United Kingdom medical device regulations

Official sources linked

MHRA rules and guidance for medical device market access, registration, and lifecycle compliance. Each code opens a metadata page with the official source and readable HTML when available.

17 documents6 categories100% linked to official sourcesauthority MHRA11 readable HTML
CodeRegulationCategoryEffectiveProvenance
SI 2026/197
The Medical Devices (Fees Amendment) Regulations 2026
Official SI 2026/197, made 25 February 2026 and commencing 1 April 2026. Full eight-page raw original retained. Public projection preserves unchanged physical pages 1–6 and 8; physical page 7 is independently confirmed blank. Original images govern, including the ISBN/barcode imprint missing from native search text. GB and Northern Ireland amendment branches remain distinct.
RegistrationApril 1, 2026Official
MHRA Medical Device CFS Guidance
Certificate of Free Sale for exporting medical devices
Complete original MHRA guidance for medical-device Certificates of Free Sale captured on 20 October 2026, including applicant and device eligibility, market scope, electronic certificates, fees, ordering and qualified processing times. Official page first published 7 November 2016 and publicly updated 1 April 2026. Source-only capture; currentness has not been independently established.
Registration—Official
MHRA CFS Reference Guide April 2026 v1
Certificates of Free Sale Reference Guide — April 2026 v1
Complete official 34-physical-page CFS reference guide, printed version April 2026 v1, captured on 20 October 2026. Retains ordering, product and manufacturer updates, certificate and schedule examples, payment warnings and Annex I workflow. Original page images are authoritative; native text is a qualified incomplete search layer. Illustrative screenshots do not establish current fees. Source-only capture; currentness has not been independently established.
Registration—Official
MHRA Device Registration Fee Guidance, April 2026
Guidance: Device registration fee — April 2026
Complete official 24-page fee guidance, published April 2026. All original page images and native search text retained, including the literal GMDN copyright/database and licence notices on pages 16–17. Government registration-fee procedures are preserved; this does not grant standalone GMDN-data reuse rights.
Registration—Official
MHRA Account Management Reference Guide, April 2026 v1
DORS Account Management Reference Guide — April 2026
Complete official 47-page April 2026 account-management guide, preserving organisation identity, addresses, designation, suspension and closure workflows. Original page images govern; native text is an incomplete search layer and can omit screenshot text. Account identity changes, physical manufacturing-site addresses and underlying conformity responsibilities remain distinct.
Registration—Official
MHRA Device Registration Reference Guide, April 2026 v1
DORS Device Registration Reference Guide — selected procedures, April 2026
Selected unchanged government procedural pages from the official 102-page April 2026 guide: physical pages 1, 2, 51, 68, 69, 73–76, 85, 88, 93 and 102. Includes update limitations, product name/model/REF fields, separate-application rules, history, representation and importer screens. Full raw original is archived and official full-PDF link retained. Other pages, including screenshots with third-party GMDN definitions, are outside this public projection; this is not the complete guide text.
Registration—Official
MHRA Registration Attributes April 2026
Manufacturer and Device and Product and Importer Attributes — April 2026
Complete literal-cell projection of the official MHRA registration-attributes workbook captured on 20 October 2026, with April 2026 identified by its official attachment filename. Preserves all four worksheets, row and cell coordinates, field definitions, validation text, editability and fee-update columns. The original XLSX remains unchanged; no exact publication day or independent currentness finding is inferred.
Registration—Official
MHRA Registration Guidance
Register medical devices to place on the market
Complete original MHRA registration guidance body captured on 20 October 2026, including Great Britain and Northern Ireland requirements, DORS, registration changes, annual fees and reference-guide notices. Official page first published 31 December 2020 and publicly updated 16 October 2026; later CMS maintenance metadata is distinct. Source-only capture; currentness has not been independently established.
Registration—Official
MHRA UK Medical Devices Regulatory Guidance
Regulating medical devices in the UK
Complete original MHRA regulatory-framework guidance body captured on 20 October 2026, preserving separate Great Britain, Northern Ireland and EU requirements, conformity routes, economic operators, labelling and surveillance. Official page first published 31 December 2020 and publicly updated 20 February 2026. Source-only capture; currentness has not been independently established.
Registration—Official
Exports carry source + retrieval date

Changes after registration in United Kingdom

Source-linked rules for common changes to a registered IVD. Read the scope and conditions with each rule; unverified details are listed separately.

Does changing a device brand name or logo require an update?

A change to the registered brand name and a change to logo artwork can follow different rules. The legal manufacturer, device design and intended purpose remain the same.

  • Filing route and prior approval

Choose the GB or Northern Ireland registration route

Fact-level verified

All devices, including IVDs, must be registered with MHRA before placement on the Great Britain market, and the registered information must be accurate and current. Since 28 May 2026 ordinary MD and IVD devices for Northern Ireland use EUDAMED instead of MHRA registration; qualifying custom-made medical devices under MDR retain the separate MHRA branch. A manufacturer outside the UK appoints one UK Responsible Person for GB. A manufacturer outside the EU or Northern Ireland, including one in GB, needs an EU- or Northern Ireland-based authorised representative for NI.

Conditions
  • Separate GB and NI portfolios, and ordinary commercial devices from custom-made or performance-study devices. Registration is not MHRA product approval.
  • Filing route and prior approval

Distinguish registered brand/model from catalogue REF

Fact-level verified

The current DORS attributes distinguish brand/trade name, model/version, catalogue/REF and UDI. Brand name and model/version are not editable product fields; catalogue/REF is editable. In the MD sheet REF is mandatory; in the IVD sheet it is optional. Populated UDI identity fields have stricter mutability. The device guide requires removing and re-adding a product for a disabled product field, or removing and re-registering the device and its underlying products for a disabled device field. The public registration guidance expressly lists trade/brand-name and non-editable characteristic changes as re-registration triggers.

