🇪🇨 Ecuador IVD regulatory data

Explore the IVD requirements currently available for Ecuador, organized by topic and linked to official sources.

18 facts6 dimensions7 official sourceslatest verification 2026-10-26
Explore questions and sources →

Common questions about changing a registered IVD in Ecuador. Each question separates published rules from details still awaiting verification.

Change control

Does the event require internal documentation, notice, prior approval, transfer, or a new application?

Compare countries

Brand and commercial-name modification evidence

qbd.brand-and-commercial-name-evidence
Fact-level verified

Annex 4 section 4.1 treats commercial-name changes separately from section 4.18 brand change/inclusion/exclusion. Commercial-name evidence includes updated Spanish labeling (with the biomedical-equipment marking exception), owner certification of the name change and unchanged quality/specifications, the specified domestic/foreign legal evidence, updated insert/manual/technical sheet, and legalized CLV bearing the new name. Brand evidence includes updated label, a justification letter issued by the titular or manufacturer, and, for foreign products when applicable, competent-authority CLV/equivalent declaring the new brand with the stated legalization.

Conditions
  • The commercial-name and brand requirements must be applied to the actual change; neither is reduced to cosmetic NDM06 merely because the underlying device is unchanged.
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-37 — Annex 4 sections 4.1 and 4.18; also pdf-page-38, pdf-page-43. See the independently pinned evidence references for every supporting page.

Modification filing framework and document formalities

qbd.dm01-modification-framework
Fact-level verified

DM01 Annex 4 requires the sanitary-registration holder to seek ARCSA authorization immediately with technical justification, electronically through VUE/ECUAPASS. Applicant data must match the operating permit and relevant activity. Technical documents are in Spanish or English and listed in an electronically signed letter from the Ecuador technical responsible person. Foreign legal documents require the stated apostille/consular formalities unless electronically issued and verifiable on the issuer website; other languages require the specified qualified translation. Listed modification cases do not themselves create a new registration number.

Conditions
  • This is the February 2024 instruction under the 2016 norm as amended in 2023, not the norm taking effect in 2027.
  • The quoted 2018-based 10% sentence is not used here as a universal 2026 fee; see the separately sourced 2026 tariff.
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-36 — Annex 4 sections 2 and 4; also pdf-page-37. See the independently pinned evidence references for every supporting page.

Combined modification/notification filing

qbd.dm02-combined-filing-limit
Fact-level verified

DM02 permits modifications and notifications in the same application with each described in “Detalle de la modificación”; for a mixed filing do not enter NDM notification codes in “Código de referencia”, or the application is rejected. Section 5.1(d) limits a mixed filing to four modifications plus three notifications. Notification-only filings may list up to seven NDM codes separated by “/” without spaces. Mixed fees are based on the modifications; the review depends on complexity, risk and device type. Supporting documents remain required for every item.

Conditions
  • This is a same-application limit, not proof that unrelated products, manufacturers or registrations may be combined.
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-4 — Sections 3(c)-(g), 5.1(b)-(e); also pdf-page-5, pdf-page-6. See the independently pinned evidence references for every supporting page.

Manufacturer-address modification evidence

qbd.manufacturer-address-evidence
Fact-level verified

Annex 4 section 4.9 requires foreign competent-authority CLV/equivalent certifying the address change with legalization, ISO/BPM showing the new address, updated internal/external labels (biomedical marking exception), a manufacturer-address notification letter, and a domestic operating permit showing the new address for national manufacture. If the event also changes manufacturing process, the holder must file a process-update modification. Section 4.8 separately governs a product owner’s address when that owner is the principal manufacturer/site.

Conditions
  • Do not equate a legal-office address change, physical factory relocation and new manufacturing-site inclusion. Resolve the actual actor and facility facts before selecting a case.
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-40 — Annex 4 sections 4.8-4.9. See the independently pinned evidence references for every supporting page.

Manufacturer company-name evidence

qbd.manufacturer-company-name-evidence
Fact-level verified

Annex 4 section 4.6 requires updated manufacturer-name labels (subject to the biomedical marking exception), foreign competent-authority CLV/equivalent certifying the company-name change with legalization or domestic chamber/mercantile evidence, a manufacturing contract bearing the new name when applicable, and the specified BPM/ISO 13485 evidence. The supporting quality certificate must justify the company-name change without change to the facilities’ address. This is distinct from replacing the principal manufacturer under section 4.3.

