • A rule for one change does not establish a combined application, a single fee or a guaranteed total timeline.
  • A commercial SMN/REF/order number is not automatically the registered model or make.
  • The stated simultaneous manufacturer/production-site rule does not establish that every name/office-address/brand combination requires new registration.

Applies to the stated national Russian dossier-change rule. The cited clause does not establish the EAEU procedure, a universal combined filing or an empirical turnaround forecast.

Non-expertise statutory review period and excluded deficiency time

non.expertise.authority.phases
Fact-level verified

Clause 113 limits the non-expertise dossier-change procedure to 18 working days from receipt of the application and the documents required by clause 112. Clause 114 provides a five-working-day completeness/accuracy check; clause 117 provides a three-working-day start decision after the check or rectification; and clause 118 provides a ten-working-day decision/register-update phase after the start decision. Under clause 115, the interval from deficiency notification until receipt of the response or notice that no response was provided is excluded from the procedure time.

Conditions
  • National Russian registration under Decree 1684, only when the actual change qualifies for the clause 111 non-expertise route.
  • The 18-working-day limit is stated in clause 113, rather than inferred by adding the phases. Preparation before submission and the clause 115 excluded deficiency interval are not a guaranteed elapsed approval time.
  • The deficiency provisions allow 30 working days for rectification; failure to rectify can terminate consideration under clause 116.
  • This statutory clock supplies no empirical best/expected/worst forecast and proves no universal combined filing, single fee, stock grace period or unrelated-holder transfer route.

Applies to the stated national Russian dossier-change rule. The cited clause does not establish the EAEU procedure, a universal combined filing or an empirical turnaround forecast.

Authorized-representative administrative changes

representative.administrative.information.change
Fact-level verified

Clause 111(г) includes authorized-representative legal-person reorganization, name and address-of-location changes among non-expertise dossier changes. The foreign manufacturer must maintain current representative information under clause 29.

Conditions
  • Apply only to a national Russian registration dossier under Decree1684.
  • A rule for one change does not establish a combined application, a single fee or a guaranteed total timeline.
  • This is evidence for the stated representative-information cases, not proof that a registration can be transferred to an unrelated new legal manufacturer/holder, recovered from a partner, or split among multiple partners.

Applies to the stated national Russian dossier-change rule. The cited clause does not establish the EAEU procedure, a universal combined filing or an empirical turnaround forecast.

Dossier requirements

Which evidence, local deltas, formats, and signatures are required?

Compare countries

Evidence supporting dossier changes

change.application.supporting.documents
Fact-level verified

Clause 112 requires the change application and describes supporting evidence including authorized-representative powers and confirmation that changes do not impair device safety/quality/efficacy or change intended purpose/principle of operation. Additional QMS documents are conditional on the specified low-risk/QMS route. Expertise-route documents are specified separately in clause 120.

Conditions
  • Apply only to a national Russian registration dossier under Decree1684.
  • A rule for one change does not establish a combined application, a single fee or a guaranteed total timeline.
  • Use the actual clause111 or119 route and applicable document conditions.
  • This does not imply that a CFS, legalization or new QMS certificate is required for every administrative change.

Applies to the stated national Russian dossier-change rule. The cited clause does not establish the EAEU procedure, a universal combined filing or an empirical turnaround forecast.

Economic-operator roles

Which manufacturer, representative, holder, applicant, importer, and distributor roles apply?

Compare countries

Authorized representative of foreign manufacturer

market-entry.local-role
Fact-level verified

Foreign manufacturers must appoint an authorized representative of the manufacturer (уполномоченный представитель производителя/изготовителя) who bears responsibility for the device's circulation in Russia, including conformity-assessment and state-registration procedures; a copy of the document(s) confirming the representative's powers is a mandatory dossier item for foreign manufacturers.

Conditions
  • A foreign manufacturer must appoint an authorised representative for the national Russian registration route.
  • For the clause 112/120 change-dossier powers document, submission is mandatory for foreign manufacturers; a domestic manufacturer submits it if it has an authorised representative.

Effective from 2025-03-01

The cited clauses establish representative appointment and powers-document requirements. They do not decide separate importer/distributor licensing, product-holder transfer or recovery arrangements.

Authorized-representative information actualization deadline

market-entry.representative-change
Fact-level verified

Clause 29 requires a foreign medical-device manufacturer to appoint an authorised representative acting during the device’s circulation in Russia and keep the dossier information current. If that representative information is absent or outdated, the clause states 1 September 2027 as the deadline to appoint a representative or update the information, reflecting the 2025 No. 1206 and 2026 No. 974 amendments.

Conditions
  • National Russian medical-device/IVD registration dossiers governed by Decree 1684; apply its transitional provisions to earlier applications.
  • The 1 September 2027 date concerns absent/outdated authorised-representative information. It is not a general registration-expiry date or permission to transfer a registration to an unrelated holder.