Conditions
  • Identify the actual product and device field; do not equate a commercial REF with a model, GMDN or UDI-DI.
  • The current statutory fee guidance governs whether re-registration creates a chargeable new GMDN category.
  • Filing route and prior approval

IVD notified-body change gates

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, where Annex IX conformity assessment applies, notify the body that approved the QMS of planned substantial QMS or covered device-range changes; it assesses the proposal and any necessary audits. Under Annex IX 4.11, a change to an approved device that could affect safety, performance or prescribed conditions of use needs the issuing body’s approval. The body decides whether a new conformity assessment or a certificate supplement is required. An identity-data edit does not settle that decision.

Conditions
  • Applies only where the relevant notified/designated-body certificate and change trigger exist. A cosmetic logo, same-person office change or REF-only edit is not automatically a substantial change.
  • No universal authority or notified-body review duration or fee is established by these clauses.
  • Labels and identifiers

English language requirement - IVDs

Fact-level verified

For in vitro diagnostic medical devices, the information to be provided under Section 8 of Part B of Annex I (labelling and instructions for use) must be in English.

Conditions
  • Great Britain E+W+S expression of reg 35(2); do not substitute for the NI EU-law branch.

Schedule 2 A para 25(f) also requires the UK responsible person's name and address on the label/IFU for IVDs of manufacturers without a UK registered place of business.

  • Supporting documents

Evidence for the actual DORS change route

Fact-level verified

For a manufacturer or device registration/re-registration, provide the legal-entity name and address as shown on the labelling, applicable legislation and class, GMDN, brand, model/version, REF and UDI information where applicable, and the relevant conformity certificate or declaration. A represented organisation also needs the applicable designation evidence. For an ordinary same-person organisation edit use the change-specific designation-letter rules; this is not a universal new technical dossier for every REF correction.

Conditions
  • Certificate/declaration evidence is required when the actual registration/re-registration or conformity-document route is triggered.
  • MHRA may request further technical evidence; a data update is not conformity approval.
  • Change-specific government fees

Free DORS updates and chargeable re-registration branches

Fact-level verified

From 1 April 2026 no separate statutory fee is charged for organisation name/address updates or new designation-letter uploads. Adding/removing products under registered devices, updating editable fields and obsolete GMDN data, and adding/deactivating importers are also uncharged as specified. Re-registration following a changed characteristic is charged only where it adds a chargeable new GMDN category under the current fee rules; this is not a per-REF or per-brand variation tariff.

Conditions
  • Apply the current annual/category and pro-rata rules to the actual registration object. UKRP replacement/new represented manufacturer is a new-registration branch, not a free same-person edit.
  • Labels and identifiers

Keep the correct GB and NI label identities

Fact-level verified

For GB choose the actual UKCA or accepted-CE conformity route. When UKCA is affixed, including a dual-marked device, the UKRP name and address must be on the product labelling, outer packaging or instructions for use. NI devices must meet EU labelling requirements. A brand artwork revision does not remove mandatory device/manufacturer/representative identifiers or validate a mismatched conformity record.

Conditions
  • A CE-only GB label follows the accepted CE route; do not impose the UKCA UKRP-label rule indiscriminately.
  • No sell-through period follows merely from keeping these particulars aligned.
  • Free sale certificate impact

Update the CFS source data before an export order

Fact-level verified

A CFS is an export service requested where the destination requires it, not a routine prerequisite for every identity update. It may be ordered by a UK-based manufacturer, UKRP or NI authorised representative for DORS-registered eligible devices with valid conformity evidence; performance-evaluation IVDs are excluded. Correct product name, model/version and REF before ordering. A pending managed-device application blocks CFS for affected devices. The published target is 10 working days from receipt, longer during busy periods; that is a CFS service target, not a variation-review deadline.

Conditions
  • No automatic duty to reissue every existing CFS is established. Check the destination and whether the existing certificate still describes the exported devices.
  • The CFS schedule uses product data, not merely a GMDN; up to 1000 products per order is not a DORS registration limit.
  • Labels and identifiers

IVD name/model changes and catalogue REF are distinct

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, a new UDI-DI is required for a change that could cause misidentification or traceability ambiguity, including a device name/trade-name or version/model change. An actual catalogue/REF change is not automatically the same thing: assess its effect on identity and traceability. A change to an existing UDI database element that does not require a new UDI-DI must be updated within 30 days. Cosmetic artwork alone is not listed as an automatic new-UDI trigger.

Conditions
  • NI/EU route only; this fact does not create a universal GB UDI mandate.
  • 30 days is an applicant data-update deadline, not an authority review period.
  • Labels and identifiers

IVD label identity after a change

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, the label identifies the device name/trade name and the manufacturer’s name, registered trade name or trademark and registered-place-of-business address, plus the representative where required and the applicable UDI carrier. Keep the actual post-change device/manufacturer identities consistent with the conformity and registration records. This label-content rule is not an old-stock grace or a permission to use another company’s certificate.

Conditions
  • NI/EU route only; use the actual GB marking branch separately.
  • Change-specific government fees

Medical Device Registration Statutory Fee

Source-verified

£300

Per chargeable GMDN category, annual (pro rata for registration during the fee year)

Level 2, or Level 1 where no Level 2 exists; fee year 1 April 2026 to 31 March 2027. After MHRA recalculation, outstanding annual fees for devices placed on the market are due 6 November 2026; unpaid accounts are suspended thereafter. This is not a per-product or ordinary variation fee.

Effective from 2026-04-01

Conditions for this question

  • Only if this identity change requires a genuinely new chargeable GMDN category under the current GB/remaining applicable MHRA branch; apply category grouping and pro-rata rules. It is not a per-change or per-REF fee, nor an NI EUDAMED fee.
  • Change-specific government fees

Conditions for this question

  • Only if a new MHRA export CFS order is actually required after the change; the exact row covers the first 10 certificates and its stated additional-certificate rule. Not a routine variation fee.