Conditions
  • Actor is the manufacturer; do not substitute product-owner or registration-holder company-name requirements.
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-39 — Annex 4 section 4.6; also pdf-page-40. See the independently pinned evidence references for every supporting page.

Model-change modification boundary

qbd.model-change-risk-boundary
Fact-level verified

Annex 4 section 4.12 permits model-change modification only without affecting risk level, use or approved characteristics. It requires the manufacturer’s technical justification expressly addressing those points, updated internal/external labeling with the biomedical marking exception, and an insert with the new model. Annex 4 section 3 requires a new registration for the specified type/model/intended-use/conditions/composition/energy changes generating a risk-level change and for the stated formulated/drug-combination formula changes. DM02 explicitly defines its “Código de referencia” as an ARCSA service/form code unrelated to manufacturer product codes/references. No pure internal SMN/reference-change route is established by these passages.

Conditions
  • The model must actually change; do not label an internal customer SMN as a model without product/dossier evidence.
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-37 — Annex 4 section 3 and section 4.12; also pdf-page-42. See the independently pinned evidence references for every supporting page.

Office-address notifications depend on actor

qbd.ndm05-ndm07-office-actor-address
Fact-level verified

DM02 NDM05 covers a product-owner address change only when that owner is not the principal manufacturer or manufacturing site. Domestic evidence includes explanatory letter and operating permit at the new address; foreign evidence includes explanatory letter, competent-authority CLV/equivalent when applicable with the stated legalization/verifiable-electronic exception, and a power stating the new address and the holder’s powers. NDM07 covers a domestic registration-holder/applicant address only when it is not the principal manufacturer/site, with notification letter and updated operating permit. The corresponding manufacturer/site address cases are modifications under DM01.

Conditions
  • An “office address” description alone does not establish notification eligibility. Resolve product owner versus holder/applicant versus actual manufacturer/site.
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-8 — Table 1, NDM05 and NDM07; also pdf-page-9. See the independently pinned evidence references for every supporting page.

NDM06 cosmetic label/IFU notification

qbd.ndm06-form-only-artwork
Fact-level verified

DM02 NDM06 covers label/insert/manual updates that do not modify approved information content and concern form only, except updates requested by ARCSA as stated in the table. For insert/manual and label changes, supply a detailed justification, a current/proposed comparison and the proposed insert/manual or label respectively. This source does not convert changed brand/manufacturer/model content into a cosmetic notification.

Conditions
  • The no-impact notification definition and conditions in DM02 sections 2-5 apply.
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-9 — Table 1, NDM06. See the independently pinned evidence references for every supporting page.

Stock notification after previously approved changes

qbd.ndm09-previously-approved-stock
Fact-level verified

DM02 NDM09 concerns stock exhaustion for changes previously approved. The table requires a justification letter describing the reason and the requirements established by the applicable current rule. This passage proves a conditional notification channel after change approval; it does not establish an automatic stock sell-through grace period, the permitted duration, or permission to implement an unapproved change.

Conditions
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-9 — Table 1, NDM09; also pdf-page-10. See the independently pinned evidence references for every supporting page.

Economic-operator roles

Which manufacturer, representative, holder, applicant, importer, and distributor roles apply?

Compare countries

ARCSA authorization for registration use/transfer

qbd.comex-authorized-use-transfer
Fact-level verified

COMEX 017-2025 replaces Article 4 to require ARCSA express authorization before using or modifying the relevant sanitary registration/notification for import. It prohibits endorsements, assignments and transfers that have not been processed and authorized by ARCSA. It does not prohibit every authorized transfer. The transitional provision required affected operators to complete the applicable authority procedure even where holder consent previously existed.

Conditions
  • This is the import/use authorization layer; it is not itself proof of registration-holder recoverability or multiple independent holders.
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-6 — Article 1 replacing Article 4; transitional provision; also pdf-page-7. See the independently pinned evidence references for every supporting page.