Replaces the obsolete 1 September 2026 deadline and the old pre-2025-only condition with the actual clause 29 text in the official government consolidation. The amendment’s separate commencement date is not inferred from the general 2025 regime date.

Reliance & recognition

Which foreign decisions or evidence can be reused, with what effect and conditions?

Compare countries

Existing records include acceptance, restrictions and evidence reuse. The applicable product scope and conditions are preserved below; a reference does not by itself confer market access.

Russia approvals and evidence in other markets (11 records)

ArmeniaBrazil ANVISA · EU conformity assessment · US FDA · Health Canada · Singapore HSA · UK MHRA · South Korea MFDS · Japan MHLW / PMDA · NMPA · Australia TGA · IMDRF member authorities · WHO

Prior approval / evidence record

Simplified national registration for a device registered in an IMDRF member country or prequalified by WHO

reliance.imdrf-member-or-who-prequalified-simplified-registration
Fact-level verified
Statement

Armenia registers medical devices under two national procedures: a general procedure and a simplified procedure. The simplified procedure applies to a device that is registered in a country that is a member of the International Medical Device Regulators Forum (IMDRF) or that is prequalified by the World Health Organization. IMDRF members include the United States (FDA), the EU member states (CE marking), Japan (PMDA), Korea (MFDS), China (NMPA), Canada, Australia, Brazil, Singapore and the United Kingdom, so a US FDA 510(k)/PMA clearance, an EU CE mark, a Japan PMDA approval and a Korea MFDS approval all satisfy the gate. The device is still registered in Armenia by the Armenian authority — the foreign registration opens the shortened procedure, it does not replace the Armenian certificate.

Source Approval

Registration in any IMDRF member country — US FDA 510(k)/PMA, EU CE marking under MDR/IVDR (any EU member state), Japan PMDA, Korea MFDS, China NMPA, Health Canada, Australia TGA, Brazil ANVISA, Singapore HSA, UK MHRA — or WHO prequalification of the device

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Premarket Review

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • The provision is written as an annex to part 3 of the same article: 'under the procedure laid down by part 3 of this article' devices are registered under the general and simplified procedures. Part 3 is itself the Armenia-only route, and a footnote to Article 46 states that part 3 operates only until the end of the transitional period set by Article 11 of the EAEU Agreement of 23 December 2014 (footnote inserted by Article 3(2) of the Law of 5 December 2024 ՀՕ-502-Ն). The transitional period was extended by the Protocol signed in Moscow on 29 December 2025, which replaced '2025' with '2027' in the first paragraph of Article 11 of that Agreement and '2026' with '2028' in the third and fourth paragraphs.
  • Bounded claim: the Law names IMDRF membership and WHO prequalification as the gate for the simplified procedure. It does not name the FDA, the CE mark, the PMDA or the MFDS; those follow from their countries' IMDRF membership, not from the text.
  • The operational content of the 'simplified procedure' (dossier list, clocks, certificate form) sits in the Government procedure that Article 46(3) contemplates; that instrument was not located in this session (see unresolved: registration-pathways).
  • Devices are registered by the Ministry of Health (the լիազոր մարմին / authorised body) with the examination carried out by the expert organisation designated by Government decision (Article 46(2), (5)).

Language: Armenian. Verbatim, Article 46(4): "Սույն հոդվածի 3-րդ մասով սահմանված կարգի համաձայն՝ բժշկական արտադրատեսակները գրանցվում են ընդհանուր և պարզեցված ընթացակարգերով: Պարզեցված ընթացակարգը կիրառվում է բժշկական արտադրատեսակների շրջանառությունը կարգավորող մարմինների միջազգային համաժողովի (IMDRF) անդամ երկրում գրանցված կամ Առողջապահության համաշխարհային կազմակերպության նախաորակավորած բժշկական արտադրատեսակների համար:" Translation: "Under the procedure laid down by part 3 of this article, medical devices are registered under general and simplified procedures. The simplified procedure applies to medical devices registered in a member country of the International Medical Device Regulators Forum (IMDRF) or prequalified by the World Health Organization." Article 46 was amended by the Law of 5 December 2024 ՀՕ-502-Ն; the consolidated expression is the version in force from 31.05.2026.

