Explore the IVD requirements currently available for India, organized by topic and linked to official sources.

17 facts9 dimensions9 official sourceslatest verification 2026-10-21
Explore questions and sources →

Common questions about changing a registered IVD in India. Each question separates published rules from details still awaiting verification.

Change control

Does the event require internal documentation, notice, prior approval, transfer, or a new application?

Compare countries

IVD constitution change is a fresh licence route

answers.constitution-import-ivd
Fact-level verified

For an import licence, a qualifying change in the constitution of the authorised agent or overseas manufacturer that executed the power of attorney requires written notice within 30 days and a fresh licence application within 180 days. Rule 39 preserves the existing licence pending issue or rejection of the fresh licence. Manufacture uses Rule27 with45-day notice and180-day application. This licence continuity does not itself authorise continued use of an inaccurate old label.

Conditions
  • Apply the legal definition of constitution; a same-entity rename alone is different.
  • Older MD FAQ Q109 prints45 days for the import event; the statutory import provision and later Q6 state30. Preserve that source conflict rather than rewriting the old FAQ.

IVD brand approval and actual-site rename

answers.ivd-brand-approval
Fact-level verified

For already approved Class C/D IVDs, CDSCO FAQ Addendum02 Q24 directs a separate online application for endorsement of a brand or additional brand and CLA approval. Q19 says a name-only change of the actual manufacturing site without a constitution change uses PAC approval rather than a fresh licence. Q24 does not establish the identical operational route for every Class A/B IVD or for a new legal manufacturer.

Conditions
  • FAQ is public-awareness guidance; statutory provisions and applicable clarifications control.

Brand content versus label artwork

answers.label-major-boundary
Fact-level verified

For licensed devices and IVDs, a substantive change to approved label information is a major change under Sixth Schedule A7. Changes only to font size, font type, colour or label design are excluded from that major-change entry. An exclusion from A7 is not, by itself, a positive notification exemption or a classification of every logo edit as a listed minor change.

Conditions
  • Class A non-sterile/non-measuring registration follows its separate rules.
  • Map a proposed brand/logo to the actual approved label and device identity.

Same entity manufacturer name or office address

answers.name-office-major
Fact-level verified

For a licensed product, Sixth Schedule A6 lists a change in the name or address of the overseas manufacturer or its manufacturing site, domestic manufacturer or its site, and Indian authorised agent as a major change. A same-entity office-address or company-name change is assessed through this change procedure; it is not automatically a factory relocation or a change in constitution.

Conditions
  • Keep the actual factory and legal person unchanged when using the name/address scenario.

Dossier requirements

Which evidence, local deltas, formats, and signatures are required?

Compare countries

Materials for an import label PAC

answers.pac-label-materials
Fact-level verified

The official Import License(Form MD-14) PAC label table marks the old/new label comparison with reasons, revised label and PAC application as mandatory. It marks an existing licence/retention with cover letter, foreign approval and packaging-validation evidence as No or conditional as printed. This is the listed import-PAC branch, not a dedicated subsequent-importer checklist.

Materials when the registered model number changes

answers.pac-model-materials
Fact-level verified

The import-PAC model-number table requires proposed details, old/new comparison with justification, existing licence/retention, predicate or similar-device reference, relevant DMF and Essential Principles changes, risk report, proposed label, applicable test reports, three-batch release/analysis evidence and notarised FSC/quality certificates. Use this table only if the changed order code is actually the registered model field.

Materials for a same-entity manufacturer name PAC

answers.pac-name-materials
Fact-level verified

The import PAC table for overseas manufacturer or manufacturing-site name without a constitution change requires the PAC application and reason, existing licence/retention documentation, authenticated power of attorney with agent undertaking, origin site/establishment registration and quality certificates showing the name change. Its FSC row is marked No; several label/PMF/undertaking rows are also marked No. Do not replace those flags with a universal all-certificates-first requirement.

Materials for a legal manufacturer office-address PAC

answers.pac-office-materials
Fact-level verified

For an overseas legal-manufacturer address change without constitution change, the import-PAC table requires the application/reason, authenticated power of attorney and agent undertaking, manufacturer no-constitution-change undertaking, notarised origin site/establishment registration, a notarised reference-country FSC and quality-certificate evidence. This office branch is distinct from physically relocating a manufacturing site.

Economic-operator roles

Which manufacturer, representative, holder, applicant, importer, and distributor roles apply?

Compare countries

IVD subsequent importer has a same-manufacturer condition

answers.subsequent-importer-ivd
Fact-level verified

IVD FAQ Addendum02 Q7 directs a new authorised agent to the Subsequent Importer portal application only for an already licensed IVD where the legal and actual manufacturing site remain the same. It does not grant a transfer to a different foreign manufacturer or establish the contents of the separately referenced final subsequent-importer checklist.