How do I update a manufacturer's name without changing the legal entity?

The same legal manufacturer changes its name, potentially alongside the device brand. A transfer to another legal entity is a separate question; permission to combine filings needs its own evidence.

  • Filing route and prior approval

Choose the GB or Northern Ireland registration route

Fact-level verified

All devices, including IVDs, must be registered with MHRA before placement on the Great Britain market, and the registered information must be accurate and current. Since 28 May 2026 ordinary MD and IVD devices for Northern Ireland use EUDAMED instead of MHRA registration; qualifying custom-made medical devices under MDR retain the separate MHRA branch. A manufacturer outside the UK appoints one UK Responsible Person for GB. A manufacturer outside the EU or Northern Ireland, including one in GB, needs an EU- or Northern Ireland-based authorised representative for NI.

Conditions
  • Separate GB and NI portfolios, and ordinary commercial devices from custom-made or performance-study devices. Registration is not MHRA product approval.
  • Filing route and prior approval
  • Supporting documents

Update a name or office address only for the same legal person

Fact-level verified

DORS organisation name and address edits are permitted only where the legal entity remains the same, and are free. A UKRP or NI representative changing its own details must supply updated designation letters for represented manufacturers; a representative editing a represented manufacturer must upload a new designation letter. Those designation-dependent changes are applied after MHRA review and acceptance. A UK manufacturer acting for itself does not need a designation letter.

Conditions
  • An office address edit does not establish approval of a factory move. This editor cannot be used to transfer a manufacturer to a different legal person.
  • Filing route and prior approval

Distinguish registered brand/model from catalogue REF

Fact-level verified

The current DORS attributes distinguish brand/trade name, model/version, catalogue/REF and UDI. Brand name and model/version are not editable product fields; catalogue/REF is editable. In the MD sheet REF is mandatory; in the IVD sheet it is optional. Populated UDI identity fields have stricter mutability. The device guide requires removing and re-adding a product for a disabled product field, or removing and re-registering the device and its underlying products for a disabled device field. The public registration guidance expressly lists trade/brand-name and non-editable characteristic changes as re-registration triggers.

Conditions
  • Identify the actual product and device field; do not equate a commercial REF with a model, GMDN or UDI-DI.
  • The current statutory fee guidance governs whether re-registration creates a chargeable new GMDN category.
  • Filing route and prior approval

Northern Ireland actor updates use the EU actor route

Fact-level verified

For ordinary NI devices under MDR or IVDR, the economic operator updates changed registered actor data within one week under MDR 31(4) or IVDR 28(4). A Local Actor Administrator creates and registers a new version of the actor details. Tax/trade-ID changes have a separate authority-validated request and pending-request lock. A person not already registered must satisfy the applicable premarket actor-registration requirement; an existing-actor edit does not itself prove succession to another legal manufacturer’s conformity evidence.

Conditions
  • NI ordinary-device branch only; this is not the GB DORS actor path or an agency review deadline.
  • Filing route and prior approval

IVD notified-body change gates

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, where Annex IX conformity assessment applies, notify the body that approved the QMS of planned substantial QMS or covered device-range changes; it assesses the proposal and any necessary audits. Under Annex IX 4.11, a change to an approved device that could affect safety, performance or prescribed conditions of use needs the issuing body’s approval. The body decides whether a new conformity assessment or a certificate supplement is required. An identity-data edit does not settle that decision.

Conditions
  • Applies only where the relevant notified/designated-body certificate and change trigger exist. A cosmetic logo, same-person office change or REF-only edit is not automatically a substantial change.
  • No universal authority or notified-body review duration or fee is established by these clauses.
  • Filing route and prior approval

IVD legacy administrative changes need the actual certificate basis

Fact-level verified

MDCG 2022-6 distinguishes administrative manufacturer name/address/legal-form changes, including its merger/acquisition example, from significant design/intended-purpose changes for legacy IVDs. This design-change guidance does not itself establish a transfer of conformity responsibility or a universal certificate-succession procedure. Observe the actual certificate conditions and agreed notified-body change procedure; current transition dates must come from amended law, not the old dates printed in this 2022 guide.

Conditions
  • Legacy-device scope only, with all actual current transition conditions satisfied.
  • This is not permission for every true new-legal-manufacturer combination or a current transition deadline.

Conditions for this question

  • Legacy conformity branch only; the fact is not a universal new-manufacturer transfer procedure.
  • Supporting documents

Evidence for the actual DORS change route

Fact-level verified

For a manufacturer or device registration/re-registration, provide the legal-entity name and address as shown on the labelling, applicable legislation and class, GMDN, brand, model/version, REF and UDI information where applicable, and the relevant conformity certificate or declaration. A represented organisation also needs the applicable designation evidence. For an ordinary same-person organisation edit use the change-specific designation-letter rules; this is not a universal new technical dossier for every REF correction.

Conditions
  • Certificate/declaration evidence is required when the actual registration/re-registration or conformity-document route is triggered.
  • MHRA may request further technical evidence; a data update is not conformity approval.
  • Change-specific government fees

Free DORS updates and chargeable re-registration branches

Fact-level verified

From 1 April 2026 no separate statutory fee is charged for organisation name/address updates or new designation-letter uploads. Adding/removing products under registered devices, updating editable fields and obsolete GMDN data, and adding/deactivating importers are also uncharged as specified. Re-registration following a changed characteristic is charged only where it adds a chargeable new GMDN category under the current fee rules; this is not a per-REF or per-brand variation tariff.

Conditions
  • Apply the current annual/category and pro-rata rules to the actual registration object. UKRP replacement/new represented manufacturer is a new-registration branch, not a free same-person edit.
  • Labels and identifiers

Keep the correct GB and NI label identities

Fact-level verified

For GB choose the actual UKCA or accepted-CE conformity route. When UKCA is affixed, including a dual-marked device, the UKRP name and address must be on the product labelling, outer packaging or instructions for use. NI devices must meet EU labelling requirements. A brand artwork revision does not remove mandatory device/manufacturer/representative identifiers or validate a mismatched conformity record.