LebanonIMDRF member authorities

Prior approval / evidence record

GHTF/IMDRF member countries — free-sale certificate (شهادة بيع حر)

Authority

GHTF/IMDRF member countries — free-sale certificate (شهادة بيع حر)

Accepted

Yes

Benefit

A mandatory document rather than a route: Decision 1506/1 Art. Four requires the registration application to carry a free-sale certificate issued by a country belonging to the GHTF/IMDRF group, meaning the device is sold in one of those countries whatever its origin, and the Ministry's service records require a free-sale certificate with an ISO 13485 certificate where the goods are made outside the reference states or the listed documents cannot be produced. No examination period is shortened by it.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

MexicoNMPA · US FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA · UK MHRA · Swissmedic · South Korea MFDS · Singapore HSA · Brazil ANVISA · IMDRF member authorities · MDSAP · WHO

Prior approval / evidence record

Vía regulatoria abreviada — ARR equivalence (COFEPRIS-04-050)

Name

Vía regulatoria abreviada — ARR equivalence (COFEPRIS-04-050)

Description

The single reliance route since the DOF Acuerdo of 18 July 2025, which took effect 30 working days after publication and abrogated every individual equivalence agreement issued between 2012 and June 2025 — the separate FDA, Swissmedic, European Commission, Health Canada, TGA and WHO prequalification acuerdos no longer exist. COFEPRIS now recognises the decision of any Reference Regulatory Authority from a country sitting on the IMDRF Management Committee, and MDSAP full membership for GMP certificates. The dossier drops to an administrative-legal section plus technical information under RIS Art. 31.

Timeline Reduction

Resolution within a maximum of 30 working days

Eligibility
  • Prior authorisation by a Reference Regulatory Authority in one of the twelve IMDRF Management Committee jurisdictions: Australia, Brazil, Canada, China, the European Union, Japan, Russia, Singapore, South Korea, Switzerland, the United Kingdom, the United States
  • MDSAP full members are recognised for the good-manufacturing-practice certificate
  • Filed under homoclave COFEPRIS-04-050 rather than the ordinary COFEPRIS-04-060 registration
  • Post-market surveillance or technovigilance evidence from the reference market, translated into Spanish under RIS Art. 153

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

MexicoUS FDA · Health Canada · Japan MHLW / PMDA · IMDRF member authorities · MDSAP

Prior approval / evidence record

Class II

Id

class-ii

Class Name

Class II

Risk Level

medium

Pathway

Registration or Equivalency

Official Timeline
Min

35

Max

35

Unit

days

Verification Status

official-verified

Official Source
Title

RIS Art. 179 — statutory resolution periods by class (COFEPRIS)

Url

https://salud.gob.mx/unidades/cdi/nom/compi/ris.html

Accessed Date

2026-08-25

Realistic Timeline
Min

15

Max

31

Unit

months

With Prior Approval
Min

8

Max

16

Unit

months

Prior Approval Benefits
  • FDA 510(k)
  • Health Canada
  • Japan PMDA
  • IMDRF Committee Members
  • MDSAP
Notes
  • RIS Art. 179: Clase II resolved within 35 calendar days (15 days with a favourable Tercero Autorizado dictamen)
  • ARR equivalence route (COFEPRIS-04-050, Lineamientos DOF 18-07-2025): maximum 30 working days
  • Equivalency route much faster if FDA approved
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

MexicoUS FDA · Health Canada · Japan MHLW / PMDA · IMDRF member authorities · MDSAP

Prior approval / evidence record

Class III

Id

class-iii

Class Name

Class III

Risk Level

highest

Pathway

Registration or Equivalency

Official Timeline
Min

60

Max

60

Unit

days

Verification Status

official-verified

Official Source
Title

RIS Art. 179 — statutory resolution periods by class (COFEPRIS)

Url

https://salud.gob.mx/unidades/cdi/nom/compi/ris.html

Accessed Date

2026-08-25

Realistic Timeline
Min

15

Max

31

Unit

months

With Prior Approval
Min

8

Max

16

Unit

months

Prior Approval Benefits
  • FDA 510(k)/PMA
  • Health Canada
  • Japan PMDA
  • IMDRF Committee Members
  • MDSAP
Notes
  • RIS Art. 179: Clase III resolved within 60 calendar days (15 days with a favourable Tercero Autorizado dictamen)
  • ARR equivalence route (COFEPRIS-04-050): maximum 30 working days
  • Equivalency pathway best option with FDA
Clinical Trial Required

Yes

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

ParaguayWHO · IMDRF member authorities

Prior approval / evidence record

PAHO/WHO reference authority or IMDRF Management Committee member authority

Authority

PAHO/WHO reference authority or IMDRF Management Committee member authority

Accepted

Yes

Benefit

Operative for class II, III and IV IVDs, which enter the Proceso Simplificado de Registro Sanitario under Resolución DINAVISA N° 44/2024 with a 15-business-day validation and approval by positive silence.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

ParaguayBrazil ANVISA · EU conformity assessment · US FDA · Health Canada · Singapore HSA · IMDRF member authorities · Portugal · Spain · South Korea MFDS · Japan MHLW / PMDA · UK MHRA · NMPA · Taiwan TFDA · Australia TGA · WHO

Prior approval / evidence record

IVD simplified registration (PSR) for products authorised by PAHO/WHO, IMDRF or bilateral-partner authorities

reliance.ivd-psr-reference-authorities
Fact-level verified
Statement

Class II, III and IV in-vitro diagnostic products already authorised and commercialised in countries regulated by PAHO/WHO reference regulatory authorities, by authorities of IMDRF Management Committee member countries, or by authorities with bilateral agreements with DINAVISA enter the Proceso Simplificado de Registro Sanitario (PSR). DINAVISA has 15 business days to validate and decide, and approval follows by positive silence if no report is issued.