Conditions
  • Only a new Indian agent seeking this same-manufacturer route triggers the unresolved dedicated checklist; same-entity brand/name/office edits do not trigger it merely by being PACs.

Label, IFU & language

Which label and IFU elements, languages, symbols, and translation controls apply?

Compare countries

IVD label identity and importer details

answers.ivd-label-particulars
Fact-level verified

For an IVD under the Medical Devices Rules, the label must identify the device, manufacturer and actual manufacturing premises and carry the applicable particulars in Rule 44. For imported products, specified missing importer name/address, import-licence number, manufacturing-premises address and manufacturing date may be added by stickering under Rule 44(n). That limited provision is not a blanket permission to relabel old stock after any transfer.

Registration pathways

Which registration, notification, listing, reliance, or conformity route applies?

Compare countries

Class A nonsterile/nonmeasuring route boundary

answers.classa-route-boundary
Fact-level verified

Class A non-sterile and non-measuring devices have a separate online registration framework under Rules19G-19M and the stated Eighth Schedule exemptions. The licensed major/minor PAC and constitution-clock answers must therefore not be applied to that subset without checking the registration route and its actual recorded information.

Conditions
  • Risk class, sterile/measuring status and current registration/licence type must be supplied.
  • Later applicable2026 QMS changes must be read for this subset; this record is a scope boundary, not its complete post-registration-change workflow.

Reliance & recognition

Which foreign decisions or evidence can be reused, with what effect and conditions?

Compare countries

Existing records include acceptance, restrictions and evidence reuse. The applicable product scope and conditions are preserved below; a reference does not by itself confer market access.

India approvals and evidence in other markets (0 records)

No linked record is currently maintained. This does not establish that a pathway is unavailable.

Prior approvals and evidence for India (13 records)

IndiaJapan MHLW / PMDA

Prior approval / evidence record

Japan PMDA

Authority

Japan PMDA

Accepted

Yes

Benefit

Named in Rule 36(3) and in the Rule 63(1) proviso, so both the clinical investigation waiver and the no-predicate relief are available.

Timeline Savings

None published — the saving is the clinical investigation, not the clock

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndiaGHTF

Prior approval / evidence record

Class C

Id

class-c

Class Name

Class C

Risk Level

high

Pathway

Registration with clinical data

Official Timeline
Min

90

Max

90

Unit

days

Verification Status

official-verified

Official Source
Title

CDSCO Medical Devices Rules, 2017 (G.S.R. 78(E))

Url

https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

Accessed Date

2026-08-25

Realistic Timeline
Min

12

Max

18

Unit

months

With Prior Approval
Min

6

Max

9

Unit

months

Prior Approval Benefits
  • GHTF country approval with predicate
Notes
  • Rules 21(4)/25(1): CDSCO document scrutiny within 45 days (restartable on deficiency rectification), then licence decision within 45 days of the site-inspection report
  • Clinical data required
Clinical Trial Required

Yes

Local Testing Required

Yes

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndiaUS FDA · EU conformity assessment · Australia TGA · Japan MHLW / PMDA · GHTF

Prior approval / evidence record

Class A

Id

class-a

Class Name

Class A

Risk Level

low

Pathway

Registration

Official Timeline
Min

45

Max

45

Unit

days

Verification Status

official-verified

Official Source
Title

CDSCO Medical Devices Rules, 2017 (G.S.R. 78(E))

Url

https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

Accessed Date

2026-08-25

Realistic Timeline
Min

6

Max

9

Unit

months

With Prior Approval
Min

3

Max

5

Unit

months

Prior Approval Benefits
  • GHTF country approval (FDA, EU, Japan, Australia, UK)
Notes
  • Licence granted or rejected within 45 days of application (Rule 20(4)); Notified Body audit happens within 120 days after licence grant, not before (Rule 20(4) proviso (ii))
  • GHTF approval provides significant advantage
Clinical Trial Required

No

Local Testing Required

Yes

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndiaGHTF

Prior approval / evidence record

Class D

Id

class-d

Class Name

Class D

Risk Level

highest

Pathway

Registration with extensive clinical

Official Timeline
Min

90

Max

90

Unit

days

Verification Status

official-verified

Official Source
Title

CDSCO Medical Devices Rules, 2017 (G.S.R. 78(E))

Url

https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

Accessed Date

2026-08-25

Realistic Timeline
Min

15

Max

24

Unit

months

With Prior Approval
Min

8

Max

12

Unit

months

Prior Approval Benefits
  • GHTF country approval
Notes
  • Rules 21(4)/25(1): CDSCO document scrutiny within 45 days (restartable on deficiency rectification), then licence decision within 45 days of the site-inspection report
  • Indian clinical trials may be required
Clinical Trial Required

Yes

Local Testing Required

Yes

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndiaUK MHRA

Prior approval / evidence record

UK MHRA

Authority

UK MHRA

Accepted

Yes

Benefit

Named only in the Rule 63(1) proviso, which waives clinical investigation results for a no-predicate device marketed in the UK for at least two years. The UK is not in the Rule 36(3) list.