Conditions
  • A CE-only GB label follows the accepted CE route; do not impose the UKCA UKRP-label rule indiscriminately.
  • No sell-through period follows merely from keeping these particulars aligned.
  • Free sale certificate impact

Update the CFS source data before an export order

Fact-level verified

A CFS is an export service requested where the destination requires it, not a routine prerequisite for every identity update. It may be ordered by a UK-based manufacturer, UKRP or NI authorised representative for DORS-registered eligible devices with valid conformity evidence; performance-evaluation IVDs are excluded. Correct product name, model/version and REF before ordering. A pending managed-device application blocks CFS for affected devices. The published target is 10 working days from receipt, longer during busy periods; that is a CFS service target, not a variation-review deadline.

Conditions
  • No automatic duty to reissue every existing CFS is established. Check the destination and whether the existing certificate still describes the exported devices.
  • The CFS schedule uses product data, not merely a GMDN; up to 1000 products per order is not a DORS registration limit.
  • Labels and identifiers

IVD name/model changes and catalogue REF are distinct

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, a new UDI-DI is required for a change that could cause misidentification or traceability ambiguity, including a device name/trade-name or version/model change. An actual catalogue/REF change is not automatically the same thing: assess its effect on identity and traceability. A change to an existing UDI database element that does not require a new UDI-DI must be updated within 30 days. Cosmetic artwork alone is not listed as an automatic new-UDI trigger.

Conditions
  • NI/EU route only; this fact does not create a universal GB UDI mandate.
  • 30 days is an applicant data-update deadline, not an authority review period.
  • Labels and identifiers

IVD label identity after a change

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, the label identifies the device name/trade name and the manufacturer’s name, registered trade name or trademark and registered-place-of-business address, plus the representative where required and the applicable UDI carrier. Keep the actual post-change device/manufacturer identities consistent with the conformity and registration records. This label-content rule is not an old-stock grace or a permission to use another company’s certificate.

Conditions
  • NI/EU route only; use the actual GB marking branch separately.
  • Labels and identifiers

English language requirement - IVDs

Fact-level verified

For in vitro diagnostic medical devices, the information to be provided under Section 8 of Part B of Annex I (labelling and instructions for use) must be in English.

Conditions
  • Great Britain E+W+S expression of reg 35(2); do not substitute for the NI EU-law branch.

Schedule 2 A para 25(f) also requires the UK responsible person's name and address on the label/IFU for IVDs of manufacturers without a UK registered place of business.

  • Change-specific government fees

Medical Device Registration Statutory Fee

Source-verified

£300

Per chargeable GMDN category, annual (pro rata for registration during the fee year)

Level 2, or Level 1 where no Level 2 exists; fee year 1 April 2026 to 31 March 2027. After MHRA recalculation, outstanding annual fees for devices placed on the market are due 6 November 2026; unpaid accounts are suspended thereafter. This is not a per-product or ordinary variation fee.

Effective from 2026-04-01

Conditions for this question

  • Only if this identity change requires a genuinely new chargeable GMDN category under the current GB/remaining applicable MHRA branch; apply category grouping and pro-rata rules. It is not a per-change or per-REF fee, nor an NI EUDAMED fee.
  • Change-specific government fees

Conditions for this question

  • Only if a new MHRA export CFS order is actually required after the change; the exact row covers the first 10 certificates and its stated additional-certificate rule. Not a routine variation fee.

What changes when the manufacturer's office address changes?

An office address change for the same legal entity, potentially alongside name and brand changes. This does not cover moving a factory or manufacturing site, and a combined filing cannot be assumed.

  • Filing route and prior approval

Choose the GB or Northern Ireland registration route

Fact-level verified

All devices, including IVDs, must be registered with MHRA before placement on the Great Britain market, and the registered information must be accurate and current. Since 28 May 2026 ordinary MD and IVD devices for Northern Ireland use EUDAMED instead of MHRA registration; qualifying custom-made medical devices under MDR retain the separate MHRA branch. A manufacturer outside the UK appoints one UK Responsible Person for GB. A manufacturer outside the EU or Northern Ireland, including one in GB, needs an EU- or Northern Ireland-based authorised representative for NI.

Conditions
  • Separate GB and NI portfolios, and ordinary commercial devices from custom-made or performance-study devices. Registration is not MHRA product approval.
  • Filing route and prior approval
  • Supporting documents

Update a name or office address only for the same legal person

Fact-level verified

DORS organisation name and address edits are permitted only where the legal entity remains the same, and are free. A UKRP or NI representative changing its own details must supply updated designation letters for represented manufacturers; a representative editing a represented manufacturer must upload a new designation letter. Those designation-dependent changes are applied after MHRA review and acceptance. A UK manufacturer acting for itself does not need a designation letter.

Conditions
  • An office address edit does not establish approval of a factory move. This editor cannot be used to transfer a manufacturer to a different legal person.
  • Filing route and prior approval

Distinguish registered brand/model from catalogue REF

Fact-level verified

The current DORS attributes distinguish brand/trade name, model/version, catalogue/REF and UDI. Brand name and model/version are not editable product fields; catalogue/REF is editable. In the MD sheet REF is mandatory; in the IVD sheet it is optional. Populated UDI identity fields have stricter mutability. The device guide requires removing and re-adding a product for a disabled product field, or removing and re-registering the device and its underlying products for a disabled device field. The public registration guidance expressly lists trade/brand-name and non-editable characteristic changes as re-registration triggers.

