Pure Global AI
TGA · structured requirements

🇦🇺 Australia IVD regulatory data

Explore the IVD requirements currently available for Australia, organized by topic and linked to official sources.

12 facts8 dimensions10 official sourceslatest verification 2026-10-21
Explore questions and sources →

Common questions about changing a registered IVD in Australia. Each question separates published rules from details still awaiting verification.

Change control

Does the event require internal documentation, notice, prior approval, transfer, or a new application?

Compare countries

DCR batching is limited to the same request

secondary.answers.dcr-batching
Fact-level verified

One DCR may cover up to ten ARTG entries only where the requested change is the same for each entry. This is distinct from listing additional entries in a sponsor-transfer form and from the conditional 215-dollar evidence-update tariff. Separate approval prerequisites still apply; these provisions do not establish every transfer-plus-name/brand/REF event as one filing.

Same kind and unchanged manufacturer responsibility

secondary.answers.same-kind-and-manufacturer
Fact-level verified

An ARTG variation must preserve the same kind of device: sponsor, manufacturer, GMDN, class and the applicable prescribed characteristics/UPI. For manufacturer name or address changes, TGA checks whether responsibility for design, production, packaging and labelling and the QMS remain the same. A same-entity office/name update with unchanged QMS can be accepted; changed control or QMS can require a new inclusion. A sponsor transfer is a different legal event.

Conditions
  • Map a brand or REF to actual public and dossier-held information; do not equate a commercial order code with UPI, model or UDI.
  • Retain the actual class and UPI criteria; the guidance allows selected UPI variations but does not remove the statutory same-kind test.
  • Unchanged QMS alone does not establish that a different responsible legal manufacturer is the same manufacturer for the statutory kind test. If that same-manufacturer criterion is no longer met, an ARTG variation cannot be used to bypass a new inclusion.

Dossier requirements

Which evidence, local deltas, formats, and signatures are required?

Compare countries

Separate optional export certificate after ARTG accuracy

secondary.answers.export-certificate
Fact-level verified

If the importing country needs a certificate, the sponsor/authorised agent may seek CFS for current Australian-supply ARTG devices or Export Certificate for export-only entries. Correct ARTG details before applying; the exports team does not amend them. The form needs sponsor/client and ARTG/GMDN details, declaration and the applicable payment; one form/one optional schedule per certificate. More than five devices need a schedule, not automatic rejection. TGA aims for ten working days; this is a separate export service, not ordinary change approval or automatic certificate reissue.

Conditions
  • The current webpage and January 2026 form differ on payment method/details; use the current operational instructions before submitting.

Transfer documents and manufacturer-evidence sequence

secondary.answers.sponsor-documents-sequence
Fact-level verified

The transfer form identifies both sponsors/client IDs, affected ARTG entries, event/date and declarations by authorised representatives; extra entries use an attached list. Submit after the transfer has occurred and identify pending variations. Manufacturer Evidence is not transferred with sponsorship: the new sponsor first obtains accepted evidence, then submits DCR to link it to the ARTG. That order is separate from the legal event transferring sponsorship.

IVD variation versus DCR

secondary.answers.variation-documents-ivd
Fact-level verified

For Class 1-3 IVDs, adding a device of the kind that meets the notification criteria uses the IVD Variation form and the manufacturer Australian DoC. Existing ARTG information/performance or intended-purpose updates use DCR; the specified combined additional-device/intended-purpose case includes the replacement-purpose document. Do not delete an active IVD unless supply has ceased. Class 4 IVDs have a single UPI and do not use that IVD variation form. Wait for approval before implementing a variation.

Economic-operator roles

Which manufacturer, representative, holder, applicant, importer, and distributor roles apply?

Compare countries

Sponsor transfer and same-person name change

secondary.answers.sponsor-transfer
Fact-level verified

Under Regulations 10F/10FA, qualifying sponsor succession/assignment and sponsor-name events are notified in the approved manner within three months. An agreed business/interest transfer carries the inclusion to the new sponsor when the event occurs; it is not a discretionary transfer by a foreign manufacturer. A same sponsor with unchanged ABN/ACN uses the name form; changed ABN/ACN requires new organisation identification and the applicable transfer process.

Conditions
  • A true new legal manufacturer is separately assessed under manufacturer/QMS and same-kind rules; do not substitute a local sponsor change for that assessment.

Fees

What regulator fee applies to this class, route, event, and date?

Compare countries

IVD variation fees by class and narrow exception

secondary.answers.variation-fee-ivd
Fact-level verified

Current Schedule 9 item 2A (d) sets AUD 1982 for Class 3/4 IVD variation and (e) AUD 1133 for other IVD variation, each excluding the (g) case and data evaluation. Item (g) statutorily covers a medical device meeting all three EU-transition/sole-evidence-update/same-evidence conditions at AUD 215 per ten entries or part. The summary table describes (g) as non-IVD; preserve that discrepancy rather than impose the summary exclusion over the statute.

Label, IFU & language

Which label and IFU elements, languages, symbols, and translation controls apply?

Compare countries

IVD manufacturer and sponsor label identity

secondary.answers.ivd-label
Fact-level verified

The shared medical-device provisions expressly cover IVDs: Essential Principle 13 requires the applicable manufacturer name/trading name, address, device identity, lot/date and English information. Regulation 10.2 also requires readily identifiable sponsor name/address; added sponsor labels must not adulterate the device or obscure manufacturer information. These are information/location requirements, not a general old-label stock grace.

Reliance & recognition

Which foreign decisions or evidence can be reused, with what effect and conditions?

Compare countries

Existing records include acceptance, restrictions and evidence reuse. The applicable product scope and conditions are preserved below; a reference does not by itself confer market access.

Australia approvals and evidence in other markets (49 records)

ArmeniaBrazil ANVISA · EU conformity assessment · US FDA · Health Canada · Singapore HSA · UK MHRA · South Korea MFDS · Japan MHLW / PMDA · NMPA · Australia TGA · Türkiye conformity assessment

Prior approval / evidence record

Import of an unregistered device requires proof of authorisation in the country of manufacture or in an IMDRF member country

reliance.origin-country-or-imdrf-authorisation-evidence-for-unregistered-import
Fact-level verified
Statement

A device that has no Armenian state registration can still be imported, but the import (conformity) certificate file must contain a document confirming that the device is authorised for use — state registration or free circulation — in its country of manufacture or in an IMDRF member country, or a reasoned reference to such a document, including a Declaration of Conformity with references to the relevant registers. This is the origin-country free-sale certificate requirement: a Turkish, Korean, Chinese or European exporter must produce its home-country registration/free-sale evidence (or an IMDRF-country equivalent) even though the device is not registered in Armenia. The document is mandatory paperwork; the Armenian expert organisation still examines the file and the Ministry of Health still decides.

Source Approval

Registration / free circulation in the country of manufacture, or in any IMDRF member country — US FDA, EU member states (CE), Japan PMDA, Korea MFDS, China NMPA, Canada, Australia, Brazil, Singapore, UK — evidenced by a registration certificate, free-sale certificate or a Declaration of Conformity with register references

Edge Type

Origin Country Evidence

Market Access Strength

Evidence Only

Regulatory Layer

Import Control

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • The obligation attaches to the import (conformity) certificate for devices without Armenian state registration; for devices that do have Armenian state registration the equivalent item is a copy of the Armenian registration certificate, and the origin-country evidence is not required.
  • One of the ten statutory grounds for refusing the import certificate is that the imported device has no Armenian state registration, except in the cases the Law allows (Article 47.2(7)(8)). The origin/IMDRF evidence does not cure that ground where registration was required; it is the supporting document for the unregistered-device import cases the Law permits.
  • The file also requires, for every batch, a quality conformity certificate or equivalent, and the instructions for use or a reference to their electronic version (Annex 2, para. 2(2)(b), (c)).
  • The document list was extended by Government Decree No. 1731-Ն of 12.10.2023 and No. 1619-Ն of 17.10.2024.

Language: Armenian. Verbatim, Government Decree No. 429-Ն of 30.03.2023, Annex No. 2, para. 2(2)(a): "պետական գրանցում չունեցող բժշկական արտադրատեսակների համար՝ ա. բժշկական արտադրատեսակի արտադրող երկրում կամ Բժշկական արտադրատեսակների կարգավորող մարմինների միջազգային համաժողովի (IMDRF) անդամ երկրներում կիրառման թույլտվությունը (պետական գրանցումը, ազատ կիրառումը) հավաստող փաստաթուղթ կամ դրա վերաբերյալ փաստարկված տեղեկանք, այդ թվում՝ բժշկական արտադրատեսակների համապատասխանության հայտարարագիր (Declaration of Conformity) համապատասխան գրանցամատյանների հղումով". Translation: "for medical devices without state registration: a. a document confirming authorisation for use (state registration, free circulation) in the country of manufacture of the medical device or in member countries of the International Medical Device Regulators Forum (IMDRF), or a reasoned reference thereto, including a Declaration of Conformity for the medical device with reference to the relevant registers".

ArmeniaBrazil ANVISA · EU conformity assessment · US FDA · Health Canada · Singapore HSA · UK MHRA · South Korea MFDS · Japan MHLW / PMDA · NMPA · Australia TGA · IMDRF member authorities · WHO

Prior approval / evidence record

Simplified national registration for a device registered in an IMDRF member country or prequalified by WHO

reliance.imdrf-member-or-who-prequalified-simplified-registration
Fact-level verified
Statement

Armenia registers medical devices under two national procedures: a general procedure and a simplified procedure. The simplified procedure applies to a device that is registered in a country that is a member of the International Medical Device Regulators Forum (IMDRF) or that is prequalified by the World Health Organization. IMDRF members include the United States (FDA), the EU member states (CE marking), Japan (PMDA), Korea (MFDS), China (NMPA), Canada, Australia, Brazil, Singapore and the United Kingdom, so a US FDA 510(k)/PMA clearance, an EU CE mark, a Japan PMDA approval and a Korea MFDS approval all satisfy the gate. The device is still registered in Armenia by the Armenian authority — the foreign registration opens the shortened procedure, it does not replace the Armenian certificate.

Source Approval

Registration in any IMDRF member country — US FDA 510(k)/PMA, EU CE marking under MDR/IVDR (any EU member state), Japan PMDA, Korea MFDS, China NMPA, Health Canada, Australia TGA, Brazil ANVISA, Singapore HSA, UK MHRA — or WHO prequalification of the device

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Premarket Review

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • The provision is written as an annex to part 3 of the same article: 'under the procedure laid down by part 3 of this article' devices are registered under the general and simplified procedures. Part 3 is itself the Armenia-only route, and a footnote to Article 46 states that part 3 operates only until the end of the transitional period set by Article 11 of the EAEU Agreement of 23 December 2014 (footnote inserted by Article 3(2) of the Law of 5 December 2024 ՀՕ-502-Ն). The transitional period was extended by the Protocol signed in Moscow on 29 December 2025, which replaced '2025' with '2027' in the first paragraph of Article 11 of that Agreement and '2026' with '2028' in the third and fourth paragraphs.
  • Bounded claim: the Law names IMDRF membership and WHO prequalification as the gate for the simplified procedure. It does not name the FDA, the CE mark, the PMDA or the MFDS; those follow from their countries' IMDRF membership, not from the text.
  • The operational content of the 'simplified procedure' (dossier list, clocks, certificate form) sits in the Government procedure that Article 46(3) contemplates; that instrument was not located in this session (see unresolved: registration-pathways).
  • Devices are registered by the Ministry of Health (the լիազոր մարմին / authorised body) with the examination carried out by the expert organisation designated by Government decision (Article 46(2), (5)).

Language: Armenian. Verbatim, Article 46(4): "Սույն հոդվածի 3-րդ մասով սահմանված կարգի համաձայն՝ բժշկական արտադրատեսակները գրանցվում են ընդհանուր և պարզեցված ընթացակարգերով: Պարզեցված ընթացակարգը կիրառվում է բժշկական արտադրատեսակների շրջանառությունը կարգավորող մարմինների միջազգային համաժողովի (IMDRF) անդամ երկրում գրանցված կամ Առողջապահության համաշխարհային կազմակերպության նախաորակավորած բժշկական արտադրատեսակների համար:" Translation: "Under the procedure laid down by part 3 of this article, medical devices are registered under general and simplified procedures. The simplified procedure applies to medical devices registered in a member country of the International Medical Device Regulators Forum (IMDRF) or prequalified by the World Health Organization." Article 46 was amended by the Law of 5 December 2024 ՀՕ-502-Ն; the consolidated expression is the version in force from 31.05.2026.