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndiaEU conformity assessment

Prior approval / evidence record

EU CE Mark

Authority

EU CE Mark

Accepted

Yes

Benefit

A free sale certificate from an EU country triggers the same Rule 36(3) waiver of clinical investigation. The EU is not named in the Rule 63(1) proviso, so a no-predicate device still needs the MD-27 route.

Timeline Savings

None published — the saving is the clinical investigation, not the clock

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndiaNMPA

Prior approval / evidence record

BRICS regulators cooperation MoU (NMPA party): no recognition effect

reliance.china-nmpa-bilateral-programme
Pending verification
Statement

CDSCO and China's NMPA are both parties to the BRICS regulators' Memorandum of Understanding (together with ANVISA, the Russian Ministry of Health, Ministry of Industry and Trade and Roszdravnadzor, and SAHPRA), listed by CDSCO as valid until 31 January 2029. The instrument is a cooperation MoU and creates no recognition of NMPA approvals in India.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

india

China Classes In Scope

Not class-specific; regulatory-cooperation instrument only

Edge Type

Cooperation Mou

Market Access Strength

none

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Multilateral (BRICS), not a bilateral India-China device MoU
  • Cooperation/information-exchange instrument; no legal effect on licensing
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Source passage: 04 BRAZIL, RUSSIA, INDIA, CHINA, SOUTH AFRICA (BRICS) Memorandum of Understanding (MoU) ... The Central Drugs Standard Control Organization of The Republic of INDIA (CDSCO) ... The National Medical Products Administration of The People's Republic of CHINA (NMPA) ... 31.01.2029

IndiaUS FDA

Prior approval / evidence record

FDA 510(k)/PMA

Authority

FDA 510(k)/PMA

Accepted

Yes

Benefit

A free sale certificate from the United States triggers Rule 36(3): the import licence is granted without clinical investigation. Rule 63(1) fourth proviso additionally waives clinical investigation results for a device without a predicate if it has been US-approved and marketed there for at least two years. Neither provision shortens the nine-month Rule 36(1) decision period.

Timeline Savings

None published — the saving is the clinical investigation, not the clock

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndiaAustralia TGA

Prior approval / evidence record

Australia TGA

Authority

Australia TGA

Accepted

Yes

Benefit

Named in Rule 36(3) and in the Rule 63(1) proviso, so both the clinical investigation waiver and the no-predicate relief are available.

Timeline Savings

None published — the saving is the clinical investigation, not the clock

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndiaNMPA

Prior approval / evidence record

China NMPA approval: origin evidence only

reliance.china-nmpa-origin-evidence
Pending verification
Statement

For devices imported from countries outside the rule 36(3) list, Class C and D licences may be granted only after safety and effectiveness are established through clinical investigation in India, and Class A and B licences after published safety and performance data or a clinical investigation in the country of origin plus a free sale certificate from the country of origin. A China NMPA registration certificate or the NMPA-based free sale certificate therefore serves only as the country-of-origin documentary evidence required with Form MD-14/MD-15; it gives no abridged review.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

india

China Classes In Scope

NMPA Class II/III certificate or Class I filing used as country-of-origin evidence; Indian Class A/B (rule 36(5)) and Class C/D (rule 36(4))

Edge Type

Origin Country Evidence

Market Access Strength

Evidence Only

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Class C/D: clinical investigation in India under Chapter VII (rule 36(4))
  • Class A/B: published safety/performance data or clinical investigation in the country of origin, and a free sale certificate from the country of origin (rule 36(5))
  • Import application must also include notarized copy of overseas manufacturing site registration in the country of origin and the Free Sale Certificate 'issued by the National Regulatory Authority or equivalent competent authority of the country concerned as referred under rule 36'
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Effective from 2018-01-01

Source passage: (4) Where a medical device is imported from countries other than those referred to in sub-rule (3), the licence in case of Class C and Class D medical devices may be granted after its safety and effectiveness has been established through clinical investigation in India as specified under provisions of Chapter VII of these rules. (5) Where a medical device, is imported from countries other than those referred to in sub-rule (3), the licence in case of Class A or Class B medical devices may be granted after its safety and performance has been established through published safety and performance data or through clinical investigation in the country of origin and a free sale certificate from the country of origin is furnished.