Conditions
  • Identify the actual product and device field; do not equate a commercial REF with a model, GMDN or UDI-DI.
  • The current statutory fee guidance governs whether re-registration creates a chargeable new GMDN category.
  • Filing route and prior approval

Northern Ireland actor updates use the EU actor route

Fact-level verified

For ordinary NI devices under MDR or IVDR, the economic operator updates changed registered actor data within one week under MDR 31(4) or IVDR 28(4). A Local Actor Administrator creates and registers a new version of the actor details. Tax/trade-ID changes have a separate authority-validated request and pending-request lock. A person not already registered must satisfy the applicable premarket actor-registration requirement; an existing-actor edit does not itself prove succession to another legal manufacturer’s conformity evidence.

Conditions
  • NI ordinary-device branch only; this is not the GB DORS actor path or an agency review deadline.
  • Filing route and prior approval

IVD notified-body change gates

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, where Annex IX conformity assessment applies, notify the body that approved the QMS of planned substantial QMS or covered device-range changes; it assesses the proposal and any necessary audits. Under Annex IX 4.11, a change to an approved device that could affect safety, performance or prescribed conditions of use needs the issuing body’s approval. The body decides whether a new conformity assessment or a certificate supplement is required. An identity-data edit does not settle that decision.

Conditions
  • Applies only where the relevant notified/designated-body certificate and change trigger exist. A cosmetic logo, same-person office change or REF-only edit is not automatically a substantial change.
  • No universal authority or notified-body review duration or fee is established by these clauses.
  • Filing route and prior approval

IVD legacy administrative changes need the actual certificate basis

Fact-level verified

MDCG 2022-6 distinguishes administrative manufacturer name/address/legal-form changes, including its merger/acquisition example, from significant design/intended-purpose changes for legacy IVDs. This design-change guidance does not itself establish a transfer of conformity responsibility or a universal certificate-succession procedure. Observe the actual certificate conditions and agreed notified-body change procedure; current transition dates must come from amended law, not the old dates printed in this 2022 guide.

Conditions
  • Legacy-device scope only, with all actual current transition conditions satisfied.
  • This is not permission for every true new-legal-manufacturer combination or a current transition deadline.

Conditions for this question

  • Legacy conformity branch only; the fact is not a universal new-manufacturer transfer procedure.
  • Supporting documents

Evidence for the actual DORS change route

Fact-level verified

For a manufacturer or device registration/re-registration, provide the legal-entity name and address as shown on the labelling, applicable legislation and class, GMDN, brand, model/version, REF and UDI information where applicable, and the relevant conformity certificate or declaration. A represented organisation also needs the applicable designation evidence. For an ordinary same-person organisation edit use the change-specific designation-letter rules; this is not a universal new technical dossier for every REF correction.

Conditions
  • Certificate/declaration evidence is required when the actual registration/re-registration or conformity-document route is triggered.
  • MHRA may request further technical evidence; a data update is not conformity approval.
  • Change-specific government fees

Free DORS updates and chargeable re-registration branches

Fact-level verified

From 1 April 2026 no separate statutory fee is charged for organisation name/address updates or new designation-letter uploads. Adding/removing products under registered devices, updating editable fields and obsolete GMDN data, and adding/deactivating importers are also uncharged as specified. Re-registration following a changed characteristic is charged only where it adds a chargeable new GMDN category under the current fee rules; this is not a per-REF or per-brand variation tariff.

Conditions
  • Apply the current annual/category and pro-rata rules to the actual registration object. UKRP replacement/new represented manufacturer is a new-registration branch, not a free same-person edit.
  • Labels and identifiers

Keep the correct GB and NI label identities

Fact-level verified

For GB choose the actual UKCA or accepted-CE conformity route. When UKCA is affixed, including a dual-marked device, the UKRP name and address must be on the product labelling, outer packaging or instructions for use. NI devices must meet EU labelling requirements. A brand artwork revision does not remove mandatory device/manufacturer/representative identifiers or validate a mismatched conformity record.

Conditions
  • A CE-only GB label follows the accepted CE route; do not impose the UKCA UKRP-label rule indiscriminately.
  • No sell-through period follows merely from keeping these particulars aligned.
  • Free sale certificate impact

Update the CFS source data before an export order

Fact-level verified

A CFS is an export service requested where the destination requires it, not a routine prerequisite for every identity update. It may be ordered by a UK-based manufacturer, UKRP or NI authorised representative for DORS-registered eligible devices with valid conformity evidence; performance-evaluation IVDs are excluded. Correct product name, model/version and REF before ordering. A pending managed-device application blocks CFS for affected devices. The published target is 10 working days from receipt, longer during busy periods; that is a CFS service target, not a variation-review deadline.

Conditions
  • No automatic duty to reissue every existing CFS is established. Check the destination and whether the existing certificate still describes the exported devices.
  • The CFS schedule uses product data, not merely a GMDN; up to 1000 products per order is not a DORS registration limit.
  • Labels and identifiers

IVD name/model changes and catalogue REF are distinct

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, a new UDI-DI is required for a change that could cause misidentification or traceability ambiguity, including a device name/trade-name or version/model change. An actual catalogue/REF change is not automatically the same thing: assess its effect on identity and traceability. A change to an existing UDI database element that does not require a new UDI-DI must be updated within 30 days. Cosmetic artwork alone is not listed as an automatic new-UDI trigger.

Conditions
  • NI/EU route only; this fact does not create a universal GB UDI mandate.
  • 30 days is an applicant data-update deadline, not an authority review period.
  • Labels and identifiers

IVD label identity after a change

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, the label identifies the device name/trade name and the manufacturer’s name, registered trade name or trademark and registered-place-of-business address, plus the representative where required and the applicable UDI carrier. Keep the actual post-change device/manufacturer identities consistent with the conformity and registration records. This label-content rule is not an old-stock grace or a permission to use another company’s certificate.

Conditions
  • NI/EU route only; use the actual GB marking branch separately.
  • Labels and identifiers

English language requirement - IVDs

Fact-level verified

For in vitro diagnostic medical devices, the information to be provided under Section 8 of Part B of Annex I (labelling and instructions for use) must be in English.