BahrainSaudi Arabia · US FDA · Australia TGA · Japan MHLW / PMDA · UK MHRA · Ireland · Swissmedic · France · Germany

Prior approval / evidence record

Registration certificate from SFDA (Saudi Arabia), US FDA, TGA (Australia), PMDA (Japan), MHRA (United Kingdom), HPRA (Ireland), Swissmedic (Switzerland), ANSM (France) or BfArM (Germany)

Authority

Registration certificate from SFDA (Saudi Arabia), US FDA, TGA (Australia), PMDA (Japan), MHRA (United Kingdom), HPRA (Ireland), Swissmedic (Switzerland), ANSM (France) or BfArM (Germany)

Accepted

Yes

Benefit

Article (7) bis of Resolution No. (48) of 2020, inserted by Resolution No. (69) of 2024, provides by way of exception to the ordinary Article (7) prohibition that medical devices and products registered in one of the nine listed reference countries 'shall also be registered in the Kingdom of Bahrain' on the basis of that country's registration certificate, on payment of the applicable fee. This is a documentary registration route rather than a full de novo technical assessment, but the application is still filed by a licensed Bahraini Authorized Representative and NHRA retains the decision.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

BahrainUS FDA · Japan MHLW / PMDA · Saudi Arabia · Australia TGA · UK MHRA · Ireland · Swissmedic · France · Germany

Prior approval / evidence record

Registration certificate from a listed reference country is an alternative basis for Bahraini registration (Art. 7 bis)

reliance.reference-country-registration-certificate-route
Fact-level verified
Statement

Article (7) bis of Resolution No. (48) of 2020, inserted by Resolution No. (69) of 2024, provides that by way of exception to the ordinary Article (7) prohibition, medical devices and products registered in one of nine listed reference countries 'shall also be registered in the Kingdom of Bahrain' on the basis of the registration certificate issued by that country, following payment of the applicable fee. The listed reference authorities are SFDA (Saudi Arabia), FDA (United States), TGA (Australia), PMDA (Japan), MHRA (United Kingdom), HPRA (Ireland), Swissmedic (Switzerland), ANSM (France) and BfArM (Germany). A holder of a US FDA or Japan PMDA registration therefore has a documentary registration route in Bahrain rather than a full de novo technical assessment.

Source Approval

US FDA registration/clearance; Japan PMDA approval; Saudi SFDA registration; Australia TGA registration; UK MHRA registration; Ireland HPRA registration; Swissmedic (Switzerland) registration; France ANSM registration; Germany BfArM registration

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Only the nine authorities named in Article (7) bis are listed; Korea MFDS, Health Canada and China NMPA are not among them
  • The EU CE mark as such is not listed; the Union route appears only through three named national authorities (HPRA, ANSM, BfArM)
  • The application is still made by a Bahraini licensed Authorized Representative and the applicable fee must be paid
  • In force from the day following publication in the Official Gazette (No. 3781, 21 November 2024)
  • Resolution No. (69) of 2024 replaced Article (10) of Resolution No. (48) of 2020 so that marketing authorisation runs for three years

Effective from 2024-11-22

Verbatim source passage: "By way of exception to Article (7) of this Resolution, medical devices and products registered in one of the following reference countries shall also be registered in the Kingdom of Bahrain: 1. Kingdom of Saudi Arabia - Represented by the Saudi Food and Drug Authority (SFDA). 2. United States of America - Represented by the Food and Drug Administration (FDA). 3. Australia - Represented by the Therapeutic Goods Administration (TGA). 4. Japan - Represented by the Pharmaceuticals and Medical Devices Agency (PMDA). 5. United Kingdom - Represented by the Medicines and Healthcare products Regulatory Agency (MHRA). 6. Ireland - Represented by the Health Products Regulatory Authority (HPRA). 7. Switzerland - Represented by Swissmedic(Swissmedic). 8. France - Represented by the French National Agency for Medicines and Health Products Safety (ANSM). 9. Germany - Represented by the Federal Institute for Drugs and Medical Devices (BfArM). Based on the registration certificate issued by one of these countries, following the payment of the applicable fee."

BahrainSaudi Arabia · US FDA · UK MHRA · Australia TGA · Health Canada · Japan MHLW / PMDA · Swissmedic · Ireland · Denmark · New Zealand · France · Netherlands · Belgium

Prior approval / evidence record

Free Sale Certificate may be issued by the country of origin or by a reference country

reliance.free-sale-certificate-origin-or-reference-country
Fact-level verified
Statement

At renewal, Bahrain requires a Free Sale Certificate (FSC) or Certificate to Foreign Government issued by the regulatory authority of the country of origin or of a reference country (Registration Guideline 9.1, section 10 item 4). The guideline's glossary (item 14) defines reference countries as Saudi Arabia, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland and Belgium, so a certificate from one of those authorities substitutes for an origin-country free-sale certificate. A Free Sale Certificate is also required on manufacturer acquisition (section 12).

Source Approval

Free Sale Certificate or Certificate to Foreign Government from the country of origin or from a reference country (Saudi Arabia, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland, Belgium)

Edge Type

Origin Country Evidence

Market Access Strength

Evidence Only

Regulatory Layer

Premarket Review

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • The classification guideline separately accepts a Free Sale Certificate OR Registration Certificate from one of the reference countries as a classification requirement
  • The reference-country list in the registration guideline glossary includes Canada, Denmark, the Netherlands ('Holland') and New Zealand, which are not in the Article (7) bis list of Resolution No. (69) of 2024, and omits Germany
  • Free Sale Certificate is also required on manufacturer acquisition (Section 12)

Verbatim source passages: "Free Sale Certificate (FSC) or Certificate to foreign government issued by the regulatory authority of the country of origin or a reference country." and "Reference countries: Saudi Arabia, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland, Belgium."

BahrainSaudi Arabia · US FDA · UK MHRA · Australia TGA · Health Canada · Japan MHLW / PMDA · Swissmedic · Ireland · Denmark · New Zealand · France · Netherlands · Belgium

Prior approval / evidence record

Classification of a product as a medical device may be evidenced by a registration certificate from a reference country

reliance.reference-country-certificate-for-classification
Fact-level verified
Statement

To obtain an NHRA classification letter, the applicant must supply a Free Sale Certificate or Registration Certificate issued by the competent authority of the country of origin or by one of the reference countries listed in the classification guideline (SFDA, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland, Belgium), together with a recognised quality assurance certificate such as ISO 13485, CE mark or FDA.

Source Approval

Registration certificate or Free Sale Certificate from SFDA (Saudi Arabia), US FDA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, New Zealand, France, Holland or Belgium

Edge Type

Origin Country Evidence

Market Access Strength

Evidence Only

Regulatory Layer

classification

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • A classification letter is a separate, fee-bearing NHRA service that precedes importation and registration
  • The reference-country list here is broader than the Article (7) bis list in Resolution No. (69) of 2024 and omits Germany and includes Canada, Denmark, the Netherlands and New Zealand
  • For class I products without CE marking a manufacturer's Declaration of Conformity may be provided

Verbatim source passage: "1. Free Sale Certificate OR Registration Certificate issued from competent authority in country of origin classifies the product as a medical device. 2. Free Sale Certificate OR Registration Certificate issued from one of the following reference countries: (SFDA, USA, UK, Australia, Canada, Japan, Switzerland, Ireland, Denmark, Newzland, France, and Holland, Belgium.)."

BotswanaEU conformity assessment · US FDA · Japan MHLW / PMDA · Swissmedic · Health Canada · Australia TGA · Iceland · Norway

Prior approval / evidence record

Stringent Regulatory Authority approval — ICH members (European Commission, US FDA, Japan MHLW/PMDA), ICH observers (Swissmedic, Health Canada) or an authority associated with an ICH member through a legally binding mutual recognition agreement (including Australia, Iceland, Liechtenstein and Norway)

Authority

Stringent Regulatory Authority approval — ICH members (European Commission, US FDA, Japan MHLW/PMDA), ICH observers (Swissmedic, Health Canada) or an authority associated with an ICH member through a legally binding mutual recognition agreement (including Australia, Iceland, Liechtenstein and Norway)

Accepted

No

Benefit

The SRA definition in G01 s.3.1.26 is used for dossier and reference purposes rather than as an automatic grant: an SRA approval does not by itself confer a Botswana registration, and it benefits the applicant only through the abridged pathway's proof-of-reference-agency requirement, which is defined by the separate Reference Regulatory Authority concept.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

BrazilUS FDA · Australia TGA · Health Canada · Japan MHLW / PMDA · Brazil ANVISA

Prior approval / evidence record

Optimized analysis (procedimento otimizado, IN 290/2024)

Name

Optimized analysis (procedimento otimizado, IN 290/2024)

Description

ANVISA uses the analysis already performed by a recognised equivalent foreign regulator as the sole or complementary reference for the registration decision. It shortens how much ANVISA has to re-review, not where the petition sits in the queue: Art. 10 states expressly that adopting the optimized procedure does not change the chronological order of petitions, and ANVISA publishes no reduced deadline for it.

Timeline Reduction

No published reduction — the RDC 743/2022 deadlines still apply and Art. 10 preserves queue position

Eligibility
  • Class III and IV primary registration petitions only, medical devices and IVDs (Art. 2)
  • Prior authorisation by one of the four recognised AREE: TGA via the ARTG, Health Canada via a Medical Device Licence, US FDA via 510(k), PMA or De Novo, or Japan MHLW via Shonin (Art. 6)
  • The device must be essentially identical to the one the AREE authorised, down to composition, indications, manufacturer and manufacturing process (Art. 5 II)
  • Excluded if the AREE itself granted the authorisation through an abbreviated route similar to ANVISA optimized analysis (Art. 4)
  • Filed as a secondary aditamento petition; omit it and the registration is analysed by the ordinary route (Art. 11)
  • ANVISA keeps the right to revert to ordinary analysis with reasons (Art. 9) and to decide differently from the AREE (Art. 13)

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

BrazilAustralia TGA

Prior approval / evidence record

TGA (Australia)

Authority

TGA (Australia)

Accepted

Yes

Benefit

Recognised AREE for IN 290/2024, evidenced by the ARTG entry

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

ColombiaUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA

Prior approval / evidence record

Class IIa

Id

class-iia

Class Name

Class IIa

Risk Level

medium

Pathway

Registration (NSO)

Official Timeline
Min

2

Max

2

Unit

working-days

Verification Status

official-verified

Official Source
Title

Decreto 4725 de 2005 — INVIMA statutory review clocks (Arts. 21-22)

Url

https://normograma.invima.gov.co/normograma/compilacion/docs/decreto_4725_2005.htm

Accessed Date

2026-08-25

Realistic Timeline
Min

1

Max

2

Unit

months

With Prior Approval
Min

2

Max

3

Unit

months

Prior Approval Benefits
  • EU CE Mark
  • FDA
  • Health Canada
  • Japan MHLW
  • Australia TGA
Notes
  • Decreto 4725/2005 Arts. 21-22: automatic registro administrative act within 2 working days once documents are complete (Classes I and IIa)
  • Reliance pathway available for recognized approvals
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

ColombiaUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA

Prior approval / evidence record

Class IIb

Id

class-iib

Class Name

Class IIb

Risk Level

high

Pathway

Registration (RSO)

Official Timeline
Min

90

Max

90

Unit

working-days

Verification Status

official-verified

Official Source
Title

Decreto 4725 de 2005 — INVIMA statutory review clocks (Arts. 21-22)

Url

https://normograma.invima.gov.co/normograma/compilacion/docs/decreto_4725_2005.htm

Accessed Date

2026-08-25

Realistic Timeline
Min

6

Max

12

Unit

months

With Prior Approval
Min

4

Max

6

Unit

months

Prior Approval Benefits
  • EU CE Mark
  • FDA
  • Health Canada
  • Japan MHLW
  • Australia TGA
Notes
  • Decreto 4725/2005: 90 working days for registro sanitario evaluation once documents are complete (Classes IIb and III)
  • Incomplete dossiers are rejected at filing (not suspended)
  • Technical documentation review required
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

ColombiaUS FDA · Australia TGA · Health Canada · Japan MHLW / PMDA

Prior approval / evidence record

Homologación for reference countries (Decreto 3275/2009)

Name

Homologación for reference countries (Decreto 3275/2009)

Description

Colombia does not run a separate reliance queue. What a reference-country approval buys is the right to skip specific numerals of the Formato Único: for Class I and IIa the technical studies and analytical checks, the sterilisation method and the disposal method; for Class IIb those three plus the clinical studies. Class III is outside the homologation and has to file all twenty numerals.