IndiaGHTF

Prior approval / evidence record

Class B

Id

class-b

Class Name

Class B

Risk Level

medium

Pathway

Registration

Official Timeline
Min

45

Max

140

Unit

days

Verification Status

official-verified

Official Source
Title

CDSCO Medical Devices Rules, 2017 (G.S.R. 78(E))

Url

https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

Accessed Date

2026-08-25

Realistic Timeline
Min

8

Max

12

Unit

months

With Prior Approval
Min

4

Max

7

Unit

months

Prior Approval Benefits
  • GHTF country approval
Notes
  • Rule 20(6) chain: Notified Body audit within 90 days of application, NB report to the licensing authority within 30 days of the audit, licence decision within 20 days of the audit report (140 days total)
  • Licence grant-or-reject clock is 45 days where it applies (Rule 20(4))
  • NABL testing may be required
Clinical Trial Required

No

Local Testing Required

Yes

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndiaHealth Canada

Prior approval / evidence record

Health Canada

Authority

Health Canada

Accepted

Yes

Benefit

Named in Rule 36(3) and in the Rule 63(1) proviso, so both the clinical investigation waiver and the no-predicate relief are available.

Timeline Savings

None published — the saving is the clinical investigation, not the clock

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndiaNMPA

Regulatory reference

China NMPA approval: not reference

reliance.china-nmpa-not-reference
Pending verification
Statement

Rule 36(3) of the Medical Devices Rules, 2017 grants the import licence without an Indian clinical investigation only where a free sale certificate has been issued by the national regulatory authority of Australia, Canada, Japan, an EU country or the United States (the United Kingdom was added by amendment in 2022). China/NMPA is not on the list, so an NMPA registration certificate does not trigger the clinical-investigation waiver.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

india

China Classes In Scope

All NMPA classes; concerns Indian import licence (Form MD-15) for Class A-D devices

Edge Type

No Reliance Route

Market Access Strength

none

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • List is exhaustive: Australia, Canada, Japan, European Union countries, United States of America (plus United Kingdom by the 2022 amendment, not reflected in the consolidated CDSCO text quoted)
  • Devices from other countries fall under rule 36(4) (Class C/D) or 36(5) (Class A/B)
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Effective from 2018-01-01

Source passage: (3) Where, a free sale certificate has already been issued in respect of any medical device by the national regulatory authority or other competent authority of any of the countries namely, Australia, Canada, Japan, European Union Countries, or the United States of America, a licence shall be granted under sub-rule (1) to the applicant without carrying out clinical investigation.

Review clocks

What review, validation, and applicant stop-clock rules apply?

Compare countries

IVD major-change and notification deadlines

answers.change-clocks-ivd
Fact-level verified

For an IVD under the licensed regime, prior approval is required for a listed major change: the manufacture rule specifies45 days and the import rule60 days for approval/rejection, each with the printed deemed-approval-on-silence provision. Listed minor changes are notified within 30 days after the change. The rules do not convert dossier preparation or applicant response time into part of that authority clock.

Submission communications & appeal

How are deficiencies answered, clocks handled, and decisions escalated or appealed?

Compare countries

IVD PAC reminders and product-count advice

answers.ivd-query-reminders
Fact-level verified

The 13 March2026 IVD FAQ gives a30-day first-reminder interval for PAC responses and further reminders at30-day intervals; nonresponse after the fourth reminder leads to deemed withdrawal. This is an applicant response policy, not a30-day PAC approval promise. The same FAQ advises no more than15 products in one import/manufacturing application to avoid delays; it does not make15 a statutory cap or permit merging arbitrary PAC changes.

UDI, listing & traceability

Which UDI, device listing, database, and traceability duties apply?

Compare countries

IVD UDI commencement is order dependent

answers.ivd-udi-order
Fact-level verified

The substituted Rule 46 requires the unique device identification from the date and in the manner specified by a Central Government order. The substituted provision itself does not establish a calendar commencement date. A commercial REF or catalogue renumbering must first be mapped to the actual recorded product/label field; it is not automatically a UDI change.

Related structured datasets

Classification systems · 1 record
Open Risk classes A through D. →
Id

india:medical-device

Jurisdiction Id

india

Domains
  • medical-device
  • ivd
Name

Risk classes A through D.

Classes
  • Id

    a

    Label

    Class A

    Normalized Risk Level

    low

  • Id

    b

    Label

    Class B

    Normalized Risk Level

    medium

  • Id

    c

    Label

    Class C

    Normalized Risk Level

    high

  • Id

    d

    Label

    Class D

    Normalized Risk Level

    highest

Regulator

CDSCO

Official Domains
  • cdsco.gov.in
Source Status

official-source-verified

Source Title

Medical Devices Rules 2017, Rule 4 and First Schedule (Parts I-II)

Source Url

https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

Verified On

2026-08-25

Source Supports
  • class-axis
  • classification-rules
  • ivd-class-axis
  • ivd-classification-rules
Notes
  • Rule 4(1): Class A low risk, B low moderate, C moderate high, D high risk per Part I of the First Schedule; Rule 4(2): IVDs classified per Part II.
  • Rule 4(3): the Central Licensing Authority classifies devices based on intended use and First Schedule parameters (binding classification).
  • Only Class A devices that are non-sterile and non-measuring take the registration-only route (Sixth Amendment Rules 2022, Chapter IIIB, rule 19G(1)).
Registration pathways, prior approvals and planning · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open India registration →
Country