Conditions
  • Great Britain E+W+S expression of reg 35(2); do not substitute for the NI EU-law branch.

Schedule 2 A para 25(f) also requires the UK responsible person's name and address on the label/IFU for IVDs of manufacturers without a UK registered place of business.

  • Change-specific government fees

Medical Device Registration Statutory Fee

Source-verified

£300

Per chargeable GMDN category, annual (pro rata for registration during the fee year)

Level 2, or Level 1 where no Level 2 exists; fee year 1 April 2026 to 31 March 2027. After MHRA recalculation, outstanding annual fees for devices placed on the market are due 6 November 2026; unpaid accounts are suspended thereafter. This is not a per-product or ordinary variation fee.

Effective from 2026-04-01

Conditions for this question

  • Only if this identity change requires a genuinely new chargeable GMDN category under the current GB/remaining applicable MHRA branch; apply category grouping and pro-rata rules. It is not a per-change or per-REF fee, nor an NI EUDAMED fee.
  • Change-specific government fees

Conditions for this question

  • Only if a new MHRA export CFS order is actually required after the change; the exact row covers the first 10 certificates and its stated additional-certificate rule. Not a routine variation fee.

Does a catalogue or reference code change affect registration or UDI?

Check which identifier actually changes. A commercial catalogue, REF or order code is not automatically a device model, registered identifier, GTIN or UDI-DI. Rules for one identifier do not establish the route for another.

  • Filing route and prior approval

Choose the GB or Northern Ireland registration route

Fact-level verified

All devices, including IVDs, must be registered with MHRA before placement on the Great Britain market, and the registered information must be accurate and current. Since 28 May 2026 ordinary MD and IVD devices for Northern Ireland use EUDAMED instead of MHRA registration; qualifying custom-made medical devices under MDR retain the separate MHRA branch. A manufacturer outside the UK appoints one UK Responsible Person for GB. A manufacturer outside the EU or Northern Ireland, including one in GB, needs an EU- or Northern Ireland-based authorised representative for NI.

Conditions
  • Separate GB and NI portfolios, and ordinary commercial devices from custom-made or performance-study devices. Registration is not MHRA product approval.
  • Filing route and prior approval

Distinguish registered brand/model from catalogue REF

Fact-level verified

The current DORS attributes distinguish brand/trade name, model/version, catalogue/REF and UDI. Brand name and model/version are not editable product fields; catalogue/REF is editable. In the MD sheet REF is mandatory; in the IVD sheet it is optional. Populated UDI identity fields have stricter mutability. The device guide requires removing and re-adding a product for a disabled product field, or removing and re-registering the device and its underlying products for a disabled device field. The public registration guidance expressly lists trade/brand-name and non-editable characteristic changes as re-registration triggers.

Conditions
  • Identify the actual product and device field; do not equate a commercial REF with a model, GMDN or UDI-DI.
  • The current statutory fee guidance governs whether re-registration creates a chargeable new GMDN category.
  • Filing route and prior approval

IVD notified-body change gates

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, where Annex IX conformity assessment applies, notify the body that approved the QMS of planned substantial QMS or covered device-range changes; it assesses the proposal and any necessary audits. Under Annex IX 4.11, a change to an approved device that could affect safety, performance or prescribed conditions of use needs the issuing body’s approval. The body decides whether a new conformity assessment or a certificate supplement is required. An identity-data edit does not settle that decision.

Conditions
  • Applies only where the relevant notified/designated-body certificate and change trigger exist. A cosmetic logo, same-person office change or REF-only edit is not automatically a substantial change.
  • No universal authority or notified-body review duration or fee is established by these clauses.
  • Supporting documents

Evidence for the actual DORS change route

Fact-level verified

For a manufacturer or device registration/re-registration, provide the legal-entity name and address as shown on the labelling, applicable legislation and class, GMDN, brand, model/version, REF and UDI information where applicable, and the relevant conformity certificate or declaration. A represented organisation also needs the applicable designation evidence. For an ordinary same-person organisation edit use the change-specific designation-letter rules; this is not a universal new technical dossier for every REF correction.

Conditions
  • Certificate/declaration evidence is required when the actual registration/re-registration or conformity-document route is triggered.
  • MHRA may request further technical evidence; a data update is not conformity approval.
  • Change-specific government fees

Free DORS updates and chargeable re-registration branches

Fact-level verified

From 1 April 2026 no separate statutory fee is charged for organisation name/address updates or new designation-letter uploads. Adding/removing products under registered devices, updating editable fields and obsolete GMDN data, and adding/deactivating importers are also uncharged as specified. Re-registration following a changed characteristic is charged only where it adds a chargeable new GMDN category under the current fee rules; this is not a per-REF or per-brand variation tariff.

Conditions
  • Apply the current annual/category and pro-rata rules to the actual registration object. UKRP replacement/new represented manufacturer is a new-registration branch, not a free same-person edit.
  • Labels and identifiers

Keep the correct GB and NI label identities

Fact-level verified

For GB choose the actual UKCA or accepted-CE conformity route. When UKCA is affixed, including a dual-marked device, the UKRP name and address must be on the product labelling, outer packaging or instructions for use. NI devices must meet EU labelling requirements. A brand artwork revision does not remove mandatory device/manufacturer/representative identifiers or validate a mismatched conformity record.

Conditions
  • A CE-only GB label follows the accepted CE route; do not impose the UKCA UKRP-label rule indiscriminately.
  • No sell-through period follows merely from keeping these particulars aligned.
  • Free sale certificate impact

Update the CFS source data before an export order

Fact-level verified

A CFS is an export service requested where the destination requires it, not a routine prerequisite for every identity update. It may be ordered by a UK-based manufacturer, UKRP or NI authorised representative for DORS-registered eligible devices with valid conformity evidence; performance-evaluation IVDs are excluded. Correct product name, model/version and REF before ordering. A pending managed-device application blocks CFS for affected devices. The published target is 10 working days from receipt, longer during busy periods; that is a CFS service target, not a variation-review deadline.