Importer/distributor inclusion: transition versus continuing fees

qbd.importer-inclusion-transition-and-continuing
Fact-level verified

ARCSA 2026-001A covers inclusion of importers/distributors in RS/NS/NSO. During the specified 120-day transition, the 10% charge was per application and an application could include one or multiple importers/distributors. After that transition, Article 2 specifies 10% for each additional importer under Ministerial Agreement 112 Article 4. The operative subsequent sentence says additional importer; do not silently extend its per-importer wording to every distributor. This regime concerns authorized commercial operators, not ownership of the sanitary registration.

Conditions
  • The transition is historical as of 2026-10-26; retain the regime sequence rather than presenting its per-application rule as indefinitely current.
  • Signed/effective January 6, 2026; the text refers to the COMEX transition ending April 9, 2026.
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-6 — Articles 1-2; effectiveness by signature; also pdf-page-7. See the independently pinned evidence references for every supporting page.

Separate product-owner and applicant transfer branches

qbd.owner-and-applicant-transfer-evidence
Fact-level verified

Annex 4 section 4.2 product-owner change requires a reason document from owner/manufacturer, a public deed transferring product ownership from old to new owner or equivalent with recognized signatures and foreign legalization, and new-owner authorization/power for the current applicant. Sections 4.4 and its continuation address changing the applicant: product-owner authorization/power to the new applicant and revocation of the former applicant’s power (foreign legalization), updated labels when applicable, technical-representative notification/employment evidence, and notarized registration-rights assignment/equivalent. Include a responsible-technical-person change in the modification detail when involved. These are distinct actors and evidence sets.

Conditions
  • These document lists do not independently establish recovery rights, multiple-holder eligibility, or that a representative appointment alone transfers the registration.
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-38 — Annex 4 sections 4.2 and 4.4; also pdf-page-39. See the independently pinned evidence references for every supporting page.

Owner or holder company-name-only changes

qbd.owner-holder-name-only-changes
Fact-level verified

Annex 4 section 4.5 separately governs a product owner’s company-name change with explanatory letter, applicable labels and marking exception, domestic mercantile proof and updated power. Section 4.7 registration-holder/applicant company-name change requires updated owner power bearing the new name, updated labels when applicable, and a public deed for the name change with recognized signatures or legalized equivalent. A company-name-only event is therefore distinct from assigning ownership or registration rights.

Conditions
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-39 — Annex 4 sections 4.5 and 4.7; also pdf-page-40. See the independently pinned evidence references for every supporting page.

Fees

What regulator fee applies to this class, route, event, and date?

Compare countries

2026 modification fee classification

qbd.tariff2026-modification-classification
Fact-level verified

The 2026 tariff page 10 distinguishes specified modifications requiring technical analysis at 20% of the applicable registration amount from enumerated non-analytical modifications at 10%; its 10% list includes product-name and manufacturer/applicant company-name changes. The residual provision charges modifications not listed in Ministerial Agreement 112 at 10%, subject to the express exceptions, including sanitary-registration notifications governed by the applicable product rules. The listed brand-inclusion row expressly concerns foods/cosmetics and is not a device-brand fee rule. Page 9 lists different registration bases: national medical devices USD 678.25, foreign medical devices USD 904.34 and biochemical reagents USD 678.25; it does not establish that every IVD is a biochemical reagent.

Conditions
  • Classify the actual change and applicable product/registration base before calculating; do not assume every modification is 10% or every combined filing has one fee.
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-10 — Modification groups 3-4; residual rule; registration base rows; also pdf-page-9. See the independently pinned evidence references for every supporting page.

Label, IFU & language

Which label and IFU elements, languages, symbols, and translation controls apply?

Compare countries

Label/insert modification and unchanged remainder

qbd.substantive-label-insert-update
Fact-level verified

Annex 4 allows only the label/insert changes corresponding to the modification requested; the rest must match prior approval, otherwise file the corresponding label/insert update. Section 4.15 requires changed insert/manual, an explanatory letter and current/proposed comparison. Section 4.16 requires new label artwork, explanatory letter (with the stated biomedical-equipment exception/marking requirements) and current/proposed comparison. Cosmetic-only NDM06 is a separate narrow notification branch.