Timeline Reduction

None — the Decreto 4725/2005 clocks are unchanged; the saving is in dossier assembly

Eligibility
  • Device originates from a reference country: the European Community, the United States, Canada, Japan, Australia, or a country with a mutual recognition agreement
  • Risk class I, IIa or IIb — Class III devices file the complete Formato Único regardless of origin

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Dominican RepublicUS FDA · Health Canada · Japan MHLW / PMDA · Australia TGA · EU conformity assessment · Iceland · Norway

Prior approval / evidence record

WHO Strict Authorities — US FDA, Health Canada, Japan NIHS, Australia TGA, EMA, Icelandic Medicines Agency, Norwegian Medicines Agency, Liechtenstein AML

Authority

WHO Strict Authorities — US FDA, Health Canada, Japan NIHS, Australia TGA, EMA, Icelandic Medicines Agency, Norwegian Medicines Agency, Liechtenstein AML

Accepted

Yes

Benefit

Eligibility for the simplified registration procedure under Resolución No. 000004 on a Certificate of Free Sale plus a good-manufacturing-practices certificate from that authority, for new registrations and renewals of imported products; the remaining legal requirements still apply and DIGEMAPS retains the decision.

Timeline Savings

No published period: the MSP service listing shows 'Inmediatamente' rather than a number of days.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

EthiopiaUS FDA · Japan MHLW / PMDA · Australia TGA · Singapore HSA · EU conformity assessment · UK MHRA · Health Canada · South Korea MFDS

Prior approval / evidence record

Registration route for devices approved by a listed Stringent Regulatory Authority (FDA, EU, Japan, Australia, Singapore, UK, Canada, Korea)

reliance.sra-verification-route
Fact-level verified
Statement

A medical device that already holds a marketing authorisation, registration certificate or free-sale certificate from one of the authorities EFDA lists as a Stringent Regulatory Authority can be filed on the SRA route (EFDA/GDL/046 §9.4). The applicant still submits the complete administrative and technical dossier required by the ordinary registration guidelines, but EFDA reviews it as a verification exercise focused on the genuineness of the submitted information rather than a de-novo assessment, so the application takes shorter processing time. EFDA keeps the right to conduct a full assessment and to request further information whenever it deems necessary.

Source Approval

US FDA marketing authorisation / registration certificate / free-sale certificate; Japan MHLW; Australia TGA; Singapore HSA; a competent authority of one of the 27 EU Member States (Directive 93/42/EEC, Directive 98/79/EC, MDR 745/2017, IVDR 746/2017); UK MHRA; Health Canada; South Korea MFDS

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Premarket Review

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • The route is a shortened review, not a dossier waiver: 'The applicants should submit all documentation required for registration of medical devices through full assessment path way, but it takes shorter processing time as it will focus on verifiying genuinness of submitted information.'
  • The claim must be supported by a marketing authorisation certificate or registration certificate or free-sale certificate issued by the listed authority (or by WHO/UNFPA prequalification evidence).
  • EFDA determines eligibility for the SRA route after screening the eRIS submission; an application filed as SRA on misleading information is rejected at screening or evaluation.
  • EFDA may still carry out a full assessment and request further information where a foreign approval does not provide the same assurance for use in Ethiopia.
  • The SRA list is published in GDL/046 §9.4 and 'will be updated as deemed necessary'.
  • The same list also appears (without Singapore HSA, UK MHRA and Korea MFDS) in the non-IVD and IVD registration requirement guidelines, which state that registration with those authorities is 'considered to be registered with Stringent Regulatory Authority approved devices registration procedures'.

Effective from 2026-06-25

Verbatim: 'At present, EFDA accepts the registration submissions as an SRA or prequalified products application if they're supported by one or more of the following document(s). 1. A Marketing Authorization certificate or registration certificate or free sale certificate issued by one of the following competent national regulatory authority: US Food and Drug Administration; Ministry of Health, Labour and Welfare, Japan; Trapeutic Goods Administration, Australia; Health Science Authority, Singapore; Competent Authorities from one of the 27 Member States of the European Union who are responsible in Europe for the oversight of Directive 93/42/EEC, Directive 98/79/EC, MDR 745/2017, IVDR 746/2017.; Medicine and Healthcare Products Regulatory Agency, UK; Health Canada, Canada; Ministry of Food and Drug Safety, South Korea'. The passage also states: 'The applicants should submit all documentation required for registration of medical devices through full assessment path way, but it takes shorter processing time as it will focus on verifiying genuinness of submitted information.' Typographical errors are reproduced as they appear in the source.

IndiaUS FDA · EU conformity assessment · Australia TGA · Japan MHLW / PMDA · GHTF

Prior approval / evidence record

Class A

Id

class-a

Class Name

Class A

Risk Level

low

Pathway

Registration

Official Timeline
Min

45

Max

45

Unit

days

Verification Status

official-verified

Official Source
Title

CDSCO Medical Devices Rules, 2017 (G.S.R. 78(E))

Url

https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2022/m_device/Medical%20Devices%20Rules%2C%202017.pdf

Accessed Date

2026-08-25

Realistic Timeline
Min

6

Max

9

Unit

months

With Prior Approval
Min

3

Max

5

Unit

months

Prior Approval Benefits
  • GHTF country approval (FDA, EU, Japan, Australia, UK)
Notes
  • Licence granted or rejected within 45 days of application (Rule 20(4)); Notified Body audit happens within 120 days after licence grant, not before (Rule 20(4) proviso (ii))
  • GHTF approval provides significant advantage
Clinical Trial Required

No

Local Testing Required

Yes

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IndiaAustralia TGA

Prior approval / evidence record

Australia TGA

Authority

Australia TGA

Accepted

Yes

Benefit

Named in Rule 36(3) and in the Rule 63(1) proviso, so both the clinical investigation waiver and the no-predicate relief are available.

Timeline Savings

None published — the saving is the clinical investigation, not the clock

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

IsraelUS FDA · UK MHRA · Germany · France · Italy · Spain · Netherlands · Belgium · Ireland · Denmark · Sweden · Finland · Norway · Iceland · Greece · Portugal · Austria · Swissmedic · Australia TGA · Health Canada · New Zealand

Regulatory reference

Recognized-state registration route

reliance.recognized-country-route
Fact-level verified

A device registered, or approved for marketing and actually marketed, in a 'recognized state' (מדינה מוכרת) is registered in the Israeli register for a period not exceeding the period it may be marketed in that state, capped at 5 years from Israeli registration. Recognized states are listed in the First Addendum to the law — 21 states including the US, UK, Germany, France, Italy, Spain, the Netherlands, Belgium, Ireland, Denmark, Sweden, Finland, Norway, Iceland, Greece, Portugal, Austria, Switzerland, Australia, Canada and New Zealand (no single 'EU' entry: member states are listed individually).

Conditions
  • Device must be registered or permitted for marketing and actually marketed in the recognized state.

First Addendum list per the official Sefer HaChukim text of the law held from gov.il.

IsraelAustralia TGA · Health Canada

Prior approval / evidence record

Health Canada or Australia TGA

Authority

Health Canada or Australia TGA

Accepted

Yes

Benefit

REG-2024/06 restricts the regulatory certificate types that can be entered on an application to CE, FDA, FDA K510, Australian Ministry of Health and Canadian Ministry of Health, so a Canadian or Australian approval is a first-class reference certificate on the same terms as a US or EU one.

Timeline Savings

Same track placement as the US and EU certificates above

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

KuwaitHealth Canada · Japan MHLW / PMDA · UK MHRA · Brazil ANVISA · Australia TGA

Prior approval / evidence record

Health Canada, Japan MHLW/PMDA, UK MHRA, Brazil ANVISA and Australia TGA approvals

Authority

Health Canada, Japan MHLW/PMDA, UK MHRA, Brazil ANVISA and Australia TGA approvals

Accepted

Yes

Benefit

Also recognised reference authorities under the Abridged Review Pathway, on the same terms as the FDA and EU routes: identical product required, national requirements still assessed, final acceptance decided by the Administration.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

LebanonIMDRF member authorities

Prior approval / evidence record

GHTF/IMDRF member countries — free-sale certificate (شهادة بيع حر)

Authority

GHTF/IMDRF member countries — free-sale certificate (شهادة بيع حر)

Accepted

Yes

Benefit

A mandatory document rather than a route: Decision 1506/1 Art. Four requires the registration application to carry a free-sale certificate issued by a country belonging to the GHTF/IMDRF group, meaning the device is sold in one of those countries whatever its origin, and the Ministry's service records require a free-sale certificate with an ISO 13485 certificate where the goods are made outside the reference states or the listed documents cannot be produced. No examination period is shortened by it.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

LebanonUS FDA · EU conformity assessment · Swissmedic · Australia TGA · Health Canada · Japan MHLW / PMDA

Prior approval / evidence record

US FDA (510(k)/PMA approval or approval letter), European Union (conformity certificate issued by the manufacturer and by EU-notified bodies), Swiss, Australian, Canadian or Japanese competent health authorities (conformity certificate)

Authority

US FDA (510(k)/PMA approval or approval letter), European Union (conformity certificate issued by the manufacturer and by EU-notified bodies), Swiss, Australian, Canadian or Japanese competent health authorities (conformity certificate)

Accepted

Yes

Benefit

Accepted and in practice required for an implantable device: Decision 455/1 Art. Six lists them as the conformity documents that go into the registration file — a European-standards certificate from the manufacturer and EU-notified bodies, an FDA marketing approval or approval letter, or a conformity certificate from the health authorities of Switzerland, Australia, Canada or Japan — and Decision 1506/1 Art. Four requires a GHTF/IMDRF free-sale certificate on top. The Ministry's service records for sterile instruments and laboratory reagents require FDA evidence for US-origin goods and CE evidence for European ones. The benefit is documentary, not a shortened clock: none of the instruments publishes an abbreviated examination period, and the registration certificate is still issued on the Health Technology Unit's advice to the Minister.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

MalaysiaUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA · UK MHRA · Singapore HSA · TFDA (jurisdiction not specified in this entry)

Prior approval / evidence record

Class B

Id

class-b

Class Name

Class B

Risk Level

medium

Pathway

Registration

Official Timeline
Min

30

Max

30

Unit

working-days

Verification Status

official-verified

Official Source
Title

MDA/GD/0070 — registration submission guide, cl. 6.4 turnaround

Url

https://www.mda.gov.my/index.php/documents/ukk/3766-final-medical-device-registration-submission-guide-for-conformity-assessment-by-way-of-verification-process-and-submission-of-application-in-medc-st-second-edition

Accessed Date

2026-08-25

Realistic Timeline
Min

3

Max

6

Unit

months

With Prior Approval
Min

3

Max

6

Unit

months

Prior Approval Benefits
  • EU CE Mark
  • MHLW
  • TGA
  • Health Canada
  • FDA
  • MHRA
  • HSA
  • TFDA
Notes
  • MDA/GD/0070 cl. 6.4: Class B registration turn-around 30 working days from complete documents and cleared payment
  • ASEAN harmonization applies
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

MalaysiaUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA · UK MHRA · Singapore HSA · TFDA (jurisdiction not specified in this entry)

Prior approval / evidence record

Class C

Id

class-c

Class Name

Class C

Risk Level

high

Pathway

Registration

Official Timeline
Min

30

Max

30

Unit

working-days

Verification Status

official-verified

Official Source
Title

MDA/GD/0070 — registration submission guide, cl. 6.4 turnaround

Url

https://www.mda.gov.my/index.php/documents/ukk/3766-final-medical-device-registration-submission-guide-for-conformity-assessment-by-way-of-verification-process-and-submission-of-application-in-medc-st-second-edition

Accessed Date

2026-08-25

Realistic Timeline
Min

3

Max

6

Unit

months

With Prior Approval
Min

3

Max

6

Unit

months

Prior Approval Benefits
  • EU CE Mark
  • MHLW
  • TGA
  • Health Canada
  • FDA
  • MHRA
  • HSA
  • TFDA
Notes
  • MDA/GD/0070 cl. 6.4: Class C registration turn-around 30 working days from complete documents and cleared payment
  • Clinical data may be required
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

MalaysiaUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA · UK MHRA · Singapore HSA · TFDA (jurisdiction not specified in this entry)

Prior approval / evidence record

Class D

Id

class-d

Class Name

Class D

Risk Level

highest

Pathway

Full Registration

Official Timeline
Min

30

Max

30

Unit

working-days

Verification Status

official-verified

Official Source
Title

MDA/GD/0070 — registration submission guide, cl. 6.4 turnaround

Url

https://www.mda.gov.my/index.php/documents/ukk/3766-final-medical-device-registration-submission-guide-for-conformity-assessment-by-way-of-verification-process-and-submission-of-application-in-medc-st-second-edition

Accessed Date

2026-08-25

Realistic Timeline
Min

3

Max

6

Unit

months

With Prior Approval
Min

3

Max

6

Unit

months

Prior Approval Benefits
  • EU CE Mark
  • MHLW
  • TGA
  • Health Canada
  • FDA
  • MHRA
  • HSA
  • TFDA
Notes
  • MDA/GD/0070 cl. 6.4: Class D registration turn-around 30 working days from complete documents and cleared payment
  • Clinical evidence required
Clinical Trial Required

Yes

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

MexicoNMPA · US FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA · UK MHRA · Swissmedic · South Korea MFDS · Singapore HSA · Brazil ANVISA · IMDRF member authorities · MDSAP · WHO

Prior approval / evidence record

Vía regulatoria abreviada — ARR equivalence (COFEPRIS-04-050)

Name

Vía regulatoria abreviada — ARR equivalence (COFEPRIS-04-050)

Description

The single reliance route since the DOF Acuerdo of 18 July 2025, which took effect 30 working days after publication and abrogated every individual equivalence agreement issued between 2012 and June 2025 — the separate FDA, Swissmedic, European Commission, Health Canada, TGA and WHO prequalification acuerdos no longer exist. COFEPRIS now recognises the decision of any Reference Regulatory Authority from a country sitting on the IMDRF Management Committee, and MDSAP full membership for GMP certificates. The dossier drops to an administrative-legal section plus technical information under RIS Art. 31.

