Swissmedic · structured requirements

🇨🇭 Switzerland IVD regulatory data

Explore the IVD requirements currently available for Switzerland, organized by topic and linked to official sources.

17 facts8 dimensions17 official sourceslatest verification 2026-10-21
Explore questions and sources →

Common questions about changing a registered IVD in Switzerland. Each question separates published rules from details still awaiting verification.

Change control

Does the event require internal documentation, notice, prior approval, transfer, or a new application?

Compare countries

Report IVD actor changes within one week

ivd.actor.data.one.week
Fact-level verified

An economic operator registered under IvDO Article 48 must report changes to its registered actor data to Swissmedic within one week. Swissmedic verifies the actor data and issues a CHRN. This is the operator’s reporting duty, not a one-week authority review period.

Conditions
  • Applies to the registered manufacturer, authorised representative or importer; the manufacturer and factory must not be conflated.

IVD notified-body change gates

ivd.conformity.change.and.certificate.gates
Fact-level verified

Under IvDO 19 and incorporated IVDR Annex IX, where Annex IX conformity assessment applies, notify the body that approved the QMS of planned substantial QMS or covered device-range changes; it assesses the proposal and any necessary audits. Under Annex IX 4.11, a change to an approved device that could affect safety, performance or prescribed conditions of use needs the issuing body’s approval. The body decides whether a new conformity assessment or a certificate supplement is required. An identity-data edit does not settle that decision.

Conditions
  • Applies only where the relevant notified/designated-body certificate and change trigger exist. A cosmetic logo, same-person office change or REF-only edit is not automatically a substantial change.
  • No universal authority or notified-body review duration or fee is established by these clauses.

IVD legacy stock requires lawful placement

ivd.lawfully.placed.legacy.stock.boundary
Fact-level verified

Qualifying devices lawfully placed on the market under the former rules or the applicable transition may continue to be made available or put into service under the stated provisions. This is a lawful-placement condition for the actual units; it does not authorise newly changed or non-conforming devices, or automatically extend an old manufacturer/brand label after a transfer. Establish placement date, conformity/legacy status, batch identity and whether the proposed alteration changes that basis.

Conditions
  • Manufacturer-held inventory is not automatically already placed on the market. No general manufacturer-name, brand, REF or successor sell-through period is established.

IVD legacy administrative changes need the actual certificate basis

ivd.legacy.administrative.change.scope
Fact-level verified

MDCG 2022-6 distinguishes administrative manufacturer name/address/legal-form changes, including its merger/acquisition example, from significant design/intended-purpose changes for legacy IVDs. This design-change guidance does not itself establish a transfer of conformity responsibility or a universal certificate-succession procedure. Observe the actual certificate conditions and agreed notified-body change procedure; current transition dates must come from amended law, not the old dates printed in this 2022 guide.

Conditions
  • Legacy-device scope only, with all actual current transition conditions satisfied.
  • This is not permission for every true new-legal-manufacturer combination or a current transition deadline.

Dossier requirements

Which evidence, local deltas, formats, and signatures are required?

Compare countries

Align export-certificate evidence after a company or product change

export.certificate.after.identity.changes
Fact-level verified

If an export destination requires an FSC, a Swiss-established manufacturer or authorised representative may apply. Update supporting documents for a company name/address change before ordering. The product list identifies the single legal manufacturer, CHRN, product name, REF/article number, class and applicable UDI/conformity references. Current-law and legacy-law products require separate orders; no mixed order is accepted. Orders processed from 1 January 2027 require the listed devices to be registered in swissdamed. This export requirement is separate from an ordinary identity-data update.

Conditions
  • Systems/procedure packs have the stated multi-manufacturer exception. Use the applicable current-law or legacy/IVD form.

IVD export-certificate evidence is separate from the MD table

ivd.export.certificate.conformity.documents
Fact-level verified

For an FSC covering current IvDO devices, supply the declaration of conformity. The table shows no body certificate for non-sterile Class A; sterile A and Classes B, C and D use the relevant class/route certificate entries, with the stated self-test, near-patient and companion-diagnostic conditions. For legacy IVDs, identify both old and current classes and the applicable transition evidence, including the prior 26 May 2022 declaration. The guide does not issue an old-law FSC for an Other-class IVDD device that is now IVDR Class A.

Conditions
  • FSC order only; do not use the MD classification matrix or treat this as an ordinary variation dossier.
  • The current transition conditions and actual conformity evidence must still be satisfied; no old printed date is extended by inference.

Economic-operator roles

Which manufacturer, representative, holder, applicant, importer, and distributor roles apply?

Compare countries

CH-REP transfer does not substitute a new legal manufacturer

actual.manufacturer.versus.chrep.transfer
Fact-level verified

A foreign manufacturer needs a Swiss-established CH-REP under a written mandate: MedDO Article 51 for MD and IvDO Article 44 for IVD. The responsible manufacturer must meet the applicable conformity, technical-documentation and declaration duties. A different legal manufacturer cannot be substituted by editing the non-updateable manufacturer SRN. A transfer of the CH-REP mandate and its registered devices is a separate representative workflow and does not itself prove that a new manufacturer succeeds to the former manufacturer’s conformity evidence.

Conditions
  • Establish whether the legal person truly changes or merely its name/legal form; identify the person covered by each declaration/certificate.
  • No automatic manufacturer succession, fresh certificate in every case, or universal combined Swiss transfer application is claimed.

Transfer a registered CH-REP mandate with its registered devices

registered.chrep.mandate.transfer
Fact-level verified

For a CH-REP-only change, the current actor, incoming actor and mandate must all have Registered status. The outgoing CH-REP requests the transfer; the incoming CH-REP must approve before the selected transfer date or the request cancels. On that date the mandate and all its registered devices transfer. The legal handover agreement also addresses mandate dates, the date until which the outgoing representative may appear in information, document transfer and complaint forwarding under the incorporated Article 12 rules.

Conditions
  • The legal manufacturer remains the same in this representative-transfer workflow. It is not a true legal-manufacturer succession procedure.

Fees

What regulator fee applies to this class, route, event, and date?

Compare countries

Swiss export-certificate fee and processing conditions

export.certificate.fee.and.service.period
Fact-level verified

A new FSC costs CHF 200 per certificate. Additional resources for product lists over 100 pages may be charged; cancellation/withdrawal or an uncompleted rejected order can attract the stated CHF 100 reduced charge. Swissmedic issues an FSC within 30 days after receipt of the order and complete, appropriate electronic documentation. An incomplete application has a separate 30-day response deadline. These are FSC conditions, not a general identity-variation fee or review clock.

Conditions
  • Only when a fresh export-certificate order is actually needed. The regular validity is 3 years,5 for Thailand; validity is not a review period.

Existing-data updates differ from new device registrations

swissdamed.data.update.fees
Fact-level verified

Updating data for an already registered UDI-DI incurs no new device registration fee. Re-registering a discarded UDI-DI is fee-bearing under the current registration schedule. Actor data changes are usually free, but Swissmedic intervention may be charged on actual time at CHF 270 per hour. A new UDI/device registration resulting from the actual identity change must not be presented as a free data update.

Conditions
  • The exact canonical first/additional-UDI fees apply only to a genuinely chargeable new or re-registered UDI, with per-manufacturer/year limits; they are not an ordinary variation tariff.

Label, IFU & language

Which label and IFU elements, languages, symbols, and translation controls apply?

Compare countries

IVD label identity after a change

ivd.changed.label.manufacturer.and.device.identity
Fact-level verified

Under IvDO 15 and incorporated IVDR label requirements, the label identifies the device name/trade name and the manufacturer’s name, registered trade name or trademark and registered-place-of-business address, plus the representative where required and the applicable UDI carrier. Keep the actual post-change device/manufacturer identities consistent with the conformity and registration records. This label-content rule is not an old-stock grace or a permission to use another company’s certificate.

Conditions
  • Apply the national language and representative particulars, including the actual device and use conditions.