Conditions
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-37 — Annex 4 section 2, sections 4.15-4.16; also pdf-page-42, pdf-page-43. See the independently pinned evidence references for every supporting page.

Reliance & recognition

Which foreign decisions or evidence can be reused, with what effect and conditions?

Compare countries

Existing records include acceptance, restrictions and evidence reuse. The applicable product scope and conditions are preserved below; a reference does not by itself confer market access.

Ecuador approvals and evidence in other markets (0 records)

No linked record is currently maintained. This does not establish that a pathway is unavailable.

Prior approvals and evidence for Ecuador (4 records)

EcuadorEU conformity assessment

Prior approval / evidence record

CE Mark (EU MDR)

Authority

CE Mark (EU MDR)

Accepted

Yes

Benefit

A device already registered in a reference country is what makes the ARCSA homologation route available, and the certificate of free sale plus that reference registration are the substance of the dossier. ARCSA publishes its periods per procedure, not per origin.

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

Prior approval / evidence record

China NMPA approval: origin evidence only

reliance.china-nmpa-origin-evidence
Fact-level verified
Statement

From 28 January 2027, ARCSA-DE-2026-003 Article 25(a) accepts a Certificate of Free Sale (CLV), or qualifying equivalent, from the country of manufacture/export, a high-vigilance agency or an IMDRF member country. A China-issued document must itself satisfy the required declaration of marketing and product/manufacturer/site particulars; an NMPA registration certificate or Class I filing is not automatically a compliant CLV. This is documentary evidence for registration, not an automatic simplified-review benefit.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China-issued Certificate of Free Sale or equivalent satisfying Article 25(a), not merely any NMPA certificate

Target Jurisdiction Id

ecuador

China Classes In Scope

Imported devices within the regulation; qualifying free-sale documentation must be established for the particular product

Edge Type

Origin Country Evidence

Market Access Strength

Evidence Only

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • CLV must state commercial name, product codes, manufacturer(s) and manufacturing sites with addresses
  • Documents apostilled or consularized; electronic CLVs verifiable online accepted (Disposición General Octava)
  • If no validity date, ARCSA treats the CLV as valid for two years from issuance
  • Future rule: not effective before 2027-01-28.

Effective from 2027-01-28

Source passage: Se aceptará el CLV proveniente del país donde se fabrica o exporta, o proveniente de una Agencia de alta vigilancia, o de un país miembro del IMDRF.

EcuadorUS FDA

Prior approval / evidence record

FDA 510(k)/PMA

Authority

FDA 510(k)/PMA

Accepted

Yes

Benefit

Serves the same role as any other reference-country registration for the homologation route.

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

Regulatory reference

Future ARCSA simplified route: dual authority conditions; NMPA eligibility unconfirmed

reliance.china-nmpa-not-reference
Pending verification
Statement

From 28 January 2027, Articles 31-32 require both high-vigilance authority status and IMDRF membership, plus the stated device marketing/class conditions. IMDRF membership alone is insufficient. NMPA appears on WHO's July 2026 transitional medical-device authority list, but ARCSA's treatment of this listing under its high-vigilance definition has not been established; the maintained evidence therefore does not resolve NMPA-only eligibility.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

ecuador

China Classes In Scope

NMPA Class II/III certificates; target ARCSA simplified registration (Arts. 31–34) for classes I–IV

Edge Type

Eligibility Unresolved

Market Access Strength

undetermined

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Both Article 31 authority conditions must hold; Article 32 does not waive the high-vigilance requirement.
  • Not effective before 2027-01-28.
  • Obtain ARCSA confirmation of the effect of WHO transitional medical-device listing; a medicines/biologics authority list does not settle device eligibility.
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Effective from 2027-01-28

Source passage: Agencias de alta vigilancia. - Son autoridades reguladoras que han sido calificadas por la OPS / OMS como Autoridades de Referencia Regional; así como aquellas agencias consideradas como Autoridades Reguladoras Estrictas por la OMS. ... Art. 31.- La ARCSA, reconocerá el registro sanitario o su equivalente, proveniente de Agencias de alta vigilancia y que, a su vez, sean miembros del IMDRF.

Review clocks

What review, validation, and applicant stop-clock rules apply?