India

Jurisdiction Id

india

Country Code

IN

Flag

🇮🇳

Regulator

Central Drugs Standard Control Organisation

Regulator Abbrev

CDSCO

Region

asia-pacific

Tier

2

Risk Classification

4-class (A, B, C, D)

Classes
  • Id

    class-a

    Class Name

    Class A

    Risk Level

    low

    Pathway

    Registration

    Official Timeline
    Min

    45

    Max

    45

    Unit

    days

    Verification Status

    official-verified

    Official Source
    Title

    CDSCO Medical Devices Rules, 2017 (G.S.R. 78(E))

    Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    6

    Max

    9

    Unit

    months

    With Prior Approval
    Min

    3

    Max

    5

    Unit

    months

    Prior Approval Benefits
    • GHTF country approval (FDA, EU, Japan, Australia, UK)
    Notes
    • Licence granted or rejected within 45 days of application (Rule 20(4)); Notified Body audit happens within 120 days after licence grant, not before (Rule 20(4) proviso (ii))
    • GHTF approval provides significant advantage
    Clinical Trial Required

    No

    Local Testing Required

    Yes

    Gmp Required

    Yes

  • Id

    class-b

    Class Name

    Class B

    Risk Level

    medium

    Pathway

    Registration

    Official Timeline
    Min

    45

    Max

    140

    Unit

    days

    Verification Status

    official-verified

    Official Source
    Title

    CDSCO Medical Devices Rules, 2017 (G.S.R. 78(E))

    Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    8

    Max

    12

    Unit

    months

    With Prior Approval
    Min

    4

    Max

    7

    Unit

    months

    Prior Approval Benefits
    • GHTF country approval
    Notes
    • Rule 20(6) chain: Notified Body audit within 90 days of application, NB report to the licensing authority within 30 days of the audit, licence decision within 20 days of the audit report (140 days total)
    • Licence grant-or-reject clock is 45 days where it applies (Rule 20(4))
    • NABL testing may be required
    Clinical Trial Required

    No

    Local Testing Required

    Yes

    Gmp Required

    Yes

  • Id

    class-c

    Class Name

    Class C

    Risk Level

    high

    Pathway

    Registration with clinical data

    Official Timeline
    Min

    90

    Max

    90

    Unit

    days

    Verification Status

    official-verified

    Official Source
    Title

    CDSCO Medical Devices Rules, 2017 (G.S.R. 78(E))

    Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    12

    Max

    18

    Unit

    months

    With Prior Approval
    Min

    6

    Max

    9

    Unit

    months

    Prior Approval Benefits
    • GHTF country approval with predicate
    Notes
    • Rules 21(4)/25(1): CDSCO document scrutiny within 45 days (restartable on deficiency rectification), then licence decision within 45 days of the site-inspection report
    • Clinical data required
    Clinical Trial Required

    Yes

    Local Testing Required

    Yes

    Gmp Required

    Yes

  • Id

    class-d

    Class Name

    Class D

    Risk Level

    highest

    Pathway

    Registration with extensive clinical

    Official Timeline
    Min

    90

    Max

    90

    Unit

    days

    Verification Status

    official-verified

    Official Source
    Title

    CDSCO Medical Devices Rules, 2017 (G.S.R. 78(E))

    Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Accessed Date

    2026-08-25

    Realistic Timeline
    Min

    15

    Max

    24

    Unit

    months

    With Prior Approval
    Min

    8

    Max

    12

    Unit

    months

    Prior Approval Benefits
    • GHTF country approval
    Notes
    • Rules 21(4)/25(1): CDSCO document scrutiny within 45 days (restartable on deficiency rectification), then licence decision within 45 days of the site-inspection report
    • Indian clinical trials may be required
    Clinical Trial Required

    Yes

    Local Testing Required

    Yes

    Gmp Required

    Yes

Prior Approvals
  • Authority

    FDA 510(k)/PMA

    Accepted

    Yes

    Benefit

    A free sale certificate from the United States triggers Rule 36(3): the import licence is granted without clinical investigation. Rule 63(1) fourth proviso additionally waives clinical investigation results for a device without a predicate if it has been US-approved and marketed there for at least two years. Neither provision shortens the nine-month Rule 36(1) decision period.

    Timeline Savings

    None published — the saving is the clinical investigation, not the clock

  • Authority

    EU CE Mark

    Accepted

    Yes

    Benefit

    A free sale certificate from an EU country triggers the same Rule 36(3) waiver of clinical investigation. The EU is not named in the Rule 63(1) proviso, so a no-predicate device still needs the MD-27 route.