IVD language requirements and their limited exceptions

ivd.three.languages.and.narrow.exceptions
Fact-level verified

IVD labelling and instructions are generally required in German, French and Italian. The fewer-languages/English exception requires all IvDO 15(3) conditions, including the stated professional-use or Art 9 branch, the user’s professional/language ability and agreement, assured protection and no jeopardy to effective use. For self-testing and near-patient testing, Article 15(8) requires the specified information to be easily understandable in all three languages. The limited placement of CH-REP particulars on an accompanying document has its own Article 15(9) conditions.

Conditions
  • Do not inherit the MD custom-made exception into IVD or use professional supply alone to waive the cumulative conditions.

Official forms & portals

Which current form, portal, account, and submission package must be used?

Compare countries

Keep the Swiss actor, mandate and device records consistent

actor.mandate.and.device.record.prerequisites
Fact-level verified

Manufacturers and their CH-REP register the actual device/UDI data in swissdamed under the applicable legislation. The manufacturer SRN in the mandate and device data must match. Device registration became mandatory 1 July 2026 with the published transition to 31 December 2026; specified vigilance cases require immediate registration. The Swiss actor, mandate and device records remain separate from an EU EUDAMED registration, and cannot be assumed to update one another automatically.

Conditions
  • The registration transition is not a manufacturer-name, new-entity or old-label stock grace period.

Update Swiss company data through the correct actor path

same.person.company.and.actor.update
Fact-level verified

For a company synchronised with the UID register/Zefix, correct the company name or address there first; swissdamed receives the update automatically. For a manually registered company or one that opted out of synchronisation, the current FAQ describes the Registered companies pencil editor and directs company-name changes to Swissmedic support. The Actors guide separately directs support for company data that cannot be edited through the actor profile. Other actor details are updated by the authorised user and confirmed. A name/address edit does not establish transfer to a different manufacturer.

Conditions
  • This is the existing company/actor data path. A factory move, new legal person or device identity change needs its own assessment.

Reliance & recognition

Which foreign decisions or evidence can be reused, with what effect and conditions?

Compare countries

Existing records include acceptance, restrictions and evidence reuse. The applicable product scope and conditions are preserved below; a reference does not by itself confer market access.

Switzerland approvals and evidence in other markets (23 records)

ArmeniaBrazil ANVISA · EU conformity assessment · US FDA · Health Canada · Singapore HSA · UK MHRA · South Korea MFDS · Japan MHLW / PMDA · NMPA · Australia TGA · IMDRF member authorities · WHO

Prior approval / evidence record

Simplified national registration for a device registered in an IMDRF member country or prequalified by WHO

reliance.imdrf-member-or-who-prequalified-simplified-registration
Fact-level verified
Statement

Armenia registers medical devices under two national procedures: a general procedure and a simplified procedure. The simplified procedure applies to a device that is registered in a country that is a member of the International Medical Device Regulators Forum (IMDRF) or that is prequalified by the World Health Organization. IMDRF members include the United States (FDA), the EU member states (CE marking), Japan (PMDA), Korea (MFDS), China (NMPA), Canada, Australia, Brazil, Singapore and the United Kingdom, so a US FDA 510(k)/PMA clearance, an EU CE mark, a Japan PMDA approval and a Korea MFDS approval all satisfy the gate. The device is still registered in Armenia by the Armenian authority — the foreign registration opens the shortened procedure, it does not replace the Armenian certificate.

Source Approval

Registration in any IMDRF member country — US FDA 510(k)/PMA, EU CE marking under MDR/IVDR (any EU member state), Japan PMDA, Korea MFDS, China NMPA, Health Canada, Australia TGA, Brazil ANVISA, Singapore HSA, UK MHRA — or WHO prequalification of the device

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Premarket Review

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • The provision is written as an annex to part 3 of the same article: 'under the procedure laid down by part 3 of this article' devices are registered under the general and simplified procedures. Part 3 is itself the Armenia-only route, and a footnote to Article 46 states that part 3 operates only until the end of the transitional period set by Article 11 of the EAEU Agreement of 23 December 2014 (footnote inserted by Article 3(2) of the Law of 5 December 2024 ՀՕ-502-Ն). The transitional period was extended by the Protocol signed in Moscow on 29 December 2025, which replaced '2025' with '2027' in the first paragraph of Article 11 of that Agreement and '2026' with '2028' in the third and fourth paragraphs.
  • Bounded claim: the Law names IMDRF membership and WHO prequalification as the gate for the simplified procedure. It does not name the FDA, the CE mark, the PMDA or the MFDS; those follow from their countries' IMDRF membership, not from the text.
  • The operational content of the 'simplified procedure' (dossier list, clocks, certificate form) sits in the Government procedure that Article 46(3) contemplates; that instrument was not located in this session (see unresolved: registration-pathways).
  • Devices are registered by the Ministry of Health (the լիազոր մարմին / authorised body) with the examination carried out by the expert organisation designated by Government decision (Article 46(2), (5)).

Language: Armenian. Verbatim, Article 46(4): "Սույն հոդվածի 3-րդ մասով սահմանված կարգի համաձայն՝ բժշկական արտադրատեսակները գրանցվում են ընդհանուր և պարզեցված ընթացակարգերով: Պարզեցված ընթացակարգը կիրառվում է բժշկական արտադրատեսակների շրջանառությունը կարգավորող մարմինների միջազգային համաժողովի (IMDRF) անդամ երկրում գրանցված կամ Առողջապահության համաշխարհային կազմակերպության նախաորակավորած բժշկական արտադրատեսակների համար:" Translation: "Under the procedure laid down by part 3 of this article, medical devices are registered under general and simplified procedures. The simplified procedure applies to medical devices registered in a member country of the International Medical Device Regulators Forum (IMDRF) or prequalified by the World Health Organization." Article 46 was amended by the Law of 5 December 2024 ՀՕ-502-Ն; the consolidated expression is the version in force from 31.05.2026.

BahrainSaudi Arabia · US FDA · Australia TGA · Japan MHLW / PMDA · UK MHRA · Ireland · Swissmedic · France · Germany

Prior approval / evidence record

Registration certificate from SFDA (Saudi Arabia), US FDA, TGA (Australia), PMDA (Japan), MHRA (United Kingdom), HPRA (Ireland), Swissmedic (Switzerland), ANSM (France) or BfArM (Germany)

Authority

Registration certificate from SFDA (Saudi Arabia), US FDA, TGA (Australia), PMDA (Japan), MHRA (United Kingdom), HPRA (Ireland), Swissmedic (Switzerland), ANSM (France) or BfArM (Germany)

Accepted

Yes

Benefit

Article (7) bis of Resolution No. (48) of 2020, inserted by Resolution No. (69) of 2024, provides by way of exception to the ordinary Article (7) prohibition that medical devices and products registered in one of the nine listed reference countries 'shall also be registered in the Kingdom of Bahrain' on the basis of that country's registration certificate, on payment of the applicable fee. This is a documentary registration route rather than a full de novo technical assessment, but the application is still filed by a licensed Bahraini Authorized Representative and NHRA retains the decision.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

BahrainUS FDA · Japan MHLW / PMDA · Saudi Arabia · Australia TGA · UK MHRA · Ireland · Swissmedic · France · Germany

Prior approval / evidence record

Registration certificate from a listed reference country is an alternative basis for Bahraini registration (Art. 7 bis)

reliance.reference-country-registration-certificate-route
Fact-level verified
Statement

Article (7) bis of Resolution No. (48) of 2020, inserted by Resolution No. (69) of 2024, provides that by way of exception to the ordinary Article (7) prohibition, medical devices and products registered in one of nine listed reference countries 'shall also be registered in the Kingdom of Bahrain' on the basis of the registration certificate issued by that country, following payment of the applicable fee. The listed reference authorities are SFDA (Saudi Arabia), FDA (United States), TGA (Australia), PMDA (Japan), MHRA (United Kingdom), HPRA (Ireland), Swissmedic (Switzerland), ANSM (France) and BfArM (Germany). A holder of a US FDA or Japan PMDA registration therefore has a documentary registration route in Bahrain rather than a full de novo technical assessment.