Source Approval

PAHO/WHO reference regulatory authorities; authorities of IMDRF Management Committee member countries (the IMDRF committee includes US FDA, EU, Japan PMDA/MHLW, Health Canada, TGA, ANVISA, China NMPA, Korea MFDS, Singapore HSA and UK MHRA); authorities with bilateral agreements with DINAVISA

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Premarket Review

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Applies to class II, III and IV IVDs (Res. 44/2024 Art. 7: 'Clases II, III y IV'); class I IVDs are registered automatically under Res. 266/2022 Art. 14.
  • The IVD must actually be commercialised in one of those countries and the foreign registration must correspond specifically to the same product being registered.
  • 15 business days to validate; approval by positive silence.
  • Objections may be answered only once (Art. 3), with 30 business days to rectify.
  • Not applicable to the cases in Arts. 13 and 16 of Res. DINAVISA N° 266/2022 (Art. 11).

Effective from 2024-01-30

Source language: Spanish. Translation of the operative part of Art. 2: 'Applications for sanitary registration of In Vitro Diagnostic Products ... that are authorised and commercialised in countries regulated by: PAHO/WHO Reference Regulatory Authorities; Regulatory Authorities of member countries of the Committee of the International Medical Device Regulators Forum - IMDRF; Regulatory Authorities with which bilateral agreements with DINAVISA exist, shall be subject to the Simplified Sanitary Registration Process (PSR).' and 'DINAVISA shall have a term of 15 (fifteen) business days to validate the required documentation and decide thereon. Once the term has elapsed with no report from DINAVISA, the application shall be approved.' Article numbering confirmed by reading the page image of the official PDF.

ParaguayWHO · IMDRF member authorities

Prior approval / evidence record

Proceso simplificado de Registro Sanitario (PSR) para productos para diagnóstico de uso in vitro

Name

Proceso simplificado de Registro Sanitario (PSR) para productos para diagnóstico de uso in vitro

Description

Class II, III and IV IVDs (Res. 44/2024 Art. 7) already authorised and commercialised by a recognised reference authority are validated in 15 business days with approval by positive silence, instead of the 30, 45 or 120 business days of the ordinary IVD route.

Timeline Reduction

From 30-120 business days down to 15 business days

Eligibility
  • Product classified as class II, III or IV under Resolución DINAVISA N° 266/2022 Art. 9
  • Current authorisation and commercialisation in a country regulated by a PAHO/WHO reference regulatory authority, by an IMDRF Management Committee member authority, or by an authority with a bilateral agreement with DINAVISA
  • The foreign registration must correspond specifically to the same product being registered in Paraguay

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

UgandaWHO · IMDRF member authorities · ISO

Prior approval / evidence record

WHO Prequalification, or quality-system certification from a certification body accredited by a regulatory authority in an IMDRF founding-member country, or any other international organisation recognised by NDA

Authority

WHO Prequalification, or quality-system certification from a certification body accredited by a regulatory authority in an IMDRF founding-member country, or any other international organisation recognised by NDA

Accepted

Yes

Benefit

Track 2: accepted as a documentary substitution for an IMDRF-country licence under DAR/GDL/028 §4.12.2 — the foreign element that unlocks the route is the quality-system certification and its accreditation rather than a product approval, and no shortened period is published for it.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

UgandaAustralia TGA · Health Canada · EU conformity assessment · Japan MHLW / PMDA · US FDA · IMDRF member authorities

Prior approval / evidence record

Licence or approval from an IMDRF founding-member country — Australia, Canada, the European Union, Japan, the United States

Authority

Licence or approval from an IMDRF founding-member country — Australia, Canada, the European Union, Japan, the United States

Accepted

Yes

Benefit

Track 1: an abridged evaluation in which the foreign approval carries the technical assessment, with a much smaller submission (attestation, abbreviated product and manufacturer details, notarized copy of the licence, Certificate of Analysis to an accepted quality-system standard, evidence of repeat sales and product samples where practicable). It is an expedited review with shortened registration timelines, but NDA publishes no number for the shortening and the final decision remains NDA's.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

ZambiaIMDRF member authorities · WHO · US FDA · EU conformity assessment

Prior approval / evidence record

WHO Prequalification; US FDA approval; EU CE marking; approval in any IMDRF member state

Authority

WHO Prequalification; US FDA approval; EU CE marking; approval in any IMDRF member state

Accepted

No

Benefit

Disclosure only. The Annex I application form requires the applicant to state whether the device is WHO prequalified, US FDA approved, CE marked or approved in any IMDRF member state and to attach valid documentation, but neither the non-IVD nor the IVD guideline attaches any procedural consequence — no shortened clock, no abridged dossier and no substitute for ZAMRA's own evaluation.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