    Timeline Savings

    None published — the saving is the clinical investigation, not the clock

  • Authority

    Japan PMDA

    Accepted

    Yes

    Benefit

    Named in Rule 36(3) and in the Rule 63(1) proviso, so both the clinical investigation waiver and the no-predicate relief are available.

    Timeline Savings

    None published — the saving is the clinical investigation, not the clock

  • Authority

    Australia TGA

    Accepted

    Yes

    Benefit

    Named in Rule 36(3) and in the Rule 63(1) proviso, so both the clinical investigation waiver and the no-predicate relief are available.

    Timeline Savings

    None published — the saving is the clinical investigation, not the clock

  • Authority

    Health Canada

    Accepted

    Yes

    Benefit

    Named in Rule 36(3) and in the Rule 63(1) proviso, so both the clinical investigation waiver and the no-predicate relief are available.

    Timeline Savings

    None published — the saving is the clinical investigation, not the clock

  • Authority

    UK MHRA

    Accepted

    Yes

    Benefit

    Named only in the Rule 63(1) proviso, which waives clinical investigation results for a no-predicate device marketed in the UK for at least two years. The UK is not in the Rule 36(3) list.

    Timeline Savings

    None published

Key Requirements
  • Indian Authorized Agent (for foreign manufacturers)
  • Rule 7: the device must meet the applicable Bureau of Indian Standards standard, or an ISO, IEC or pharmacopoeial standard where BIS has not laid one down. This is a conformity obligation in the dossier, not a separate BIS licence, which is only compulsory for the narrow set of products covered by a Quality Control Order.
  • NABL testing from accredited labs
  • Import license
  • Free Sale Certificate from home country
  • Import licence (Rule 36(1)): CLA grants or rejects within nine months of application; clinical investigation permission within 90 days of the MD-22 application (Rule 52(ii))
  • No-predicate permission (MD-26) 120 days extendable by 30; new IVD permission (MD-28) 90 days extendable by 30; test licence (MD-16) 30 days; personal-use import (MD-20) 7 days; sale registration (MD-41) 10 days
Local Rep Required

Yes

Local Rep Name

Indian Authorized Agent

Mdsap Accepted

No

Mdsap Benefit

Not accepted. The relief for foreign approvals runs through the Rule 36(3) free sale certificate list, not through MDSAP.

Registration Validity

Perpetual, not a fixed term. A manufacturing licence or loan licence in Form MD-5, MD-6, MD-9 or MD-10 stays valid indefinitely under Rule 29(1) and an import licence in Form MD-15 under Rule 37, in each case provided the retention fee is paid before every fifth anniversary. Miss it and the licence is deemed cancelled, after a 180-day late window with a 2% monthly surcharge for manufacturing licences and a 90-day window for import licences.

Verification Status

official-verified

Official Sources
  • Title

    CDSCO Medical Devices Rules, 2017 (G.S.R. 78(E))

    Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Accessed Date

    2026-08-25

Reviewed On

2026-08-30

Last Updated

2026-08-30

Process
Template Id

default

Status

planning-template

Usage Rule

Maintained planning steps and estimates, not verified regulator deadlines.

Steps
  • Step

    1

    Title

    Pre-Submission Preparation

    Description

    Gather technical documentation, quality management system (QMS) records, and clinical data required by CDSCO. Appoint a Indian Authorized Agent.

    Duration

    1-3 months

    Icon

    📋

  • Step

    2

    Title

    Application Submission

    Description

    Submit complete registration dossier to CDSCO including device description, risk analysis, and testing reports.

    Duration

    1-2 weeks

    Icon

    📤

  • Step

    3

    Title

    Administrative Review

    Description

    CDSCO reviews application for completeness. May request additional documents or clarifications.

    Duration

    2-4 weeks

    Icon

    🔍

  • Step

    4

    Title

    Technical Evaluation

    Description

    In-depth technical assessment of device safety and performance. CDSCO may issue questions.

    Duration

    45 days

    Icon

    ⚙️

  • Step

    5

    Title

    Registration Approval

    Description

    Upon successful review, CDSCO issues India market authorization certificate.