Source Approval

US FDA registration/clearance; Japan PMDA approval; Saudi SFDA registration; Australia TGA registration; UK MHRA registration; Ireland HPRA registration; Swissmedic (Switzerland) registration; France ANSM registration; Germany BfArM registration

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Only the nine authorities named in Article (7) bis are listed; Korea MFDS, Health Canada and China NMPA are not among them
  • The EU CE mark as such is not listed; the Union route appears only through three named national authorities (HPRA, ANSM, BfArM)
  • The application is still made by a Bahraini licensed Authorized Representative and the applicable fee must be paid
  • In force from the day following publication in the Official Gazette (No. 3781, 21 November 2024)
  • Resolution No. (69) of 2024 replaced Article (10) of Resolution No. (48) of 2020 so that marketing authorisation runs for three years

Effective from 2024-11-22

Verbatim source passage: "By way of exception to Article (7) of this Resolution, medical devices and products registered in one of the following reference countries shall also be registered in the Kingdom of Bahrain: 1. Kingdom of Saudi Arabia - Represented by the Saudi Food and Drug Authority (SFDA). 2. United States of America - Represented by the Food and Drug Administration (FDA). 3. Australia - Represented by the Therapeutic Goods Administration (TGA). 4. Japan - Represented by the Pharmaceuticals and Medical Devices Agency (PMDA). 5. United Kingdom - Represented by the Medicines and Healthcare products Regulatory Agency (MHRA). 6. Ireland - Represented by the Health Products Regulatory Authority (HPRA). 7. Switzerland - Represented by Swissmedic(Swissmedic). 8. France - Represented by the French National Agency for Medicines and Health Products Safety (ANSM). 9. Germany - Represented by the Federal Institute for Drugs and Medical Devices (BfArM). Based on the registration certificate issued by one of these countries, following the payment of the applicable fee."

BahrainSaudi Arabia · US FDA · UK MHRA · Australia TGA · Health Canada · Japan MHLW / PMDA · Swissmedic · Ireland · Denmark · New Zealand · France · Netherlands · Belgium

Prior approval / evidence record

Free Sale Certificate may be issued by the country of origin or by a reference country

reliance.free-sale-certificate-origin-or-reference-country
Fact-level verified
Statement

At renewal, Bahrain requires a Free Sale Certificate (FSC) or Certificate to Foreign Government issued by the regulatory authority of the country of origin or of a reference country (Registration Guideline 9.1, section 10 item 4). The guideline's glossary (item 14) defines reference countries as Saudi Arabia, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland and Belgium, so a certificate from one of those authorities substitutes for an origin-country free-sale certificate. A Free Sale Certificate is also required on manufacturer acquisition (section 12).

Source Approval

Free Sale Certificate or Certificate to Foreign Government from the country of origin or from a reference country (Saudi Arabia, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland, Belgium)

Edge Type

Origin Country Evidence

Market Access Strength

Evidence Only

Regulatory Layer

Premarket Review

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • The classification guideline separately accepts a Free Sale Certificate OR Registration Certificate from one of the reference countries as a classification requirement
  • The reference-country list in the registration guideline glossary includes Canada, Denmark, the Netherlands ('Holland') and New Zealand, which are not in the Article (7) bis list of Resolution No. (69) of 2024, and omits Germany
  • Free Sale Certificate is also required on manufacturer acquisition (Section 12)

Verbatim source passages: "Free Sale Certificate (FSC) or Certificate to foreign government issued by the regulatory authority of the country of origin or a reference country." and "Reference countries: Saudi Arabia, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland, Belgium."

BahrainSaudi Arabia · US FDA · UK MHRA · Australia TGA · Health Canada · Japan MHLW / PMDA · Swissmedic · Ireland · Denmark · New Zealand · France · Netherlands · Belgium

Prior approval / evidence record

Classification of a product as a medical device may be evidenced by a registration certificate from a reference country

reliance.reference-country-certificate-for-classification
Fact-level verified
Statement

To obtain an NHRA classification letter, the applicant must supply a Free Sale Certificate or Registration Certificate issued by the competent authority of the country of origin or by one of the reference countries listed in the classification guideline (SFDA, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland, Belgium), together with a recognised quality assurance certificate such as ISO 13485, CE mark or FDA.

Source Approval

Registration certificate or Free Sale Certificate from SFDA (Saudi Arabia), US FDA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland or Belgium

Edge Type

Origin Country Evidence

Market Access Strength

Evidence Only

Regulatory Layer

classification

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • A classification letter is a separate, fee-bearing NHRA service that precedes importation and registration
  • The reference-country list here is broader than the Article (7) bis list in Resolution No. (69) of 2024 and omits Germany and includes Canada, Denmark, the Netherlands and New Zealand
  • For class I products without CE marking a manufacturer's Declaration of Conformity may be provided

Verbatim source passage: "1. Free Sale Certificate OR Registration Certificate issued from competent authority in country of origin classifies the product as a medical device. 2. Free Sale Certificate OR Registration Certificate issued from one of the following reference countries: (SFDA, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, Newzland, France, and Holland, Belgium.)."

BotswanaEU conformity assessment · US FDA · Japan MHLW / PMDA · Swissmedic · Health Canada · Australia TGA · Iceland · Norway

Prior approval / evidence record

Stringent Regulatory Authority approval — ICH members (European Commission, US FDA, Japan MHLW/PMDA), ICH observers (Swissmedic, Health Canada) or an authority associated with an ICH member through a legally binding mutual recognition agreement (including Australia, Iceland, Liechtenstein and Norway)

Authority

Stringent Regulatory Authority approval — ICH members (European Commission, US FDA, Japan MHLW/PMDA), ICH observers (Swissmedic, Health Canada) or an authority associated with an ICH member through a legally binding mutual recognition agreement (including Australia, Iceland, Liechtenstein and Norway)

Accepted

No

Benefit

The SRA definition in G01 s.3.1.26 is used for dossier and reference purposes rather than as an automatic grant: an SRA approval does not by itself confer a Botswana registration, and it benefits the applicant only through the abridged pathway's proof-of-reference-agency requirement, which is defined by the separate Reference Regulatory Authority concept.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

IsraelUS FDA · UK MHRA · Germany · France · Italy · Spain · Netherlands · Belgium · Ireland · Denmark · Sweden · Finland · Norway · Iceland · Greece · Portugal · Austria · Swissmedic · Australia TGA · Health Canada · New Zealand

Regulatory reference

Recognized-state registration route

reliance.recognized-country-route
Fact-level verified

A device registered, or approved for marketing and actually marketed, in a 'recognized state' (מדינה מוכרת) is registered in the Israeli register for a period not exceeding the period it may be marketed in that state, capped at 5 years from Israeli registration. Recognized states are listed in the First Addendum to the law — 21 states including the US, UK, Germany, France, Italy, Spain, the Netherlands, Belgium, Ireland, Denmark, Sweden, Finland, Norway, Iceland, Greece, Portugal, Austria, Switzerland, Australia, Canada and New Zealand (no single 'EU' entry: member states are listed individually).

Conditions
  • Device must be registered or permitted for marketing and actually marketed in the recognized state.

First Addendum list per the official Sefer HaChukim text of the law held from gov.il.