Compare countries

Notification deadline and applicant cure period

qbd.dm02-notification-deadline-and-cure
Fact-level verified

DM02 requires listed notification events to be notified within sixty working days after they occur; “término” is defined as working days. An unfavorable notification-document review is returned once for applicant correction within five working days, otherwise the application is cancelled; a second unfavorable review also requires abandonment/re-filing as specified. Mixed-filing review depends on complexity, risk and device type. These are notification filing/cure rules, not an empirical best/expected/worst approval duration or a fixed authority review promise.

Conditions
  • Only qualifying DM02 notifications and their stated review steps; not a deadline to authorize substantive modifications.
  • Bounded reading of the cited official original; no claim that every question aspect or later amendment has been exhausted.

Passage context: pdf-page-4 — Sections 3(b),(g), 4 and 5.1(g)-(i); also pdf-page-5, pdf-page-6. See the independently pinned evidence references for every supporting page.

Related structured datasets

Classification systems · 1 record
Open Risk classes I through IV. →
Id

ecuador:medical-device

Jurisdiction Id

ecuador

Domains
  • medical-device
  • ivd
Name

Risk classes I through IV.

Classes
  • Id

    i

    Label

    Class I

    Normalized Risk Level

    low

  • Id

    ii

    Label

    Class II

    Normalized Risk Level

    medium

  • Id

    iii

    Label

    Class III

    Normalized Risk Level

    high

  • Id

    iv

    Label

    Class IV

    Normalized Risk Level

    highest

Regulator

ARCSA

Official Domains
  • controlsanitario.gob.ec
Source Status

official-source-verified

Source Title

Resolución ARCSA-DE-2026-003-DASP, Art. 21 (nivel de riesgo de los dispositivos médicos)

Source Url

https://www.controlsanitario.gob.ec/wp-content/plugins/download-monitor/download.php?id=15580&force=0

Verified On

2026-08-25

Source Supports
  • class-axis
Notes
  • Identical list under Resolución ARCSA-DE-026-2016-YMIH Art. 11 (R.O. 921) and the 2026 substitute Art. 21: Nivel de Riesgo I (bajo), II (moderado bajo), III (moderado alto), IV (alto).
  • Devices also divide into 4 types: active (DMA), invasive (DMI), non-invasive (DMNI) and IVD (DMDIV).
  • Risk classification applies ARCSA instructivos based on GHTF/IMDRF rules; standalone SaMD is classified per the IMDRF/N12 risk matrix (Art. 22, Disposición General Tercera).
  • Sets/kits: expiry is that of the shortest-shelf-life component; risk class is that of the highest-risk component device (Art. 16).
Registration pathways, prior approvals and planning · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Ecuador registration →
Country

Ecuador

Jurisdiction Id

ecuador

Country Code

EC

Flag

🇪🇨

Regulator

National Agency for Regulation, Control and Sanitary Surveillance

Regulator Abbrev

ARCSA

Region

americas

Tier

3

Risk Classification

4-class (I, II, III, IV)

Classes
  • Id

    standard-registration

    Class Name

    Standard registration (all classes)

    Pathway

    ARCSA sanitary registration through the VUE single window

    Official Timeline
    Min

    152

    Max

    152

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    ARCSA response times by trámite — device registration 1,216 working hours, recognition route 760 working hours

    Url

    https://www.controlsanitario.gob.ec/tiempos-de-respuesta-por-tramite-de-arcsa

    Accessed Date

    2026-08-29

    Realistic Timeline
    Min

    8

    Max

    14

    Unit

    months

    Notes
    • ARCSA publishes trámite ARCSA-004-12-01 at 1,216 working hours, shown here as 152 working days on Ecuador’s 8-hour statutory workday
    • The clock is route-wide: ARCSA-DE-026-2016 sets the same procedure for Class I to IV and differentiates only the dossier, not the review period
    • The published figure excludes the time the applicant spends answering objections
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

  • Id

    recognition-registration

    Class Name

    Recognition route (all classes)

    Pathway

    Simplified registration recognising an IMDRF high-surveillance authority approval

    Official Timeline
    Min

    95

    Max

    95

    Unit

    working-days

    Verification Status

    official-verified

    Official Source
    Title

    ARCSA response times by trámite — device registration 1,216 working hours, recognition route 760 working hours