Timeline Reduction

Resolution within a maximum of 30 working days

Eligibility
  • Prior authorisation by a Reference Regulatory Authority in one of the twelve IMDRF Management Committee jurisdictions: Australia, Brazil, Canada, China, the European Union, Japan, Russia, Singapore, South Korea, Switzerland, the United Kingdom, the United States
  • MDSAP full members are recognised for the good-manufacturing-practice certificate
  • Filed under homoclave COFEPRIS-04-050 rather than the ordinary COFEPRIS-04-060 registration
  • Post-market surveillance or technovigilance evidence from the reference market, translated into Spanish under RIS Art. 153

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

MexicoUS FDA · Health Canada · Japan MHLW / PMDA · IMDRF member authorities · MDSAP

Prior approval / evidence record

Class II

Id

class-ii

Class Name

Class II

Risk Level

medium

Pathway

Registration or Equivalency

Official Timeline
Min

35

Max

35

Unit

days

Verification Status

official-verified

Official Source
Title

RIS Art. 179 — statutory resolution periods by class (COFEPRIS)

Url

https://salud.gob.mx/unidades/cdi/nom/compi/ris.html

Accessed Date

2026-08-25

Realistic Timeline
Min

15

Max

31

Unit

months

With Prior Approval
Min

8

Max

16

Unit

months

Prior Approval Benefits
  • FDA 510(k)
  • Health Canada
  • Japan PMDA
  • IMDRF Committee Members
  • MDSAP
Notes
  • RIS Art. 179: Clase II resolved within 35 calendar days (15 days with a favourable Tercero Autorizado dictamen)
  • ARR equivalence route (COFEPRIS-04-050, Lineamientos DOF 18-07-2025): maximum 30 working days
  • Equivalency route much faster if FDA approved
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

MexicoUS FDA · Health Canada · Japan MHLW / PMDA · IMDRF member authorities · MDSAP

Prior approval / evidence record

Class III

Id

class-iii

Class Name

Class III

Risk Level

highest

Pathway

Registration or Equivalency

Official Timeline
Min

60

Max

60

Unit

days

Verification Status

official-verified

Official Source
Title

RIS Art. 179 — statutory resolution periods by class (COFEPRIS)

Url

https://salud.gob.mx/unidades/cdi/nom/compi/ris.html

Accessed Date

2026-08-25

Realistic Timeline
Min

15

Max

31

Unit

months

With Prior Approval
Min

8

Max

16

Unit

months

Prior Approval Benefits
  • FDA 510(k)/PMA
  • Health Canada
  • Japan PMDA
  • IMDRF Committee Members
  • MDSAP
Notes
  • RIS Art. 179: Clase III resolved within 60 calendar days (15 days with a favourable Tercero Autorizado dictamen)
  • ARR equivalence route (COFEPRIS-04-050): maximum 30 working days
  • Equivalency pathway best option with FDA
Clinical Trial Required

Yes

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

ParaguayBrazil ANVISA · US FDA · Health Canada · Austria · Belgium · Denmark · Finland · France · Germany · Hungary · Ireland · Italy · Netherlands · New Zealand · Norway · Spain · Sweden · Japan MHLW / PMDA · UK MHRA · Swissmedic · Australia TGA · WHO

Prior approval / evidence record

IVD Annex I reference-country list drives the 30/45/120-business-day clocks

reliance.ivd-anexo-i-reference-country-list
Fact-level verified
Statement

For IVDs, DINAVISA decides within 30 business days for products manufactured nationally or in a country of Annex I, within 45 business days for products manufactured outside Annex I but registered and commercialised in an Annex I country (which must additionally file the foreign registration certificate and free-sale certificate), and within 120 business days for other imports. Annex I names high-surveillance and adequate-surveillance authorities and the full MERCOSUR members; the United States and Japan are on it and Korea is not.

Source Approval

Sanitary registration and free-sale certificate in force in an Annex I country — high-surveillance authorities (Germany, Austria, Belgium, Canada, Denmark, Spain, United States, France, Israel, Italy, Japan, Netherlands, United Kingdom, Sweden, Switzerland) or adequate-surveillance authorities (Australia, Chile, Cuba, Finland, Hungary, Ireland, Luxembourg, Mexico, Norway, New Zealand); Annex I also lists full MERCOSUR members (Argentina, Brazil, Paraguay, Uruguay) and PAHO/WHO reference authorities

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Premarket Review

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • The Annex I list derives from Ley N° 3283/2007 Art. 11 (high and adequate sanitary surveillance).
  • A product made outside Annex I is admitted on a 45-day clock only if it is registered and commercialised in an Annex I country.
  • Korea (MFDS) does not appear in Annex I, so a Korean-only IVD authorisation does not by itself reach the 30/45-day clocks.
  • A Paraguayan sanitary registration is still required in every case.

Source language: Spanish. Translation of the quoted Annex I heading and entries: 'Countries with High and Adequate Sanitary Surveillance Authorities (Law 3283/07 Article 11)' / 'Countries with high sanitary surveillance regulatory authorities: Germany, Austria, Belgium, Canada, Denmark, Spain, United States, France, Israel, Italy, Japan, Netherlands, United Kingdom, Sweden and Switzerland.' / 'Full MERCOSUR members: Argentina, Brazil, Paraguay and Uruguay.' The excerpt is taken from the resolution's own Annex I; Art. 6 supplies the 30- and 45-business-day clocks.

ParaguayWHO · IMDRF member authorities

Prior approval / evidence record

PAHO/WHO reference authority or IMDRF Management Committee member authority

Authority

PAHO/WHO reference authority or IMDRF Management Committee member authority

Accepted

Yes

Benefit

Operative for class II, III and IV IVDs, which enter the Proceso Simplificado de Registro Sanitario under Resolución DINAVISA N° 44/2024 with a 15-business-day validation and approval by positive silence.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

ParaguayBrazil ANVISA · EU conformity assessment · US FDA · Health Canada · Singapore HSA · IMDRF member authorities · Portugal · Spain · South Korea MFDS · Japan MHLW / PMDA · UK MHRA · NMPA · Taiwan TFDA · Australia TGA · WHO

Prior approval / evidence record

IVD simplified registration (PSR) for products authorised by PAHO/WHO, IMDRF or bilateral-partner authorities

reliance.ivd-psr-reference-authorities
Fact-level verified
Statement

Class II, III and IV in-vitro diagnostic products already authorised and commercialised in countries regulated by PAHO/WHO reference regulatory authorities, by authorities of IMDRF Management Committee member countries, or by authorities with bilateral agreements with DINAVISA enter the Proceso Simplificado de Registro Sanitario (PSR). DINAVISA has 15 business days to validate and decide, and approval follows by positive silence if no report is issued.

Source Approval

PAHO/WHO reference regulatory authorities; authorities of IMDRF Management Committee member countries (the IMDRF committee includes US FDA, EU, Japan PMDA/MHLW, Health Canada, TGA, ANVISA, China NMPA, Korea MFDS, Singapore HSA and UK MHRA); authorities with bilateral agreements with DINAVISA

Edge Type

Abridged Or Recognition Route

Market Access Strength

facilitated

Regulatory Layer

Premarket Review

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Applies to class II, III and IV IVDs (Res. 44/2024 Art. 7: 'Clases II, III y IV'); class I IVDs are registered automatically under Res. 266/2022 Art. 14.
  • The IVD must actually be commercialised in one of those countries and the foreign registration must correspond specifically to the same product being registered.
  • 15 business days to validate; approval by positive silence.
  • Objections may be answered only once (Art. 3), with 30 business days to rectify.
  • Not applicable to the cases in Arts. 13 and 16 of Res. DINAVISA N° 266/2022 (Art. 11).

Effective from 2024-01-30

Source language: Spanish. Translation of the operative part of Art. 2: 'Applications for sanitary registration of In Vitro Diagnostic Products ... that are authorised and commercialised in countries regulated by: PAHO/WHO Reference Regulatory Authorities; Regulatory Authorities of member countries of the Committee of the International Medical Device Regulators Forum - IMDRF; Regulatory Authorities with which bilateral agreements with DINAVISA exist, shall be subject to the Simplified Sanitary Registration Process (PSR).' and 'DINAVISA shall have a term of 15 (fifteen) business days to validate the required documentation and decide thereon. Once the term has elapsed with no report from DINAVISA, the application shall be approved.' Article numbering confirmed by reading the page image of the official PDF.

ParaguayWHO · IMDRF member authorities

Prior approval / evidence record

Proceso simplificado de Registro Sanitario (PSR) para productos para diagnóstico de uso in vitro

Name

Proceso simplificado de Registro Sanitario (PSR) para productos para diagnóstico de uso in vitro

Description

Class II, III and IV IVDs (Res. 44/2024 Art. 7) already authorised and commercialised by a recognised reference authority are validated in 15 business days with approval by positive silence, instead of the 30, 45 or 120 business days of the ordinary IVD route.

Timeline Reduction

From 30-120 business days down to 15 business days

Eligibility
  • Product classified as class II, III or IV under Resolución DINAVISA N° 266/2022 Art. 9
  • Current authorisation and commercialisation in a country regulated by a PAHO/WHO reference regulatory authority, by an IMDRF Management Committee member authority, or by an authority with a bilateral agreement with DINAVISA
  • The foreign registration must correspond specifically to the same product being registered in Paraguay

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

PeruUS FDA · UK MHRA · Health Canada · Japan MHLW / PMDA · Australia TGA · Swissmedic · South Korea MFDS · Germany · France · Netherlands · Spain · Italy · Belgium · Portugal · Ireland · Austria · Hungary · Denmark · Norway · Sweden

Prior approval / evidence record

País de Alta Vigilancia Sanitaria (art. 9 D.S. 016-2011-SA): United States, Canada, Japan, Australia, Switzerland, South Korea, United Kingdom, Germany, France, Netherlands, Spain, Italy, Belgium, Portugal, Ireland, Austria, Hungary, Denmark, Norway, Sweden

Authority

País de Alta Vigilancia Sanitaria (art. 9 D.S. 016-2011-SA): United States, Canada, Japan, Australia, Switzerland, South Korea, United Kingdom, Germany, France, Netherlands, Spain, Italy, Belgium, Portugal, Ireland, Austria, Hungary, Denmark, Norway, Sweden

Accepted

Yes

Benefit

The Peruvian route to relying on a foreign approval is documentary, not procedural. A certificate of free sale or a GMP certificate issued by a high-surveillance authority is accepted directly, and if the issuing authority appears on the Resolución Directoral 169-2014-DIGEMID list the document does not need an apostille. Everything else still goes through DIGEMID certification.