LebanonIMDRF member authorities

Prior approval / evidence record

GHTF/IMDRF member countries — free-sale certificate (شهادة بيع حر)

Authority

GHTF/IMDRF member countries — free-sale certificate (شهادة بيع حر)

Accepted

Yes

Benefit

A mandatory document rather than a route: Decision 1506/1 Art. Four requires the registration application to carry a free-sale certificate issued by a country belonging to the GHTF/IMDRF group, meaning the device is sold in one of those countries whatever its origin, and the Ministry's service records require a free-sale certificate with an ISO 13485 certificate where the goods are made outside the reference states or the listed documents cannot be produced. No examination period is shortened by it.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

LebanonUS FDA · EU conformity assessment · Swissmedic · Australia TGA · Health Canada · Japan MHLW / PMDA

Prior approval / evidence record

US FDA (510(k)/PMA approval or approval letter), European Union (conformity certificate issued by the manufacturer and by EU-notified bodies), Swiss, Australian, Canadian or Japanese competent health authorities (conformity certificate)

Authority

US FDA (510(k)/PMA approval or approval letter), European Union (conformity certificate issued by the manufacturer and by EU-notified bodies), Swiss, Australian, Canadian or Japanese competent health authorities (conformity certificate)

Accepted

Yes

Benefit

Accepted and in practice required for an implantable device: Decision 455/1 Art. Six lists them as the conformity documents that go into the registration file — a European-standards certificate from the manufacturer and EU-notified bodies, an FDA marketing approval or approval letter, or a conformity certificate from the health authorities of Switzerland, Australia, Canada or Japan — and Decision 1506/1 Art. Four requires a GHTF/IMDRF free-sale certificate on top. The Ministry's service records for sterile instruments and laboratory reagents require FDA evidence for US-origin goods and CE evidence for European ones. The benefit is documentary, not a shortened clock: none of the instruments publishes an abbreviated examination period, and the registration certificate is still issued on the Health Technology Unit's advice to the Minister.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

LiechtensteinSwissmedic

Regulatory reference

Swiss authorisations valid across the customs territory

reliance.swiss-customs-territory-recognition
Fact-level verified

Exceptional authorisations (Ausnahmebewilligungen) for non-compliant devices issued by Swissmedic under the Swiss-side rules (Art. 22 Abs. 1 MepV, Art. 18 Abs. 1/61 Abs. 3 IvDV) are valid for the whole customs territory (Switzerland/Liechtenstein); Liechtenstein accepts Swissmedic acts for the customs-union channel.

Effective from 2024-02-27

MexicoNMPA · US FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA · UK MHRA · Swissmedic · South Korea MFDS · Singapore HSA · Brazil ANVISA · IMDRF member authorities · MDSAP · WHO

Prior approval / evidence record

Vía regulatoria abreviada — ARR equivalence (COFEPRIS-04-050)

Name

Vía regulatoria abreviada — ARR equivalence (COFEPRIS-04-050)

Description

The single reliance route since the DOF Acuerdo of 18 July 2025, which took effect 30 working days after publication and abrogated every individual equivalence agreement issued between 2012 and June 2025 — the separate FDA, Swissmedic, European Commission, Health Canada, TGA and WHO prequalification acuerdos no longer exist. COFEPRIS now recognises the decision of any Reference Regulatory Authority from a country sitting on the IMDRF Management Committee, and MDSAP full membership for GMP certificates. The dossier drops to an administrative-legal section plus technical information under RIS Art. 31.

Timeline Reduction

Resolution within a maximum of 30 working days

Eligibility
  • Prior authorisation by a Reference Regulatory Authority in one of the twelve IMDRF Management Committee jurisdictions: Australia, Brazil, Canada, China, the European Union, Japan, Russia, Singapore, South Korea, Switzerland, the United Kingdom, the United States
  • MDSAP full members are recognised for the good-manufacturing-practice certificate
  • Filed under homoclave COFEPRIS-04-050 rather than the ordinary COFEPRIS-04-060 registration
  • Post-market surveillance or technovigilance evidence from the reference market, translated into Spanish under RIS Art. 153

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

MexicoUS FDA · Health Canada · Japan MHLW / PMDA · IMDRF member authorities · MDSAP

Prior approval / evidence record

Class II

Id

class-ii

Class Name

Class II

Risk Level

medium

Pathway

Registration or Equivalency

Official Timeline
Min

35

Max

35

Unit

days

Verification Status

official-verified

Official Source
Title

RIS Art. 179 — statutory resolution periods by class (COFEPRIS)

Url

https://salud.gob.mx/unidades/cdi/nom/compi/ris.html

Accessed Date

2026-08-25

Realistic Timeline
Min

15

Max

31

Unit

months

With Prior Approval
Min

8

Max

16

Unit

months

Prior Approval Benefits
  • FDA 510(k)
  • Health Canada
  • Japan PMDA
  • IMDRF Committee Members
  • MDSAP
Notes
  • RIS Art. 179: Clase II resolved within 35 calendar days (15 days with a favourable Tercero Autorizado dictamen)
  • ARR equivalence route (COFEPRIS-04-050, Lineamientos DOF 18-07-2025): maximum 30 working days
  • Equivalency route much faster if FDA approved
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

MexicoUS FDA · Health Canada · Japan MHLW / PMDA · IMDRF member authorities · MDSAP

Prior approval / evidence record

Class III

Id

class-iii

Class Name

Class III

Risk Level

highest

Pathway

Registration or Equivalency

Official Timeline
Min

60

Max

60

Unit

days

Verification Status

official-verified

Official Source
Title

RIS Art. 179 — statutory resolution periods by class (COFEPRIS)

Url

https://salud.gob.mx/unidades/cdi/nom/compi/ris.html

Accessed Date

2026-08-25

Realistic Timeline
Min

15

Max

31

Unit

months

With Prior Approval
Min

8

Max

16

Unit

months

Prior Approval Benefits
  • FDA 510(k)/PMA
  • Health Canada
  • Japan PMDA
  • IMDRF Committee Members
  • MDSAP
Notes
  • RIS Art. 179: Clase III resolved within 60 calendar days (15 days with a favourable Tercero Autorizado dictamen)
  • ARR equivalence route (COFEPRIS-04-050): maximum 30 working days
  • Equivalency pathway best option with FDA
Clinical Trial Required

Yes

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

ParaguayBrazil ANVISA · US FDA · Health Canada · Austria · Belgium · Denmark · Finland · France · Germany · Hungary · Ireland · Italy · Netherlands · New Zealand · Norway · Spain · Sweden · Japan MHLW / PMDA · UK MHRA · Swissmedic · Australia TGA · WHO

Prior approval / evidence record

IVD Annex I reference-country list drives the 30/45/120-business-day clocks

reliance.ivd-anexo-i-reference-country-list
Fact-level verified
Statement

For IVDs, DINAVISA decides within 30 business days for products manufactured nationally or in a country of Annex I, within 45 business days for products manufactured outside Annex I but registered and commercialised in an Annex I country (which must additionally file the foreign registration certificate and free-sale certificate), and within 120 business days for other imports. Annex I names high-surveillance and adequate-surveillance authorities and the full MERCOSUR members; the United States and Japan are on it and Korea is not.

Source Approval

Sanitary registration and free-sale certificate in force in an Annex I country — high-surveillance authorities (Germany, Austria, Belgium, Canada, Denmark, Spain, United States, France, Israel, Italy, Japan, Netherlands, United Kingdom, Sweden, Switzerland) or adequate-surveillance authorities (Australia, Chile, Cuba, Finland, Hungary, Ireland, Luxembourg, Mexico, Norway, New Zealand); Annex I also lists full MERCOSUR members (Argentina, Brazil, Paraguay, Uruguay) and PAHO/WHO reference authorities

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Premarket Review

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • The Annex I list derives from Ley N° 3283/2007 Art. 11 (high and adequate sanitary surveillance).
  • A product made outside Annex I is admitted on a 45-day clock only if it is registered and commercialised in an Annex I country.
  • Korea (MFDS) does not appear in Annex I, so a Korean-only IVD authorisation does not by itself reach the 30/45-day clocks.
  • A Paraguayan sanitary registration is still required in every case.