    Url

    https://www.controlsanitario.gob.ec/tiempos-de-respuesta-por-tramite-de-arcsa

    Accessed Date

    2026-08-29

    Realistic Timeline
    Min

    5

    Max

    9

    Unit

    months

    Notes
    • ARCSA publishes trámite ARCSA-004-01-10 at 760 working hours, shown here as 95 working days
    • Open to Class I to IV devices already registered by an authority that sits on the IMDRF Management Committee, capped at 100 items per application
    • The remaining standard-route documents are due within 30 días término of approval or ARCSA cancels the registration
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

Key Requirements
  • Local Authorized Representative required for foreign manufacturers
  • Post-approval clocks published by ARCSA: modification (ARCSA-004-12-02) 240 working hours; reinscripción (ARCSA-004-19-02) 120 hours, filed at least 90 días término before expiry
  • Sanitary permit valid 5 years; renewals and modifications follow the same review track
  • Resolución ARCSA-DE-2026-003-DASP replaces ARCSA-DE-026-2016 on 28 Jan 2027 and changes the picture: art. 33 keeps a 30 días término clock for the recognition route, but the standard route loses its fixed period because art. 26 makes evaluation time depend on risk level, device type and complexity
Prior Approvals
  • Authority

    CE Mark (EU MDR)

    Accepted

    Yes

    Benefit

    A device already registered in a reference country is what makes the ARCSA homologation route available, and the certificate of free sale plus that reference registration are the substance of the dossier. ARCSA publishes its periods per procedure, not per origin.

    Timeline Savings

    None published

  • Authority

    FDA 510(k)/PMA

    Accepted

    Yes

    Benefit

    Serves the same role as any other reference-country registration for the homologation route.

    Timeline Savings

    None published

Local Rep Required

Yes

Local Rep Name

Ecuadorian Authorized Representative

Mdsap Accepted

No

Mdsap Benefit

Not accepted

Registration Validity

5 years

Last Updated

2026-08-29

Verification Status

official-verified

Official Sources
  • Title

    ARCSA response times by trámite — device registration 1,216 working hours, recognition route 760 working hours

    Url

    https://www.controlsanitario.gob.ec/tiempos-de-respuesta-por-tramite-de-arcsa

    Accessed Date

    2026-08-29

  • Title

    Resolución ARCSA-DE-2026-003-DASP — substitute device regulation, art. 33 recognition-route clock, in force 2027-01-28

    Url

    https://www.controlsanitario.gob.ec/wp-content/plugins/download-monitor/download.php?id=15580&force=0

    Effective Date

    2027-01-28

    Accessed Date

    2026-08-29

Reviewed On

2026-08-29

Process
Template Id

default

Status

planning-template

Usage Rule

Maintained planning steps and estimates, not verified regulator deadlines.

Steps
  • Step

    1

    Title

    Pre-Submission Preparation

    Description

    Gather technical documentation, quality management system (QMS) records, and clinical data required by ARCSA. Appoint a Ecuadorian Authorized Representative.

    Duration

    1-3 months

    Icon

    📋

  • Step

    2

    Title

    Application Submission

    Description

    Submit complete registration dossier to ARCSA including device description, risk analysis, and testing reports.

    Duration

    1-2 weeks

    Icon

    📤

  • Step

    3

    Title

    Administrative Review

    Description

    ARCSA reviews application for completeness. May request additional documents or clarifications.

    Duration

    2-4 weeks

    Icon

    🔍

  • Step

    4

    Title

    Technical Evaluation

    Description

    In-depth technical assessment of device safety and performance. ARCSA may issue questions.

    Duration

    152 working days

    Icon

    ⚙️

  • Step

    5

    Title

    Registration Approval

    Description

    Upon successful review, ARCSA issues Ecuador market authorization certificate.