Prior approvals and evidence for Russia (5 records)

RussiaNMPA

Regulatory reference

China NMPA approval: no recognition regime

reliance.china-nmpa-no-recognition-route
Pending verification
Statement

National registration is conducted on the basis of technical tests, toxicological studies and clinical trials carried out as Russian conformity-assessment forms, production inspection, analysis of the manufacturer's evidence materials and Roszdravnadzor's expertise of quality, effectiveness and safety. The Rules recognise no foreign approval and the dossier list contains no country-of-origin registration certificate item, so a China NMPA certificate has no legal effect in the national procedure.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

russia

China Classes In Scope

All classes under the national registration route (Government Decree No. 1684)

Edge Type

No Reliance Route

Market Access Strength

none

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Decree 1684 in force from 1 March 2025; national route extended to 31 December 2028 (Decree 2214 of 30.12.2025; applications until 31 December 2027)
  • Foreign manufacturers must act through an authorised representative in Russia
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Effective from 2025-03-01

Source passage: Государственная регистрация медицинских изделий проводится на основании результатов технических испытаний, токсикологических исследований, клинических испытаний, представляющих собой формы оценки соответствия медицинских изделий с учетом классификации в зависимости от потенциального риска их применения, инспектирования производства, анализа доказательных материалов производителя (изготовителя) медицинского изделия, экспертизы качества, эффективности и безопасности медицинских изделий...

RussiaEU conformity assessment

Prior approval / evidence record

CE Mark (EU MDR)

Authority

CE Mark (EU MDR)

Accepted

Yes

Benefit

Technical and QMS documentation can be reused, but the EAEU route requires its own expertise at VNIIMT and its own Russian-language dossier. Roszdravnadzor publishes no reduced period for CE-marked devices.

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

RussiaNMPA

Prior approval / evidence record

BRICS regulators cooperation MoU (NMPA party): no recognition effect

reliance.china-nmpa-bilateral-programme
Pending verification
Statement

Roszdravnadzor, the Russian Ministry of Health and the Ministry of Industry and Trade are parties with China's NMPA to the BRICS regulators' Memorandum of Understanding listed by CDSCO as valid until 31 January 2029. It is a cooperation MoU and gives NMPA registrations no legal effect in Russia; no Russia-China mutual-recognition arrangement for medical devices exists.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

russia

China Classes In Scope

Not class-specific; cooperation instruments only

Edge Type

Cooperation Mou

Market Access Strength

none

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Multilateral cooperation instrument (BRICS)
  • No recognition of NMPA approvals under Decree 1684 or EAEU rules
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Source passage: The Ministry of Health of The Russian Federation, The Ministry of Industry and Trade of The Russian Federation, The Federal Service for Surveillance in Healthcare (Russian Federation) ... The National Medical Products Administration of The People's Republic of CHINA (NMPA) ... 31.01.2029

RussiaNMPA

Prior approval / evidence record

China NMPA certificate: optional origin-registration item in the EAEU dossier

reliance.china-nmpa-origin-evidence-eaeu
Pending verification
Statement

In the EAEU registration dossier filed with Roszdravnadzor, a copy of the registration certificate, free-sale certificate or export certificate issued in the manufacturer's country is an 'if available' item, together with information on registration in other countries. A China NMPA certificate is submitted under this item and does not abridge the EAEU expertise.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

russia

China Classes In Scope

All classes under the EAEU common registration route (Council Decision No. 46)

Edge Type

Origin Country Evidence

Market Access Strength

Evidence Only

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Item 10 is 'при наличии' (if available)
  • Export certificate not accepted for devices first produced in an EAEU member state
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Effective from 2016-02-12

Source passage: 10. Копия регистрационного удостоверения медицинского изделия (сертификата свободной продажи, сертификата на экспорт (за исключением медицинских изделий, впервые произведенных на территории государства-члена)), выданного в стране производителя (при наличии)

Related structured datasets

Classification systems · 1 record
Open Risk classes 1, 2a, 2b, and 3. →
Id

russia:medical-device

Jurisdiction Id

russia

Domains
  • medical-device
  • ivd
Name

Risk classes 1, 2a, 2b, and 3.