Source language: Spanish. Translation of the quoted Annex I heading and entries: 'Countries with High and Adequate Sanitary Surveillance Authorities (Law 3283/07 Article 11)' / 'Countries with high sanitary surveillance regulatory authorities: Germany, Austria, Belgium, Canada, Denmark, Spain, United States, France, Israel, Italy, Japan, Netherlands, United Kingdom, Sweden and Switzerland.' / 'Full MERCOSUR members: Argentina, Brazil, Paraguay and Uruguay.' The excerpt is taken from the resolution's own Annex I; Art. 6 supplies the 30- and 45-business-day clocks.

ParaguayWHO · IMDRF member authorities

Prior approval / evidence record

PAHO/WHO reference authority or IMDRF Management Committee member authority

Authority

PAHO/WHO reference authority or IMDRF Management Committee member authority

Accepted

Yes

Benefit

Operative for class II, III and IV IVDs, which enter the Proceso Simplificado de Registro Sanitario under Resolución DINAVISA N° 44/2024 with a 15-business-day validation and approval by positive silence.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

ParaguayBrazil ANVISA · EU conformity assessment · US FDA · Health Canada · Singapore HSA · IMDRF member authorities · Portugal · Spain · South Korea MFDS · Japan MHLW / PMDA · UK MHRA · NMPA · Taiwan TFDA · Australia TGA · WHO

Prior approval / evidence record

IVD simplified registration (PSR) for products authorised by PAHO/WHO, IMDRF or bilateral-partner authorities

reliance.ivd-psr-reference-authorities
Fact-level verified
Statement

Class II, III and IV in-vitro diagnostic products already authorised and commercialised in countries regulated by PAHO/WHO reference regulatory authorities, by authorities of IMDRF Management Committee member countries, or by authorities with bilateral agreements with DINAVISA enter the Proceso Simplificado de Registro Sanitario (PSR). DINAVISA has 15 business days to validate and decide, and approval follows by positive silence if no report is issued.

Source Approval

PAHO/WHO reference regulatory authorities; authorities of IMDRF Management Committee member countries (the IMDRF committee includes US FDA, EU, Japan PMDA/MHLW, Health Canada, TGA, ANVISA, China NMPA, Korea MFDS, Singapore HSA and UK MHRA); authorities with bilateral agreements with DINAVISA

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Premarket Review

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Applies to class II, III and IV IVDs (Res. 44/2024 Art. 7: 'Clases II, III y IV'); class I IVDs are registered automatically under Res. 266/2022 Art. 14.
  • The IVD must actually be commercialised in one of those countries and the foreign registration must correspond specifically to the same product being registered.
  • 15 business days to validate; approval by positive silence.
  • Objections may be answered only once (Art. 3), with 30 business days to rectify.
  • Not applicable to the cases in Arts. 13 and 16 of Res. DINAVISA N° 266/2022 (Art. 11).

Effective from 2024-01-30

Source language: Spanish. Translation of the operative part of Art. 2: 'Applications for sanitary registration of In Vitro Diagnostic Products ... that are authorised and commercialised in countries regulated by: PAHO/WHO Reference Regulatory Authorities; Regulatory Authorities of member countries of the Committee of the International Medical Device Regulators Forum - IMDRF; Regulatory Authorities with which bilateral agreements with DINAVISA exist, shall be subject to the Simplified Sanitary Registration Process (PSR).' and 'DINAVISA shall have a term of 15 (fifteen) business days to validate the required documentation and decide thereon. Once the term has elapsed with no report from DINAVISA, the application shall be approved.' Article numbering confirmed by reading the page image of the official PDF.

ParaguayWHO · IMDRF member authorities

Prior approval / evidence record

Proceso simplificado de Registro Sanitario (PSR) para productos para diagnóstico de uso in vitro

Name

Proceso simplificado de Registro Sanitario (PSR) para productos para diagnóstico de uso in vitro

Description

Class II, III and IV IVDs (Res. 44/2024 Art. 7) already authorised and commercialised by a recognised reference authority are validated in 15 business days with approval by positive silence, instead of the 30, 45 or 120 business days of the ordinary IVD route.

Timeline Reduction

From 30-120 business days down to 15 business days

Eligibility
  • Product classified as class II, III or IV under Resolución DINAVISA N° 266/2022 Art. 9
  • Current authorisation and commercialisation in a country regulated by a PAHO/WHO reference regulatory authority, by an IMDRF Management Committee member authority, or by an authority with a bilateral agreement with DINAVISA
  • The foreign registration must correspond specifically to the same product being registered in Paraguay

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

PeruUS FDA · UK MHRA · Health Canada · Japan MHLW / PMDA · Australia TGA · Swissmedic · South Korea MFDS · Germany · France · Netherlands · Spain · Italy · Belgium · Portugal · Ireland · Austria · Hungary · Denmark · Norway · Sweden

Prior approval / evidence record

País de Alta Vigilancia Sanitaria (art. 9 D.S. 016-2011-SA): United States, Canada, Japan, Australia, Switzerland, South Korea, United Kingdom, Germany, France, Netherlands, Spain, Italy, Belgium, Portugal, Ireland, Austria, Hungary, Denmark, Norway, Sweden

Authority

País de Alta Vigilancia Sanitaria (art. 9 D.S. 016-2011-SA): United States, Canada, Japan, Australia, Switzerland, South Korea, United Kingdom, Germany, France, Netherlands, Spain, Italy, Belgium, Portugal, Ireland, Austria, Hungary, Denmark, Norway, Sweden

Accepted

Yes

Benefit

The Peruvian route to relying on a foreign approval is documentary, not procedural. A certificate of free sale or a GMP certificate issued by a high-surveillance authority is accepted directly, and if the issuing authority appears on the Resolución Directoral 169-2014-DIGEMID list the document does not need an apostille. Everything else still goes through DIGEMID certification.

Timeline Savings

None published — the TUPA period is set by class, not by origin

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

UgandaWHO · IMDRF member authorities · ISO

Prior approval / evidence record

WHO Prequalification, or quality-system certification from a certification body accredited by a regulatory authority in an IMDRF founding-member country, or any other international organisation recognised by NDA

Authority

WHO Prequalification, or quality-system certification from a certification body accredited by a regulatory authority in an IMDRF founding-member country, or any other international organisation recognised by NDA

Accepted

Yes

Benefit

Track 2: accepted as a documentary substitution for an IMDRF-country licence under DAR/GDL/028 §4.12.2 — the foreign element that unlocks the route is the quality-system certification and its accreditation rather than a product approval, and no shortened period is published for it.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

UgandaAustralia TGA · Health Canada · EU conformity assessment · Japan MHLW / PMDA · US FDA · IMDRF member authorities

Prior approval / evidence record

Licence or approval from an IMDRF founding-member country — Australia, Canada, the European Union, Japan, the United States

Authority

Licence or approval from an IMDRF founding-member country — Australia, Canada, the European Union, Japan, the United States

Accepted

Yes

Benefit

Track 1: an abridged evaluation in which the foreign approval carries the technical assessment, with a much smaller submission (attestation, abbreviated product and manufacturer details, notarized copy of the licence, Certificate of Analysis to an accepted quality-system standard, evidence of repeat sales and product samples where practicable). It is an expedited review with shortened registration timelines, but NDA publishes no number for the shortening and the final decision remains NDA's.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

VietnamNMPA · US FDA · EU conformity assessment · Australia TGA · Japan MHLW / PMDA · UK MHRA · Swissmedic · South Korea MFDS

Prior approval / evidence record

Class C

Id

class-c

Class Name

Class C

Risk Level

high

Pathway

Circulation number registration (Ministry of Health)

Official Timeline
Min

45

Max

45

Unit

days

Verification Status

official-verified

Official Source
Title

Nghị định 98/2021/NĐ-CP on the management of medical devices (as amended by Decrees 07/2023, 96/2023, 04/2025) — Art. 28 declaration for Class A/B and Art. 32 appraisal periods for Class C/D