    Duration

    1-2 weeks

    Icon

    ✅

Government fee schedules · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Ecuador fees →
Country

Ecuador

Jurisdiction Id

ecuador

Country Code

EC

Authority

ARCSA

Low Risk Local
Min

678.3

Max

678.3

Currency

USD

Per Product

Yes

High Risk Local
Min

904.34

Max

904.34

Currency

USD

Per Product

Yes

Details
  • Id

    domestic-device-registration

    Item

    Domestic medical-device registration

    Currency

    USD

    Amount Local

    678.3

    Amount Local Formatted

    USD 678.30

    Frequency

    Per registration

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.gob.ec/arcsa/tramites/inscripcion-registro-sanitario-dispositivos-medicos-fabricacion-nacional-extranjera

    Source Title

    ARCSA trámite: inscripción registro sanitario de dispositivos médicos (nacional/extranjera)

    Verified On

    2026-08-25

  • Id

    foreign-device-registration

    Item

    Foreign medical-device registration

    Currency

    USD

    Amount Local

    904.34

    Amount Local Formatted

    USD 904.34

    Frequency

    Per registration

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.gob.ec/arcsa/tramites/inscripcion-registro-sanitario-dispositivos-medicos-fabricacion-nacional-extranjera

    Source Title

    ARCSA trámite: inscripción registro sanitario de dispositivos médicos (nacional/extranjera)

    Verified On

    2026-08-25

  • Id

    domestic-device-re-registration

    Item

    Domestic medical-device re-registration (reinscripción)

    Currency

    USD

    Amount Local

    135.66

    Amount Local Formatted

    USD 135.66

    Frequency

    Per re-registration; 20% of the registration amount

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.gob.ec/arcsa/tramites/reinscripcion-certificado-registro-sanitario-medicamentos-sintesis-quimica-medicamentos-biologicos-dispositivos-medicos-fabricacion-nacional-extranjera

    Source Title

    ARCSA trámite: reinscripción certificado registro sanitario

    Verified On

    2026-08-25

  • Id

    foreign-device-re-registration

    Item

    Foreign medical-device re-registration (reinscripción)

    Currency

    USD

    Amount Local

    180.87

    Amount Local Formatted

    USD 180.87

    Frequency

    Per re-registration; 20% of the registration amount

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.gob.ec/arcsa/tramites/reinscripcion-certificado-registro-sanitario-medicamentos-sintesis-quimica-medicamentos-biologicos-dispositivos-medicos-fabricacion-nacional-extranjera

    Source Title

    ARCSA trámite: reinscripción certificado registro sanitario

    Verified On

    2026-08-25

  • Id

    domestic-device-modification

    Item

    Domestic registration modification

    Currency

    USD

    Amount Local

    67.83

    Amount Local Formatted

    USD 67.83

    Frequency

    Per modification; 10% of the registration amount

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.gob.ec/arcsa/tramites/modificacion-registro-sanitario-dispositivos-medicos-fabricacion-nacional-extranjera

    Source Title

    ARCSA trámite: modificación registro sanitario de dispositivos médicos

    Verified On

    2026-08-25

  • Id

    foreign-device-modification

    Item

    Foreign registration modification

    Currency

    USD

    Amount Local

    90.43

    Amount Local Formatted

    USD 90.43

    Frequency

    Per modification; 10% of the registration amount

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.gob.ec/arcsa/tramites/modificacion-registro-sanitario-dispositivos-medicos-fabricacion-nacional-extranjera

    Source Title

    ARCSA trámite: modificación registro sanitario de dispositivos médicos

    Verified On

    2026-08-25

Notes
  • Amounts are published on the official ARCSA trámite pages (checked 25 August 2026).
  • Re-registration collects 20% and each modification event 10% of the fixed registration amount; notifications to the registration are free of charge.
  • Recheck ARCSA before payment because amounts follow the resolution in force.
Last Updated

2026-08-25

Verification Status

official-verified

Official Sources
  • Title

    ARCSA trámite: inscripción registro sanitario de dispositivos médicos (nacional/extranjera)

    Url

    https://www.gob.ec/arcsa/tramites/inscripcion-registro-sanitario-dispositivos-medicos-fabricacion-nacional-extranjera

    Accessed Date

    2026-08-25

  • Title

    ARCSA trámite: reinscripción certificado registro sanitario

    Url

    https://www.gob.ec/arcsa/tramites/reinscripcion-certificado-registro-sanitario-medicamentos-sintesis-quimica-medicamentos-biologicos-dispositivos-medicos-fabricacion-nacional-extranjera