Timeline Savings

None published — the TUPA period is set by class, not by origin

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

SingaporeUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA

Prior approval / evidence record

Class C

Id

class-c

Class Name

Class C

Risk Level

high

Pathway

IBR/Expedited/Full Route

Official Timeline
Min

0

Max

220

Unit

working-days

Verification Status

official-verified

Official Source
Title

HSA fees and turnaround time for medical devices

Url

https://www.hsa.gov.sg/medical-devices/fees/

Effective Date

2026-03-26

Accessed Date

2026-08-13

Realistic Timeline
Min

6

Max

12

Unit

months

With Prior Approval
Min

0.03

Max

0.03

Unit

months

Prior Approval Benefits
  • TGA
  • Health Canada
  • MHLW
  • FDA
  • EU NB
Notes
  • Five Class C routes: an immediate route confined to standalone medical mobile applications, expedited (ECR) at 120 working days, abridged at 160, full at 220, and full under the Priority Review Scheme at 143
  • ECR-1 needs one reference approval plus three years of marketing there; ECR-2 needs two reference approvals. Both require no prior rejection or withdrawal by HSA or a reference agency
  • Hip, knee and shoulder joint replacement non-bioactive implants are barred from ECR and must go full or abridged
  • HSA turnaround times exclude applicant stop-clock time
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

SingaporeUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA

Prior approval / evidence record

Class B

Id

class-b

Class Name

Class B

Risk Level

medium

Pathway

IBR/Expedited/Full Route

Official Timeline
Min

0

Max

160

Unit

working-days

Verification Status

official-verified

Official Source
Title

HSA fees and turnaround time for medical devices

Url

https://www.hsa.gov.sg/medical-devices/fees/

Effective Date

2026-03-26

Accessed Date

2026-08-13

Realistic Timeline
Min

3

Max

9

Unit

months

With Prior Approval
Min

0.03

Max

0.03

Unit

months

Prior Approval Benefits
  • TGA
  • Health Canada
  • MHLW
  • FDA
  • EU NB
Notes
  • Four Class B routes, and which one applies is decided by how many reference approvals the device already holds rather than by choice: Immediate Class B Registration is granted on submission, abridged takes 100 working days, full takes 160, and full under the Priority Review Scheme takes 104
  • IBR needs either one reference approval plus three years of marketing in that jurisdiction, or two reference approvals, and in both cases no global safety issue and no prior rejection or withdrawal. A standalone medical mobile application qualifies for an immediate route on one reference approval with no marketing history
  • Abridged needs at least one reference approval; a device with none must go full route
  • HSA turnaround times exclude applicant stop-clock time
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

SingaporeUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA

Prior approval / evidence record

Two reference approvals (any two of US FDA, EU Notified Body, Health Canada, Australia TGA, Japan MHLW)

Authority

Two reference approvals (any two of US FDA, EU Notified Body, Health Canada, Australia TGA, Japan MHLW)

Accepted

Yes

Benefit

At Class B this is the IBR trigger and registration is granted on submission. At Class C and D it is the ECR-2 or EDR trigger.

Timeline Savings

Class B immediate against 160 working days; Class C 120 against 220; Class D 180 against 310

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

SingaporeAustralia TGA · Health Canada · Japan MHLW / PMDA

Prior approval / evidence record

Health Canada, Australia TGA, Japan MHLW

Authority

Health Canada, Australia TGA, Japan MHLW

Accepted

Yes

Benefit

A Canadian or Australian device registration licence, or a Japanese Ninsho certification or Shonin approval, counts as a reference approval on the same terms as the US and EU.

Timeline Savings

Same as the abridged route figures above; two of them together trigger IBR, ECR-2 or EDR

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

SingaporeUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA

Prior approval / evidence record

Class D

Id

class-d

Class Name

Class D

Risk Level

highest

Pathway

IBR/Expedited/Full Route

Official Timeline
Min

180

Max

310

Unit

working-days

Verification Status

official-verified

Official Source
Title

HSA fees and turnaround time for medical devices

Url

https://www.hsa.gov.sg/medical-devices/fees/

Effective Date

2026-03-26

Accessed Date

2026-08-13

Realistic Timeline
Min

9

Max

18

Unit

months

With Prior Approval
Min

6

Max

12

Unit

months

Prior Approval Benefits
  • TGA
  • Health Canada
  • MHLW
  • FDA
  • EU NB
Notes
  • Four Class D routes: expedited (EDR) at 180 working days, abridged at 220, full at 310, and full under the Priority Review Scheme at 202. There is no immediate route at Class D
  • EDR needs two reference approvals and no prior rejection or withdrawal. Active implantables, devices contacting the central circulatory or central nervous system, bioactive joint replacements, devices with a registrable drug in a secondary role, and HIV or donor-compatibility IVD assays are barred from EDR
  • A Class D device with a registrable drug in a secondary role has only abridged at 220 and full at 310, and is excluded from the Priority Review Scheme
  • HSA turnaround times exclude applicant stop-clock time
Clinical Trial Required

Yes

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

ThailandUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA · WHO

Prior approval / evidence record

Class 3

Id

class-3

Class Name

Class 3

Risk Level

high

Pathway

Registration

Official Timeline
Min

250

Max

250

Unit

days

Verification Status

official-verified

Official Source
Title

Thailand Ministerial Regulation B.E. 2563 — statutory review clocks (ข้อ ๕-๖)

Url

https://medical.fda.moph.go.th/media.php?id=478924090456023040&name=กฏกระทรวงการแจ้งรายการละเอียดและการการออกใบรับแจ้งรายการละเอียดผลิตหรือนำเข้าเครื่องมือแพทย์ พ.ศ. 2563.pdf

Accessed Date

2026-08-25

Realistic Timeline
Min

4

Max

5

Unit

months

Prior Approval Benefits
  • FDA
  • EU CE Mark
  • Japan
  • TGA
  • Health Canada
  • WHO (for IVD)
Notes
  • Statutory cap: class-3 registration decided within 250 days (Ministerial Regulation B.E. 2563 ข้อ ๖)
  • Fee payment due within 60 days of the invoice; submission receipt issued within 7 days of a complete file
  • Clinical data may be required
Clinical Trial Required

No

Local Testing Required

Yes

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

ThailandUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA · WHO

Prior approval / evidence record

Class 2

Id

class-2

Class Name

Class 2

Risk Level

medium

Pathway

Registration

Official Timeline
Min

250

Max

250

Unit

days

Verification Status

official-verified

Official Source
Title

Thailand Ministerial Regulation B.E. 2563 — statutory review clocks (ข้อ ๕-๖)

Url

https://medical.fda.moph.go.th/media.php?id=478924090456023040&name=กฏกระทรวงการแจ้งรายการละเอียดและการการออกใบรับแจ้งรายการละเอียดผลิตหรือนำเข้าเครื่องมือแพทย์ พ.ศ. 2563.pdf

Accessed Date

2026-08-25

Realistic Timeline
Min

4

Max

5

Unit

months

Prior Approval Benefits
  • FDA
  • EU CE Mark
  • Japan
  • TGA
  • Health Canada
  • WHO (for IVD)
Notes
  • Statutory cap: class-2 registration decided within 250 days (Ministerial Regulation B.E. 2563 ข้อ ๖)
  • Fee payment due within 60 days of the invoice; submission receipt issued within 7 days of a complete file
  • ASEAN harmonization applies
Clinical Trial Required

No

Local Testing Required

Yes

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

ThailandUS FDA · EU conformity assessment · Australia TGA · Health Canada · Japan MHLW / PMDA · WHO

Prior approval / evidence record

Class 4

Id

class-4

Class Name

Class 4

Risk Level

highest

Pathway

Full Registration

Official Timeline
Min

300

Max

300

Unit

days

Verification Status

official-verified

Official Source
Title

Thailand Ministerial Regulation B.E. 2563 — statutory review clocks (ข้อ ๕-๖)

Url

https://medical.fda.moph.go.th/media.php?id=478924090456023040&name=กฏกระทรวงการแจ้งรายการละเอียดและการการออกใบรับแจ้งรายการละเอียดผลิตหรือนำเข้าเครื่องมือแพทย์ พ.ศ. 2563.pdf

Accessed Date

2026-08-25

Realistic Timeline
Min

4

Max

5

Unit

months

Prior Approval Benefits
  • FDA
  • EU CE Mark
  • Japan
  • TGA
  • Health Canada
  • WHO (for IVD)
Notes
  • Statutory cap: class-4 registration decided within 300 days (Ministerial Regulation B.E. 2563 ข้อ ๖)
  • Fee payment due within 60 days of the invoice; submission receipt issued within 7 days of a complete file
  • Clinical evidence required
Clinical Trial Required

Yes

Local Testing Required

Yes

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

United Arab EmiratesAustralia TGA · Health Canada

Prior approval / evidence record

Health Canada or Australia TGA

Authority

Health Canada or Australia TGA

Accepted

Yes

Benefit

A Canadian or Australian registration can serve as the free sale or registration certificate from the country of origin, but EDE names EC, 510(k) and PMA as the conformity certificate types.

Timeline Savings

None published

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

UgandaWHO · IMDRF member authorities · ISO

Prior approval / evidence record

WHO Prequalification, or quality-system certification from a certification body accredited by a regulatory authority in an IMDRF founding-member country, or any other international organisation recognised by NDA

Authority

WHO Prequalification, or quality-system certification from a certification body accredited by a regulatory authority in an IMDRF founding-member country, or any other international organisation recognised by NDA

Accepted

Yes

Benefit

Track 2: accepted as a documentary substitution for an IMDRF-country licence under DAR/GDL/028 §4.12.2 — the foreign element that unlocks the route is the quality-system certification and its accreditation rather than a product approval, and no shortened period is published for it.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

UgandaAustralia TGA · Health Canada · EU conformity assessment · Japan MHLW / PMDA · US FDA · IMDRF member authorities

Prior approval / evidence record

Licence or approval from an IMDRF founding-member country — Australia, Canada, the European Union, Japan, the United States

Authority

Licence or approval from an IMDRF founding-member country — Australia, Canada, the European Union, Japan, the United States

Accepted

Yes

Benefit

Track 1: an abridged evaluation in which the foreign approval carries the technical assessment, with a much smaller submission (attestation, abbreviated product and manufacturer details, notarized copy of the licence, Certificate of Analysis to an accepted quality-system standard, evidence of repeat sales and product samples where practicable). It is an expedited review with shortened registration timelines, but NDA publishes no number for the shortening and the final decision remains NDA's.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

United KingdomUS FDA · EU conformity assessment · Australia TGA · Health Canada · UK MHRA

Prior approval / evidence record

International Reliance (draft legislation, not in force)

Name

International Reliance (draft legislation, not in force)

Description

Stand-alone route giving a Certificate of International Reliance to devices already authorised by the FDA, Health Canada or the TGA, without UKCA marking; the EU is deliberately excluded pending the separate CE-recognition decision

Timeline Reduction

Not quantified — the draft fixes no review period

Eligibility
  • Draft Medical Devices (Amendment) Regulations 2026, WTO-notified 8 May 2026
  • Route 1 lowest-risk: self-declaration and registration only
  • Routes 2 and 3: certificate from a UK approved body first
  • Anticipated in force mid-2028

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

VietnamNMPA · US FDA · EU conformity assessment · Australia TGA · Japan MHLW / PMDA · UK MHRA · Swissmedic · South Korea MFDS

Prior approval / evidence record

Class C

Id

class-c

Class Name

Class C

Risk Level

high

Pathway

Circulation number registration (Ministry of Health)

Official Timeline
Min

45

Max

45

Unit

days

Verification Status

official-verified

Official Source
Title

Nghị định 98/2021/NĐ-CP on the management of medical devices (as amended by Decrees 07/2023, 96/2023, 04/2025) — Art. 28 declaration for Class A/B and Art. 32 appraisal periods for Class C/D

Url

https://vanban.chinhphu.vn/?pageid=27160&docid=204442

Effective Date

2022-01-01

Accessed Date

2026-08-29

Realistic Timeline
Min

4

Max

5

Unit

months

With Prior Approval
Min

10

Max

10

Unit

working-days

Prior Approval Benefits
  • FDA
  • TGA
  • MHLW
  • EU
  • MHRA
  • Swissmedic
  • NMPA
  • MFDS
Notes
  • Decree 98/2021 Art. 32(5): the Minister of Health must appraise and grant the circulation number within 45 days of a complete and valid dossier, counted only once the appraisal fee receipt is in, and must issue any request for supplementation within 40 days
  • Art. 32(2) shortens this to 30 days, with a 25-day supplementation notice, where a corresponding national technical regulation (QCVN) exists for the device, and the same 30 days applies to measuring instruments that need pattern approval
  • Art. 32(3): the fast track for devices already authorised in a reference country, or previously import-licensed, is decided in 10 working days with a 10-working-day supplementation notice — here the Ministry appraises only the legal documents and verifies the rest after the number is granted
  • Art. 32(4): the emergency route is also 10 working days, with an 8-working-day supplementation notice
  • The clock does not simply run out in the applicant favour: if the dossier is not corrected within 90 days of a request, or fails after five rounds of supplementation, the procedure restarts from the beginning
  • Art. 32(6): the granted number and the dossier are published on the medical device portal within 1 working day
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