    Duration

    1-2 weeks

    Icon

    ✅

Government fee schedules · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open India fees →
Country

India

Jurisdiction Id

india

Country Code

IN

Authority

CDSCO

Low Risk Local
Min

50

Max

1000

Currency

USD

Per Product

Yes

High Risk Local
Min

1500

Max

3000

Currency

USD

Per Product

Yes

Details
  • Id

    import-licence-class-a-site

    Item

    Import licence - Class A manufacturing site

    Currency

    USD

    Amount Local

    1000

    Amount Local Formatted

    USD 1,000

    Frequency

    Per site

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    import-licence-class-a-device

    Item

    Import licence - Class A device

    Currency

    USD

    Amount Local

    50

    Amount Local Formatted

    USD 50

    Frequency

    Per distinct device

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    import-licence-class-b-site

    Item

    Import licence - Class B manufacturing site

    Currency

    USD

    Amount Local

    2000

    Amount Local Formatted

    USD 2,000

    Frequency

    Per site

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    import-licence-class-b-device

    Item

    Import licence - Class B device

    Currency

    USD

    Amount Local

    1000

    Amount Local Formatted

    USD 1,000

    Frequency

    Per distinct device

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    import-licence-class-c-d-site

    Item

    Import licence - Class C/D manufacturing site

    Currency

    USD

    Amount Local

    3000

    Amount Local Formatted

    USD 3,000

    Frequency

    Per site

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    import-licence-class-c-d-device

    Item

    Import licence - Class C/D device

    Currency

    USD

    Amount Local

    1500

    Amount Local Formatted

    USD 1,500

    Frequency

    Per distinct device

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    import-licence-ivd-class-a-b-site

    Item

    Import licence - IVD Class A/B manufacturing site

    Currency

    USD

    Amount Local

    1000

    Amount Local Formatted

    USD 1,000

    Frequency

    Per site

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    import-licence-ivd-class-a-b-device

    Item

    Import licence - IVD Class A/B device

    Currency

    USD

    Amount Local

    10

    Amount Local Formatted

    USD 10

    Frequency

    Per distinct device

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    import-licence-ivd-class-c-d-site

    Item

    Import licence - IVD Class C/D manufacturing site

    Currency

    USD

    Amount Local

    3000

    Amount Local Formatted

    USD 3,000

    Frequency

    Per site

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    import-licence-ivd-class-c-d-device

    Item

    Import licence - IVD Class C/D device

    Currency

    USD

    Amount Local

    500

    Amount Local Formatted

    USD 500

    Frequency

    Per distinct device

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    manufacture-licence-class-a-b-site

    Item

    Manufacture licence - Class A/B site

    Currency

    INR

    Amount Local

    5000

    Amount Local Formatted

    INR 5,000

    Frequency

    Per site; retention fee is the same

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    manufacture-licence-class-a-b-device

    Item

    Manufacture licence - Class A/B device

    Currency

    INR

    Amount Local

    500

    Amount Local Formatted

    INR 500

    Frequency

    Per distinct device

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    manufacture-licence-class-c-d-site

    Item

    Manufacture licence - Class C/D site

    Currency

    INR

    Amount Local

    50000

    Amount Local Formatted

    INR 50,000

    Frequency

    Per site; retention fee is the same

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    manufacture-licence-class-c-d-device

    Item

    Manufacture licence - Class C/D device

    Currency

    INR

    Amount Local

    1000

    Amount Local Formatted

    INR 1,000

    Frequency

    Per distinct device

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    overseas-site-inspection

    Item

    Inspection of an overseas manufacturing site

    Currency

    USD

    Amount Local

    6000

    Amount Local Formatted

    USD 6,000

    Frequency

    Per inspection

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    pilot-clinical-investigation-permission

    Item

    Permission to conduct a pilot clinical investigation

    Currency

    INR

    Amount Local

    100000

    Amount Local Formatted

    INR 100,000

    Frequency

    Per permission

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    pivotal-clinical-investigation-permission

    Item

    Permission to conduct a pivotal clinical investigation

    Currency

    INR

    Amount Local

    100000

    Amount Local Formatted

    INR 100,000

    Frequency

    Per permission

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    clinical-performance-evaluation-permission

    Item

    Permission for clinical performance evaluation

    Currency

    INR

    Amount Local

    25000

    Amount Local Formatted

    INR 25,000

    Frequency

    Per permission

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    no-predicate-device-permission

    Item

    Permission to investigate a new device with no predicate

    Currency

    INR

    Amount Local

    50000

    Amount Local Formatted

    INR 50,000

    Frequency

    Per permission

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    new-ivd-permission

    Item

    Permission for a new in-vitro diagnostic device

    Currency

    INR

    Amount Local

    25000

    Amount Local Formatted

    INR 25,000

    Frequency

    Per permission

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

  • Id

    export-certificate

    Item

    Export certificate for a medical device

    Currency

    INR

    Amount Local

    1000

    Amount Local Formatted

    INR 1,000

    Frequency

    Per certificate

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Source Title

    India Medical Devices Rules 2017, fee schedule

    Verified On

    2026-08-25

Notes
  • Import-licence fees are denominated in USD under the Medical Devices Rules; domestic-manufacture, investigation, and certificate fees are in INR.
  • Site and per-device fees are cumulative; IVD import fees are lower than the non-IVD schedule (e.g. USD 10 per Class A/B IVD device).
  • Test/evaluation/demonstration imports cost USD 100 per device; government-hospital patient-treatment imports cost INR 500 per device.
Last Updated