Url

https://vanban.chinhphu.vn/?pageid=27160&docid=204442

Effective Date

2022-01-01

Accessed Date

2026-08-29

Realistic Timeline
Min

4

Max

5

Unit

months

With Prior Approval
Min

10

Max

10

Unit

working-days

Prior Approval Benefits
  • FDA
  • TGA
  • MHLW
  • EU
  • MHRA
  • Swissmedic
  • NMPA
  • MFDS
Notes
  • Decree 98/2021 Art. 32(5): the Minister of Health must appraise and grant the circulation number within 45 days of a complete and valid dossier, counted only once the appraisal fee receipt is in, and must issue any request for supplementation within 40 days
  • Art. 32(2) shortens this to 30 days, with a 25-day supplementation notice, where a corresponding national technical regulation (QCVN) exists for the device, and the same 30 days applies to measuring instruments that need pattern approval
  • Art. 32(3): the fast track for devices already authorised in a reference country, or previously import-licensed, is decided in 10 working days with a 10-working-day supplementation notice — here the Ministry appraises only the legal documents and verifies the rest after the number is granted
  • Art. 32(4): the emergency route is also 10 working days, with an 8-working-day supplementation notice
  • The clock does not simply run out in the applicant favour: if the dossier is not corrected within 90 days of a request, or fails after five rounds of supplementation, the procedure restarts from the beginning
  • Art. 32(6): the granted number and the dossier are published on the medical device portal within 1 working day
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

VietnamNMPA · US FDA · EU conformity assessment · Australia TGA · Japan MHLW / PMDA · UK MHRA · Swissmedic · South Korea MFDS

Prior approval / evidence record

Class D

Id

class-d

Class Name

Class D

Risk Level

highest

Pathway

Circulation number registration (Ministry of Health)

Official Timeline
Min

45

Max

45

Unit

days

Verification Status

official-verified

Official Source
Title

Nghị định 98/2021/NĐ-CP on the management of medical devices (as amended by Decrees 07/2023, 96/2023, 04/2025) — Art. 28 declaration for Class A/B and Art. 32 appraisal periods for Class C/D

Url

https://vanban.chinhphu.vn/?pageid=27160&docid=204442

Effective Date

2022-01-01

Accessed Date

2026-08-29

Realistic Timeline
Min

4

Max

5

Unit

months

With Prior Approval
Min

10

Max

10

Unit

working-days

Prior Approval Benefits
  • FDA
  • TGA
  • MHLW
  • EU
  • MHRA
  • Swissmedic
  • NMPA
  • MFDS
Notes
  • Decree 98/2021 Art. 32(5): the Minister of Health must appraise and grant the circulation number within 45 days of a complete and valid dossier, counted only once the appraisal fee receipt is in, and must issue any request for supplementation within 40 days
  • Art. 32(2) shortens this to 30 days, with a 25-day supplementation notice, where a corresponding national technical regulation (QCVN) exists for the device, and the same 30 days applies to measuring instruments that need pattern approval
  • Art. 32(3): the fast track for devices already authorised in a reference country, or previously import-licensed, is decided in 10 working days with a 10-working-day supplementation notice — here the Ministry appraises only the legal documents and verifies the rest after the number is granted
  • Art. 32(4): the emergency route is also 10 working days, with an 8-working-day supplementation notice
  • The clock does not simply run out in the applicant favour: if the dossier is not corrected within 90 days of a request, or fails after five rounds of supplementation, the procedure restarts from the beginning
  • Art. 32(6): the granted number and the dossier are published on the medical device portal within 1 working day
Clinical Trial Required

Yes

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

ZambiaIMDRF member authorities · WHO · US FDA · EU conformity assessment

Prior approval / evidence record

WHO Prequalification; US FDA approval; EU CE marking; approval in any IMDRF member state

Authority

WHO Prequalification; US FDA approval; EU CE marking; approval in any IMDRF member state

Accepted

No

Benefit

Disclosure only. The Annex I application form requires the applicant to state whether the device is WHO prequalified, US FDA approved, CE marked or approved in any IMDRF member state and to attach valid documentation, but neither the non-IVD nor the IVD guideline attaches any procedural consequence — no shortened clock, no abridged dossier and no substitute for ZAMRA's own evaluation.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

Prior approvals and evidence for Switzerland (7 records)

SwitzerlandEU conformity assessment

Prior approval / evidence record

EU CE Mark (MDR)

Authority

EU CE Mark (MDR)

Accepted

Yes

Benefit

Full recognition - immediate Swiss market access

Timeline Savings

Immediate

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

SwitzerlandEU conformity assessment · Swissmedic

Prior approval / evidence record

Unilateral recognition of EU/EEA certificates

Name

Unilateral recognition of EU/EEA certificates

Description

The device chapter of the EU-Switzerland MRA has not applied since 26 May 2021 and has not been restored; Switzerland instead recognises EU/EEA notified-body certificates one-way under MepV Art. 25(4), which is why a CH-REP and Swiss labelling are still required

Timeline Reduction

No second conformity assessment

Eligibility
  • Certificate from a body designated under EU law and established in an EU or EEA state
  • Procedures must meet Swiss requirements and the body must hold equivalent qualification

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

Prior approval / evidence record

FDA 510(k)

Authority

FDA 510(k)

Accepted

No

Benefit

Documentation can support Swiss application

Timeline Savings

Indirect

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

SwitzerlandEU conformity assessment · Swissmedic

Prior approval / evidence record

Class I

Id

class-i

Class Name

Class I

Risk Level

low

Pathway

Self-declaration (no conformity assessment body)

Official Timeline
Min

0

Max

0

Unit

days

Verification Status

official-verified

Official Source
Title

Medizinprodukteverordnung (MepV, SR 812.213), Stand 1. Juli 2026 — Art. 23 conformity assessment by reference to EU MDR, Art. 25(4) equivalence of EU/EEA notified-body certificates

Url

https://www.fedlex.admin.ch/eli/cc/2020/552/de

Effective Date

2026-07-01

Accessed Date

2026-08-29

Realistic Timeline
Min

3

Max

5

Unit

months

With Prior Approval
Min

1

Max

2

Unit

months

Prior Approval Benefits
  • EU CE Mark (recognised unilaterally by Switzerland)
Notes
  • MepV Art. 23 applies the EU MDR conformity assessment procedures, so a Class I device that is not sterile, has no measuring function and is not a reusable surgical instrument is self-declared with no body involved and no review clock
  • Swissmedic runs no product approval at all: the only Swiss step is swissdamed registration, which carries a filing deadline rather than a review period
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

No

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

Regulatory reference

China NMPA approval: no recognition regime

reliance.china-nmpa-no-recognition-route
Pending verification
Statement

Switzerland recognises only the EU CE conformity route (unilateral recognition of CE-marked devices under the MedDO/IvDO); a China NMPA registration certificate gives no route to the Swiss market and is not assessed. A Chinese manufacturer needs CE conformity assessment plus a Swiss authorised representative and Swissmedic registration duties.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

switzerland

China Classes In Scope

All NMPA classes -> all Swiss classes

Edge Type

No Reliance Route

Market Access Strength

none

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Medical devices and IVDs subject to route-specific scope.
  • Swiss representative, registration, language and surveillance obligations apply.
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Source passage: Switzerland unilaterally recognises the CE label for medical devices (existing verified source section).

SwitzerlandEU conformity assessment

Prior approval / evidence record

EU CE Mark (MDD)

Authority

EU CE Mark (MDD)

Accepted

Yes

Benefit

Recognized until certificate expiry

Timeline Savings

Immediate

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

SwitzerlandEU conformity assessment

Prior approval / evidence record

EU CE Mark Recognition

Name

EU CE Mark Recognition

Description

Full recognition of EU CE marked devices until 2028

Timeline Reduction

Immediate market access

Eligibility
  • Valid EU MDR CE Mark
  • MDD CE Mark until expiry

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

UDI, listing & traceability

Which UDI, device listing, database, and traceability duties apply?