VietnamNMPA · US FDA · EU conformity assessment · Australia TGA · Japan MHLW / PMDA · UK MHRA · Swissmedic · South Korea MFDS

Prior approval / evidence record

Class D

Id

class-d

Class Name

Class D

Risk Level

highest

Pathway

Circulation number registration (Ministry of Health)

Official Timeline
Min

45

Max

45

Unit

days

Verification Status

official-verified

Official Source
Title

Nghị định 98/2021/NĐ-CP on the management of medical devices (as amended by Decrees 07/2023, 96/2023, 04/2025) — Art. 28 declaration for Class A/B and Art. 32 appraisal periods for Class C/D

Url

https://vanban.chinhphu.vn/?pageid=27160&docid=204442

Effective Date

2022-01-01

Accessed Date

2026-08-29

Realistic Timeline
Min

4

Max

5

Unit

months

With Prior Approval
Min

10

Max

10

Unit

working-days

Prior Approval Benefits
  • FDA
  • TGA
  • MHLW
  • EU
  • MHRA
  • Swissmedic
  • NMPA
  • MFDS
Notes
  • Decree 98/2021 Art. 32(5): the Minister of Health must appraise and grant the circulation number within 45 days of a complete and valid dossier, counted only once the appraisal fee receipt is in, and must issue any request for supplementation within 40 days
  • Art. 32(2) shortens this to 30 days, with a 25-day supplementation notice, where a corresponding national technical regulation (QCVN) exists for the device, and the same 30 days applies to measuring instruments that need pattern approval
  • Art. 32(3): the fast track for devices already authorised in a reference country, or previously import-licensed, is decided in 10 working days with a 10-working-day supplementation notice — here the Ministry appraises only the legal documents and verifies the rest after the number is granted
  • Art. 32(4): the emergency route is also 10 working days, with an 8-working-day supplementation notice
  • The clock does not simply run out in the applicant favour: if the dossier is not corrected within 90 days of a request, or fails after five rounds of supplementation, the procedure restarts from the beginning
  • Art. 32(6): the granted number and the dossier are published on the medical device portal within 1 working day
Clinical Trial Required

Yes

Local Testing Required

No

Gmp Required

Yes

Maintained reference · planning ranges remain estimates · reviewed 2026-08-29

ZambiaIMDRF member authorities · WHO · US FDA · EU conformity assessment

Prior approval / evidence record

WHO Prequalification; US FDA approval; EU CE marking; approval in any IMDRF member state

Authority

WHO Prequalification; US FDA approval; EU CE marking; approval in any IMDRF member state

Accepted

No

Benefit

Disclosure only. The Annex I application form requires the applicant to state whether the device is WHO prequalified, US FDA approved, CE marked or approved in any IMDRF member state and to attach valid documentation, but neither the non-IVD nor the IVD guideline attaches any procedural consequence — no shortened clock, no abridged dossier and no substitute for ZAMRA's own evaluation.

Maintained reference · planning ranges remain estimates · reviewed 2026-09-16

Prior approvals and evidence for Australia (8 records)

Prior approval / evidence record

MDSAP

Authority

MDSAP

Accepted

Yes

Benefit

Eliminates separate QMS audits

Timeline Savings

2-3 months

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

AustraliaUS FDA · EU conformity assessment · MDSAP

Prior approval / evidence record

Class III — ARTG inclusion

Id

class-iii

Class Name

Class III — ARTG inclusion

Risk Level

highest

Pathway

Inclusion in the ARTG on overseas or TGA conformity assessment evidence

Official Timeline
Min

20

Max

20

Unit

working-days

Realistic Timeline
Min

6

Max

9

Unit

months

With Prior Approval
Min

3

Max

5

Unit

months

Prior Approval Benefits
  • EU CE Mark (30-50% faster)
  • FDA 510(k)
  • MDSAP
Notes
  • The inclusion decision itself still carries the 20-working-day statutory clock. What makes Class III slow is what sits either side of it: an application audit, which has no legislated timeframe, and where TGA conformity assessment is required, a separate certificate that must be obtained first
  • Non-IVD Level 2 application audits: TGA expects the first round of specialist assessment within 100 working days, and will explain a revised timeframe if it cannot meet that
  • Sponsor responses to a s41JA notice or a Proposal to Refuse are due within 20-40 business days depending on complexity, and that time does not count against TGA targets
  • The published targets replaced an older blanket 60-working-day audit target that TGA says was arbitrary and not consistently met
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Verification Status

official-verified

Official Source
Title

TGA — processing your application for inclusion of a medical device: the decision is notified within 20 working days after the day the application fee was paid

Url

https://www.tga.gov.au/products/medical-devices/application-and-market-authorisation/supply-medical-device/medical-device-inclusion-process/processing-your-application-inclusion-medical-device

Accessed Date

2026-08-30

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Regulatory reference

China NMPA approval: not reference

reliance.china-nmpa-not-reference
Pending verification
Statement

TGA accepts market-authorisation evidence only from its designated comparable overseas regulators and assessment bodies: EU notified bodies (MDD/AIMDD/IVDD, MDR, IVDR), US FDA (PMA and 510(k)), Health Canada, Japan MHLW/PMDA, Singapore HSA, and UK approved bodies designated by MHRA, with MDSAP certificates as manufacturer evidence. China's NMPA is not a comparable overseas regulator, so an NMPA registration certificate cannot be used as manufacturer evidence or product-assessment evidence for ARTG inclusion or to abridge a TGA conformity assessment.

Source Jurisdiction Id

china

Source Authority Id

nmpa

Source Approval

China NMPA medical device registration certificate (Class II/III) or Class I filing record

Target Jurisdiction Id

australia

China Classes In Scope

All classes; ARTG inclusion applications and abridgement of TGA conformity assessment

Edge Type

No Reliance Route

Market Access Strength

none

Regulatory Layer

Product Marketing

Local Application Required

Yes

Target Decision Retained

Yes

Conditions
  • Evidence must come from the listed regulators/bodies and be for the same device and manufacturer
  • Australian sponsor and Australian requirements remain
  • Independent verification is incomplete. Treat this record as a research lead, not a confirmed benefit or exclusion; see the 2026-09-15 independent review.

Effective from 2018-10-01

Source passage: Premarket approval and 510K decisions of the United States Food and Drug Administration (FDA) • Approvals and medical device licences issued by Health Canada • Pre-market approvals or certification from Japan (issued by the Ministry of Health, Labour and Welfare (MHLW), Pharmaceutical and Medical Devices Agency (PMDA) • Registrations or approvals of the Singapore Health Sciences Authority (HSA)

AustraliaUS FDA · EU conformity assessment · MDSAP

Prior approval / evidence record

Class IIb — ARTG inclusion

Id

class-iib

Class Name

Class IIb — ARTG inclusion

Risk Level

high

Pathway

Inclusion in the ARTG on overseas or TGA conformity assessment evidence

Official Timeline
Min

20

Max

20

Unit

working-days

Realistic Timeline
Min

1.5

Max

2

Unit

months

With Prior Approval
Min

1

Max

1.5

Unit

months

Prior Approval Benefits
  • EU CE Mark
  • FDA 510(k)
  • MDSAP
Notes
  • Same 20-working-day statutory inclusion clock; audit selection is more likely at this class and suspends any published limit
  • Level 1 compulsory audits ran to a median of 14 and a mean of 27 TGA days in the published snapshot, and Level 2 compulsory audits to a median of 161 and a mean of 174
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Verification Status

official-verified

Official Source
Title

TGA — processing your application for inclusion of a medical device: the decision is notified within 20 working days after the day the application fee was paid

Url

https://www.tga.gov.au/products/medical-devices/application-and-market-authorisation/supply-medical-device/medical-device-inclusion-process/processing-your-application-inclusion-medical-device

Accessed Date

2026-08-30

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

AustraliaUS FDA

Prior approval / evidence record

FDA 510(k)

Authority

FDA 510(k)

Accepted

Yes

Benefit

Accepted for Class IIa/IIb

Timeline Savings

20-30%

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

AustraliaUS FDA · EU conformity assessment · MDSAP

Prior approval / evidence record

Class IIa — ARTG inclusion

Id

class-iia

Class Name

Class IIa — ARTG inclusion

Risk Level

medium

Pathway

Inclusion in the ARTG on overseas or TGA conformity assessment evidence

Official Timeline
Min

20

Max

20

Unit

working-days

Realistic Timeline
Min

1

Max

1.5

Unit

months

With Prior Approval
Min

0.75

Max

1

Unit

months

Prior Approval Benefits
  • EU CE Mark
  • FDA 510(k)
  • MDSAP
Notes
  • The 20-working-day statutory clock runs to the inclusion decision, not to the end of an audit: if the application is selected for audit there is no legislated limit on what follows
  • Non-Class I applications completed without audit took a median of 9 and a mean of 10 TGA days in the published snapshot
  • Australia is unusual in splitting conformity assessment from market authorisation, which is what allows overseas certification to be used in place of a TGA assessment at this class
Clinical Trial Required

No

Local Testing Required

No

Gmp Required

Yes

Verification Status

official-verified

Official Source
Title

TGA — processing your application for inclusion of a medical device: the decision is notified within 20 working days after the day the application fee was paid

Url

https://www.tga.gov.au/products/medical-devices/application-and-market-authorisation/supply-medical-device/medical-device-inclusion-process/processing-your-application-inclusion-medical-device

Accessed Date

2026-08-30

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

AustraliaSource authority not specified

Prior approval / evidence record

Priority Review Pathway

Name

Priority Review Pathway

Description

Front-of-queue placement for novel devices addressing unmet needs

Timeline Reduction

30-50% faster

Eligibility
  • Novel device
  • Addresses unmet clinical need
  • No marketed alternatives

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

AustraliaEU conformity assessment

Prior approval / evidence record

EU CE Mark

Authority

EU CE Mark

Accepted

Yes

Benefit

Abridged assessment route - major timeline reduction

Timeline Savings

30-50%

Maintained reference · planning ranges remain estimates · reviewed 2026-08-30

Review clocks

What review, validation, and applicant stop-clock rules apply?

Compare countries

Sponsor-notification processing target

secondary.answers.sponsor-processing
Fact-level verified

There is no fee for the sponsor transfer/name notification itself. TGA says it normally updates the ARTG within ten working days of a fully completed, signed notification and any required information; the form describes this as an aim. This administrative processing target does not defer legal sponsorship responsibility until the record is updated or price additional ARTG variations.

Variation evaluation has no general legislated clock

secondary.answers.variation-clock-boundary
Fact-level verified

The current-linked TGA variation guide expressly says there is no legislated timeframe for evaluating an ARTG variation. Its 3-5-business-day target applies only to the stated Class III programming-software variation after the new hardware is approved. The reasonable information-request response period of at least ten working days is an applicant deadline, not a general evaluation SLA.

Conditions
  • Do not substitute initial ARTG inclusion or conformity-assessment timelines for a variation.

UDI, listing & traceability

Which UDI, device listing, database, and traceability duties apply?

Compare countries

Brand/model UDI changes are distinct from an order REF

secondary.answers.udi-ref
Fact-level verified

For a UDI medical device, Essential Principle 13C.2 treats the listed brand-name, model/version and packaging changes as a different type of model, with corresponding database identifiers. Changes made in the same production change are treated together; the pre-change device remains its prior type. Database timing depends on the applicable general start day and supply event. A commercial REF must first be mapped to these fields; UDI consequences do not themselves determine ARTG new-kind or variation classification.

Conditions
  • Select the actual UDI scope, class, start-day and supply dates; do not apply all future stages as current.