Classes
  • Id

    1

    Label

    Class 1

    Normalized Risk Level

    low

  • Id

    2a

    Label

    Class 2a

    Normalized Risk Level

    medium

  • Id

    2b

    Label

    Class 2b

    Normalized Risk Level

    high

  • Id

    3

    Label

    Class 3

    Normalized Risk Level

    highest

Regulator

Roszdravnadzor

Official Domains
  • roszdravnadzor.gov.ru
Source Status

official-source-verified

Source Title

Roszdravnadzor — state registration of medical devices (risk class per MOH nomenclature classification)

Source Url

https://roszdravnadzor.gov.ru/medproducts/registration/registration

Verified On

2026-08-25

Source Supports
  • class-axis
Notes
  • Classes follow the nomenclature classification of medical devices by potential risk of application approved by the Ministry of Health (Order No. 4н of 06.06.2012, appendices amended by Order No. 686н of 2020); the applicant states the class in the registration application.
  • State fees are differentiated by class: class 1 - 72,000 RUB; class 2а - 104,000; class 2б - 136,000; class 3 - 184,000 RUB.
  • The EAEU route uses its own classification rules (EEC Board Decision No. 173, as amended by EEC Collegium Decision No. 18 of 18.02.2025; IVD classification guidance in EEC Recommendation No. 20 of 2024).
Registration pathways, prior approvals and planning · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Russia registration →
Country

Russia

Jurisdiction Id

russia

Country Code

RU

Flag

🇷🇺

Regulator

Federal Service for Surveillance in Healthcare

Regulator Abbrev

Roszdravnadzor

Region

europe

Tier

3

Risk Classification

4-class (1, 2a, 2b, 3) — EAEU classification

Classes
  • Id

    class-1

    Class Name

    Class 1

    Risk Level

    low

    Pathway

    RZN state registration (EAEU route)

    Official Timeline
    Min

    31

    Max

    50

    Unit

    working-days

    Realistic Timeline
    Min

    2

    Max

    4

    Unit

    months

    Notes
    • Staged VNIIMT expertise 25 wd; single 15-wd query round; sponsor response ≤35 wd excluded from the clock; conclusion 10 wd; RZN decision 5 wd; RU language check 5 wd; deficit-list checks ≤5 wd
    Clinical Trial Required

    No

    Local Testing Required

    Yes

    Gmp Required

    Yes

    Verification Status

    official-verified

    Official Source
    Title

    Roszdravnadzor — registration timelines from 01.03.2025 (50/31 working days)

    Url

    https://roszdravnadzor.gov.ru/medproducts/registration/registration

    Effective Date

    2025-03-01

    Accessed Date

    2026-08-25

  • Id

    class-2a

    Class Name

    Class 2a

    Risk Level

    medium

    Pathway

    RZN state registration (EAEU route)

    Official Timeline
    Min

    31

    Max

    50

    Unit

    working-days

    Realistic Timeline
    Min

    3

    Max

    5

    Unit

    months

    Notes
    • Staged VNIIMT expertise 25 wd; single 15-wd query round; sponsor response ≤35 wd excluded from the clock; conclusion 10 wd; RZN decision 5 wd
    Clinical Trial Required

    Yes

    Local Testing Required

    Yes

    Gmp Required

    Yes

    Verification Status

    official-verified

    Official Source
    Title

    Roszdravnadzor — registration timelines from 01.03.2025 (50/31 working days)

    Url

    https://roszdravnadzor.gov.ru/medproducts/registration/registration

    Effective Date

    2025-03-01

    Accessed Date

    2026-08-25

  • Id

    class-2b

    Class Name

    Class 2b

    Risk Level

    high

    Pathway

    RZN state registration (EAEU route)

    Official Timeline
    Min

    31

    Max

    50

    Unit

    working-days

    Realistic Timeline
    Min

    4

    Max

    7

    Unit

    months

    Notes
    • Staged VNIIMT expertise 25 wd; single 15-wd query round; sponsor response ≤35 wd excluded from the clock; conclusion 10 wd; RZN decision 5 wd
    Clinical Trial Required

    Yes

    Local Testing Required

    Yes

    Gmp Required

    Yes

    Verification Status

    official-verified

    Official Source
    Title

    Roszdravnadzor — registration timelines from 01.03.2025 (50/31 working days)

    Url

    https://roszdravnadzor.gov.ru/medproducts/registration/registration

    Effective Date

    2025-03-01

    Accessed Date

    2026-08-25

  • Id

    class-3

    Class Name

    Class 3

    Risk Level

    highest

    Pathway

    RZN state registration (EAEU route)

    Official Timeline
    Min

    31

    Max

    50

    Unit

    working-days

    Realistic Timeline
    Min

    5

    Max

    9

    Unit

    months

    Notes
    • Staged VNIIMT expertise 25 wd; single 15-wd query round; sponsor response ≤35 wd excluded from the clock; conclusion 10 wd; RZN decision 5 wd
    Clinical Trial Required

    Yes

    Local Testing Required

    Yes

    Gmp Required

    Yes

    Verification Status

    official-verified

    Official Source
    Title

    Roszdravnadzor — registration timelines from 01.03.2025 (50/31 working days)

    Url

    https://roszdravnadzor.gov.ru/medproducts/registration/registration

    Effective Date

    2025-03-01

    Accessed Date

    2026-08-25

Key Requirements
  • Local Authorized Representative accredited in Russia (EAEU route)
  • EAEU technical/documentation expertise, including clinical and toxicological evidence for Class 2b/3
  • GOST-R/EAC marking after registration; national register entry (RZN) required before sale
  • Re-registration within 150 working days when substantive changes occur
Prior Approvals
  • Authority

    CE Mark (EU MDR)

    Accepted

    Yes

    Benefit

    Technical and QMS documentation can be reused, but the EAEU route requires its own expertise at VNIIMT and its own Russian-language dossier. Roszdravnadzor publishes no reduced period for CE-marked devices.