Compare countries

IVD name/model changes and catalogue REF are distinct

ivd.name.model.udi.and.ref.assessment
Fact-level verified

Under IvDO 16(4) and its incorporated IVDR clauses, a new UDI-DI is required for a change that could cause misidentification or traceability ambiguity, including a device name/trade-name or version/model change. An actual catalogue/REF change is not automatically the same thing: assess its effect on identity and traceability. A change to an existing UDI database element that does not require a new UDI-DI must be updated within 30 days. Cosmetic artwork alone is not listed as an automatic new-UDI trigger.

Conditions
  • Only as incorporated by the relevant current national law; custom-made, legacy and special-device identifier rules require their own scope assessment.
  • 30 days is an applicant data-update deadline, not an authority review period.

Technical field rules do not override UDI law

swissdamed.identity.fields.and.ref.mutability
Fact-level verified

The current swissdamed data dictionary distinguishes manufacturer SRN, Basic UDI-DI, UDI-DI, model/name, trade names and catalogue/REF. Its Updateable legend says blank means No: the REF field FLD-UDID-163 is not updateable in this version for either MD or IVD, while trade names have explicit edit/add/delete permission. The manufacturer SRN is expressly non-updateable in the business rules. Discarding a registered UDI is only for incorrect data that cannot be corrected by updating, not a general workaround for a real commercial or legal-manufacturer change.

Conditions
  • REF alone is not automatically a legal new-UDI trigger, but the actual traceability consequences and accepted operational route must be resolved.
  • Do not treat technically editable trade names as permission to retain an identifier when incorporated UDI law requires a new one.

Related structured datasets

Classification systems · 1 record
Open Switzerland IVD risk classification →
Id

switzerland:ivd

Jurisdiction Id

switzerland

Domains
  • ivd
Name

Switzerland IVD risk classification

Classes
  • Id

    a

    Label

    IVD Class A

    Normalized Risk Level

    low

  • Id

    b

    Label

    IVD Class B

    Normalized Risk Level

    medium

  • Id

    c

    Label

    IVD Class C

    Normalized Risk Level

    high

  • Id

    d

    Label

    IVD Class D

    Normalized Risk Level

    highest

Regulator

Swissmedic

Official Domains
  • swissmedic.ch
  • fedlex.admin.ch
Source Status

official-source-verified

Source Title

Ordinance on In Vitro Diagnostic Medical Devices (IvDO), Art. 14 (Klassifizierung)

Source Url

https://www.fedlex.admin.ch/filestore/fedlex.data.admin.ch/eli/cc/2022/291/20250101/20250101/de/pdf/a.pdf

Verified On

2026-08-25

Source Supports
  • ivd-class-axis
  • ivd-classification-rules
Notes
  • Art. 14 IvDO: IVDs are classified considering intended purpose and associated risks into classes A, B, C and D; Annex VIII EU-IVDR (incorporated by reference into IvDO) is decisive for classification.
Registration pathways, prior approvals and planning · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Switzerland registration →
Country

Switzerland

Jurisdiction Id

switzerland

Country Code

CH

Flag

🇨🇭

Regulator

Swiss Agency for Therapeutic Products

Regulator Abbrev

Swissmedic

Region

europe

Tier

2

Risk Classification

4-class (I, IIa, IIb, III) - EU MDR aligned

Classes
  • Id

    class-i

    Class Name

    Class I

    Risk Level

    low

    Pathway

    Self-declaration (no conformity assessment body)

    Official Timeline
    Min

    0

    Max

    0

    Unit

    days

    Verification Status

    official-verified

    Official Source
    Title

    Medizinprodukteverordnung (MepV, SR 812.213), Stand 1. Juli 2026 — Art. 23 conformity assessment by reference to EU MDR, Art. 25(4) equivalence of EU/EEA notified-body certificates

    Url

    https://www.fedlex.admin.ch/eli/cc/2020/552/de

    Effective Date

    2026-07-01

    Accessed Date

    2026-08-29

    Realistic Timeline
    Min

    3

    Max

    5

    Unit

    months

    With Prior Approval
    Min

    1

    Max

    2

    Unit

    months

    Prior Approval Benefits
    • EU CE Mark (recognised unilaterally by Switzerland)
    Notes
    • MepV Art. 23 applies the EU MDR conformity assessment procedures, so a Class I device that is not sterile, has no measuring function and is not a reusable surgical instrument is self-declared with no body involved and no review clock
    • Swissmedic runs no product approval at all: the only Swiss step is swissdamed registration, which carries a filing deadline rather than a review period
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    No

Expedited Pathways
  • Name

    EU CE Mark Recognition

    Description

    Full recognition of EU CE marked devices until 2028

    Timeline Reduction

    Immediate market access

    Eligibility
    • Valid EU MDR CE Mark
    • MDD CE Mark until expiry
  • Name

    Unilateral recognition of EU/EEA certificates

    Description

    The device chapter of the EU-Switzerland MRA has not applied since 26 May 2021 and has not been restored; Switzerland instead recognises EU/EEA notified-body certificates one-way under MepV Art. 25(4), which is why a CH-REP and Swiss labelling are still required

    Timeline Reduction

    No second conformity assessment

    Eligibility
    • Certificate from a body designated under EU law and established in an EU or EEA state
    • Procedures must meet Swiss requirements and the body must hold equivalent qualification
Prior Approvals
  • Authority

    EU CE Mark (MDR)

    Accepted

    Yes

    Benefit

    Full recognition - immediate Swiss market access

    Timeline Savings

    Immediate

  • Authority

    EU CE Mark (MDD)

    Accepted

    Yes

    Benefit

    Recognized until certificate expiry

    Timeline Savings

    Immediate

  • Authority

    FDA 510(k)

    Accepted

    No

    Benefit

    Documentation can support Swiss application

    Timeline Savings

    Indirect

Key Requirements
  • Swiss Authorized Representative (CH-REP) for foreign manufacturers
  • EU MDR technical documentation
  • Swiss specific labeling requirements
  • Swissmedic registration
  • Actor registration: manufacturers, CH-REPs and importers register in swissdamed within 3 months after first placing a product on the market (MedDO Art. 55; IvDO Art. 88)
  • Device registration: swissdamed UDI Devices module live since 18.08.2025; registration mandatory from 1 July 2026, with a transition period until 31 December 2026 for devices placed on the market before then
  • Clinical investigations (KlinV-Mep Art. 12): the ethics committee confirms receipt and flags formal deficiencies within 10 days, gives the sponsor 10 days to cure them, and decides within 40 days of confirming a formally correct dossier; the clock is suspended while additional information is outstanding
  • Multicentre investigations (KlinV-Mep Art. 13): the lead committee confirms receipt within 10 days (each deadline extendable by 5 days on request), the participating committees report their local assessment within 15 days, and the lead committee decides within 40 days
  • Swissmedic authorisation (KlinV-Mep Art. 19): receipt confirmed within 10 days, 10-day cure period extendable by 20 days, decision within 45 days of a formally correct dossier and only after the ethics committee has approved, extendable by up to 20 days for first-in-human use or a new manufacturing process; substantial amendments are decided within 38 days (+7)
Local Rep Required

Yes

Local Rep Name

Swiss Authorized Representative (CH-REP)

Mdsap Accepted

No

Mdsap Benefit

Not accepted but ISO 13485 required

Registration Validity

5 years (aligned with CE certificate)

Last Updated

2026-08-29

Verification Status

partially-verified

Official Sources
  • Title

    Medizinprodukteverordnung (MepV, SR 812.213), Stand 1. Juli 2026 — Art. 23 conformity assessment by reference to EU MDR, Art. 25(4) equivalence of EU/EEA notified-body certificates

    Url

    https://www.fedlex.admin.ch/eli/cc/2020/552/de

    Effective Date

    2026-07-01

    Accessed Date

    2026-08-29

  • Title

    Verordnung über klinische Versuche mit Medizinprodukten (KlinV-Mep, SR 810.306), Stand 20. Mai 2025 — Art. 12, 13 and 19 procedural deadlines

    Url

    https://www.fedlex.admin.ch/eli/cc/2020/553/de

    Effective Date

    2025-05-20

    Accessed Date

    2026-08-29

Reviewed On

2026-08-29

Process
Template Id

default

Status

planning-template

Usage Rule

Maintained planning steps and estimates, not verified regulator deadlines.