Related structured datasets

Classification systems · 1 record
Open Australia IVD risk classification →
Id

australia:ivd

Jurisdiction Id

australia

Domains
  • ivd
Name

Australia IVD risk classification

Classes
  • Id

    1

    Label

    IVD Class 1

    Normalized Risk Level

    low

  • Id

    2

    Label

    IVD Class 2

    Normalized Risk Level

    medium

  • Id

    3

    Label

    IVD Class 3

    Normalized Risk Level

    high

  • Id

    4

    Label

    IVD Class 4

    Normalized Risk Level

    highest

Regulator

TGA

Official Domains
  • tga.gov.au
Source Status

official-source-verified

Source Title

TGA overview of medical devices and IVD regulation

Source Url

https://www.tga.gov.au/products/medical-devices/overview/overview-medical-devices-and-ivd-regulation

Verified On

2026-08-13

Source Supports
  • ivd-class-axis
  • ivd-classification-rules
Registration pathways, prior approvals and planning · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Australia registration →
Country

Australia

Jurisdiction Id

australia

Country Code

AU

Flag

🇦🇺

Regulator

Therapeutic Goods Administration

Regulator Abbrev

TGA

Region

asia-pacific

Tier

1

Risk Classification

5-class (I, IIa, IIb, III, AIMD)

Classes
  • Id

    class-i

    Class Name

    Class I — ARTG inclusion

    Risk Level

    low

    Pathway

    Inclusion in the Australian Register of Therapeutic Goods

    Official Timeline
    Min

    20

    Max

    20

    Unit

    working-days

    Realistic Timeline
    Min

    0.5

    Max

    1

    Unit

    months

    Notes
    • TGA notifies the inclusion decision within 20 working days after the day the application fee was paid, the same statutory clock that applies to every class
    • Since October 2020 Class I applications have run through the same assessment and audit-selection process as every other class, so "notification only" no longer describes the route
    • In the published half-yearly snapshot, Class I applications completed without audit took a median of 1 and a mean of 3 TGA days; those selected for audit took a median of 11 and a mean of 31
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    No

    Verification Status

    official-verified

    Official Source
    Title

    TGA — processing your application for inclusion of a medical device: the decision is notified within 20 working days after the day the application fee was paid

    Url

    https://www.tga.gov.au/products/medical-devices/application-and-market-authorisation/supply-medical-device/medical-device-inclusion-process/processing-your-application-inclusion-medical-device

    Accessed Date

    2026-08-30

  • Id

    class-iia

    Class Name

    Class IIa — ARTG inclusion

    Risk Level

    medium

    Pathway

    Inclusion in the ARTG on overseas or TGA conformity assessment evidence

    Official Timeline
    Min

    20

    Max

    20

    Unit

    working-days

    Realistic Timeline
    Min

    1

    Max

    1.5

    Unit

    months

    With Prior Approval
    Min

    0.75

    Max

    1

    Unit

    months

    Prior Approval Benefits
    • EU CE Mark
    • FDA 510(k)
    • MDSAP
    Notes
    • The 20-working-day statutory clock runs to the inclusion decision, not to the end of an audit: if the application is selected for audit there is no legislated limit on what follows
    • Non-Class I applications completed without audit took a median of 9 and a mean of 10 TGA days in the published snapshot
    • Australia is unusual in splitting conformity assessment from market authorisation, which is what allows overseas certification to be used in place of a TGA assessment at this class
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

    Verification Status

    official-verified

    Official Source
    Title

    TGA — processing your application for inclusion of a medical device: the decision is notified within 20 working days after the day the application fee was paid

    Url

    https://www.tga.gov.au/products/medical-devices/application-and-market-authorisation/supply-medical-device/medical-device-inclusion-process/processing-your-application-inclusion-medical-device

    Accessed Date

    2026-08-30

  • Id

    class-iib

    Class Name

    Class IIb — ARTG inclusion

    Risk Level

    high

    Pathway

    Inclusion in the ARTG on overseas or TGA conformity assessment evidence

    Official Timeline
    Min

    20

    Max

    20

    Unit

    working-days

    Realistic Timeline
    Min

    1.5

    Max

    2

    Unit

    months

    With Prior Approval
    Min

    1

    Max

    1.5

    Unit

    months

    Prior Approval Benefits
    • EU CE Mark
    • FDA 510(k)
    • MDSAP
    Notes
    • Same 20-working-day statutory inclusion clock; audit selection is more likely at this class and suspends any published limit
    • Level 1 compulsory audits ran to a median of 14 and a mean of 27 TGA days in the published snapshot, and Level 2 compulsory audits to a median of 161 and a mean of 174
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

    Verification Status

    official-verified

    Official Source
    Title

    TGA — processing your application for inclusion of a medical device: the decision is notified within 20 working days after the day the application fee was paid

    Url

    https://www.tga.gov.au/products/medical-devices/application-and-market-authorisation/supply-medical-device/medical-device-inclusion-process/processing-your-application-inclusion-medical-device

    Accessed Date

    2026-08-30

  • Id

    class-iii

    Class Name

    Class III — ARTG inclusion

    Risk Level

    highest

    Pathway

    Inclusion in the ARTG on overseas or TGA conformity assessment evidence

    Official Timeline
    Min

    20

    Max

    20

    Unit

    working-days

    Realistic Timeline
    Min

    6

    Max

    9

    Unit

    months

    With Prior Approval
    Min

    3

    Max

    5

    Unit

    months

    Prior Approval Benefits
    • EU CE Mark (30-50% faster)
    • FDA 510(k)
    • MDSAP
    Notes
    • The inclusion decision itself still carries the 20-working-day statutory clock. What makes Class III slow is what sits either side of it: an application audit, which has no legislated timeframe, and where TGA conformity assessment is required, a separate certificate that must be obtained first
    • Non-IVD Level 2 application audits: TGA expects the first round of specialist assessment within 100 working days, and will explain a revised timeframe if it cannot meet that
    • Sponsor responses to a s41JA notice or a Proposal to Refuse are due within 20-40 business days depending on complexity, and that time does not count against TGA targets
    • The published targets replaced an older blanket 60-working-day audit target that TGA says was arbitrary and not consistently met
    Clinical Trial Required

    No

    Local Testing Required

    No

    Gmp Required

    Yes

    Verification Status

    official-verified

    Official Source
    Title

    TGA — processing your application for inclusion of a medical device: the decision is notified within 20 working days after the day the application fee was paid

    Url

    https://www.tga.gov.au/products/medical-devices/application-and-market-authorisation/supply-medical-device/medical-device-inclusion-process/processing-your-application-inclusion-medical-device

    Accessed Date

    2026-08-30

  • Id

    conformity-assessment-new

    Class Name

    TGA Conformity Assessment certificate — new or substantial change

    Pathway

    TGA-issued conformity assessment certification

    Official Timeline
    Min

    160

    Max

    200

    Unit

    working-days

    Realistic Timeline
    Min

    10

    Max

    15

    Unit

    months

    Notes
    • Regulation 4.3 caps every Conformity Assessment application, including IVDs, at 255 working days. Within that cap TGA publishes a non-legislated target of 160 business days and a maximum of 200, aiming to meet the target in 90% of applications
    • These are TGA days only. TGA estimates 40-100 business days of applicant time on top, for a total of roughly 10-15 months
    • Segments inside the clock: pre-assessment within 30 business days, then all component areas aim to finish round 1 within 100 business days, with 20-40 business days for each applicant response round
    • The application clock stops while the applicant holds the file, and can be stopped by mutual agreement in extenuating circumstances — but not for holidays, missed TGA requests, or time spent translating documents or generating missing information
    • An on-site QMS audit report is issued within 60 days of the audit; MDSAP or EU notified-body audit evidence can substitute a desktop assessment for the on-site audit
    Gmp Required

    Yes

    Verification Status

    official-verified

    Official Source
    Title

    TGA — applying for a Conformity Assessment certificate for medical devices: Regulation 4.3 requires processing within 255 working days, with non-legislated targets of 160/200 business days for new or substantial applications, 190/225 where ACMD advice or an on-site audit is needed, and 80/150 for recertification

    Url

    https://www.tga.gov.au/resources/guidance/applying-conformity-assessment-certificate-medical-devices

    Accessed Date

    2026-08-30

  • Id

    conformity-assessment-committee-or-onsite-audit

    Class Name

    TGA Conformity Assessment certificate — ACMD advice or on-site facility audit

    Pathway

    TGA-issued conformity assessment certification with committee referral or on-site audit

    Official Timeline
    Min

    190

    Max

    225

    Unit

    working-days

    Realistic Timeline
    Min

    12

    Max

    16

    Unit

    months

    Notes
    • Applies where the Advisory Committee on Medical Devices is consulted or an on-site TGA audit of manufacturing facilities is required, both of which are most common for high-risk or novel devices
    • Target 190 business days with TGA and a maximum of 225, plus 40-100 business days of applicant time, for roughly 12-16 months in total
    • Still bounded by the same 255-working-day Regulation 4.3 cap
    Gmp Required

    Yes

    Verification Status

    official-verified

    Official Source
    Title

    TGA — applying for a Conformity Assessment certificate for medical devices: Regulation 4.3 requires processing within 255 working days, with non-legislated targets of 160/200 business days for new or substantial applications, 190/225 where ACMD advice or an on-site audit is needed, and 80/150 for recertification

    Url

    https://www.tga.gov.au/resources/guidance/applying-conformity-assessment-certificate-medical-devices

    Accessed Date

    2026-08-30

  • Id

    conformity-assessment-recertification

    Class Name

    TGA Conformity Assessment certificate — recertification

    Pathway

    Recertification of an expiring TGA conformity assessment certificate

    Official Timeline
    Min

    80

    Max

    150

    Unit

    working-days

    Realistic Timeline
    Min

    6

    Max

    10

    Unit

    months

    Notes
    • Target 80 business days with TGA and a maximum of 150, plus 40-60 business days of applicant time, for roughly 6-10 months in total
    • Substantial changes to an existing certificate are not recertifications: they run on the new-or-substantial target of 160/200 business days
    Gmp Required

    Yes

    Verification Status

    official-verified

    Official Source
    Title

    TGA — applying for a Conformity Assessment certificate for medical devices: Regulation 4.3 requires processing within 255 working days, with non-legislated targets of 160/200 business days for new or substantial applications, 190/225 where ACMD advice or an on-site audit is needed, and 80/150 for recertification

    Url

    https://www.tga.gov.au/resources/guidance/applying-conformity-assessment-certificate-medical-devices

    Accessed Date

    2026-08-30

Expedited Pathways
  • Name

    Priority Review Pathway

    Description

    Front-of-queue placement for novel devices addressing unmet needs

    Timeline Reduction

    30-50% faster

    Eligibility
    • Novel device
    • Addresses unmet clinical need
    • No marketed alternatives
Prior Approvals
  • Authority

    EU CE Mark

    Accepted

    Yes

    Benefit

    Abridged assessment route - major timeline reduction

    Timeline Savings

    30-50%

  • Authority

    FDA 510(k)

    Accepted

    Yes

    Benefit

    Accepted for Class IIa/IIb

    Timeline Savings

    20-30%

  • Authority

    MDSAP

    Accepted

    Yes

    Benefit

    Eliminates separate QMS audits

    Timeline Savings

    2-3 months

Key Requirements
  • Australian Sponsor (for foreign manufacturers)
  • ARTG (Australian Register of Therapeutic Goods) listing
  • Conformity assessment documentation
  • MDSAP or TGA audit
  • Two statutory clocks and nothing else: 20 working days to notify the inclusion decision, and 255 working days to process a Conformity Assessment application under Regulation 4.3. Application audits have no legislated timeframe, and neither does evaluating an application to vary an ARTG entry
  • Failing preliminary assessment is a refusal, not a request for more information, and the application fee is neither refunded nor transferable to a resubmission
  • Audit fees are payable only for mandatory audits, and TGA cannot start the audit until they are paid; non-mandatory audit selection currently carries no fee
  • Software variations that transition a software device to new intended hardware carry their own 3-5 business day target
Local Rep Required

Yes

Local Rep Name

Australian Sponsor

Mdsap Accepted

Yes

Mdsap Benefit

Eliminates separate TGA QMS audits, 5-year certificate validity

Registration Validity

5 years

Last Updated

2026-08-30

Verification Status

official-verified

Reviewed On

2026-08-30

Official Sources
  • Title

    TGA — processing your application for inclusion of a medical device: the decision is notified within 20 working days after the day the application fee was paid

    Url

    https://www.tga.gov.au/products/medical-devices/application-and-market-authorisation/supply-medical-device/medical-device-inclusion-process/processing-your-application-inclusion-medical-device

    Accessed Date

    2026-08-30

  • Title

    TGA — applying for a Conformity Assessment certificate for medical devices: Regulation 4.3 requires processing within 255 working days, with non-legislated targets of 160/200 business days for new or substantial applications, 190/225 where ACMD advice or an on-site audit is needed, and 80/150 for recertification

    Url

    https://www.tga.gov.au/resources/guidance/applying-conformity-assessment-certificate-medical-devices

    Accessed Date

    2026-08-30

  • Title

    TGA — understanding the medical device application audit process: statutory 20-working-day preliminary assessment, non-IVD Level 2 first-round specialist assessment expected in 100 working days, sponsor responses due in 20-40 business days

    Url

    https://www.tga.gov.au/resources/guidance/understanding-medical-device-application-audit-process

    Accessed Date

    2026-08-30

  • Title

    TGA — understanding processing timeframes for medical device applications: published half-yearly snapshot of actual processing times by application and audit type

    Url

    https://www.tga.gov.au/resources/guidance/understanding-processing-timeframes-medical-device-applications

    Accessed Date

    2026-08-30

  • Title

    TGA — varying your medical device or IVD ARTG entry: software transition variations carry a 3-5 business day target, and section 41JA or section 31 notices allow not less than 10 working days to respond

    Url

    https://www.tga.gov.au/resources/resources/user-guide/varying-your-medical-device-or-ivd-artg-entry

    Accessed Date

    2026-08-30

Process
Template Id

default

Status

planning-template

Usage Rule

Maintained planning steps and estimates, not verified regulator deadlines.