    Accessed Date

    2026-08-25

  • Title

    ARCSA trámite: modificación registro sanitario de dispositivos médicos

    Url

    https://www.gob.ec/arcsa/tramites/modificacion-registro-sanitario-dispositivos-medicos-fabricacion-nacional-extranjera

    Accessed Date

    2026-08-25

Reviewed On

2026-08-25

Amount Status

exact-current

Range Available

Yes

Additional certifications and licences · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Ecuador certifications →
Slug

ecuador

Market

Ecuador

Confidence

low

Schemes
  • Id

    arcsa-electrical-emc-evidence

    Name

    ARCSA electrical/EMC evidence

    Category

    Medical Electrical Safety

    Status

    Included In Medical Registration

    Object

    Finished Device

    Note

    Art. 24 Resolución ARCSA-DE-2026-003-DASP: ISO 10993 biocompatibility studies only for invasive class III/IV devices (num. 2.1); IEC compliance certificate for electromedical or laboratory DMA/DMDIV at all risk levels (num. 2.7; Art. 49 requires IEC 60601/80601/ISO 80601 marking for biomedical equipment); IVD performance studies (precision, sensitivity, specificity) at all levels (num. 2.4); sterility certificate for sterile devices and stability study for expiry-dated devices.

    Source Urls
    • https://www.controlsanitario.gob.ec/wp-content/plugins/download-monitor/download.php?id=15580&force=0
    Source Binding

    reviewed

  • Id

    arcotel-terminal-equipment-homologation

    Name

    ARCOTEL terminal-equipment homologation

    Category

    Radio Spectrum

    Status

    Mandatory If Feature

    Object

    Listed Telecommunications Terminal

    Note

    Reglamento para homologación y certificación de equipos terminales de telecomunicaciones, Resolución 03-03-ARCOTEL-2017 of 10 May 2017, published in Registro Oficial No. 15 of 15 June 2017, as quoted on the official gob.ec ARCOTEL procedure. Art. 1 sets the requirements and procedures for homologating and certifying telecommunications terminal equipment that uses radio spectrum and connects to public networks, including revocation grounds and the application of fees. Art. 2 applies that reglamento to commercialisation and use of such terminals of any class, make and model, and to persons with or without enabling titles. The same official procedure defines the Certificado de Homologación as the ARCOTEL document that allows a terminal meeting the technical homologation norms to be commercialised and used in Ecuador and connected to telecommunications networks. Resolución 01-01-ARCOTEL-2021 later made the Art. 13 fee free; that later fee change is not re-read here.

    Source Binding

    unresolved

  • Id

    radiation-legal-metrology

    Name

    Radiation/legal metrology

    Category

    Ionizing Radiation

    Status

    Mandatory If Feature

    Object

    Product And Use Site

    Note

    Resolución ARCSA-DE-2026-003-DASP Art. 24 num. 2.6: active devices using ionizing radiation require an authorization or license from the competent nuclear-applications control authority in the applicant's name, at all risk levels. The nuclear-authority instrument itself was not opened in this pass.

    Source Binding

    unresolved

Official Sources
  • Authority

    Gob.ec / ARCOTEL

    Topic

    Official procedure quoting Resolución 03-03-ARCOTEL-2017 Arts. 1-2 homologation object, scope and certificate

    Url

    https://www.gob.ec/arcotel/tramites/atencion-solicitudes-homologacion-equipos-terminales-telecomunicaciones

  • Authority

    ARCOTEL

    Topic

    terminal-equipment homologation requirements

    Url

    https://www.arcotel.gob.ec/homologacion-requisitos-para-homologar-un-equipo-terminal-de-telecomunicaciones/

  • Authority

    ARCSA

    Topic

    Resolución ARCSA-DE-2026-003-DASP dossier testing and radiation requirements

    Url

    https://www.controlsanitario.gob.ec/wp-content/plugins/download-monitor/download.php?id=15580&force=0

Keep researching this market
View the Ecuador resource hub →
Compare regulatory data across all countries →
DJ Fang
Need to apply Ecuador's requirements to a product?
Confirm the current ARCSA pathway, evidence, and local-role details for the actual device before filing.
Pricing