Steps
  • Step

    1

    Title

    Pre-Submission Preparation

    Description

    Gather technical documentation, quality management system (QMS) records, and clinical data required by Swissmedic. Appoint a Swiss Authorized Representative (CH-REP).

    Duration

    1-3 months

    Icon

    📋

  • Step

    2

    Title

    Application Submission

    Description

    Submit complete registration dossier to Swissmedic including device description, risk analysis, and testing reports.

    Duration

    1-2 weeks

    Icon

    📤

  • Step

    3

    Title

    Administrative Review

    Description

    Swissmedic reviews application for completeness. May request additional documents or clarifications.

    Duration

    2-4 weeks

    Icon

    🔍

  • Step

    4

    Title

    Technical Evaluation

    Description

    In-depth technical assessment of device safety and performance. Swissmedic may issue questions.

    Duration

    0 days

    Icon

    ⚙️

  • Step

    5

    Title

    Registration Approval

    Description

    Upon successful review, Swissmedic issues Switzerland market authorization certificate.

    Duration

    1-2 weeks

    Icon

    ✅

Government fee schedules · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Switzerland fees →
Country

Switzerland

Jurisdiction Id

switzerland

Country Code

CH

Authority

Swissmedic

Low Risk Local
Min

200

Max

200

Currency

CHF

Per Product

No

High Risk Local
Min

200

Max

10000

Currency

CHF

Per Product

No

Details
  • Id

    swissdamed-basic-fee-first-udi-di

    Item

    swissdamed Basic Fee - First UDI-DI

    Currency

    CHF

    Amount Local

    200

    Amount Local Formatted

    CHF 200

    Frequency

    First UDI-DI per manufacturer per calendar year

    Verification Status

    official-verified

    Source Url

    https://www.swissmedic.ch/swissmedic/en/home/medical-devices/medizinprodukte-datenbank/registrierungsgebuehr.html

    Source Title

    Swissmedic swissdamed device registration fee

    Source Effective Date

    2026-07-01

    Verified On

    2026-08-25

    Amount Status

    exact-current

  • Id

    swissdamed-unit-fee-additional-udi-di

    Item

    swissdamed Unit Fee - Additional UDI-DI

    Currency

    CHF

    Amount Local

    20

    Amount Local Formatted

    CHF 20

    Frequency

    Each additional UDI-DI in the same calendar year

    Notes

    Maximum CHF 10,000 per manufacturer per year

    Verification Status

    official-verified

    Source Url

    https://www.swissmedic.ch/swissmedic/en/home/medical-devices/medizinprodukte-datenbank/registrierungsgebuehr.html

    Source Title

    Swissmedic swissdamed device registration fee

    Source Effective Date

    2026-07-01

    Verified On

    2026-08-25

    Amount Status

    exact-current

  • Id

    swissdamed-registration-first-device

    Item

    swissdamed device registration - first device per year

    Currency

    CHF

    Amount Local

    200

    Amount Local Formatted

    CHF 200

    Frequency

    Annual, per manufacturer

    Notes

    Effective 1 July 2026; capped at CHF 10,000 per year per manufacturer or assembler of systems/procedure packs (no additional cost from 492 devices per year).

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.swissmedic.ch/swissmedic/en/home/medical-devices/medizinprodukte-datenbank/registrierungsgebuehr.html

    Source Title

    Swissmedic swissdamed device registration fee

    Source Effective Date

    2026-07-01

    Verified On

    2026-08-25

  • Id

    swissdamed-registration-additional-device

    Item

    swissdamed device registration - each additional device

    Currency

    CHF

    Amount Local

    20

    Amount Local Formatted

    CHF 20

    Frequency

    Annual, per additional device

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.swissmedic.ch/swissmedic/en/home/medical-devices/medizinprodukte-datenbank/registrierungsgebuehr.html

    Source Title

    Swissmedic swissdamed device registration fee

    Source Effective Date

    2026-07-01

    Verified On

    2026-08-25

Notes
  • Mandatory swissdamed device registration took effect July 1, 2026
  • CHF 200 for the first UDI-DI registered in the calendar year
  • CHF 20 for every additional UDI-DI
  • Maximum CHF 10,000 per manufacturer per calendar year
  • Invoicing rules: the first fee event is 31 December 2026 for all UDI-DIs with status "On the market"; devices marked "No longer placed on the market" as at that date incur no fee; from 1 January 2027 only new registrations are invoiced, each device once, in the year of its initial registration; data updates are free; re-registering a discarded device is fee-bearing.
  • Fees are charged per manufacturer (not per CH-REP); invoices for foreign manufacturers are issued to the CH-REP.
  • Reactivation of an inactivated swissdamed actor registration is possible after a Swissmedic review and is fee-bearing on a time-and-material basis (Art. 4(2) FeeO-Swissmedic, SR 812.214.5).
Last Updated

2026-08-25

Verification Status

official-verified

Official Sources
  • Title

    Swissmedic swissdamed device registration fee

    Url

    https://www.swissmedic.ch/swissmedic/en/home/medical-devices/medizinprodukte-datenbank/registrierungsgebuehr.html

    Effective Date

    2026-07-01

    Accessed Date

    2026-08-25

Reviewed On

2026-08-25

Amount Status

exact-current

Range Available

Yes

Additional certifications and licences · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Switzerland certifications →
Slug

switzerland

Market

Switzerland

Confidence

low

Schemes
  • Id

    swiss-medical-electrical-emc-evidence

    Name

    Swiss medical electrical/EMC evidence

    Category

    Medical Electrical Safety

    Status

    Included In Medical Registration

    Object

    Finished Device

    Note

    Designated-body certificates (per Annexes IX-XI EU-MDR) must be issued in one of the three official languages or English, contain at least the Annex XII information and are valid for a maximum of 5 years, extendable on application by up to 5 years per re-assessment and repeatedly (MedDO Art. 25-26; identical rule for IVD certificates IvDO Art. 22); ancillary federal law (Electricity Act, Metrology Act, Product Safety Act, Radiation Protection Act) is triggered alongside device certification per the MedDO preamble.

    Source Binding

    unresolved

  • Id

    bakom-ofcom-radio-market-conformity

    Name

    BAKOM/OFCOM radio market conformity

    Category

    Radio Spectrum

    Status

    Mandatory If Feature

    Object

    Radio Equipment

    Source Urls
    • https://www.bakom.admin.ch/en/market-access-of-radiocommunications-equipment
    Source Binding

    candidate

  • Id

    swiss-or-accepted-ce-conformity-marking

    Name

    Swiss or accepted CE conformity marking

    Category

    Radio Spectrum

    Status

    Evidence Only No Separate Certificate

    Object

    Radio Equipment

    Source Urls
    • https://www.bakom.admin.ch/en/equipment-marking
    Source Binding

    candidate

  • Id

    foph-radiation-authorization

    Name

    FOPH radiation authorization

    Category

    Ionizing Radiation

    Status

    Mandatory If Feature

    Object

    Product Or Use Site

    Source Binding

    unresolved

Official Sources
  • Authority

    BAKOM/OFCOM

    Topic

    market access for radiocommunications equipment

    Url

    https://www.bakom.admin.ch/en/market-access-of-radiocommunications-equipment

  • Authority

    BAKOM/OFCOM

    Topic

    equipment marking

    Url

    https://www.bakom.admin.ch/en/equipment-marking

  • Authority

    Fedlex

    Topic

    MedDO designated-body certificate rules and ancillary legal bases

    Url

    https://www.fedlex.admin.ch/filestore/fedlex.data.admin.ch/eli/cc/2020/552/20231101/20231101/de/pdf/a.pdf