2026-08-25

Verification Status

official-verified

Official Sources
  • Title

    India Medical Devices Rules 2017, fee schedule

    Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

    Accessed Date

    2026-08-25

Reviewed On

2026-08-25

Amount Status

exact-current

Range Available

Yes

Additional certifications and licences · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open India certifications →
Slug

india

Market

India

Confidence

high

Schemes
  • Id

    cdsco-electrical-emc-evidence

    Name

    CDSCO electrical/EMC evidence

    Category

    Medical Electrical Safety

    Status

    Included In Medical Registration

    Object

    Finished Device

    Note

    Standards hierarchy (Rule 7, Medical Devices Rules 2017): BIS or MoHFW-notified standards, else ISO/IEC or pharmacopoeial standards, else the validated manufacturer's own standards. Chapter IIIA registration (G.S.R. 102(E) rule 19B(2)(iii)) requires an ISO 13485 certificate accredited by NABCB or the International Accreditation Forum; IVD licence dossiers need a performance evaluation report from a central device testing laboratory or a rule 83(3)-registered laboratory; device testing laboratories register with the CLA (rule 81).

    Source Urls
    • https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/mdr%2C%202017%20%281%29.pdf
    Source Binding

    reviewed

  • Id

    wpc-eta

    Name

    WPC ETA

    Category

    Radio Spectrum

    Status

    Mandatory If Feature

    Object

    Wireless Equipment In Licence Exempt Band

    Source Urls
    • https://www.deveservices.dot.gov.in/equipment-type-approval-eta
    Source Binding

    candidate

  • Id

    tec-mtcte

    Name

    TEC MTCTE

    Category

    Telecom Network Attachment

    Status

    Mandatory If Feature

    Object

    Notified Telecom Equipment

    Source Urls
    • https://www.deveservices.dot.gov.in/equipment-type-approval-eta
    Source Binding

    candidate

  • Id

    bis-compulsory-scheme

    Name

    BIS compulsory scheme

    Category

    General Electrical Components

    Status

    Mandatory If Feature

    Object

    Notified Product Or Component

    Source Binding

    unresolved

  • Id

    aerb-x-ray-equipment-type-approval-and-supplier-import-permissions

    Name

    AERB X-ray equipment Type Approval and supplier/import permissions

    Category

    Ionizing Radiation

    Status

    Mandatory If Feature

    Object

    Product And Supplier

    Source Urls
    • https://www.aerb.gov.in/english/regulatory-facilities/radiation-facilities/application-in-medicine/diagnostic-radiology
    • https://aerb.gov.in/index.php/regulatory-requirements-and-guidelines-for-suppliers-of-x-ray-equipment
    Source Binding

    candidate

  • Id

    aerb-facility-operation-registration-licence

    Name

    AERB facility operation registration/licence

    Category

    Ionizing Radiation

    Status

    Mandatory For Use Site

    Object

    facility

    Source Urls
    • https://aerb.gov.in/index.php/regulatory-requirements-and-guidelines-for-suppliers-of-x-ray-equipment
    Source Binding

    candidate

  • Id

    legal-metrology-model-approval-registration

    Name

    Legal Metrology model approval/registration

    Category

    Legal Metrology

    Status

    Mandatory If Feature

    Object

    Regulated Measuring Instrument Or Packaged Product

    Source Urls
    • https://lm.doca.gov.in/index.aspx
    Source Binding

    candidate

Official Sources
  • Authority

    Department of Telecommunications

    Topic

    WPC equipment type approval

    Url

    https://www.deveservices.dot.gov.in/equipment-type-approval-eta

  • Authority

    AERB

    Topic

    diagnostic radiology type approval and licensing

    Url

    https://www.aerb.gov.in/english/regulatory-facilities/radiation-facilities/application-in-medicine/diagnostic-radiology

  • Authority

    AERB

    Topic

    supplier requirements for X-ray equipment

    Url

    https://aerb.gov.in/index.php/regulatory-requirements-and-guidelines-for-suppliers-of-x-ray-equipment

  • Authority

    Department of Consumer Affairs

    Topic

    legal metrology

    Url

    https://lm.doca.gov.in/index.aspx

  • Authority

    CDSCO

    Topic

    Medical Devices Rules 2017 standards hierarchy and testing laboratories

    Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/mdr%2C%202017%20%281%29.pdf

  • Authority

    CDSCO

    Topic

    G.S.R. 102(E) registration NABCB accredited ISO 13485

    Url

    https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/MDR_G.S.R.%20102%28E%29%20dt_11.02.2020_Registration%20of%20certain%20medical%20devices.pdf

Keep researching this market
View the India resource hub →
Compare regulatory data across all countries →
DJ Fang
Need to apply India's requirements to a product?
Confirm the current CDSCO pathway, evidence, and local-role details for the actual device before filing.