Steps
  • Step

    1

    Title

    Pre-Submission Preparation

    Description

    Gather technical documentation, quality management system (QMS) records, and clinical data required by TGA. Appoint a Australian Sponsor.

    Duration

    1-3 months

    Icon

    📋

  • Step

    2

    Title

    Application Submission

    Description

    Submit complete registration dossier to TGA including device description, risk analysis, and testing reports.

    Duration

    1-2 weeks

    Icon

    📤

  • Step

    3

    Title

    Administrative Review

    Description

    TGA reviews application for completeness. May request additional documents or clarifications.

    Duration

    2-4 weeks

    Icon

    🔍

  • Step

    4

    Title

    Technical Evaluation

    Description

    In-depth technical assessment of device safety and performance. TGA may issue questions.

    Duration

    20 working days

    Icon

    ⚙️

  • Step

    5

    Title

    Registration Approval

    Description

    Upon successful review, TGA issues Australia market authorization certificate.

    Duration

    1-2 weeks

    Icon

    ✅

Government fee schedules · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Australia fees →
Country

Australia

Jurisdiction Id

australia

Country Code

AU

Authority

TGA

Low Risk Local
Min

121

Max

879

Currency

AUD

Per Product

Yes

High Risk Local
Min

1244

Max

1662

Currency

AUD

Per Product

Yes

Details
  • Id

    annual-charge-class-i-other

    Item

    Annual Charge - Class I (other)

    Currency

    AUD

    Amount Local

    121

    Amount Local Formatted

    AUD 121

    Frequency

    Annual per ARTG entry

    Verification Status

    official-verified

    Source Url

    https://www.tga.gov.au/sites/default/files/2026-06/fees-charges-summary-1-july-2026.pdf

    Source Title

    TGA fees and charges summary from 1 July 2026

    Source Effective Date

    2026-07-01

    Verified On

    2026-08-25

    Amount Status

    exact-current

  • Id

    annual-charge-class-i-measuring-sterile

    Item

    Annual Charge - Class I (measuring/sterile)

    Currency

    AUD

    Amount Local

    879

    Amount Local Formatted

    AUD 879

    Frequency

    Annual per ARTG entry

    Verification Status

    official-verified

    Source Url

    https://www.tga.gov.au/sites/default/files/2026-06/fees-charges-summary-1-july-2026.pdf

    Source Title

    TGA fees and charges summary from 1 July 2026

    Source Effective Date

    2026-07-01

    Verified On

    2026-08-25

    Amount Status

    exact-current

  • Id

    annual-charge-class-iia-iib

    Item

    Annual Charge - Class IIa/IIb

    Currency

    AUD

    Amount Local

    1305

    Amount Local Formatted

    AUD 1,305

    Frequency

    Annual per ARTG entry

    Verification Status

    official-verified

    Source Url

    https://www.tga.gov.au/sites/default/files/2026-06/fees-charges-summary-1-july-2026.pdf

    Source Title

    TGA fees and charges summary from 1 July 2026

    Source Effective Date

    2026-07-01

    Verified On

    2026-08-25

    Amount Status

    exact-current

  • Id

    annual-charge-class-iii-aimd

    Item

    Annual Charge - Class III/AIMD

    Currency

    AUD

    Amount Local

    1662

    Amount Local Formatted

    AUD 1,662

    Frequency

    Annual per ARTG entry

    Verification Status

    official-verified

    Source Url

    https://www.tga.gov.au/sites/default/files/2026-06/fees-charges-summary-1-july-2026.pdf

    Source Title

    TGA fees and charges summary from 1 July 2026

    Source Effective Date

    2026-07-01

    Verified On

    2026-08-25

    Amount Status

    exact-current

  • Id

    application-fee-class-i

    Item

    Application Fee - Class I

    Currency

    AUD

    Amount Local

    651

    Amount Local Formatted

    AUD 651

    Frequency

    Per ARTG inclusion application

    Verification Status

    official-verified

    Source Url

    https://www.tga.gov.au/sites/default/files/2026-06/fees-charges-summary-1-july-2026.pdf

    Source Title

    TGA fees and charges summary from 1 July 2026

    Source Effective Date

    2026-07-01

    Verified On

    2026-08-25

    Amount Status

    exact-current

  • Id

    application-fee-class-iia-iib

    Item

    Application Fee - Class IIa/IIb

    Currency

    AUD

    Amount Local

    1244

    Amount Local Formatted

    AUD 1,244

    Frequency

    Per ARTG inclusion application

    Verification Status

    official-verified

    Source Url

    https://www.tga.gov.au/sites/default/files/2026-06/fees-charges-summary-1-july-2026.pdf

    Source Title

    TGA fees and charges summary from 1 July 2026

    Source Effective Date

    2026-07-01

    Verified On

    2026-08-25

    Amount Status

    exact-current

  • Id

    application-fee-class-iii

    Item

    Application Fee - Class III

    Currency

    AUD

    Amount Local

    1603

    Amount Local Formatted

    AUD 1,603

    Frequency

    Per ARTG inclusion application

    Verification Status

    official-verified

    Source Url

    https://www.tga.gov.au/sites/default/files/2026-06/fees-charges-summary-1-july-2026.pdf

    Source Title

    TGA fees and charges summary from 1 July 2026

    Source Effective Date

    2026-07-01

    Verified On

    2026-08-25

    Amount Status

    exact-current

  • Id

    ivd-annual-charge

    Item

    IVD Annual Charge

    Currency

    AUD

    Amount Local

    948

    Amount Local Formatted

    AUD 948

    Frequency

    Annual per ARTG entry

    Verification Status

    official-verified

    Source Url

    https://www.tga.gov.au/sites/default/files/2026-06/fees-charges-summary-1-july-2026.pdf

    Source Title

    TGA fees and charges summary from 1 July 2026

    Source Effective Date

    2026-07-01

    Verified On

    2026-08-25

    Amount Status

    exact-current

  • Id

    ivd-application-fee

    Item

    IVD Application Fee

    Currency

    AUD

    Amount Local

    1244

    Amount Local Formatted

    AUD 1,244

    Frequency

    Per ARTG inclusion application

    Verification Status

    official-verified

    Source Url

    https://www.tga.gov.au/sites/default/files/2026-06/fees-charges-summary-1-july-2026.pdf

    Source Title

    TGA fees and charges summary from 1 July 2026

    Source Effective Date

    2026-07-01

    Verified On

    2026-08-25

    Amount Status

    exact-current

  • Id

    udi-consent-to-supply-first-entry

    Item

    UDI Consent to Supply application - first ARTG entry

    Currency

    AUD

    Amount Local

    80

    Amount Local Formatted

    AUD 80

    Frequency

    Per application

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.tga.gov.au/products/medical-devices/labelling-and-advertising/unique-device-identification-udi-hub/udi-information-sponsors-and-manufacturers/unique-device-identification-consent-supply-process

    Source Title

    TGA Unique Device Identification Consent to Supply process and application fee

    Verified On

    2026-08-25

  • Id

    udi-consent-to-supply-additional-entry

    Item

    UDI Consent to Supply - each additional ARTG entry in the same application

    Currency

    AUD

    Amount Local

    10

    Amount Local Formatted

    AUD 10

    Frequency

    Per additional entry

    Amount Status

    exact-current

    Verification Status

    official-verified

    Source Url

    https://www.tga.gov.au/products/medical-devices/labelling-and-advertising/unique-device-identification-udi-hub/udi-information-sponsors-and-manufacturers/unique-device-identification-consent-supply-process

    Source Title

    TGA Unique Device Identification Consent to Supply process and application fee

    Verified On

    2026-08-25

Notes
  • Annual charges apply each financial year per ARTG entry
  • Application fees apply per ARTG inclusion application
  • Different fees for Class I subtypes
  • ARTG variation application fees follow Schedule 9, Part 2 Item 2A of the Therapeutic Goods Regulations 1990; notification of a change of sponsorship carries no fee.
  • Annual charges apply to every ARTG entry: invoices issue each July for non-ACE entries, non-payment within 20 working days is a ground for cancellation (s 41GL(f)) or suspension (s 41GA(1)(b)), and the Annual Charge Exemption applies until turnover commences. Current-year amounts sit in TGA’s downloadable fees and charges summaries.
Last Updated

2026-08-25

Verification Status

official-verified

Official Sources
  • Title

    TGA fees and charges summary from 1 July 2026

    Url

    https://www.tga.gov.au/sites/default/files/2026-06/fees-charges-summary-1-july-2026.pdf

    Effective Date

    2026-07-01

    Accessed Date

    2026-08-25

  • Title

    TGA Unique Device Identification Consent to Supply process and application fee

    Url

    https://www.tga.gov.au/products/medical-devices/labelling-and-advertising/unique-device-identification-udi-hub/udi-information-sponsors-and-manufacturers/unique-device-identification-consent-supply-process

    Accessed Date

    2026-08-25

  • Title

    TGA user guide: varying your medical device or IVD ARTG entry

    Url

    https://www.tga.gov.au/resources/resources/user-guide/varying-your-medical-device-or-ivd-artg-entry

    Accessed Date

    2026-08-25

  • Title

    TGA annual charges for therapeutic goods

    Url

    https://www.tga.gov.au/products/regulations-all-products/fees-and-payments/annual-charges/annual-charges-therapeutic-goods

    Accessed Date

    2026-08-25

Reviewed On

2026-08-25

Amount Status

exact-current

Range Available

Yes

Additional certifications and licences · 1 record

These country-level records can cover multiple product types. Check each item's scope and conditions before applying a fee, pathway or licence to an IVD.

Open Australia certifications →
Slug

australia

Market

Australia

Confidence

high

Schemes
  • Id

    tga-electrical-emc-evidence

    Name

    TGA electrical/EMC evidence

    Category

    Medical Electrical Safety

    Status

    Included In Medical Registration

    Object

    Finished Device

    Note

    TGA conformity-assessment audits the manufacturer QMS against regulatory requirements and internationally recognised standards, specifically ISO 13485:2016 (on-site or desktop assessments); microbiology and sterility assessments may reference ISO 11135 (ethylene oxide), the ISO 11137 series (radiation processing) and AAMI TIR 56.

    Source Urls
    • https://www.tga.gov.au/resources/guidance/applying-conformity-assessment-certificate-medical-devices
    Source Binding

    candidate

  • Id

    eess

    Name

    EESS

    Category

    General Electrical Components

    Status

    Mandatory If Feature

    Object

    Household Device Or Component

    Note

    Professional/TGA-regulated medical equipment is generally outside scope; household equipment and adapters require a separate scope test.

    Source Urls
    • https://www.eess.gov.au/wp-content/uploads/2020/07/20-009-General-guidance-Requirements-for-medical-equipment-the-EESS.pdf
    Source Binding

    reviewed

  • Id

    acma-rcm

    Name

    ACMA RCM

    Category

    Radio Spectrum

    Status

    Mandatory If Feature

    Object

    Radio Or Emc Equipment

    Source Urls
    • https://www.acma.gov.au/technical-standards
    Source Binding

    candidate

  • Id

    arpansa-state-radiation-licensing

    Name

    ARPANSA/state radiation licensing

    Category

    Ionizing Radiation

    Status

    Mandatory For Use Site

    Object

    Facility Or Commonwealth Entity

    Source Urls
    • https://www.arpansa.gov.au/regulation-and-licensing/licensing
    Source Binding

    candidate

Official Sources
  • Authority

    ACMA

    Topic

    technical standards and RCM

    Url

    https://www.acma.gov.au/technical-standards

  • Authority

    EESS

    Topic

    medical equipment scope guidance

    Url

    https://www.eess.gov.au/wp-content/uploads/2020/07/20-009-General-guidance-Requirements-for-medical-equipment-the-EESS.pdf

  • Authority

    ARPANSA

    Topic

    radiation regulation and licensing

    Url

    https://www.arpansa.gov.au/regulation-and-licensing/licensing

  • Authority

    TGA

    Topic

    conformity assessment certificate guidance QMS audit and sterility standards

    Url

    https://www.tga.gov.au/resources/guidance/applying-conformity-assessment-certificate-medical-devices

Keep researching this market
View the Australia resource hub →
Compare regulatory data across all countries →