Conditions
  • No automatic duty to reissue every existing CFS is established. Check the destination and whether the existing certificate still describes the exported devices.
  • The CFS schedule uses product data, not merely a GMDN; up to 1000 products per order is not a DORS registration limit.
  • Labels and identifiers

IVD name/model changes and catalogue REF are distinct

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, a new UDI-DI is required for a change that could cause misidentification or traceability ambiguity, including a device name/trade-name or version/model change. An actual catalogue/REF change is not automatically the same thing: assess its effect on identity and traceability. A change to an existing UDI database element that does not require a new UDI-DI must be updated within 30 days. Cosmetic artwork alone is not listed as an automatic new-UDI trigger.

Conditions
  • NI/EU route only; this fact does not create a universal GB UDI mandate.
  • 30 days is an applicant data-update deadline, not an authority review period.
  • Labels and identifiers

IVD label identity after a change

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, the label identifies the device name/trade name and the manufacturer’s name, registered trade name or trademark and registered-place-of-business address, plus the representative where required and the applicable UDI carrier. Keep the actual post-change device/manufacturer identities consistent with the conformity and registration records. This label-content rule is not an old-stock grace or a permission to use another company’s certificate.

Conditions
  • NI/EU route only; use the actual GB marking branch separately.
  • Labels and identifiers

English language requirement - IVDs

Fact-level verified

For in vitro diagnostic medical devices, the information to be provided under Section 8 of Part B of Annex I (labelling and instructions for use) must be in English.

Conditions
  • Great Britain E+W+S expression of reg 35(2); do not substitute for the NI EU-law branch.

Schedule 2 A para 25(f) also requires the UK responsible person's name and address on the label/IFU for IVDs of manufacturers without a UK registered place of business.

  • Change-specific government fees

Medical Device Registration Statutory Fee

Source-verified

£300

Per chargeable GMDN category, annual (pro rata for registration during the fee year)

Level 2, or Level 1 where no Level 2 exists; fee year 1 April 2026 to 31 March 2027. After MHRA recalculation, outstanding annual fees for devices placed on the market are due 6 November 2026; unpaid accounts are suspended thereafter. This is not a per-product or ordinary variation fee.

Effective from 2026-04-01

Conditions for this question

  • Only if this identity change requires a genuinely new chargeable GMDN category under the current GB/remaining applicable MHRA branch; apply category grouping and pro-rata rules. It is not a per-change or per-REF fee, nor an NI EUDAMED fee.
  • Change-specific government fees

Conditions for this question

  • Only if a new MHRA export CFS order is actually required after the change; the exact row covers the first 10 certificates and its stated additional-certificate rule. Not a routine variation fee.

Can several changes be filed together?

Changes to brand, manufacturer name, office address and reference codes may occur together, including with a real transfer to another legal manufacturer, registration holder or owner. Assess transfer with each single change, and transfer with brand plus reference, manufacturer name plus reference, or office address plus reference changes: seven transfer combinations in total. Evidence must establish the exact combination and the order of filings and approvals. These combinations and their sequence are not all verified by the rules below; a rule for one change does not prove that a single application, fee or review period covers the combination.

  • Filing route and prior approval

Choose the GB or Northern Ireland registration route

Fact-level verified

All devices, including IVDs, must be registered with MHRA before placement on the Great Britain market, and the registered information must be accurate and current. Since 28 May 2026 ordinary MD and IVD devices for Northern Ireland use EUDAMED instead of MHRA registration; qualifying custom-made medical devices under MDR retain the separate MHRA branch. A manufacturer outside the UK appoints one UK Responsible Person for GB. A manufacturer outside the EU or Northern Ireland, including one in GB, needs an EU- or Northern Ireland-based authorised representative for NI.

Conditions
  • Separate GB and NI portfolios, and ordinary commercial devices from custom-made or performance-study devices. Registration is not MHRA product approval.
  • Filing route and prior approval

Distinguish registered brand/model from catalogue REF

Fact-level verified

The current DORS attributes distinguish brand/trade name, model/version, catalogue/REF and UDI. Brand name and model/version are not editable product fields; catalogue/REF is editable. In the MD sheet REF is mandatory; in the IVD sheet it is optional. Populated UDI identity fields have stricter mutability. The device guide requires removing and re-adding a product for a disabled product field, or removing and re-registering the device and its underlying products for a disabled device field. The public registration guidance expressly lists trade/brand-name and non-editable characteristic changes as re-registration triggers.

Conditions
  • Identify the actual product and device field; do not equate a commercial REF with a model, GMDN or UDI-DI.
  • The current statutory fee guidance governs whether re-registration creates a chargeable new GMDN category.
  • Filing route and prior approval

Northern Ireland actor updates use the EU actor route

Fact-level verified

For ordinary NI devices under MDR or IVDR, the economic operator updates changed registered actor data within one week under MDR 31(4) or IVDR 28(4). A Local Actor Administrator creates and registers a new version of the actor details. Tax/trade-ID changes have a separate authority-validated request and pending-request lock. A person not already registered must satisfy the applicable premarket actor-registration requirement; an existing-actor edit does not itself prove succession to another legal manufacturer’s conformity evidence.

Conditions
  • NI ordinary-device branch only; this is not the GB DORS actor path or an agency review deadline.
  • Filing route and prior approval

IVD notified-body change gates

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, where Annex IX conformity assessment applies, notify the body that approved the QMS of planned substantial QMS or covered device-range changes; it assesses the proposal and any necessary audits. Under Annex IX 4.11, a change to an approved device that could affect safety, performance or prescribed conditions of use needs the issuing body’s approval. The body decides whether a new conformity assessment or a certificate supplement is required. An identity-data edit does not settle that decision.