    Timeline Savings

    None published

  • Authority

    FDA 510(k)/PMA

    Accepted

    Yes

    Benefit

    Usable as supporting technical evidence only. There is no recognition route and no published reduction.

    Timeline Savings

    None published

Local Rep Required

Yes

Local Rep Name

Russian Authorized Representative

Mdsap Accepted

No

Mdsap Benefit

Not accepted

Registration Validity

Indefinite (re-registration on changes)

Verification Status

official-verified

Official Sources
  • Title

    Roszdravnadzor — registration timelines from 01.03.2025 (50/31 working days)

    Url

    https://roszdravnadzor.gov.ru/medproducts/registration/registration

    Effective Date

    2025-03-01

    Accessed Date

    2026-08-25

Reviewed On

2026-08-30

Last Updated

2026-08-30

Process
Template Id

default

Status

planning-template

Usage Rule

Maintained planning steps and estimates, not verified regulator deadlines.

Steps
  • Step

    1

    Title

    Pre-Submission Preparation

    Description

    Gather technical documentation, quality management system (QMS) records, and clinical data required by Roszdravnadzor. Appoint a Russian Authorized Representative.

    Duration

    1-3 months

    Icon

    📋

  • Step

    2

    Title

    Application Submission

    Description

    Submit complete registration dossier to Roszdravnadzor including device description, risk analysis, and testing reports.

    Duration

    1-2 weeks

    Icon

    📤

  • Step

    3

    Title

    Administrative Review

    Description

    Roszdravnadzor reviews application for completeness. May request additional documents or clarifications.

    Duration

    2-4 weeks

    Icon

    🔍

  • Step

    4

    Title

    Technical Evaluation

    Description

    In-depth technical assessment of device safety and performance. Roszdravnadzor may issue questions.

    Duration

    31-50 working days

    Icon

    ⚙️

  • Step

    5

    Title

    Registration Approval

    Description

    Upon successful review, Roszdravnadzor issues Russia market authorization certificate.

    Duration

    1-2 weeks

    Icon

    ✅

Government fee schedules · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Russia fees →
Jurisdiction Id

russia

Country

Russia

Country Code

RU

Authority

Roszdravnadzor

Low Risk Local
Min

83000

Max

83000

Currency

RUB

Per Product

Yes

High Risk Local
Min

147000

Max

195000

Currency

RUB

Per Product

Yes

Details
  • Id

    registration-certificate-state-duty

    Item

    Registration certificate state duty

    Currency

    RUB

    Amount Local

    11000

    Amount Local Formatted

    RUB 11,000

    Frequency

    Per registration

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.roszdravnadzor.gov.ru/medproducts/registration/registration

    Source Title

    Roszdravnadzor medical-device registration, duties, and expertise fees

    Verified On

    2026-08-25

  • Id

    expertise-class-1

    Item

    Expertise - Class 1

    Currency

    RUB

    Amount Local

    72000

    Amount Local Formatted

    RUB 72,000

    Frequency

    Per registration

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.roszdravnadzor.gov.ru/medproducts/registration/registration

    Source Title

    Roszdravnadzor medical-device registration, duties, and expertise fees

    Verified On

    2026-08-25

  • Id

    expertise-class-2a

    Item

    Expertise - Class 2a

    Currency

    RUB

    Amount Local

    104000

    Amount Local Formatted

    RUB 104,000

    Frequency

    Per registration

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.roszdravnadzor.gov.ru/medproducts/registration/registration

    Source Title

    Roszdravnadzor medical-device registration, duties, and expertise fees

    Verified On

    2026-08-25

  • Id

    expertise-class-2b

    Item

    Expertise - Class 2b

    Currency

    RUB

    Amount Local

    136000

    Amount Local Formatted

    RUB 136,000

    Frequency

    Per registration

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.roszdravnadzor.gov.ru/medproducts/registration/registration

    Source Title

    Roszdravnadzor medical-device registration, duties, and expertise fees

    Verified On

    2026-08-25

  • Id

    expertise-class-3

    Item

    Expertise - Class 3

    Currency

    RUB

    Amount Local

    184000

    Amount Local Formatted

    RUB 184,000

    Frequency

    Per registration

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.roszdravnadzor.gov.ru/medproducts/registration/registration

    Source Title

    Roszdravnadzor medical-device registration, duties, and expertise fees

    Verified On

    2026-08-25

  • Id

    expertise-deficit-list-class-1

    Item

    Expertise - Class 1 device on the deficit list (Decree No. 552/2022)

    Currency

    RUB

    Amount Local

    32000

    Amount Local Formatted

    RUB 32,000

    Frequency

    Per registration

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.roszdravnadzor.gov.ru/medproducts/registration/registration

    Source Title

    Roszdravnadzor medical-device registration, duties, and expertise fees

    Verified On

    2026-08-25

Notes
  • Total state charges combine the RUB 11,000 certificate duty with the class-specific expertise fee.
  • State duties have applied at these levels since 1 January 2022; payment requires the УИН (unique identifier number).
  • Laboratory, samples, and inspection costs are separate.
Last Updated

2026-08-25

Verification Status

official-verified

Official Sources
  • Title

    Roszdravnadzor medical-device registration, duties, and expertise fees

    Url

    https://www.roszdravnadzor.gov.ru/medproducts/registration/registration

    Accessed Date

    2026-08-25

Reviewed On

2026-08-25

Amount Status

exact-current

Range Available

Yes

Additional certifications and licences · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Russia certifications →
Slug

russia

Market

Russia

Confidence

low

Schemes
  • Id

    eaeu-medical-device-safety-evidence

    Name

    EAEU medical-device safety evidence

    Category

    Medical Electrical Safety

    Status

    Included In Medical Registration

    Object

    Finished Device

    Note

    National conformity assessment comprises technical tests, toxicological studies and clinical trials (Minzdrav Order 885н of 30.08.2021); QMS evidence per GOST ISO 13485-2017 / ISO 13485:2016 certificates with prior inspection reports; EAEU-route testing per EAEU Decisions 25 and 26 of 17.03.2022 and Decision 29 of 12.02.2016 (edition of 24.12.2021).

    Source Urls
    • https://eec.eaeunion.org/upload/medialibrary/353/Agreement-on-common-principles-and-rules-for-the-circulation-of-medical-devices.pdf
    Source Binding

    candidate

  • Id

    eaeu-eac-lvd-emc-scope

    Name

    EAEU EAC LVD/EMC scope

    Category

    General Electrical Components

    Status

    Mandatory If Feature

    Object

    Covered Device Or Accessory

    Source Binding

    unresolved

  • Id

    radio-high-frequency-import-authorization

    Name

    Radio/high-frequency import authorization

    Category

    Radio Spectrum

    Status

    Mandatory If Feature

    Object

    Radio Equipment

    Source Urls
    • https://eec.eaeunion.org/upload/medialibrary/750/Guidance-on-the-Eurasian-Economic-Union-Trade-Regulation-2020.pdf
    Source Binding

    candidate

  • Id

    encryption-notification

    Name

    Encryption notification

    Category

    Import Conformity

    Status

    Mandatory If Feature

    Object

    Cryptographic Product

    Source Urls
    • https://eec.eaeunion.org/upload/medialibrary/750/Guidance-on-the-Eurasian-Economic-Union-Trade-Regulation-2020.pdf
    Source Binding

    candidate

  • Id

    medical-measuring-instrument-type-approval

    Name

    Medical measuring-instrument type approval

    Category

    Legal Metrology

    Status

    Mandatory If Feature

    Object

    Designated Measuring Device

    Note

    Registration dossier for measuring-instrument devices requires a draft type description of the measuring instrument and a verification method for the instrument type; the medical-device measuring-instrument list is per Minzdrav Order 257н of 29.04.2025; legal basis Federal Law 102-FZ on ensuring uniformity of measurements.

    Source Urls
    • https://roszdravnadzor.gov.ru/medproducts/registration/registration
    Source Binding

    candidate

Official Sources
  • Authority

    Eurasian Economic Commission

    Topic

    medical-device agreement and measuring instruments

    Url

    https://eec.eaeunion.org/upload/medialibrary/353/Agreement-on-common-principles-and-rules-for-the-circulation-of-medical-devices.pdf

  • Authority

    Eurasian Economic Commission

    Topic

    radio, encryption and import controls

    Url

    https://eec.eaeunion.org/upload/medialibrary/750/Guidance-on-the-Eurasian-Economic-Union-Trade-Regulation-2020.pdf

  • Authority

    Roszdravnadzor

    Topic

    registration dossier measuring-instrument type description and verification method

    Url

    https://roszdravnadzor.gov.ru/medproducts/registration/registration

  • Authority

    Roszdravnadzor

    Topic

    Order 885н technical toxicological clinical assessment rules

    Url

    https://roszdravnadzor.gov.ru/medproducts

  • Authority

    Roszdravnadzor

    Topic

    EAEU-route registration testing rules

    Url

    https://roszdravnadzor.gov.ru/spec/medproducts/registrationEAEU

Keep researching this market
View the Russia resource hub →
Compare regulatory data across all countries →
DJ Fang
Need to apply Russia's requirements to a product?
Confirm the current Roszdravnadzor pathway, evidence, and local-role details for the actual device before filing.