Conditions
  • Applies only where the relevant notified/designated-body certificate and change trigger exist. A cosmetic logo, same-person office change or REF-only edit is not automatically a substantial change.
  • No universal authority or notified-body review duration or fee is established by these clauses.
  • Supporting documents

Evidence for the actual DORS change route

Fact-level verified

For a manufacturer or device registration/re-registration, provide the legal-entity name and address as shown on the labelling, applicable legislation and class, GMDN, brand, model/version, REF and UDI information where applicable, and the relevant conformity certificate or declaration. A represented organisation also needs the applicable designation evidence. For an ordinary same-person organisation edit use the change-specific designation-letter rules; this is not a universal new technical dossier for every REF correction.

Conditions
  • Certificate/declaration evidence is required when the actual registration/re-registration or conformity-document route is triggered.
  • MHRA may request further technical evidence; a data update is not conformity approval.
  • Change-specific government fees

Free DORS updates and chargeable re-registration branches

Fact-level verified

From 1 April 2026 no separate statutory fee is charged for organisation name/address updates or new designation-letter uploads. Adding/removing products under registered devices, updating editable fields and obsolete GMDN data, and adding/deactivating importers are also uncharged as specified. Re-registration following a changed characteristic is charged only where it adds a chargeable new GMDN category under the current fee rules; this is not a per-REF or per-brand variation tariff.

Conditions
  • Apply the current annual/category and pro-rata rules to the actual registration object. UKRP replacement/new represented manufacturer is a new-registration branch, not a free same-person edit.
  • Labels and identifiers

Keep the correct GB and NI label identities

Fact-level verified

For GB choose the actual UKCA or accepted-CE conformity route. When UKCA is affixed, including a dual-marked device, the UKRP name and address must be on the product labelling, outer packaging or instructions for use. NI devices must meet EU labelling requirements. A brand artwork revision does not remove mandatory device/manufacturer/representative identifiers or validate a mismatched conformity record.

Conditions
  • A CE-only GB label follows the accepted CE route; do not impose the UKCA UKRP-label rule indiscriminately.
  • No sell-through period follows merely from keeping these particulars aligned.
  • Free sale certificate impact

Update the CFS source data before an export order

Fact-level verified

A CFS is an export service requested where the destination requires it, not a routine prerequisite for every identity update. It may be ordered by a UK-based manufacturer, UKRP or NI authorised representative for DORS-registered eligible devices with valid conformity evidence; performance-evaluation IVDs are excluded. Correct product name, model/version and REF before ordering. A pending managed-device application blocks CFS for affected devices. The published target is 10 working days from receipt, longer during busy periods; that is a CFS service target, not a variation-review deadline.

Conditions
  • No automatic duty to reissue every existing CFS is established. Check the destination and whether the existing certificate still describes the exported devices.
  • The CFS schedule uses product data, not merely a GMDN; up to 1000 products per order is not a DORS registration limit.
  • Labels and identifiers

IVD name/model changes and catalogue REF are distinct

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, a new UDI-DI is required for a change that could cause misidentification or traceability ambiguity, including a device name/trade-name or version/model change. An actual catalogue/REF change is not automatically the same thing: assess its effect on identity and traceability. A change to an existing UDI database element that does not require a new UDI-DI must be updated within 30 days. Cosmetic artwork alone is not listed as an automatic new-UDI trigger.

Conditions
  • NI/EU route only; this fact does not create a universal GB UDI mandate.
  • 30 days is an applicant data-update deadline, not an authority review period.
  • Labels and identifiers

IVD label identity after a change

Fact-level verified

For ordinary Northern Ireland devices under the EU regime, the label identifies the device name/trade name and the manufacturer’s name, registered trade name or trademark and registered-place-of-business address, plus the representative where required and the applicable UDI carrier. Keep the actual post-change device/manufacturer identities consistent with the conformity and registration records. This label-content rule is not an old-stock grace or a permission to use another company’s certificate.

Conditions
  • NI/EU route only; use the actual GB marking branch separately.
  • Labels and identifiers

English language requirement - IVDs

Fact-level verified

For in vitro diagnostic medical devices, the information to be provided under Section 8 of Part B of Annex I (labelling and instructions for use) must be in English.

Conditions
  • Great Britain E+W+S expression of reg 35(2); do not substitute for the NI EU-law branch.

Schedule 2 A para 25(f) also requires the UK responsible person's name and address on the label/IFU for IVDs of manufacturers without a UK registered place of business.

  • Change-specific government fees

Medical Device Registration Statutory Fee

Source-verified

£300

Per chargeable GMDN category, annual (pro rata for registration during the fee year)

Level 2, or Level 1 where no Level 2 exists; fee year 1 April 2026 to 31 March 2027. After MHRA recalculation, outstanding annual fees for devices placed on the market are due 6 November 2026; unpaid accounts are suspended thereafter. This is not a per-product or ordinary variation fee.

Effective from 2026-04-01

Conditions for this question

  • Only if this identity change requires a genuinely new chargeable GMDN category under the current GB/remaining applicable MHRA branch; apply category grouping and pro-rata rules. It is not a per-change or per-REF fee, nor an NI EUDAMED fee.
  • Change-specific government fees

Conditions for this question

  • Only if a new MHRA export CFS order is actually required after the change; the exact row covers the first 10 certificates and its stated additional-certificate rule. Not a routine variation fee.
Regulation codes

What these codes are

Each row above represents one regulatory instrument. Its code is the authority’s identifier. A regulation page collects the metadata and official source; when readable HTML is available, article links can open and highlight the exact cited passage.

Use this market context

Continue from regulations into practical research

Keep United Kingdom as the target market while you move into registration records, classification, cross-country product research, or pathway planning.

Keep researching this market
View the United Kingdom resource hub →
Compare regulations across all countries →
Tracked from official MHRA publications · row IDs are stable — cite as regulations:uk:approved-bodies